Software Status Bulletin


OpenLAB CDS ChemStation Edition     M83xxAA

Known Problem Report as of Feb 19 2013 9:56AM

Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software product designated above. The SSB is derived from Known Problem Reports (KPR) which result from user problems that have been classified as documentation problems or software defects. When a KPR is written, an identifying number is assigned to it, and the KPR is added to the next edition of the SSB.

User inputs that have been classified as Enhancement Requests are not documented in the SSB. User problems that have been submitted, but that have not been classified by the time the SSB is generated are not included in the SSB.

How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the problem has been reported already, and if there is a temporary workaround available for the problem, or if the problem has already been fixed by a new revision. If the problem is not listed in this SSB then you may wish to report it to the Response Center or to your field support representative.

To determine if your problem is documented in this SSB, first look in the "Keyword Index" section of the SSB. Under each keyword is a listing of one-line descriptions of related KPRs. If any of these sound like yours, locate the KPR # in the "Known Problem Reports" section of the SSB, and read the full KPR. The KPRs in the "Known Problem Reports" section are sorted by KPR #.

There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a brief description and a KPR #. A KPR may be associated with more than one keyword.

Known Problem Reports: This section contains KPRs, with all the available information relevant to the problem. KPRs in this section are sorted by KPR #.


Keyword Glossary



1200 Control
1200 Control , Method Scouting Wizard
35900 Control
6850 Control , 6890 Control
6890 Control
7890A Control
ACQ
Audit Trail
Bitmaps
Calibration
CE Control
Client Report Rendering
Common Infrastructure
Companion
Custom Fields
Custom Reports
DAD/MWD Configuration
Data Analysis
Data Analysis , Navigation Table
Data Analysis , SEQ
Documentation
Easy Sequence
ECM Integration/OpenLab Option
ELSD Driver
GC ACQ
GC ACQ , SEQ
GC Analysis
GC Autosampler
Globalization
Globalization , Sampler Configuration
Initialization
Initialization , CE Control
Installation
Instrument Control
Integration
Integration , Audit Trail
Localization,Japanese
Logbook
Method
Method , ACQ
MSD Analysis
MSD Control
Navigation Table
Output
Peak Purity
Programming
Pump Configuration
Pump Method
Report Template Design
Report Templates
Reports
Reports , Custom Fields
Reports , Data Analysis
Reports , MSD Control
Reports , SEQ
Run Control
Sampler Configuration
Sampler Configuration , Sampler Status
Sampler Driver
Sampler Driver , Sampler Status
Sampler Method
SEQ
SEQ , 1100 Control , 1200 Control
Uninstall
User Interface
User Interface , Data Analysis
User Interface , SEQ , ACQ , ECM Integration/OpenLab Option
Utilities
Utilities , Spectra Analysis , MSD Analysis

Keyword Index


Keyword: 1100 Control

Lampall off switches off lamps and pumps
KPR#:1601

Keyword: 1200 Control

Wrong firmware revision displayed in Instrument Configuration dialog after firmware upgrade
KPR#:1539
Cluster information can be overwritten from the offline session
KPR#:1720
Lampall off switches off lamps and pumps
KPR#:1601

Keyword: 1200 Control , Method Scouting Wizard

Cluster information can be overwritten from the offline session
KPR#:1720

Keyword: 35900 Control

35900E - Error The Current Event Name is the same for a special event
KPR#:1623
cannot start two instruments using the two channels of one ADC at the same time
KPR#:1677
An AD converter cannot be split into two instruments from a ChemStation client
KPR#:1702
35900: Unable to set up Timed Events
KPR#:1707
View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.
KPR#:1721
ADC can be non-responsive after long reprocessing
KPR#:1739

Keyword: 6850 Control

Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826

Keyword: 6850 Control , 6890 Control

Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826

Keyword: 6890 Control

Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826
6890 RC.NET driver: instrument diagram not updated to show configuration
KPR#:1412
Addition of new columns to the column catalog (database) is not possible after upgrade
KPR#:1464
Sequence continues while GC temperature is not ready
KPR#:1622
6890 always opens to Configuration tab when opening Edit Agilent 6890 GC Parameters
KPR#:1684
Column info is missing with 6890
KPR#:1685
6890 RC.NET: Edit a running Method by making the oven hold time longer and click ok, then the Sequence/Single run stops
KPR#:1784

Keyword: 7890A Control

7890A: ALS errors abort sequence; no Retry/Abort option offered
KPR#:914
7890A Runs cannot be started until the GC and injector are idle, view GC Status
KPR#:1239
ALS errors abort sequence; no Retry/Abort dialog box option offered
KPR#:1329
Sequence aborts if vial not found with Overlap injection enabled (7890A only)
KPR#:1569
Must restart data system for barcode reader availability change (7890A)
KPR#:1625

Keyword: ACQ

Not possible to modify data analysis parameters of method during a single run in the offline; cannot save
KPR#:1641
No special characters should be used for names of methods, data files or sequences
KPR#:1660
Terminal Service can close under high workload
KPR#:1701
Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”
KPR#:1732
Changes in sequence methods made during acquisition are not saved automatically
KPR#:1742
Queue Method vial range reflects range of loaded method, not the method browsed to
KPR#:1747
Acq.txt and acq.macaml is missing for 7890A back injector only sequences
KPR#:1811
"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763

Keyword: Audit Trail

Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured
KPR#:1749
UIB II - set a timetable and control "Method Audit Trail"
KPR#:1819
manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676

Keyword: Bitmaps

Intelligent reporting does not support exponential baselines
KPR#:13
Insufficient contrast on b/w printers running in eco mode
KPR#:30
Signal Plot thickness appears inhomogenous in case of high data rates
KPR#:31

Keyword: CE Control

The VialTable check does not occur during the startup of the CE ChemSation
KPR#:1587
CE G1600 instrument cannot be launched
KPR#:1664

Keyword: Calibration

Amount is given as zero when it actually could not be calculated
KPR#:1311
Calibration method printout does not print compound level's Heights when quantitation is based on height
KPR#:1766
Re-establish the old layout for calibration reports
KPR#:1793
Changing calibration table in Batch review generates an error
KPR#:1799
Calibration method printout does not print compound level heights when quantitation is based on height
KPR#:1807
C.01.04 HotFix03 : Amount units in calibration report printout is not updated
KPR#:1809
Sequence run for bracketing standard which is started with blank completed without creating data files
KPR#:1817
C.01.04 Hot Fix03 : incorrect calibration curves legend in calibration curves printout
KPR#:1827

Keyword: Client Report Rendering

Method information partially hidden in intelligent reports exported to XLS file format
KPR#:15
"Report Printed" time stamp in report footer is several hours off in Print Preview
KPR#:16
Localized character in saved PDF report cannot be searched in Adobe Reader
KPR#:20
Page numbering start counting at 0 in case of .doc report file format
KPR#:21
Scaling a signal to "full scale" does not add a percent sign to the axis label
KPR#:23
Report shows unexpected blank pages
KPR#:24
Report Export to .TXT or .DOC may fail if report renderer creates an out-of-memory condition
KPR#:28
Ineractively selected report parameters switch back to default when printing to PDF
KPR#:29
Interactivly entered report parameters are not always updated
KPR#:34
Large matrix reports may produce additional empty pages
KPR#:628
Calculating the standard deviation of identical numbers may return NaN
KPR#:653
Changes in report templates are not immediately shown on report printouts
KPR#:657
Rendering error when using a template containing a large result matrix
KPR#:659

Keyword: Common Infrastructure

High load on CAN may cause “Spectra runbuffer overflow” entries and a loss of spectra
KPR#:18

Keyword: Companion

Companion fails to run blank run for valved/ALS injections
KPR#:1609

Keyword: Custom Fields

Custom Field content might get lost when a sequence is edited
KPR#:1771
Custom fields are not transferred to new result set
KPR#:1789
Custom Fields table erased when opening the Sequence table while running
KPR#:1795
Missing Sample Custom fields in the report
KPR#:1810
Custom field values become blank in the sequence table
KPR#:1828
Empty Custom Fields are not printed in Intelligent Report
KPR#:1837

Keyword: Custom Reports

Relative Retention Time solution does not work for uncalibrated peaks
KPR#:980

Keyword: DAD/MWD Configuration

Installing a lamp with tag may invalidate "alway use lamp" setting in detector's control
KPR#:3
Autobalace executed although in Sequence table set to "OFF"
KPR#:26

Keyword: Data Analysis

Reprocess a sequence that is still in progress
KPR#:1356
Reprocessing copy starting as a background window.
KPR#:1578
Data System will not "Autoscale" a TCD data file to the correct peak
KPR#:1639
Summed Peaks calculation error in the Summed Peaks Report
KPR#:1651
Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation
KPR#:1652
ChemStation B.04.03 [54] will not "Autoscale" the TCD data file
KPR#:1654
Red X mark displays in data explorer of offline data analysis
KPR#:1706
Error "An additional register must be added to the list of locked registers"
KPR#:1762
Baseline not displayed properly when using ChemStation AIC’s
KPR#:1768
In Recalculate Last Result mode no reports can be printed after method modification
KPR#:1785
Report template not found when reprocessing with a master method
KPR#:1796
Data analysis: Fractions are not displayed in UIB chromatogram
KPR#:1813
Wrong method displayed in navigation table after partial reprocessing
KPR#:1678
PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688

Keyword: Data Analysis , Navigation Table

Wrong method displayed in navigation table after partial reprocessing
KPR#:1678

Keyword: Data Analysis , SEQ

PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662

Keyword: Documentation

HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher
KPR#:1575
Understanding Your ChemStation manual needs more information about peak width calculations
KPR#:1734
G2613A 7683A injector firmware is incorrect in the Firmware guide
KPR#:1783

Keyword: ECM Integration/OpenLab Option

OpenLAB Option: Difficulties to lock the ChemStation during reprocessing
KPR#:1212
Problems after moving items from ChemStation to a different LCDF path in ECM
KPR#:1215
no automatic "check-in" after deleting a checked-out method
KPR#:1267
Acquisition only runs do not write instrument and column parameters to ACAML
KPR#:1309
token "operatorname" uses full name including special characters
KPR#:1363
ACAML result file does not contain all peak performance parameters
KPR#:1428
OpenLAB Option: data not marked as modified after changing the method
KPR#:1449
chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode
KPR#:1453
Path definition in OpenLab Control Panel is not used in ChemStation
KPR#:1611
Data not transferred to ECM after resuming the queue in the ChemStation
KPR#:1629
The storage path for files in ECM is user-specific
KPR#:1659
Cannot switch current ECM authentication to a different ECM server
KPR#:1667
ChemStore legacy data doesn't get reprocessed automatically
KPR#:1714
Wrong error message when downloading a method from ECM
KPR#:1753
Audit entry generated in the ECM or ECM Queue when doing Partial Sequence
KPR#:1774
ECM user licenses not released when user is changed in OLCDS ChemStation
KPR#:1794
Incorrect file name suggested when uploading from Review view
KPR#:1797
Update methods from ECM dialog locks up
KPR#:1806
In file information, the Operator shows disconnected on client
KPR#:1814
No warning before taking over session for data analysis only instruments
KPR#:1815
User token in upload path is resolved to current user - not to submit user
KPR#:1818
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1833
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763

Keyword: ELSD Driver

In localized CDS for ELSD detector G4218A no signal data is shown or stored
KPR#:4
ELSD G4218A can not handle change of Sample Time during run
KPR#:7

Keyword: Easy Sequence

Easy Sequences can end up in unexpected locations
KPR#:1655
Queue Planner will not run sequences of differing injection source
KPR#:1670

Keyword: GC ACQ

Methods have to be in single folder
KPR#:1185
Sequence stops when pushing the GC STOP key during sequence
KPR#:1600
FetchCPVar Cannot Talk to Application Error
KPR#:1608
The ability to collect partial data for specified times is needed for RC.NET GC drivers
KPR#:1626
FetchCPVar Cannot Talk to Application Error
KPR#:1628
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1630
G1888A - Handshake Abort - Vial Status does not update
KPR#:1638
Partial storage of data is not an option for GC RC.Net drivers
KPR#:1657
Error "creation of a data object failed“ when acquiring runs with data rates 1 Hz and below with 6890 RC.NET
KPR#:1680
OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed
KPR#:1683
Pulsed Split/Splitless injection does not wait for ready 6890
KPR#:1693
7890A stops run, ChemStation keeps going; GC Connection state is MONITOR in Status
KPR#:1700
Capacity factor (k') is not calculated; the file acqres.reg on the .D directory contains no column information
KPR#:1759
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604

Keyword: GC ACQ , SEQ

HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604

Keyword: GC Analysis

Self-assembled result sets do not have full functionality for GC data
KPR#:1741

Keyword: GC Autosampler

Unclear error message: Undefined symbol: License$
KPR#:1738

Keyword: Globalization

The PDF report on a Japanese Windows XP system might look garbled.
KPR#:5
Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17

Keyword: Globalization , Sampler Configuration

Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17

Keyword: Initialization

B.04.03 [16], error, no string 8002, Abort this copy of Alice?
KPR#:1615
Sporadically launch of an online session will not proceed
KPR#:1618
User with corresponding privilege cant't launch instrument configured on the AIC
KPR#:1658
CE G1600 instrument cannot be launched
KPR#:1664

Keyword: Initialization , CE Control

CE G1600 instrument cannot be launched
KPR#:1664

Keyword: Installation

missing link and no automatic installation of the localized readme.txt of the CDS ed.
KPR#:1589
AIC ports not opened for 7890A acquisition
KPR#:1627

Keyword: Instrument Control

Methods Must be Stored on the C Drive
KPR#:90

Keyword: Integration

Determination of tangent point on the tailing part of the peak
KPR#:821
Integration unable to handle complex chromatogram for area sum integration event
KPR#:981
Peak areas may be calculated incorrectly when using "New Exponential" as tangent skim mode
KPR#:1580
Data lost when using area sum integration event, must have 0.06 minute space between groups
KPR#:1642
Misleading display of tangent peak width
KPR#:1744
Timed integration event "Height Reject" cannot be configured.
KPR#:1800
GC compound peaks might be labelled as solvent peaks
KPR#:1838
manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676

Keyword: Integration , Audit Trail

manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676

Keyword: Localization,Japanese

Japanese characters from saved pdf report are garbled when copied and pasted to another application
KPR#:1668

Keyword: Logbook

'Method not downloaded to device' message when PAL is present
KPR#:1637
LC/MS ChemStation cannot open Logbook files
KPR#:1649

Keyword: MSD Analysis

Deconvolution tools are not available for LC/MS
KPR#:1665
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616

Keyword: MSD Control

Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment
KPR#:1675
MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled
KPR#:1686
LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error
KPR#:1730
Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681

Keyword: Method

"Error occurred during loading method" after rejecting method resolution
KPR#:1343
CTC portion of method printout does not include overlap settings
KPR#:1473
Revision B.04.02 SP1 [208]: Import eMethod on 6890 RC.NET driver does not work
KPR#:1522
ChemStation hangs up during Edit Entire Method when starting this process via the UI
KPR#:1593
Method printout for RID contains typos
KPR#:1602
Undefined symbol: PrDARep error displayed when selecting Export e-Method
KPR#:1624
Open GC edit parameters and close it without changes; method modification flag is set
KPR#:1643
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1650
OpenLAB CDS C.01.02 sequence table can't find currently loaded method
KPR#:1656
No information around handling for special characters
KPR#:1663
After a single run, switching from the Method & Run Control view to the Data analysis view of the ChemStation, the method loaded in Data Analysis in different than the one used to run the sample
KPR#:1689
C.01.01, modified method flag gets set intermittently in error when loading different methods
KPR#:1691
Method marked Dirty on initial "Specify Report ..." with DEF_GC.M
KPR#:1694
Switching between Injection source sequences does not load correct method
KPR#:1704
Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa
KPR#:1710
Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified
KPR#:1748
Cannot save offline method if configuration is different: 'Method is not consistent with the configuration'
KPR#:1750
"Chemmain has encountered a problem and needs to close" when selecting back/dual inj position for GC manual inj
KPR#:1765
C.01.03 upgraded to C.01.04: 'Edit Entire Method' not working
KPR#:1802
"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671

Keyword: Method , ACQ

"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671

Keyword: Method Scouting Wizard

Cluster information can be overwritten from the offline session
KPR#:1720

Keyword: Navigation Table

Wrong method displayed in navigation table after partial reprocessing
KPR#:1678
Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure
KPR#:1752
Date Time entries can be incorrect for aborted fraction collection runs
KPR#:1836

Keyword: Output

Printing fails if printer name exceeds 100 characters
KPR#:1612

Keyword: Peak Purity

Peak Purity Report Template is not available in Intelligent report
KPR#:1829

Keyword: Programming

AIC can experience communication loss if Application Experience service is running
KPR#:1617
The command "macro" does not list *.mac files
KPR#:1755
The command "macro " cannot be used to select *.mac files
KPR#:1761
Scheduler does not work on an AIC
KPR#:1831

Keyword: Pump Configuration

Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster
KPR#:27
Cannot Check “Pump and solvent selection valve installed” Box for Flex Cube if Once Clear it
KPR#:32

Keyword: Pump Method

PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)
KPR#:34

Keyword: Report Template Design

Wrong labeling of standard deviation and variance
KPR#:11
Re-calibration does no longer update the data analysis method change time stamp
KPR#:19
Last displayed digits might be cut off, in cases where the number is zero
KPR#:25
RSD calculation based on converted report item causes template corruption
KPR#:27
Error message when adding a chromatogram to a report template
KPR#:620
Internal grouping of tables and usage of calculation variables cannot be applied at the same time
KPR#:621
Summary Calculations cannot be applied to table columns containing calculation variables
KPR#:622
Adding a template description to a previously erased and saved description field causes a KeyNotFoundException
KPR#:623
Customized composite groups should have a unique name
KPR#:627
Filter definition does not always allow usage of OR operators
KPR#:662
Reporting time format does not offer an option of (dd/mm/yyyy and time)
KPR#:668

Keyword: Report Templates

Different background colour in .pdf files generated with PDF XChange printer of ChemStation C.01.01
KPR#:6
Data field CalibCurve_CorrCoefficient is not properly labeld
KPR#:14
Warning message concerning missing MSPGothic font
KPR#:18
Export of Short_ESTD reports to .DOC produces two extra pages
KPR#:22
Quantified compound markers are shown at absolute position in case of ISTD calibrations
KPR#:26
Early Maintenance Feedback (EMF) counters may not appear on report
KPR#:33
Customized chromatogram or axis lables may not display correclty
KPR#:658

Keyword: Reports

problem with sizing of the chromatogram in a Short Report
KPR#:72
Chromatogram Output in reports: meaning of "Portrait" and "Landscape"
KPR#:276
ChemStation does not report resolution as defined in European Pharmacopoeia
KPR#:1011
Sequence summary statistics with different methods
KPR#:1324
it is not possible to print report with chromatograms that have been overlaid from different runs
KPR#:1377
acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off
KPR#:1440
Batch mode does not allow usage of intelligent reports
KPR#:1572
Report Templates newly added to a Result Set do not appear in the list of available templates
KPR#:1579
Extended performance report (IR) reports "Infinity" for plates per meter
KPR#:1581
Mismatch in drawing of peak performance between Classic Report and Intelligent Report
KPR#:1582
System may hang when ".DIF" output is choosen as report type.
KPR#:1585
Calibration Peak Areas included in the Area sum
KPR#:1614
Intelligent Reports do not contain ECM specific information in Review mode
KPR#:1619
B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work
KPR#:1634
No Chromatogram displayed on Intelligent Report for GC
KPR#:1645
CSV, DIF, and EMF files are not generated for some Classic report styles
KPR#:1673
sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01
KPR#:1679
Resolution of chromatograms in Intelligent Reporting is low
KPR#:1690
ChemMain.exe can crash when a long file path is used
KPR#:1703
Chinese/Japanese characters cannot be searched in the Report Viewer
KPR#:1705
Tangent peak width not correctly evaluated.
KPR#:1722
Mass- Based Purification (Fraction Collection) report missing information & shows inaccurate Fraction Tick marks (RC.NET only)
KPR#:1733
Reports might show different calibration time stamps for the same method.
KPR#:1743
eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles
KPR#:1751
The structure of the reports to file has changed
KPR#:1754
CP3: Number format error in: %d when printing report to screen in Data Analysis
KPR#:1756
Error 41265 occurred at end of run; Number format error in: %d shown in logbook
KPR#:1757
Result data sent to OpenLAB ECM Intelligent Reporter may not be updated.
KPR#:1758
Preview of classic reports fails with a macro error. 'Number Format error' in Sequence log file.
KPR#:1775
Chromatogram in On-Screen Classic report is not displayed properly
KPR#:1776
'Number Format error' in Sequence log file after printing report to screen
KPR#:1778
Peaks are calculated by area only or height only in a report
KPR#:1779
Sequence summary report sorts the data by injection time
KPR#:1780
A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence
KPR#:1786
Macro error occurs (Parameter 2: Table not found) when printing a calibration table
KPR#:1791
C.01.03: XML results: The 'ESTD %' and 'ISTD %' result types in the Quant Calc node are incorrectly 'ESTD' and 'ISTD'
KPR#:1801
Acquisition method name is not displayed in an intelligent report
KPR#:1803
Missing Report Style Options for LCMS and LCMS Qualitative
KPR#:1820
Classic detailed report does not show calibration curves after encountering a compound with invalid calib curve
KPR#:1821
ChemStation freezes while generating report preview
KPR#:1832
Calibration data modification time stamp is missing on reports printed from review view
KPR#:1834
Calibration data modification time stamp is set-off by 12 hours
KPR#:1835
Empty Custom Fields are not printed in Intelligent Report
KPR#:1837
report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620
Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715

Keyword: Reports , Custom Fields

Empty Custom Fields are not printed in Intelligent Report
KPR#:1837

Keyword: Reports , Data Analysis

report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620

Keyword: Reports , MSD Control

Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681

Keyword: Reports , SEQ

Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715

Keyword: Run Control

No other Windows application can run, when online ChemStation is busy
KPR#:1597

Keyword: SEQ

PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715
Change of sample type during sequence acquisition is ignored
KPR#:778
Method changes not used after pause/edit/save/resume sequence using unique folder ON
KPR#:1247
Mismatch for sequence completed time stamp in sequence and instrument logbook
KPR#:1317
ChemStation crash when trying to open Partial Sequence with non-exisiting data path in the Preferences
KPR#:1443
The path to a data file has a limited length
KPR#:1514
Cannot edit Method in Offline session while it is running in an online sequence
KPR#:1548
Missing Vial during sequences causes CTC ChemStation to hang
KPR#:1571
Cannot run partial sequence acquisition with headspace samplers
KPR#:1603
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1636
Data system uses the currently loaded method to validate the vial range in the Sequence table
KPR#:1640
Sequence parameter subdirectory overwrites sample info parameters during run
KPR#:1648
Missing ‘Reprocessing Only’ in the Dropdown List for Unique Folder Creation OFF
KPR#:1687
Acq method for 7890A back injector is not displayed in data analysis view
KPR#:1696
Missing vial in Headspace sequence obtained one, 1 less data file than expected
KPR#:1699
C.01.03: Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table
KPR#:1709
Opening the sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler
KPR#:1713
Sample statistics are not done, when the first compond in the calibration table is not detected.
KPR#:1725
Partial sequence with methods saved to a location other than C:\ does not run with headspace and PAL Samplers
KPR#:1736
GC: sequence does not continue if there is a missing vial, uploads data to ECM and goes to next sequence
KPR#:1767
Sequence stops when pushing the GC STOP key during sequence
KPR#:1787
Custom field window not prompted for Back injector on first run of the sequence
KPR#:1788
Sequence Reprocessing: Not possible to stop a sequence that is reprocessing
KPR#:1798
Sequence Import Mapping is missing LIMS ID columns
KPR#:1805
Sequence parameters: Changes to Fraction start location ignored
KPR#:1822
After upgrade, Sequence modified flag is set even if no changes are made before reprocessing
KPR#:1824
'Number Format error' in Sequence log file after reprocessing
KPR#:1830
Lampall off switches off lamps and pumps
KPR#:1601
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763

Keyword: SEQ , 1100 Control , 1200 Control

Lampall off switches off lamps and pumps
KPR#:1601

Keyword: Sampler Configuration

Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17
Control of Tray illumination missing for G1329B & G4303A
KPR#:5
Parameter mismatch for custom-defined Well Plates
KPR#:6
AFC Cluster: Edit Wellplates Dialog does not display imported wellplates.
KPR#:29
Use Vial as Fraction collections
KPR#:36
Control of Tray illumination missing for new G1329A
KPR#:9

Keyword: Sampler Configuration , Sampler Status

Control of Tray illumination missing for new G1329A
KPR#:9

Keyword: Sampler Driver

Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation
KPR#:16
Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15

Keyword: Sampler Driver , Sampler Status

Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15

Keyword: Sampler Method

ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.
KPR#:31
Dua Loop Sampler: ACQ method shows wrong loop with overlap injection
KPR#:33

Keyword: Sampler Status

Control of Tray illumination missing for new G1329A
KPR#:9
Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15

Keyword: Spectra Analysis

Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616

Keyword: Uninstall

B.04.01 - Cannot remove (de-install) an instrument once configured
KPR#:1172

Keyword: User Interface

Error Fetching CP Variable observed sometimes when starting the chemstation
KPR#:731
OpenLAB Control Panel does not prevent connecting to an instrument already in use
KPR#:1574
Slow performances observed while loading datafiles when a networked printer is not accessible
KPR#:1577
Additional information about the [ChemStation:Sequence][Save Sequence] privilege
KPR#:1583
PAL Sampling diagram always shows 'Vial Status: Normal' for every vial
KPR#:1644
7820A, 7890A missing sampling diagram when only a rear injector is installed
KPR#:1647
GC instrument diagram appears in offline ChemStation data analysis
KPR#:1653
Distributed ChemStation: some dropdown lists are displayed only after a certain delay
KPR#:1666
Graphic artifacts/software crashes caused by graphic settings
KPR#:1674
7890A, 7820A missing sampling diagram when only a rear injector is installed
KPR#:1682
Change Access Level... is not available from menu bar
KPR#:1695
490 Micro GC is not a valid option for a 7890A GC
KPR#:1698
The logged-on user is not displayed in lock dialog
KPR#:1740
Finish Queue Sequence Help is hidden behind dialog
KPR#:1745
Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA
KPR#:1746
Run-time error ‘282’: No foreign application responded to a DDE initiate
KPR#:1764
Missing dialog items when using ChemStation AIC’s
KPR#:1769
Change user, does not update the Connected user field in OLCP when using CS AIC
KPR#:1773
Values under Overview Table are not aligned with the parameter heading
KPR#:1808
Fraction Collector - F5 and F* not working
KPR#:1816
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763

Keyword: User Interface , Data Analysis

Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688

Keyword: User Interface , SEQ , ACQ , ECM Integration/OpenLab Option

In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763

Keyword: Utilities

link to the base reference file IQTREF.XML on IQT Report not working
KPR#:1118
Cleanup on Shutdown and Data Cleanup not working with Data Store or ECM backend
KPR#:1770
FPD Checkout Macros no longer work since there is no "autoint.mac" with OL CDS ChemStation C.01.04
KPR#:1772
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616

Keyword: Utilities , Spectra Analysis , MSD Analysis

Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616


KPR#: 1    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

6890 Method printout does not include units for NPD Adj offset, Equil time

Problem:
6890 Method printout does not include units for NPD 'Adjust offset' or 'Equilibration time', which should be pA and min, respectively.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1    Product: Unionsquare     Unionsquare    A.01.01

Keyword: 

One-line Description:

Intelligent Reporting does not report sequence logbook of ChemStation legacy data

Problem:
With Intelligent Reporting, it is not possible to report the sequence log book for sequences which have been acquired with ChemStation revisions prior C.01.01.
Temporary Solution:
Users who need to report the sequence log book for old sequences from B- or A- ChemStation can use the classic reporting which is still available in ChemStation C.01.01.
Fix Information:

There is currently no fix planned.


KPR#: 1    Product: Prep LC     Varian Prep LC    A.01.01

Keyword: 

One-line Description:

Pump settings not set to initial value after run and/or after loading a Method

Problem:
During run a stop is requested to the Software.
Opening the User Interface where the system is shown the Initial pump parameters were not loaded.
It was still  A: 99.84%  B: 0.16% instead of A:100% and B:0%.
Loading the Default method still the gradient is wrong displayed in UI.
Temporary Solution:
n/a
Fix Information:

there's any fix for this issue.
The simptoms was very seldom and random.
Restarting the PC and the Modules will erase the wrong information displayed on the User Interface.


KPR#: 2    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

6850/6890 Oven Cryo off should disable Oven Cryo setpoints

Problem:
When the 6850 or 6890 Oven Cryo On is unchecked (turned off) the Cryo setpoints are not grayed out.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 2    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

LC-3D, CE, and MSD technique-specific columns missing

Problem:
The columns that are missing include:

LC-3D and MSD:  Target Masses (number, range, list) and AutoBalance (blank or ‘OFF’)

CE: Voltage (number), StopTime (number), and User1-10 (1-48 or blank)
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 3    Product: Easy SamplePrep     Easy SamplePrep (G7300)    2.01

Keyword: 

One-line Description:

6890: Add\Advanced Tower Selection Syring Volume does not update

Problem:
6890: Add\Advanced Tower Selection Syring Volume does not update until session is closed/reopened.

For any Add step, select Advanced. The volume determines which tower will be used if automatically select by volume is selected (default). The Syringe Volume of that injector is displayed read-only below it.
If you go to the Configuration and change the syringe size for any tower, then add a new Add step, save and return, this syringe Volume change is not detected or displayed. It is only refreshed by closing and reopening the online session of ChemStation.
Temporary Solution:
Close and reopen the session to update the syringe volume.
Fix Information:

n/a


KPR#: 3    Product: Unionsquare     Unionsquare    A.01.01

Keyword: 

One-line Description:

Reporting of multiple internal standard amounts

Problem:
When designing a new report template with Intelligent Reporting, it is inconvenient to access the data for the internal standard amounts. 
If the user drags and drops the report item “DAInternalStandardAmounts” on the report, text will be reported. This text contains a ';'-separated list of the ISTD values. If a name for a specific value is given, a name-values pair is reported like 'name1=value1; name2=value2'.
If only a number is given for the ISTD, a 'ISTD-" prefix is added to the specific number: 'ISTD-1=value1; ISTD-2=value2’.
So the field reports the name and the amount of each ISTD, separated by semicolon.
Temporary Solution:
To report the amount of a specific ISTD, an expression needs to be set for the text field.
Example:  
       =choose(1,split(First(Sample_DAInternalStandardAmount),";"))  
reports the amount of the first ISTD in the list.
Fix Information:

There is currently no design change planned


KPR#: 4    Product: Easy SamplePrep     Easy SamplePrep (G7300)    2.01

Keyword: 

One-line Description:

Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources

Problem:
No Injection Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources, even though the Easy SamplePrep program does not run in this case.
Temporary Solution:
Supply an excess of resources to allow the sequence to be submitted to the Queue.
Fix Information:

n/a


KPR#: 5    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Sequence location directory selection is not correct

Problem:
When creating an Easy Sequence template and select the sequence location funcion, the software displays the normal windows function for selecting a directory / sub-directlry. This is wrong. The software should show the selection of sequence locations as specified in the sequence preferences.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 5    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Error Message when certain methodsare loaded onto an offline configured 7890.

Problem:
If methods ISTD_EX.m or DEF_GC.M are loaded onto a configured 7890 offline session an error message (Exception in command "translateMethod$" occured).
These 2 methods are 6890 specific and should not be loaded onto a 7890.
Temporary Solution:
n/a
Fix Information:

Fixed in GC driver 4.01 [047].  

This driver was included in the following datasystem packages:
GC ChemStation B.04.03 SP1 [87] and OpenLab CDS C.01.02
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#: 6    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

ChemStation: Vials are validated against the currently loaded method

Problem:
In the Easy Sequence Setup, a method is selected and then the vial numbers are entered. These vial numbers are passed to the data system to be validated.  

The validation occurs against the currently loaded method, not the method used in the template that will actually use the vials.
Temporary Solution:
n/a
Fix Information:

Fixed in Easy Sequence 2.4[002].


KPR#: 6    Product: Easy SamplePrep     Easy SamplePrep (G7300)    2.01

Keyword: 

One-line Description:

Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked

Problem:
Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checked
Temporary Solution:
Use a different methods for blanks that has 'Use Sample Prep program' unchecked
Fix Information:

n/a


KPR#: 7    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Custom Fields are missing in Extended Parameters dialog

Problem:
Sample Custom Fields and Compound Custom Fields are missing for Extended Parameters.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 8    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

Missing Minimum Peak Width and Save Partial Signal in Signal Source

Problem:
Minimum Peak Width and Save Partial Signal not available in Signal Source
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10.


KPR#: 9    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Gas Sampling Valve (GSV) and Multi Position Valve readiness not reported in GC Status.

Problem:
If a GSV and/or MSV is configured on the GC but the hardware is not present, the GC status will report not ready but not display the reason for the not readiness.
Temporary Solution:
No current fix.
Fix Information:

n/a


KPR#: 9    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Bracketing with unique folders off

Problem:
When bracketing is used with Easy Sequence with Unique Folder Creation OFF, data file names can not be generated for the calibration runs. The potential exists to overwrite files when using Bracketing with Unique Folder Creation OFF with Easy Sequence.
Temporary Solution:
When using Bracketing with Easy Sequence, use Unique Folder Creation ON from Preferences.
Fix Information:

n/a


KPR#: 9    Product: Unionsquare     Unionsquare    A.01.01

Keyword: 

One-line Description:

PDF Summary Reports with 0 Byte size

Problem:
It may occur that PDF Summary Reports with 0 Byte file size are created for "complex" or long (around 1000pages) summary reports. This may also occur after rendering summary reports multiple times.
Temporary Solution:
If such report are expected, it is recommended to created the report using a newly started Offline session.
Fix Information:

n/a


KPR#: 10    Product: Easy SamplePrep     Easy SamplePrep (G7300)    2.10

Keyword: 

One-line Description:

6890 Easy SamplePrep Add step fails for both by use and by volume resources

Problem:
If there is an ‘Add’ step in Easy SamplePrep program with just the ‘Solvent Wash’; even though the Back injector is to be used, it may not use the Back injector.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 10    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Transfer of 6890 methods to 7890 ignores some setpoints.

Problem:
If a 6890 method is loaded onto a 7890 with identical hardware some of the common parameters are not loaded.
Typical parameters ignored are:
Front PTV:
Split ratio
Split flow
Total flow

Column 1 (Agilent 19091J-413:
Average velocity
Column 2 (same column but not part of common sample pathway):
Pressure
Nominal initial flow
Average velocity

Front detector (NPD):
Hydrogen flow state (6890 method was off, came in as on)
Makeup flow state (6890 method was off, came in as on)

BAck detector (FID):
Hydrogen flow state (6890 method was off, came in as on)
Temporary Solution:
No current fix
Fix Information:

n/a


KPR#: 11    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Aux Detector 1 and Valve Box temp not excluded in readiness

Problem:
If the Aux Detector 1 and Valve Box temp are checked in the ignore ready checkbox, they are still included in the readiness calculation.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[23].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 11    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.02

Keyword: 

One-line Description:

Resource layout on 7890A GC instrument dialog is not translated

Problem:
Resource layout on 7890A GC instrument dialog is not translated into Chinese and Japanese.
Temporary Solution:
n/a
Fix Information:

Fixed in revision A.02.03.


KPR#: 12    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.02

Keyword: 

One-line Description:

Easy SamplePrep Method printout is not translated

Problem:
Easy SamplePrep Method printout is not translated.
Temporary Solution:
n/a
Fix Information:

Fixed in Revision A.02.03.


KPR#: 12    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Manual/Valved/HS templates cannot be loaded if inj source does not match loaded method

Problem:
6890 with no ALS or tray. Configured to a HeadSpace.
Create a front valve method, a back manual method, and a HeadSpace (front) method.

Depending on the method loaded, load an .est and this fails with:
'---------------------------
Inconsistent Easy Sequence Setup
---------------------------
Sequence requires a back injector not present on the instrument.

---------------------------
OK   
---------------------------'

The message really does not advise the user of what should be done to correct this situation.
Temporary Solution:
This is current ChemStation behavior.  The template created in Easy Sequence must use the currently loaded method.
Fix Information:

n/a


KPR#: 12    Product: LC and CE Drivers     RC 1200.NET    N/A

Keyword: 

One-line Description:

LC driver starts run with ALS Chiller turned off

Problem:
Using the RC.NET Drivers to control the Autosampler Chiller the systems is injecting with the Autosampler Chiller turned off. This is a different behavior compared to the legacy Rapid Control drivers for EZChrom which forced an ON of the Autosampler Chiller.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 12    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Column Comp subtraction options are not available for AIB detector

Problem:
Method Edit screen for AIB does not offer the 'Subtract from Signal: Column Compensation Curves' options.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[24].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 13    Product: 78xx Driver Software     7890A    1.11

Keyword: 

One-line Description:

Invalid operating parameters loaded when transfering methods between GCs with different hardware.

Problem:
Method transfer (7890 to 7890) loads the values for positional detectors/inlets even when they do not match; e.g. FID to TCD.
Invalid flows will be entered into the method.
Temporary Solution:
No current fix.
Fix Information:

n/a


KPR#: 13    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

6890 Method Edit\Injector labels are inconsistent for heater/mixer

Problem:
6890 with a G4515A BCR Mixer/Heater installed. Select Method Edit\Injector\Tray screen.

The labels for these items are misleading for the 6890.  Although the three items, reader, mixer, heater are a single unit, they are not all related to barcode reading. They are there for use with the Sample Prep Program.
The labels should not include the word 'barcode' for this reason in the title (Barcode heating and mixing) or label for Enable Heater (Enable barcode heater). Enable Mixer is already correct.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 13    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.02

Keyword: 

One-line Description:

Resources replenished on 7890A GC instrument dialog not translated

Problem:
Resources replenished on 7890A GC instrument dialog not translated
Temporary Solution:
n/a
Fix Information:

Fixed in revision A.02.03.


KPR#: 14    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.03

Keyword: 

One-line Description:

Certain combination of setpoints cause setpoint error from tower

Problem:
The software allows the user to enter invalid parameters for the draw and dispense speeds for the syringe.

Using an "Add" command it generates a "Command to XXX Tower failed: XXX Tower Prep Failure Code: 142: Autosampler setpoint error..."

The error message is limited and it takes effort to find the "invalid" parameter.
Temporary Solution:
n/a
Fix Information:

Fixed in A.03.01.


KPR#: 14    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Easy Sequence Setup does not display dual sources on method load; must Clear All to activate

Problem:
Open online session to a 6890 with dual ALS (Plasticman 6890A/7693).
DEF_GC.M is configured by default to front injector.
Set injection source to Dual and save as a new method, now loaded.

Select Easy Sequence Setup. No injection sources are displayed, (presumably because of the injection source selection of the previously loaded method had one injector.)

Load the Dual method. Still no injection sources.
Temporary Solution:
The user must select 'Clear All' to refresh the screen in order to display the Front/Back injection source checkboxes after loading the dual injector method.
Fix Information:

n/a


KPR#: 15    Product: Easy SamplePrep     Easy SamplePrep (G7300)    2.01

Keyword: 

One-line Description:

Need ability to 'shake' with Mix command

Problem:
The Mix command needs some more detailed control. The mixer to is used to shake a vial (Mix 2500 rpm 1 sec, stop 1 sec , and change direction mix 2500 1 sec, repeat), this cannot be done with the current mix command.
Temporary Solution:
n/a
Fix Information:

Fixed in revision 2.3.


KPR#: 15    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

7693A Sampler error reports identical for all errors

Problem:
The displayed sampler error is the same for all sampler error scenarios. Error appears only in the Instrument Activity log. 

Error reads "6890 GC Error # 1 Injection Error: stop pressed or sampler error. (410)"
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10.


KPR#: 15    Product: 78xx Driver Software     7890A    1.10

Keyword: 

One-line Description:

Shutdown events are not displayed on the software Status screen

Problem:
Shutdown events, such as a thermal or pneumatic shutdown, displayed on the GC Keyboard are not reported to the software and therefore not displayed on the “Agilent GC Status” screen.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

This driver is included with the following datasystems:
GC ChemStationB.04.03 DSP1 [52] and OpenLAB C.01.02
GCMSD ChemStation E.02.02 
OpenLab CDS EzChrom A.04.02


KPR#: 16    Product: 68xx Driver Software     6890    5.03

Keyword: 

One-line Description:

Confusing error message reported in log when vial is missing.

Problem:
When the 6890 GC encounters a missing vial, the IALog displays Stop key pressed error #1. The error should read Missing Vial Error.
Temporary Solution:
None
Fix Information:

None


KPR#: 17    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Easy Sequence Setup does not validate HeadSpace vial ranges

Problem:
HeadSpace Method loaded in ChemStation.
The Injection Source is set to HeadSpace Sampler

Easy Sequence Setup
Sample Information
Starting Vial Location

Any number entered into this location will error with the following error message:
The vial position could not be validated against the current configuration
Temporary Solution:
n/a
Fix Information:

Fixed in Easy Sequence 2.4[012].


KPR#: 17    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Barcode mismatch error

Problem:
When a barcode mismatch occurs with overlapped injections (option of do not inject on barcode mismatch), the mismatched vial is taken to the turret and chemstation stalls with the sequence running and waiting.  An abort is required to recover from this error.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 18    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

Signal 2 data only cannot be saved.

Problem:
The signal 2 data cannot be saved independantly as signal 1 check box cannot be unchecked.
Temporary Solution:
If only one signal needs to be saved, you must assign it as signal 1.
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 18    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.02

Keyword: 

One-line Description:

Add step fails to execute due to mismatch in # injections/volume/syringe size/solvent allowed by firmware

Problem:
Method with a single add step of 1000 uL. Syringe size is 10 uL. Vials are moved to turret and then error message displays: method not running.

This is a mismatch in # injections/volume/syringe size/solvent allowed by firmware.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 19    Product: Easy SamplePrep     Easy SamplePrep (G7300)    A.02.03

Keyword: 

One-line Description:

A method using Easy Sample Prep cannot be used to create a method not using Easy Sample Prep

Problem:
Create a method which has the "Use Sample Prep Program" checkbox enabled. Use this method to create a method which does not have the "Use Sample Prep Program" checkbox enabled. Load the method which does not have the checkbox enabled. Error message displays: Configuration mismatch: Default parameters loaded for 'Agilent 7890'.
Temporary Solution:
n/a
Fix Information:

Fixed in 3.1[21]. This revision will only fix with Openlab Chemstation C.01.03.


KPR#: 19    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

ALS faults during a run - No Pop up Errors.

Problem:
If the ALS faults during a run, no pop up error messages appear on the Chemstation screen. A message will be entered in the instrument log book and on the GC display.
Temporary Solution:
None.
Fix Information:

n/a


KPR#: 19    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Parameters active when cryo is unchecked

Problem:
When using the multimode injection port, all parameters relating to cryo are available whether cryo on is checked or unchecked.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[23].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 20    Product: 68xx Driver Software     6890    5.10

Keyword: 

One-line Description:

Instrument Diagram does not match installed column in method

Problem:
The Instrument diagram of the GC instrumentation indicates a different column configuration than is set in the method.  The GC keyboard will reflect the new changes, only the software instrument diagram is affected.
Temporary Solution:
Beginning with B.04.03 DSP1 [54], after the configuration change is made to the columns, the workaround is to save the method and then load it.  This will update the displayed column configuration.  

Another workaround is to shut down the ChemStation software and restart to reflect the new configuration changes.  

These workarounds may not work for all configurations.

Classic 68xx driver works correctly.
Fix Information:

n/a


KPR#: 21    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

Meth Edit/Config and GC Config screens disagree with 7693A Tray enable on a 6890

Problem:
Even with the 7693A tray disabled on the GC Config display, the config display when using method editing shows the tray as enabled. 
Enabling or disabling the 7693 Tray does not evoke method resolution.
Temporary Solution:
n/a
Fix Information:

Fixed in Revision 6.02.


KPR#: 22    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

In 68XX driver revision 5.02 - 6.11, it is not possible to Modify a Method During a Run to Extend the Run Time or Change the Oven Temp

Problem:
It is possible to the software edit parameters screen and change the oven hold time or oven temp and select OK.  The GC does not accept the updates.  The method in the software is changed to the new values; the GC ignores all.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 23    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

Info in Edit instrument parameters\Columns is incomplete

Problem:
Edit instrument parameters; columns

Applies to GC’s: 7890A, 6890N, 6850A

The window displayed is incomplete; it shows the column type and flow/pressure parameters, but does not give the length, ID or film thickness.  These are critical parameters, and should be displayed just as they were in the previous ChemStation revisions.  To find these parameters, the user now has to look on the configuration window.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 23    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Incorrect pressure units when auto configuring a 7890

Problem:
The 7890 pressure units are set to kPa from the GC display. If an auto configuration is performed from EZChrom the pressure units are reset to psi.
This error occurs in EZChrom only.
Temporary Solution:
Change the pressure units from EZChrom and download the configuation to the GC.
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 24    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

HeadSpace Injection Volume is accepting negative numbers; no checking; no meaning

Problem:
Create an Easy Sequence Template for a G1888/6890. There is no error checking for Injection volume for this injection source in either the Setup or Sample List. All values are accepted, including negative numbers.
There is no place to setup an 'injection volume' for a HeadSpace. There is only a time and pressure.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 25    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Easy Seq and Setup Starting Vial Location and Number of Samples do not check for limits

Problem:
Starting Vial Location and Number of Samples do not do any range checking against any method or configuration.
Temporary Solution:
n/a
Fix Information:

Fixed in revision 2.3


KPR#: 25    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

Starting run from GC to do manual injection not possible

Problem:
GC is ready; Injection source = manual

It is not possible to walk up with a sample and manually inject into the GC.  The run can not be started from the GC Keyboard start button.
Temporary Solution:
The user must first start the run from the Chemstation menu RunControl > Run Method and then press start at the GC.
Fix Information:

Fixed in driver revision 6.10.


KPR#: 26    Product: Easy Sequence     Easy Sequence    A.01.01

Keyword: 

One-line Description:

Easy Seq Setup: Sample Name text entry should check for a Counter if used as Data File name

Problem:
Easy Sequence Setup: Entry of text in the Sample Name combo box does not check for and advise that a counter token is needed if the user also uses the token 'Sample Name' for the Data File Name.

The user will find out when sending to the queue 'Error: Sample Name xxxx occurs more than once.'
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 28    Product: LC and CE Drivers     RC 1200.NET    A.02.03

Keyword: 

One-line Description:

Lamps and cells of G1314F show incorrect dates in the RFID tags

Problem:
Lamps and cells of G1314F show incorrect dates in the RFID tags
Temporary Solution:
n/a
Fix Information:

Fixed with the driver version A.02.06


KPR#: 29    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Easy Sequence Setup not refreshing the Injection Location

Problem:
7890/7693: Create methods for front, back, and dual ALS injection sources.
Easy Seqeunce Setup does not refresh to show the correct injection location of the loaded method until user loads the corresponding method in ChemStation.

Even loading a different method, the previous injection source is kept.
Temporary Solution:
If the injection source has been changed, use the "Clear All" button and reload the method.
Fix Information:

n/a


KPR#: 31    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

No error checking for vial ranges in Easy Sequence Setup; fails at runtime

Problem:
There is no error checking within Easy Sequence Setup to determine if the vial range of the loaded method's injection source is accurate.

The method can be saved and submitted to the queue.  It fails at this time with:

'---------------------------
Error
---------------------------
Invalid location found in FRONT sequence, line 1
There may be others
---------------------------
OK   
---------------------------'

This defect occurs because there is a difference between the loaded method's injection source and the method that is selected in the Easy Sequence Template.
Temporary Solution:
Load the method in ChemStation from Method > Load Method menu that will be used to create the Easy Sequence Setup Template.
Fix Information:

n/a


KPR#: 31    Product: 68xx Driver Software     6890    5.11

Keyword: 

One-line Description:

Ignore missing vial option is not present in GC

Problem:
Run a sequence with 11 runs and the 11th vial is missing.  The sequence aborts and SSIZip file was not uploaded to ECM as the 11th vial was missing.

The ignore missing vial option is not present in GC.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 32    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Pause/Resume button is in the wrong state after a sequence is edited and then pauses

Problem:
Send a sequence to the Queue and while it is running, edit it. This modified/changed sequence pauses the Queue, but the first blue button is still in a 'Pause||' state. Click the button and change it to Resume, but it does not do anything as it is waiting for the current sequence to be saved before starting the next sequence.

In this case, because the sequence has been modified, the ChemStation is not accepting sequences. If you click pause || button, you pause the queue and it does not do any thing. The message on the active queue is red and the message says the sequence has been modified.
Temporary Solution:
Save the sequence that has been modified.  Once the sequence has been saved, the queue will resume running the next sequence.
Fix Information:

n/a


KPR#: 33    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

7890A with no ALS shows Read Barcode button in Easy Sequence Sample List

Problem:
Load an Easy Sequence.es for a valve or HeadSpace Injection Source and the Sample Name cell in the Sample List shows a Read barcode button.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 33    Product: 68xx Driver Software     6890    5.10

Keyword: 

One-line Description:

Instrument Setup graph has incorrect column legend

Problem:
In the 6890, the legend for the Instrument Setup graph says Column 2, irrespective of which column is highlighted.  The graph is correct, it is just the legend that is wrong.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 34    Product: 68xx Driver Software     6890    5.10

Keyword: 

One-line Description:

Instrument Setup's column tab says a PCM is an unknown inlet

Problem:
The 6890 has an split-splitless front inlet and PCM rear inlet.  The Instrument Setup’s Inlet tab shows that the rear inlet is a PCM and that it can be controlled.  However, the column tab shows no ability to control the PCM rear inlet and it is called an Unknown Inlet.  This is incorrect and inconsistent behavior.  PCM is a legitimate inlet whose pressure can be controlled.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 34    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

HeadSpace Simple Calibration Blank type should only allow 'Blank with injection'

Problem:
HeadSpace software will not run a sequence without a vial location.  Therefore, the Easy Sequence Setup\Calibration\Simple Calibration options should only allow Blanks with Injection and not allow blanks without injection.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 35    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Calibration Blank with no injection loads with Inj Vol as 0

Problem:
Create a setup with a blank with injection and a blank with no injection.
Save and load to Easy Sequence. The Injection volume for the Blank with no injection is entered as 0. Native sequence table also shows this as zero.
This is not consistent with the volume for the blank with injection
Temporary Solution:
n/a
Fix Information:

Fixed in revision 2.2


KPR#: 35    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: 

One-line Description:

AFC & UIB II: AFC does not fractioning as expected

Problem:
System configuration:
1 DAD
1 UIB II
2 AFCs -> 1 AFC Cluster
Analog out of DAD connected to Analog In of UIB II
Simulated signal of DAD was used

Method settings for modules (expect threshold for UIB II signal): see attached method

1. Created a data file with the simulated signal recorded by the UIB II
2. Loaded data file into Fraction Preview
3. Used Fraction Preview to collect peakbased fractions with threshold > 0.12 V
4. The value 0.12 V is written into the table on the left side without conversion (the table shows mV as unit)

Using this method will cause that the AFC will fractioning all the time.

Cause:
The firmware interface expects the threshold in mV but the driver writes 0.12 (V) to the AFC
This causes that the AFC is fractioning all the time.
Temporary Solution:
n/a
Fix Information:

Fixed in UIB II signal is now delivered as mV signal such like in the UIB I


KPR#: 37    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: 

One-line Description:

Recovery volume and fraction volume: behaviour for reset volume

Problem:
I worked with one AFC.
Configured 2 wps +10 vials and 15 vial tray for recovery.

Started multiple runs. 
For each run a new Recovery location was used.
Then removed the recovery tray and entered it again.

Next run started with first recovery location --> this was already used.
Temporary Solution:
If you want to reset the Funnel Trays only, use "Reset Funnel Tray Volume"

If you want to reset Wellplates and vials (all except funnels), use "Reset Fraction Volume"
Fix Information:

Issue fixed with the LC and CE driver revision A.02.07


KPR#: 37    Product: 78xx Driver Software     7890A    1.08

Keyword: 

One-line Description:

ALS Errors Abort Sequence; no Retry/Abort dialog box option offered

Problem:
The setting is selected by opening the Tray/Overlap tab in Method Editor.  

The Instrument Status must be opened to view the dialog box.

Pause is the default setting for ChemStation and MassHunter.
Skip is the default setting for EZChrom.
Temporary Solution:
n/a
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 38    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

ALS errors abort sequence; no Retry/Abort dialog box option offered with 68xx RC.NET Driver

Problem:
6890A RC.NET driver with ALS. 

If an ALS error occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort.  The sequence simply aborts.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 39    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Calibration table clears on method reload from Samples tab

Problem:
Create a Setup with Calibration. Return to Sample tab and load the same or a different method with the same injection location. Return to Calibration tab and all has been cleared. Reloading a method does not clear entries made on the Samples tab.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 40    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Queue displays running vials incorrectly when initial HeadSpace vial not found

Problem:
Start a 3 line headspace sequence from the queue, but remove the FIRST vial in the sequence. Queue does not update correctly, showing the first vial as running, though the HeadSpace has moved on to vial 2.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 41    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Aborted HeadSpace run displays Status as 'Complete' in Queue History

Problem:
HeadSpace aborts, but ChemStation does not. Select the Abort Run button from the Waiting for ready dialog and ChemStation aborts the sequence.
Queue History shows this run as 'Complete' instead of 'Aborted'.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 43    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Use Method Inj Vol in sequence table not available for 7890A

Problem:
The Use Inject the volume from the Method is not enabled.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 44    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Easy Sequence does not advise of invalid vial locations until sequence is fully loaded into Queue

Problem:
There is no advisory about the invalid vial locations until the sequence has been added to the queue.
Temporary Solution:
n/a
Fix Information:

Fixed in revision 2.3


KPR#: 45    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Sequence submitted to queue but will not run and can not be removed/deleted

Problem:
Sequence submitted to the queue but will not start and cannot be deleted.
Temporary Solution:
As a workaround, delete the following file:  SequenceQueue.que from the C:\Chem32\4\TEMP directory.
Fix Information:

Fixed in revision 2.4.


KPR#: 45    Product: 68xx Driver Software     6890    Future

Keyword: 

One-line Description:

68xx RC.NET: When look-ahead vial is not found in a sequence using overlap, the currently running sample and sequence will abort

Problem:
Create a sequence using a method specifying sample overlap. If a non-initial vial is not found, the currently running sample will abort, the sequence will abort, and there will be only an error displayed in the logbook and GC Status.

68xx RC.NET
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 47    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Injection Volume on ES Setup should show units (uL); valved inj are in mL and user must convert

Problem:
The Easy Sequence Setup dialog should show Injection volume units, which in ChemStation always seem to be uL. 

When the user uses EasySequence with a Valved injection, the volume, which for valves is mL, must be multiplied by 1000 to get a value that will be accepted.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 48    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Easy Sequence Setup should check for Sample Name length and not allow invalid entries

Problem:
Enter a long Sample Name in Easy Sequence Setup and add a counter.
Save and load to Easy Sequence. Fill Samples. Not until then is the user warned (with a red X) in the Sample list that there is a 40 character limit.
It is possible to save this but it will not be possible to add to the Queue.
Temporary Solution:
n/a
Fix Information:

Fixed in Easy Sequence 2.4[014].


KPR#: 48    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Chiller/Heater: Off appears in method printout even though no chiller tray is installed

Problem:
Chiller/Heater:  On/Off should only appear in the method printout if the there is a chiller tray installed.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10.


KPR#: 49    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Missing a Box in Lower Right Corner in Graphical Plot

Problem:
A small box should be displayed in lower right corner in graphical plot field.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 50    Product: Easy Sequence     Easy Sequence    2.2

Keyword: 

One-line Description:

Easy Sequence Setup: Change Inj/vial to a value >1; all labels and menu bar disappear

Problem:
Select Easy Sequence Setup.  Enter a number greater than 1 in the Inj/vial cell on the right side and then tab. All labels and menu items disappear.
Temporary Solution:
Restart the session.
Fix Information:

n/a


KPR#: 51    Product: Easy Sequence     Easy Sequence    2.2

Keyword: 

One-line Description:

Vial Status shows first run as completed for Easy Sequence following Run Method

Problem:
Run a single vial using Run Method.

Send sequence to queue vials 2-4. When the run method completes and the sequence loads and then the first vial in the sequence starts, Vial status shows the first line (vial 2) as immediately 'completed'.

The system stalls at Sequence running.  The GC is in prerun state and the system cannot start another run until it is shutdown and restarted.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 51    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

No Barcode Options for 7683B Injector

Problem:
There were no barcode options in the GC Parameters for the 68xx RC.NET driver with a 7683B ALS with a barcode reader.
Temporary Solution:
Turn the barcode on from the GC.
Fix Information:

Fixed in driver revision 6.10.


KPR#: 52    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Easy sequence actions are not included in the audit trail and no separate audit log exists

Problem:
Any changes done on Easy Sequence are not included in the audit trail; there is no separate audit log present for Easy Sequence.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 53    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

Vial not found reported in software Status not consistent with instrument status

Problem:
Run an Easy Sequence using a HeadSpace injection source. When a Vial not Found event happens, the final Status for this run, which did not start a run, is reported as 'Completed'. The status being reported by the HS device is 'Run Aborted'.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 54    Product: Easy Sequence     Easy Sequence    2.01

Keyword: 

One-line Description:

PAL autosampler plate/vial locations not accepted in Setup location box

Problem:
Easy Sequence 2.01.41 on Chemstation B.04.03 DSP1 will not accept plate designations of PAL injection source in the Setup location combo box. (P1A1)
The user should be able to choose the location designation.
Temporary Solution:
n/a
Fix Information:

Resolved in revision 2.4 [12].


KPR#: 54    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Error: Code #370 occurred in module #1 in event viewer log

Problem:
68xx RC.net

Quickly setup a method and started a run.  Error: Code #370 occurred in module #1 happened because there is no signal saved.  This is not obvious.
Temporary Solution:
n/a
Fix Information:

This issue no longer occurs in driver revision 6.10.


KPR#: 55    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Opening 68xx RC.NET GC types result in 'FetchCPVar cannot talk to application'

Problem:
Configure 6850 RC.NET or 6890N RC.NET GC. 

Open an online session, initial or any subsequent try. This results in a benign pop-up error:  'FetchCPVar cannot talk to application'.
Temporary Solution:
Go to view menu and turn off the sampling and instrument diagrams.  This will keep hpgcui00.exe from starting up.
Fix Information:

Fixed in driver revision 6.02.
Driver included in OpenLab C.01.01.


KPR#: 57    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

68 RC.NET: Vial is not taken directly from the barcode reader to the turret for injection

Problem:
Bar Code support was not available in B.04.02 for 68xx RC.NET – only heating and mixing function was supported.

Bar Code support was added in B.04.02 SP1 for 68xx RC.NET and is implemented as follows:

Vial -> BCR -> vial back to tray -> injector.   
Note:  the gripper remains over the vial until it is picked up to move to the turret 

Making an intermediate stop at the tray location added from 9 to 11 seconds with the newer circular tray, and 8.2 to 9.3 seconds with the rectangular tray.
Temporary Solution:
The 68xx classic driver works as follows:

Vial -> BCR -> injector
Fix Information:

Fixed in driver revision 6.10.


KPR#: 59    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

In B.04.02 [96] it is not possible to start the 6890 RC.NET from the front panel Start key

Problem:
In B.04.02 [96] it is not possible to start the 6890 RC.NET from the front panel Start key.  Software configuration is Injection source = manual.  6890 is READY, press start and get error message "ERROR Start Key disabled".
Temporary Solution:
One workaround is for the user to start the run from the "run control" then "run Method" in the Chemstation software and then press start at the GC. 

Second workaround is for the user to switch to the "Classic" driver.
Fix Information:

Fixed in driver revision 6.10.


KPR#: 60    Product: Easy Sequence     Easy Sequence    2.4

Keyword: 

One-line Description:

Print preview screen does not appear when blank is added to calibration program

Problem:
1. Method and Run control screen> easy sequence settings tab > calibration tab
2. Select "simple calibration" for calibration mode, add Blank to the sequence. click print button to print preview easy sequence settings. 

Print preview screen does not appear.
Temporary Solution:
n/a
Fix Information:

Fixed in Easy Sequence 2.4[30].


KPR#: 66    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

CIS4 inlet is not recognized with 6890 GC RC.NET

Problem:
CIS4 inlet is not recognized with 6890 GC RC.NET
Temporary Solution:
Currently, only the 68xx Classic driver supports the CIS4 inlet.
Fix Information:

n/a


KPR#: 67    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

ChemStation does not show "Hydrogen Safety Shutdown" in 7890A status

Problem:
When the 7890A goes into hydrogen safety shutdown the ChemStation shows "Front inlet flow shutdown" and other messages, but not  "Hydrogen Safety Shutdown".

The chemstation does not report the hydrogen shutdown but it does list the other causes. In addition, the heated zones can be turned on from the chemstation. Since a hydrogen shutdown should require operator intervention AT THE GC remote control from the (any) chemstation should be disabled.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The 4.01 driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 69    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Partial data collection is not available in 68xx RC.NET driver

Problem:
It is not possible to edit when to save the start and stop time for GC signals.  On the signals screen, need the ability to collect partial data for specified times.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10 [27].


KPR#: 70    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Adjust offset of the NPD bead is not available from the ChemStation software

Problem:
Adjust offset of the NPD bead is not available from the ChemStation software
Temporary Solution:
Adjust offset of the NPD bead must be done via the GC Keyboard.
Fix Information:

n/a


KPR#: 71    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

6890 NPD parameters are off but are shown as Not Ready

Problem:
All setpoints are off for the NPD, but the Instrument Actuals continue to show GC Not Ready and lists all the temps, utility gas flow, and H2 flows for the  detector as not ready and temp 'waiting'.

This should not be seen as Not Ready.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 72    Product: 68xx Driver Software     6890    6.02

Keyword: 

One-line Description:

6890A Easy Sample Prep Add step does not read syringe volumes correctly: always 10 uL

Problem:
Select to add an Easy Sample Prep program, select an Add step. Enter any injection volume and select the Advanced>> button to open the dialog.
Syringe size is always noted as 10 uL for both front and back syringes. It is not possible to refresh this from the checkbox tower selection. Configuration and method edit show this correctly. 

This did NOT happen for a 6890N, 6890 Plus, or 7890A.
Temporary Solution:
If you change the syringe size, to get the new size reflected in Easy SamplePrep program editor, restart the instrument session.
Fix Information:

Fixed in driver revision 6.10.


KPR#: 73    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

68xx: Possible to make changes at the GC Keyboard when the Keyboard lock is On

Problem:
With the Keyboard lock set to on, it is possible to make changes to the GC set points at the GC keyboard, even during a run.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx Driver version 6.02 [11].


KPR#: 76    Product: 78xx Driver Software     7890A    2.13

Keyword: 

One-line Description:

7820 method printout includes parameters and setpoints unavailable for PP inlet

Problem:
Purged Packed Inlet includes parameters:
'Pressure  On  20.261 psi'
'Total Flow, On, 5.4077 mL/min'
'Septum Purge Flow On -0.0089673 mL/min'

Pressure, Total Flow, and Septum Purge Flow are parameters that are  not available for this GC.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

This driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 78    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

System Not Ready: HS continues when set, but GC never starts if not ready at run start

Problem:
Start a 3 line sequence using a method where the Sequence Actions, System Not Ready @0.00 min is set to Continue.

As soon as the first vial starts to process, the GC goes to a Not Ready state. First sample starts the GC, as expected, all subsequent runs are continued, but the GC never starts for the rest of the sequence. 

Seq logbook shows 'Data not available'.

The 68xx will not start a run while being controlled by Chemstation, but will start a run if it is running stand alone.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10.


KPR#: 79    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Sequence stops when pushing the GC STOP key during sequence

Problem:
n/a
Temporary Solution:
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence.  

This occurs on ChemStation ver.
- B.04.01 SP1
- B.04.02
- B.04.02 SP1
- B.04.03
Fix Information:

n/a


KPR#: 80    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

If 7890 gets into an error status (flame off) this is NOT reported within the 7890 status window

Problem:
If the 7890 gets into an error status (such as flame off) this is NOT reported within the 7890 status window 

The flame failed to ignite at the FID but the status window ONLY 
displays that the GC is Not Ready.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[009].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 81    Product: 68xx Driver Software     6890    6.02

Keyword: 

One-line Description:

6890 does not have the NPD adjust offset option

Problem:
The NPD Offset adjust to adjust the target current is not available.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 82    Product: 78xx Driver Software     7820A    2.13

Keyword: 

One-line Description:

When Using a 7820A GC, the Supported Number of Columns Does Not Match in the Hardware and Software

Problem:
On the column configuration dialog (Instrument -> Agilent 7820A Configuration -> Configuration tab -> column tab), 6 columns can be configured, but the 7820 hardware allows only 4 columns.
Temporary Solution:
Configure only 4 columns since that is what is allowed.
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 83    Product: 78xx Driver Software     7820A    2.13

Keyword: 

One-line Description:

7820: Two "Sampling Diagram" shown in Method & Run Control view

Problem:
Run online 7820 GC with 7683/7693 ALS.  Select View -->Sampling Diagram, Sampling Diagram is displayed.
Click [switch screen layout], "Agilent 7820ATray" diagram is displayed.
There are two "Sampling Diagram" in Method & Run Control view.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 83    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Only Fast Plunger Speed is available for PTV Solvent Vent multiple injections

Problem:
The fast plunger speed is only available when using multiple PTV Solvent Vent injections.  Most users of multiple PTV Solvent Vent injections need to use Slow or Variable plunger speed, not Fast.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 6.10.


KPR#: 85    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

6890 RC.NET injects before it is ready if it has been sitting idle between runs

Problem:
The 6890 RC.NET injects before it is ready only if the instrument has been sitting idle in between runs.  For example, the GC becoming Not Ready because a new method was downloaded with new thermal setpoints during the sequence.  (The sequence didn’t use the same method in every line.) 

This defect occurs while in a sequence.   

This defect does not occur if running isothermal or for systems that run the next sample soon after the last run was completed.

This defect only occurs in 68xx RC.NET Driver systems only.
Temporary Solution:
n/a
Fix Information:

Fixed in GC driver revision 6.10 [027].


KPR#: 86    Product: 78xx Driver Software     7820A    2.20

Keyword: 

One-line Description:

7820 - GC Configuration - Inlets labels have a 'Z' after type

Problem:
In the 7820 GC Configuration,
Any inlets are labelled the following way:

SSZ Inlet
PPZ Inlet
Temporary Solution:
n/a
Fix Information:

Fixed in 78XX driver revision 4.02.

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 87    Product: 68xx Driver Software     6890    6.02

Keyword: 

One-line Description:

IQ1.CFG file of 6890/6850 instrument corrupted after regional setting change

Problem:
From Control panel\Regional settings: change the decimal separator from a period to a comma.

It is not possible to open a 6890/6850 ChemStation.
Temporary Solution:
1.  Backup data, methods, sequences, column databases (found in C:\Chem32\DRIVERS), etc 

Record GC IP addresses

2.  Delete IQ1.CFG located under the Chem32\1 directory for example

3.  Uninstall Chemstation

4.  Reinstall Chemstation
Fix Information:

Period is required in Regional Settings.


KPR#: 90    Product: 68xx Driver Software     6850    5.02

Keyword: 

One-line Description:

B.04.03 error occurs (FetchCPVar cannot talk to application)

Problem:
Start up Chemstation, but no instrument or sampling diagrams are shown.  
FetchCPVar Cannot Talk to Application Error occurs.

Systems use 7683 ALS.
Temporary Solution:
n/a
Fix Information:

Fixed in revision B.04.03 DSP1.


KPR#: 90    Product: 78xx Driver Software     7890A    1.05

Keyword: 

One-line Description:

Partial data collection is not available on the signals screen

Problem:
On the signals screen, need the ability to collect partial data for specified times.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01 [31].
This driver version is included with the following data systems:
GC ChemStation B.04.03 DSP1 [54]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#: 91    Product: 78xx Driver Software     7890A    1.08

Keyword: 

One-line Description:

Need ability to do Digital switching for detector signals in Runtime events

Problem:
The ability to do digital switching of detector signals in the Runtime events table is needed.

It is only possible to do analog switching.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

Driver version 4.01 is included with the following data systems:
GC ChemStation B.04.03 DSP1 [52]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#: 92    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

Instrument Diagram does not match column 2's inlet configuration

Problem:
Column 2 inlet configuration will never be correct if column 2 is connected to anything other than inlet 2.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.10[044].


KPR#: 93    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Sample Prep 'Return All' command not working

Problem:
The 'return all' command of Sample Prep was not performed.
Temporary Solution:
Workaround is to use the "Move" command to return vials instead of the "Return all" command.
Fix Information:

This is fixed in software driver 6.10 [028].


KPR#: 94    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

68xx Method download times out

Problem:
Connect to 6890N GC.

Within seconds of successfully opening an online session, logbook logs the following simultaneously with the loading of the method.

'Logbook      Welcome to ChemStation                         12:17:12 06/28/10
   0 1da6 4c28cb1f    0
6890       1 Timed out waiting for method download to fin>  12:17:35 06/28/10
   0 1da6 4c28cb1f    0
6890       1 ish.                                           12:17:35 06/28/10
7da 41e0 4c28cb1f    0
Method       Loading Method LCV_GCCONTROL.M                 12:17:35 06/28/10'
Temporary Solution:
n/a
Fix Information:

Fixed in GC Driver revision 6.10 [027].


KPR#: 95    Product: 68xx Driver Software     6890    5.02

Keyword: 

One-line Description:

"Method modified after loading" occurs when running a sequence (6890 RC.NET)

Problem:
The "Method modified after loading" flag occurs after lauching the sequence.

Started the sequence from the ChemStation software.  The method modification status indicates that the method has changed when it was not changed and the report will indicate 'modified after loading'.

The GC had a calibrated capillary column that was changed to a packed column.  When using this method that now has a packed column, the GC reports an error in the length parameter, and that is causing the method download not to reset the 'modified after loading' flag.
Temporary Solution:
There is a workaround: bring Method into MethodEdit UI, and "reconfigure" column (Packed to Packed would be OK) to get new structure.
Fix Information:

Fixed in GC Driver revision 6.10 [027].

Note, this fix will only apply to new methods, not to current methods.


KPR#: 96    Product: 68xx Driver Software     6890    6.01

Keyword: 

One-line Description:

6890 - Set Point Out of Range Error when connected to a new 7693 ALS

Problem:
7693A is mounted on a 6890 PLUS using B.04.03 chemstation software, the driver being used is the RC.NET. Every time a sequence is started it aborts with the error "set point out of range".  This usually means the 6890 is configured as "classic" with a 7693 (not supported) but in this case the driver is RC.NET. 6890 firmware is correct (A.03.08) and all of the ALS firmware is the latest revision.
Temporary Solution:
Build a new method with no pre or post washes or sample washes or pumps, and only do a 1ul injection. The ALS ran an injection with no errors.
Methods without post washes run correctly, no errors.
Fix Information:

n/a


KPR#: 96    Product: 78xx Driver Software     7820A    2.12

Keyword: 

One-line Description:

7820A configuration shows 8 valve positions when there are only 2 valve positions

Problem:
The 7820A configuration displays 8 valve positions. The 7820A GC only supports 2 valve positions.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 97    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Runlog appends from previous run on non-datasystem-initiated start

Problem:
When a run is started by a user press of the GC's Start button or activation of the remote start input without being prepared by the data system by a call to PerformInjection(), PerformPretreatInjection(), or StartRun(), the run log contents reported at the end of the run contain the contents reported on the previous run followed by any events in the run log for the new run.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[003].


KPR#: 98    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

GC Column Catalog File cannot be opened when using the 68xx RC.NET driver

Problem:
It is not possible to add new columns to the GC column catalog (database) using the RC.NET driver.  It is possible to add columns to the same column database using the Classic driver.  User switched back to classic on same system and could add columns without error.

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb

The RC.NET driver is prevented from opening the db because you can’t have two column entries with the same model number!
Temporary Solution:
Start with a new, “stock” db column catalog file from another system.
Fix Information:

The column model number should be unique. The newer RC.NET drivers enforce this behavior.


KPR#: 98    Product: 78xx Driver Software     7890A    2.13

Keyword: 

One-line Description:

Park during sequence with Overlap causes unresponsiveness

Problem:
Running ChemStation B.04.02.96. Start a sequence with overlapped injection set to immediately after injection. While the sample is being delivered to the turret, the Park button is pressed on the 7693A Tray. After delivering the vial, the tray is put in the Park position and the Injector completes the injection. At this point the ChemStation is unresponsive. (For example, when the ChemStation is in this state, any report printed to the screen can not be closed.)  

The ChemStation responds after the Park button is pushed and vial is returned.
Temporary Solution:
The ChemStation responds after the Park button is pushed and vial is returned.
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 99    Product: 68xx Driver Software     6890    1.01

Keyword: 

One-line Description:

Sequence continues while GC temperature is not ready

Problem:
When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run.  This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period.

This could result in a (retention time) RT shift.
Temporary Solution:
n/a
Fix Information:

Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1 [54].

Fixed beginning in: B.04.03 SP1 [87] and OpenLAB C.01.03 [37].


KPR#: 100    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

A method can be created which won't download to the instrument, sequence does not run

Problem:
Method download fails, sequence aborts after start; no injection made

Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.
Temporary Solution:
n/a
Fix Information:

This fix is in the 68xx driver version 6.10.032 and 6.11.001.


KPR#: 101    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

GC Back Injector not present in edit Method screen

Problem:
After successfully running a sequence with a back method, another back method is loaded. The back injector is not present in the edit method screen.
Temporary Solution:
Change a parameter at the GC keyboard and click "upload from instrument" in the edit method screen. The back injector will show in the edit method screen.
Fix Information:

n/a


KPR#: 102    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Validation failed shown on left corner of Resolve Method screen

Problem:
Load a method which was developed for both detectors and has the front detector signal chosen under signals tab.
Load this method on a GC with no front detector - method resolution. 

Changes made to the detector signal tab are not detected, tab remains yellow (invalid).
Temporary Solution:
Click on a different tab and modify the value of any field to a valid value or click on the configuration tab and then click away. Change the signals tab to a valid value or turn it off to resolve the method.
Fix Information:

n/a


KPR#: 103    Product: 78xx Driver Software     7890A    2.13

Keyword: 

One-line Description:

RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods

Problem:
RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods.

Only Column 1 is updated
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.02[034].

This driver is included with the following datasystems:
GC OpenLab CDS C.01.01
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.04


KPR#: 104    Product: 78xx Driver Software     7820A    3.01

Keyword: 

One-line Description:

7820A does not support RTL but the Retention Time Locking (RTL) menus are shown

Problem:
RTL is not supported on a 7820A GC System.  However, the RTL menus are shown in the 7820A software user interface menus in GC driver versions prior to 4.01.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 106    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Opening the GC method parameters to only look (no change made) and selecting OK to close stops the sequence or single run

Problem:
During a run, the GC method parameters are opened via the instrument diagram icon.  Select OK after looking at the oven parameter to close the screen.  The sequence or single run will just stop and the method is dirty.  See Log File for error message.

Log File:
  67 41e2 4e69f8b3    0
Method       Method started                                 07:29:55 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 1 (front)       07:29:56 09/09/11
  67 41e2 4e69f8b4    0
Method       Instrument running sample Vial 2 (back)        07:29:56 09/09/11
   0 1da5 4e69f91a    0
6890       1 Attempt to download method to GC failed.       07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Instrument Error - Method/Sequence stopped     07:31:38 09/09/11
  67 41e2 4e69f91a    0
Method       Method aborted                                 07:31:38 09/09/11
Temporary Solution:
Click the Cancel button rather than the OK button if the intent was to check the values of some of the instrument setpoints.  This would not interfere with any run which is in progress.
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 107    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

New Abort Behavior

Problem:
Once it detected a vial not found, everything would abort, including samples that were in process (that have already been injected and are being analyzed).  

The new behavior, is that the vial that is currently being processed will be finished.  Everything else will abort, if the abort option is selected.
Temporary Solution:
n/a
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 111    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Sequence terminates if user edits parameter during sequence.

Problem:
Sequence terminates with an error if user attempts to edit a parameter such as inlet temp while sequence is running. Logbook reports: Instrument Error - Method/Sequence stopped.
Temporary Solution:
n/a
Fix Information:

Fixed in 68XX driver revision 6.11.


KPR#: 112    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Column compensation cannot be started from the Data System with 68xx RC.NET

Problem:
Column compensation (comp) cannot be started from the Data System with 68xx RC.NET
Temporary Solution:
Column compenstation must be started via the GC keyboard.
Fix Information:

n/a


KPR#: 112    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Septum purge is not supported on the 7820A GC

Problem:
The 7820A does not support the septum purge and it should be inactive.  In driver versions before 4.01 septum purge was active and should not have been.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 113    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Cannot run a packed GC column with a 6890

Problem:
Method download fails, sequence aborts after start when trying to run a 6890 with a packed column.  
  
Error reported in logbook:
Attempt to download method to GC failed.

The Data System will not allow the run to continue when the method download has failed.
Temporary Solution:
This can be run by setting the column as a capillary column but that changes the chromatography.
Fix Information:

Fixed in driver revision 6.10[032] and 6.11[001].


KPR#: 115    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

7890A: Shutdown events are not published to the software GC Status screen

Problem:
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen.  

Only the first event, which is 'not ready', is reported to the Agilent GC Status screen.  The second event, which is the shutdown, is not displayed.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 116    Product: 78xx Driver Software     7890A    2.13

Keyword: 

One-line Description:

Actuals not shown for Bar Code Reader (BCR) heater or tray temperature

Problem:
Within the chemstation software, there is no display for the BCR heater actual or tray temperature actual.  The setpoint is displayed and the instrument can be configured for that temperature to affect readiness.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 116    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Pulsed Split/Splitless injection does not wait for ready 6890.

Problem:
The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample at the same time, then starts data acquisition and increase of inlet pressure to pulsed pressure.
Temporary Solution:
If the 6890 is operated from the GC keyboard, the GC has the expected result.
Fix Information:

Fixed in 68xx driver revision 6.11[007].


KPR#: 117    Product: 78xx Driver Software     7890A    2.10

Keyword: 

One-line Description:

Upgrading B.01.03 to B.04.02 GC ChemStation: Addition of new columns to the column catalog (database) is not possible; error occurs

Problem:
Upon upgrading from B.01.03 to B.04.02 it is no longer possible to add new columns to the GC column catalog (database).

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb
Temporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this.

The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP, HP-5 in the database.  Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) 

Temporary Fix:

Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example.  Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.
Fix Information:

This is fixed in GC Driver version 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 119    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Vial left in gripper; no message shown to alert user about the problem

Problem:
After a stop, if there is a vial in the gripper it is not returned to its original location, it is kept in the gripper until the start of a new run. In the interim, if a new vial is placed in the open location, during the next run the system runs into an error because the gripper wants to return the vial to its original location which is now occupied.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 121    Product: 68xx Driver Software     6890    5.01

Keyword: 

One-line Description:

Open GC edit parameters and close it without changes; method modification flag is set

Problem:
Saved a method.  Open the GC edit parameters to review a setpoint and  made no changes or changed the view.  Selected OK and the method modification flag indicates a change was made.  No changes were made.

This does not occur with the Classic Driver.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[007].


KPR#: 122    Product: 68xx Driver Software     6890    6.10

Keyword: 

One-line Description:

Method resolution changed oven temperature program

Problem:
Load a method from a 6890 GC with a fast oven on a 6890 GC with different hardware and a normal oven. Method resolution occurs. The oven temperature program from the fast oven GC has not been loaded on the normal oven GC.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[002].


KPR#: 123    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

Column info is missing with 6890

Problem:
The column specified as being used is missing information: description, Manufacturer and Model Number.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[005].


KPR#: 124    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

Method Modification Flag is Set after Edit Method: Occurs on first connection or after upload from GC.

Problem:
1. First Connection to an online GC and save the method OR connect to an instrument and save the method, open method editor, click on ‘Upload from instrument’. Click ‘Ok’.
3. Click ‘Edit entire Method’, go on clicking ‘Ok’ without making any change.
4. Method is dirty at the end.

This only occurs on 6850 GCs with G4513A injectors.
Temporary Solution:
Save the method as a new name after initial connection or subsequent method upload.
Fix Information:

Fixed in 68xx driver revision 6.11[008].


KPR#: 124    Product: 78xx Driver Software     7890A    0.0.1.5

Keyword: 

One-line Description:

Freeze Signal Value and Resume Signal Value missing from runtime events

Problem:
Freeze Signal Value and Resume Signal Value missing from runtime events
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 126    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

The Override Column Max Temp Checkbox Behavior

Problem:
The override column max temp checkbox in the oven parameter screen is automaticaly checked by default in versions of the driver prior to 5.01.  In driver version 5.01, it will not be checked and the box will by highlighted in yellow if the column maximum temperature exceeds the value that is recommended for the column that is installed.  

To make the box not yellow, either check the "override column max temp" checkbox, or enter a value that is lower than the recommended maximum temperature for the installed column.
Temporary Solution:
n/a
Fix Information:

Change of Behavior in Driver Version 5.01

This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 127    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Cryo Trap Run Time Event Support

Problem:
When a thermal aux zone is configured as a Cryo Trap, the cooling of the trap is accomplished via a run time event entitled Cryo trap cooling with an ON an OFF state.
Temporary Solution:
n/a
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 129    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

Instrument actual value for COC inlet does not display correctly

Problem:
6890N GC with COC inlet connected to FID. The COC inlet pressure setpoint reflects the correct pressure set for the column, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.
Temporary Solution:
Enable setup actuals (inlet) in the setup method screen, the inlet pressure is properly displayed.
Fix Information:

Fixed in 68xx driver revision 6.11[018].


KPR#: 130    Product: 68xx Driver Software     6890    Monitoring

Keyword: 

One-line Description:

Instrument actual value for anode purge flow on an ECD does not display correctly

Problem:
The ECD anode purge flow setpoint reflects the correct value, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.
Temporary Solution:
Enable setup actuals in the setup method screen, the actual anode purge flow displayed properly.
Fix Information:

Fixed in 68xx driver revision 6.11[018].


KPR#: 131    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

Sampling Diagram shows tray with none present (Injection source is GSV).

Problem:
6890 configured with the classic driver. Selected valve as the injection source. Sampling diagram shows a tray when none is present. 

When injection source is set for Injector, diagram shows the front and back 7683 injectors, each with a three vial turret.
Temporary Solution:
n/a
Fix Information:

This functionality is available in the .NET driver, improvements to the classic driver will not be addressed.


KPR#: 133    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

GC Keyboard Does Not Stay Unlocked

Problem:
Edit GC Configuration, unchecked the Keyboard Lock checkbox to unlock the GC Keypad.  Save the method in the software, and then make a change on the GC. The change will upload to the software. Any subsequent changes from the keyboard are not allowed. Open Configuration screen: Keyboard Lock checkbox is rechecked. Only one change can be made from the GC before the keyboard locks. 

This occurs on Windows 7 and Windows Vista PCs.
Temporary Solution:
n/a
Fix Information:

Fixed in 68xx driver revision 6.11[10].


KPR#: 134    Product: 68xx Driver Software     6890    6.11

Keyword: 

One-line Description:

Easy SamplePrep reset of one or any group of resources creates log entry: 'User reset all resources'

Problem:
Create a resource layout. Select menu item Reset Resources... From the list, select one or several resources and OK. The logbook always logs 'User reset all resources' no matter how many resources are actually reset.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 134    Product: 78xx Driver Software     7890A    3.02

Keyword: 

One-line Description:

Running a sequence requires the arm from the tray, G4514A to be in the UN-PARKED position

Problem:
The tray "Park" status overrides the start of a sequence.
Temporary Solution:
n/a
Fix Information:

The G4514A arm from the tray must be UN-PARKED in order for a sequence to proceed. User must press Park button to UN-PARK arm.


KPR#: 135    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Control mode for column pressure/flow is turned off, the inlet pressure is turned off

Problem:
It is not clear that the check box for the Control mode for column pressure/flow is linked to the inlet pressure.  When the control mode check box in turned off in the column screen, the inlet pressure/flow is also turned off.  This is not obvious to most users that this happens.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 137    Product: 78xx Driver Software     7890A    2.20

Keyword: 

One-line Description:

Extended Parameters does not disable the BCR when no hardware is present

Problem:
Extended Parameters for both the native sequence Parameters and for Easy Sequence Setup does not disable the Bar Code Reader section when no such hardware exists.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 139    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Switched and standard mode for septum purge not available on MMI

Problem:
For the split/splitless inlet, it is possible to run the septum purge in either standard mode or "switched" mode. In the switched mode, the septum purge is turned off during the injection cycle so that the only flow out of the inlet is through the column. The MMI would benefit from having this mode as well, particularly for splitless and solvent vent modes. In both modes on the MMI, the septum purge does not remain off until the inlet switches back to the purge state.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 140    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Negative Split Splitless inlet Pressure values during Post Run Backflush

Problem:
When using a 7890A GC with Split Splitless inlet to perform backflush the Inlet Setpoint and Actual Pressure assume negative values on the GC and Chemstation display during backflush. This occurs with a splitter (>2 columns) or with a purged union (Primary column plus restrictor). The columns must be running in "Flow" mode. 

A similar symptom is the inlet pressure gets set to "0.00" PSI during backflush rather than the wizard recommended value of "1.0" PSI.
Temporary Solution:
The workaround is to set the columns to "pressure" mode then back to flow and re-save the method.
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 145    Product: 78xx Driver Software     7890A    2.12

Keyword: 

One-line Description:

Backflush parameters are not entered after selecting apply on method edit

Problem:
It is difficult to set up backflush parameters by using the backflush method editor.  After applying the settings, these are not correctly filled in the CFT and oven fields. After repeating the procedure (sometimes more than twice) it will be correct.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 3.01[38]

This driver is included in the following datasystem packages:
GC ChemStation B.04.02 SP1 [208]
Ezchrom Elite 332 SP2 
GCMSD ChemStation E.02.02


KPR#: 148    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Split Splitless Inlet Temperature Increases when run starts

Problem:
Started a run and observed that the inlet's temperature increased.

Sometimes the SS inlet’s temperature is increasing at the start of a run. Looking at the xml contents of the methods, there was an initial hold time of 0.05 min and a ramp associated with that inlet, even though this ramp did not show on either the GC’s front panel or in the chemstation’s method editor.

The behavior observed was due to using a method originally developed for an MMI on a SS Inlet in that same location. Especially given that an initial hold time of 0.05 min, a typical time at which MMI operation would be to begin ramping its temperature.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 152    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Method "Runtime Events" allow control of a valve which is not configured

Problem:
7890A Runtime Events allow entries controlling valves which are not configured.

The 7890A display doesn't show information about the Valve status (upon pressing Valve#) since the 7890A is not aware of the valve presence.
Temporary Solution:
Not Applicable
Fix Information:

Not Applicable


KPR#: 154    Product: 78xx Driver Software     7890A    1.02

Keyword: 

One-line Description:

Column compensation menu is available during the sequence

Problem:
The Column compensation menu is available during the sequence and if selected will stop the sequence.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 166    Product: 78xx Driver Software     7890A    1.08

Keyword: 

One-line Description:

Column: need to include Ramped and Constant pressure and flow modes as selection

Problem:
User interface needs Constant Flow, Constant Pressure, Ramped Pressure, and Constant Flow for the Control Mode selections.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

This driver was included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#: 167    Product: 78xx Driver Software     7890A    1.10

Keyword: 

One-line Description:

GC goes offline momentarily because of SOAP Exception

Problem:
Setup 7693 injector and Tray. Put Tray in Park and then tried to start a run. The ChemStation showed that the GC went offline momentariily. The Event Log showed the following events :

Disconnecting because a System.Web.Services.Protocols.SoapException was caught when sending a command; 29: GC refused to do an injection
The connection state of the GC at 130.30.245.60 is now Offline
Attempting to connect to 130.30.245.60
In DataChannel[1].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[2].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000
The connection state of the GC at IP address is now Online
Connected to 7890 at 130.30.245.60, using license key 

Cannot start runs until GC and injector are idle.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 170    Product: 78xx Driver Software     7890A    1.02

Keyword: 

One-line Description:

No indication of inlet pressure shutdown in software status with 7890A

Problem:
Inlet pressure shutdown advisories indicating a problem with the instrument not shown in software status user interface.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

This driver is included with the following datasystems:
GC ChemStation B.04.03 DSP1 [52] and OpenLab CDS C.01.02
GCMSD ChemStation E.02.02
OpenLab CDS EzChrom A.04.02


KPR#: 174    Product: 78xx Driver Software     7890A    1.09

Keyword: 

One-line Description:

GC Keypad Lock Behavior

Problem:
In configuration we  now offer the option to configure how the keypad lock will behave.
Temporary Solution:
n/a
Fix Information:

The driver with this new behavior is included in:

A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 182    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

GC Instrument Status for Detectors

Problem:
In previous versions of software, only the temperature of the detectors was shown in instrument status.  In driver version 5.01, we show all GC detector status in instrument status.
Temporary Solution:
n/a
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 185    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

On a 7890A with a circular tray, the 7693A Tray is shown

Problem:
7890A has the round 100 position tray but the 7693A tray is shown instead.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.02[035].

This driver is included with the following datasystems:
GC OpenLab CDS C.01.01
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.04


KPR#: 189    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

It is not possible to set a positive pressure when column outlet is set to "other"

Problem:
If the colum is configured from one inlet to other (in case for a non-purged CFT or other applications) it's not possible to define the pressure of the outlet.  It's possble to set it from the GC keyboard.
Temporary Solution:
n/a
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 190    Product: 78xx Driver Software     7890A    1.05

Keyword: 

One-line Description:

FID Flame out; no feedback in software

Problem:
The GC status and instrument actuals list the signal value as 0.0 but does not provide feedback if the flame is out.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[001]

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 195    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

uECD and TCD report printout for makeup flows is wrong

Problem:
7890A has a uECD detector with flows configured.
With the following method settings...
    Makeup Flow(N2)               :  30 mL/min
    Constant Column + Makeup : 31 mL/min

... the method printout reads:
    Makeup Flow(N2)               :  31 mL/min
    Constant Column + Makeup :    0 mL/min
Temporary Solution:
n/a
Fix Information:

Fixed in 7890A SW Driver 4.01 [048].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 196    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Control of PCM used as inlet EPC does not work in B.04.03 [52]

Problem:
If PCM is configured with the inlet heater as PCM A or PCM B controlling the column flows/pressures, the temperature and pressure/flow setpoints cannot be set. CS shows a septum purge value and the instrument goes into shutdown, due to "septum purge not ready".
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[007].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 197    Product: 78xx Driver Software     7820A    2.12

Keyword: 

One-line Description:

Column Compensation not available on 7820

Problem:
By design column compensation is not available on the 7820.
Temporary Solution:
Not Available
Fix Information:

Not Available


KPR#: 198    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Autozero at beginning of run did not occur for run 4 for Sequence using same method

Problem:
A sequence was run using auto zero at the beginning of each run.  At run 4 in the sequence, autozero did not occur.  This is due to a firmware issue wherein only 15 signal events (including auto-zero) would be processed before failure.  The number 15 is cumulative since the last reboot which means that after 15 runs, the signal event will not be done.
Temporary Solution:
n/a
Fix Information:

This issue is fixed in 7890A firmware revision A.01.12.1.  No software update is required.


KPR#: 200    Product: 78xx Driver Software     7820A    4.01

Keyword: 

One-line Description:

GC continues when STOP on 7820 is pressed

Problem:
User makes a manual injection and waits until the peaks of interest elute, then hits the STOP on the 7820 front panel the GC stops (and resets the initial GC settings) but the chemstation continues and reports the system is running data analysis but does not go into data analysis. System appears to hang up waiting on the expected run time and report is generated at that point rather than directly after STOP is pressed.
Temporary Solution:
Use STOP run from the software.
Fix Information:

Fixed in driver revision 4.02[11].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 201    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

Run Log entries remain for each run until restart of GC.

Problem:
Run Log entries (i.e. oven door opened) occurring during a run remain in the logbook until the restart of the GC.
Temporary Solution:
n/a
Fix Information:

No software update is required.  This is fixed in GC firmware revision A.01.12.1.


KPR#: 203    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Open GC edit parameters and close it without changes; method modification flag is set

Problem:
Saved a method.  Open the GC edit parameters to review a setpoint and  made no changes or changed the view.  Selected OK and the method modification flag indicates a change was made.  No changes were made.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[019].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 204    Product: 78xx Driver Software     7890A    3.02

Keyword: 

One-line Description:

7890A logbook shows 7693A turret, plunger errors as GCI1 did not start a run

Problem:
7693A ALS errors, e.g. turret, plunger, door open, etc. show a non-specific logbook entry: GCI1 did not start a run.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 209    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Unable to set equilibration time for LTM columns from Edit Parameters

Problem:
The LTM column's equilibration time is shown as 0. Change the equilibration time by entering 0.5 and downloading the method. This does not change the value at the GC.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[005] and GC Firmware revision A.01.13.

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 210    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Unable to plot the LTM column temperature

Problem:
The diagnostic signal can be selected for the LTM actual temperature. Upon download and upload it is returned as oven temperature.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[007].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 211    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

Changes made to 7890 method from GC instrument diagram result do not load.

Problem:
Open the instrument parameter from GC Instrument Diagram and confirm sample prep is off. Set inlet heater - save changes. Open the instrument parameter from the GC Instrument Diagram and confirm sample prep is off. Change the inlet heater temp and save as a new method. Load the new method  - a configuration mismatch message is displayed.  

Requires Easy Sample Prep
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[004].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 213    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

B.04.03[16] with 7890 and software will not accept the syringe size

Problem:
After modifying and saving the method the injector screen will not accept the syringe size of 10 uL.  The menu will pull down and 10uL is accepted but when the screen is OK'ed the size changes to blank.
Temporary Solution:
Enter the syringe size from the GC front panel, upload the configuration and save the method.
Fix Information:

Fixed in 78xx driver revision 4.02[019].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 214    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

LTM subtabs do not update without switching away from the Oven tab

Problem:
After making a change to an LTM ramp, switch to a different LTM subtab or oven subtab. When user goes back to the original LTM subtab the change is gone and everything has reverted to the initial settings.
Temporary Solution:
Make a change to an LTM ramp; and then switch to a different main tab (signals for example).
Fix Information:

Fixed in driver revision 4.02[001].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 218    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

Other and Unknown Inlet cannot be configured

Problem:
Other and Unknown Inlet cannot be selected on the config panel for inlets or on the instrument.
Temporary Solution:
n/a
Fix Information:

Fixed in GC Firmware revision A.01.013.  No software update is required.


KPR#: 219    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

PTV pressure reverts back to existing value

Problem:
Enter a new pressure on the PTV inlet. The number is not accepted, the value reverts back to the existing value.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[19].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 220    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

Modifying the Syringe Size When Using a 7890A GC

Problem:
After modifying and saving the method the injector screen will not accept the syringe size of 10 uL.  The menu will pull down and 10uL is accepted but when the screen is OK'ed the size changes to blank.
Temporary Solution:
Enter the syringe size from the GC front panel, upload the configuration and save the method.
Fix Information:

Fixed in 78xx driver revision 4.02[019].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 223    Product: 78xx Driver Software     7890A    0.0.1.2

Keyword: 

One-line Description:

Backflush icon is not displayed when the outlet of the column configuration is set as "Other".

Problem:
When the outlet of the column configuration is set as "Other", there is no backflush icon displayed in the 7890 Edit Parameter Screen.
Temporary Solution:
n/a
Fix Information:

Fixed in 78xx driver revision 4.02[009].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03

Please note, when the configuration is changed from backflush not available to backflush available, the user must exit the method editor and begin the edit a second time in order for the backflush icon to be shown.


KPR#: 227    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

When running Backflush from ChemStation with segmented columns only the in-segment is taken into account in the calculation.

Problem:
If you run Backflush from Chemstation with segmented columns (usually associated with LTM), only the in-segment portion of the configuration is used in the backflush flow rate and void volume calculations. This leaves the user unable to calculate the backflush time.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.02[023].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 228    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

The display of the parameter setting screen of the method setup is small.

Problem:
In the Edit Parameter screen for the oven and the inlets, the display of this screen is small. Adjusting the method setup screen will not resize this screen.
Temporary Solution:
Right click on  parameter setting screen, and go to "show graphical plots". The display will resize to a normal size.
Fix Information:

Fixed in 78xx driver revision 4.02[006].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 229    Product: 78xx Driver Software     7890A    4.01

Keyword: 

One-line Description:

After editing the GC method and selecting ok - validation error message is displayed.

Problem:
In driver revision 4.01, after editing the GC method and selecting ok, a "Validation Error" error message flashes in the left hand corner of the screen. User must select cancel to exit the GC parameter screen. 

No method resolution occurs when loading method
Software loads fine and can make runs. Methods cannot be edited in online or offline mode.  

The configuration includes 4 packed columns that flow to the Back TCD Detector. Column 1's inlet is PCM C-2 (@ const pressure); 4's inlet is Aux EPC 5 (@ const pressure); and both 2's and 3's inlet is PCM C-1. Column 2's mode is constant flow; and 3, since it is also connected to PCM C-1, is reported as being controlled by #2.
Temporary Solution:
Workaround for customers with driver revision 4.01:  replace columns #2 and #3 with a single undefined (e.g., packed) column, suitably described as being a parallel combination. The user can then set either a flow or a pressure setpoint to the new combined column #2.
Fix Information:

Fixed in driver revision 4.02.  

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03
OpenLab CDS ChemStation C.01.04


KPR#: 230    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

7820 Back Injector Only Resets Injection Source to Manual

Problem:
On a 7820, with only a Back Injector Installed the following message is displayed when launching the 7820: 

Upon launching the 7820A, the following message appears:

DEF_GC.M: 7820
Resetting to Manual injection.
For method 'DEF_GC.M, injection source 'GC Injector' conflicts with injection location.

This does not happen when there is only a Front Injector installed.
Temporary Solution:
Open the injection source and change it to GC Injector.
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 237    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

Channel Partner Detector for the 7890 GC does not work

Problem:
Start on oven page, switch to channel partner detector tab, ok method. Restart editor - editor displays as a blank window with buttons at the bottom.
Temporary Solution:
Use the instrument diagram to start the editor on any tab except the detector tab. Then proceed to navigate to the detector tab.
Fix Information:

This is fixed in 78xx driver revision 4.02[022].

This driver is included with the following datasystems:
GC ChemStation B.04.03 SP1 [87]
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.03


KPR#: 238    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

PCM Inlet Heater Shutdown

Problem:
7890 GC with PCM controlled inlet heater. Open Edit Parameters, enable PCM inlet heater checkbox and make a change to the heater settings. Download the method to the GC or hit apply and the inlet heater is turned OFF at the GC. If the user closes edit parameters, and then re-opens edit parameters the PCM inlet heater checkbox is disabled.
Temporary Solution:
Create your method, save, and download to the GC (the PCM inlet heater will shutdown). At the GC, set the PCM inlet heater to the desired setting. Upload the method from the GC (Instrument, Upload Method), and save your method as a new method (Save method as …). This method can now be used and loaded successfully in the future.

The workaround will need to be performed any time changes are made to the method.
Fix Information:

This driver is included with the following datasystems:
OpenLab GC ChemStation C.01.04
GCMSD ChemStation E.02.02 SP1
OpenLab CDS EzChrom A.04.04


KPR#: 247    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

Multiposition valve resets to position # 1 between injections

Problem:
The set position of the gas streams used for a multiposition valve is controlled in the sequence. The user can run from different gas streams by entering the different numbers in the sequence table. The position is not stored as part of the method. The MPV set position is always reset to start from position #1 between runs.
Temporary Solution:
n/a
Fix Information:

This is fixed in 7890A firmware revision A.01.14.  No software update is required.


KPR#: 252    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value

Problem:
MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value. The Total Flow combo box is active, but any entry made there reverts to its original entry.
Temporary Solution:
Total Flow is editable in this mode from the GC keypad in driver versions prior to 5.01.
Fix Information:

Fixed in the 5.01 Driver.
This driver is included in:
A.01.05 OpenLAB CDS
B.07.00 MassHunter


KPR#: 265    Product: 78xx Driver Software     7890A    5.01

Keyword: 

One-line Description:

How to Exit the 7890B GC Maintenance Screen

Problem:
Click on the Maintenance item in the Instrument drop down menu to open or close the Maintenance screen.  There is not a red X to close this menu.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 266    Product: 78xx Driver Software     7890A    5.01

Keyword: 

One-line Description:

The GC Maintenance View May Not Change Until the Software is Closed and Re-Opened

Problem:
The Maintenance View May Not Change Until Software is Closed and Re-Opened
Temporary Solution:
In the Maintenance menu, the user has the option to change the view of the menu by changing it under the Options button.  

Sometimes the change may not go into effect until the software is closed and re-opened.
Fix Information:

n/a


KPR#: 269    Product: 78xx Driver Software     7890A    5.01

Keyword: 

One-line Description:

ALS Behavior When Aborting a Sequence

Problem:
If a sequence is aborted in the middle of a sequence, the vial will be left in the turret until the next run is started.  At that time, it will be put back in the tray.
Temporary Solution:
This is expected behavior.
Fix Information:

n/a


KPR#: 273    Product: 78xx Driver Software     7890A    4.02

Keyword: 

One-line Description:

7890B GC Method Compatibility

Problem:
The7890B GC driver included in the new data systems (A.01.05 OpenLAB CDS and B.07.00 MassHunter) supports up to 12 valves.  The GC driver in previous data systems only reported 8 valves.  Because of this difference, you cannot develop a method on a new data system and run it on an old data sytem.  You can take a method loaded on an old data system and run it on the new data system.

If you try to load a new data system method on an old data system, the behavior that you will see is that after starting a run, the software and the GC just sit there waiting for prerun to start, but nothing happens.
Temporary Solution:
n/a
Fix Information:

Documented Behavior for GC Methods Developed Using A.01.05 OpenLAB CDS and B.07.00 MassHunter.


KPR#: 615    Product: RTE     RTE    A.02.04

Keyword: 

One-line Description:

The peak inflection point position does not match classical ChemStation reporting.

Problem:
The peak inflection point position is drawn different compared to the classic reporting. This is because there are multiple valid positions where the point can be drawn. 
The effect can be observed for peaks that show only very small changes of the slope over a "longer" time range and the 2nd derivative is close to 0 over that range.
Temporary Solution:
The reported peak performance values are the same in classic and Intelligent Reporting. Users who prefer the old drawing need to use classic reporting which is still available in ChemStation C.01.01.
Fix Information:

There is currently no fix planned


KPR#: 617    Product: RTE     RTE    A.02.04

Keyword: 

One-line Description:

Exponential calibration curve plot isn't shown in Calibration Report when using Intelligent Reporting.

Problem:
Exponential calibration curve plots cannot be reported with Intelligent Reporting. An empty plot is shown instead.
Temporary Solution:
If exponential calibration curves plots need to be reported, the (old) classic reporting needs to be used in ChemStation. 
EZChrom users need to use the EZChrom standard reporting.
Fix Information:

A potentially fix is currently under investigation.


KPR#: 1178    Product: 78xx Driver Software     7820A    3.01

Keyword: 

One-line Description:

Max oven temp not changed after installing column with higher max temperature

Problem:
The software will automatically lower the oven max temperature when a low temp column is installed.  This is to protect the column. The software does not rely on the User to do this to avoid damaging expensive columns. 

The requested enhancement is to automatically increase the oven maximum to match the higher oven max of the newly installed column.

Currently,  the user is required to manually update the oven max if it needs to be higher; the oven max is not automatically increased to match the higher temp column.  This is true for 68xx and 7890.
Temporary Solution:
n/a
Fix Information:

Fixed in 7890A Software Driver Version 4.01 [044].

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 1196    Product: 78xx Driver Software     7890A    3.01

Keyword: 

One-line Description:

7890A: Shutdown events are not published to the software GC Status screen

Problem:
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen.  

Only the first event, which is 'not ready', is reported to the Agilent GC Status screen.  The second event, which is the shutdown, is not displayed.
Temporary Solution:
n/a
Fix Information:

Fixed in driver revision 4.01.  7890A requires a minimum Firmware revision of A.01.12.

The driver is included in the following datasystem packages:
GC ChemStation B.04.03 DSP1 [52]
OpenLab CDS EzChrom A.04.02
GCMSD ChemStation E.02.02


KPR#: 1731    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 

One-line Description:

LC/MS spray chamber temperatures are no longer be applied to the spray chamber heated zones following a spray chamber open/close action.

Problem:
After an LC/MS spray chamber has been opened and closed, temperature will no longer be applied to the Spray Chamber heated zones until the spray chamber parameters are  updated (by clicking "ok" in the Spray Chamber Parameters dialog box or loading a method) or the system is turned "off" or to "standby" and then back to on.  This occurs with the APCI, APPI and Agilent Jet Stream spray chambers.  The MultiMode source will set the system to "Standby" if opened & closed, however.
Temporary Solution:
Turn the instrument to "off" or to "standby" in ChemStation, then back to "on" if this condition is encountered.
Fix Information:

n/a


KPR#: 1781    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: 

One-line Description:

Intelligent Report Templates generated with ChemStation C.01.03 may require update

Problem:
Intelligent Reporter templates designed for multi-signal methods may need to be adjusted after updating to ChemStation C.01.04. 
Non-calibrated peaks of signals calibrated in a main signal do no longer appear as calibrated compounds (as they are not quantified on the 2nd signal) but as un-identified peaks.

Note: Calibration tables setup to "calculate signals separatly" are not affected by this design change.
Temporary Solution:
In order to display the results of the mentioned peaks, the peak filter of the report template should be turned off completely (remove all check marks in the "show peak types" screen).
Fix Information:

In rev. C.01.05. the mentioned peaks are treated as identified, but not quantified peaks.


KPR#: 1782    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: 

One-line Description:

UNC paths are not supported as report file storage loaction

Problem:
The "copy to" function of the intelligent report method setup does not support usage of UNC paths.
Temporary Solution:
Map the network path as a drive letter and select the new drive letter as a report destination.
Fix Information:

There is no fix planned.


KPR#: 1804    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: 

One-line Description:

Queue method generated status “error” when using 35900E

Problem:
Queue method generated status “error” when using 35900E
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1823    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: 

One-line Description:

Default column information report shows incorrect GC column information and labels

Problem:
The built-in reporting snippet "Samples > Column Info" of intelligent reporting of OpenLAB CDS shows inaccurate units and lables.
The inner diameter unit is printed as milimeters, where it should be micrometers.
The column lenght unit is printed as milimeters where it should be meters. 
The lable for Inner diameter, Inventory number does not match the classical report.
Temporary Solution:
Use report template editor to modify the settings of the report items inide the Column info snippet

Inner diameter:  Change the value definition from 
=FORMATNUMBER(First(Injection_ColumnDiameter),2,True,False,False) & " mm"
to
=FORMATNUMBER(First(Injection_ColumnDiameter),2,True,False,False) & " um"
in order to replace mm by um.

Column length:   Change the value definition from 
=FormatNumber(First(Injection_ColumnLength),1,0,0,0) & " " & First(Injection_ColumnLengthUnit)
to
=FormatNumber(First(Injection_ColumnLength),1,0,0,0) & " " & "m"
in order to replace mm by m.

The columns void time can be added as an individual reporting parameter (Fields > Injection >  ColumnVoidTime)
Fix Information:

There is currently no fix planned.


KPR#: 22769    Product: OpenLAB CDS EE     OpenLAB CDS EE    A.04.02

Keyword: 

One-line Description:

OpenLAB CDS Print Services when using redirected printers.

Problem:
OpenLAB CDS print services may stop unexpectedly when a redirected printer is configured for use with OpenLAB Shared Services. This results in the inability to print reports automatically from acquisition when using this printer.
Temporary Solution:
Redirected printers are created during remote desktop or terminal service sessions and are intended only for use by the application that created them. Redirected printers are not supported with OpenLAB CDS.
Fix Information:

Use only directly configured printers when adding printers to OpenLAB Shared Services.


KPR#: 22775    Product: OpenLAB CDS EE     OpenLAB CDS EE    A.04.03

Keyword: 

One-line Description:

Printer hardware error causes lockout of print activity

Problem:
If any printer configured for use on a CDS system is in an error condition, such as a paper jam, all printing ceases on the system. Print operations back up on the print server and no print operations will take place on any printer in the system. Printing of all compiled reports occurs when the printer error is cleared.
Temporary Solution:
Check for printer errors and clear them if printing ceases to function on the CDS system.
Fix Information:

Fix is currently under investigation.


KPR#: 22780    Product: OpenLAB CDS EE     OpenLAB CDS EE    A.04.03

Keyword: 

One-line Description:

IQ fails after upgrade wizard if non-default path was used for installation

Problem:
When upgrading an A.01.02 CDS system that was installed in a location other than the default location (C:\Program Files\Agilent Technologies\) the IQ fails. This happens as a result of the A.01.03 version installing in the default location although A.01.02 was not in that location and the IQ failure is a valid result.
Temporary Solution:
If OpenLAB CDS A0102 was installed in a non-default directory do not use the automatic upgrade. Uninstall A.01.02 using the A.01.02 installation media prior to performing installation of an A.01.03 system. If the upgrade was already performed and IQ is failing uninstall all OpenLAB CDS software and perform the A.01.03 installation again.
Fix Information:

Will be fixed in next minor revision.

Update - Resolved with OpenLAB CDS A.01.04 product revisions.


KPR#: 22789    Product: OpenLAB CDS EE     OpenLAB CDS EE    A.04.04

Keyword: 

One-line Description:

Unhandled exception during printer definition on shared server with UAC enabled

Problem:
When defining printers in the OpenLAB Control Panel interface an unhandled exception may occur on a system that has User Account Control (UAC) enabled when attempting to start or stop the print service.
Temporary Solution:
Open the control panel with a right click > as administrator or disable UAC to define and enable CDS print service configurations.
Fix Information:

n/a


KPR#: 1539    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: 1200 Control

One-line Description:

Wrong firmware revision displayed in Instrument Configuration dialog after firmware upgrade

Problem:
After a firmware upgrade for LC modules using RC.NET drivers, the old firmware revision (prior to upgrade) is still displayed in the Instrument Configuration dialog.
Temporary Solution:
Remove the affected modules from the configuration and re-add them.
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1720    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 1200 Control , Method Scouting Wizard

One-line Description:

Cluster information can be overwritten from the offline session

Problem:
Changes in a TCC or pump cluster configuration (like e.g. changing a solvent name) usually cause method resolution. 
The changes will also be updated, even when the configuration dialog is closed by pressing Cancel. This is true for online and offline. This will consequently modify methods, although nothing has actually been changed.
Temporary Solution:
None
Fix Information:

Fixed in revision B.04.03 SP1.


KPR#: 1623    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 35900 Control

One-line Description:

35900E - Error The Current Event Name is the same for a special event

Problem:
B.04.03 GC ChemStation with a 35900E configured. Select the CHANGE, DEFINE EVENTS tab in device configuration and you get an error "The Current Event Name is the same for a special event. Please change the current event name." 
 
This error is reported on a clean install of B.04.03 [16].

ChemStation B.03.02 works correctly.
Temporary Solution:
n/a
Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#: 1677    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: 35900 Control

One-line Description:

cannot start two instruments using the two channels of one ADC at the same time

Problem:
When two instruments are configured that use channel A and channel B of the same ADC converter, respectively, the two instruments cannot be running at the same time. When the second instrument is started, both ChemStation instruments will freeze and can no longer be operated.
Temporary Solution:
The two instruments will work when using only one of the instruments at a time.
Fix Information:

Fixed in OpenLAB C.01.03.


KPR#: 1702    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: 35900 Control

One-line Description:

An AD converter cannot be split into two instruments from a ChemStation client

Problem:
The two channels of a 35900 ADC cannot be part of two different instruments when controlled from a ChemStation client.
Temporary Solution:
When the 35900 is connected to an AIC and its two channels need to be part of two different instruments, they need to be setup from a networked workstation.
Fix Information:

No fix available.


KPR#: 1707    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: 35900 Control

One-line Description:

35900: Unable to set up Timed Events

Problem:
In the 35900 configuration, the Timed Events table is populated with default values. 
In the method, with the injection source set to 35900 Pre-Run Events, the Setup ADC Timed Events window opened but would not permit setup of the events. All buttons and drop down menus were gray and inaccessible.
Temporary Solution:
n/a
Fix Information:

 Fixed in revision C.01.04


KPR#: 1721    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 35900 Control

One-line Description:

View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.

Problem:
The acquisition parameters of an AD converter are not listed in the method printout from Data Analysis view.
Temporary Solution:
Use the Print Method... option from Method&Run Control view.
Fix Information:

No fix available.


KPR#: 1739    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: 35900 Control

One-line Description:

ADC can be non-responsive after long reprocessing

Problem:
The ADC might become NOT READY when the software is busy for a long time (e.g sequence reprocessing with IR to screen). When the system is no longer busy it cannot communicate with the ADC.
Temporary Solution:
Software and instrument need to be restarted.
Fix Information:

Fixed in revision C.01.04.


KPR#: 1826    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: 6850 Control , 6890 Control

One-line Description:

Column compensation cannot be started from the Data System with 68xx RC.NET

Problem:
Column compensation (comp) cannot be started from the Data System with 68xx RC.NET
Temporary Solution:
Column compenstation must be started via the GC keyboard.
Fix Information:

n/a


KPR#: 1412    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: 6890 Control

One-line Description:

6890 RC.NET driver: instrument diagram not updated to show configuration

Problem:
In the 6890 RC.NET driver, changed the GC column configuration to have the back inlet go to the back column to the back detector.  This change was not reflected in the instrument diagram as it is still showing the back inlet going to the front column to the back detector.
Temporary Solution:
After the configuration change is made to the column connections, the workaround is to shut down the ChemStation software and restart to reflect the new configuration changes.

Beginning with  B.04.03 DSP1 [54], after updating the column configuration, save and reload the method will update the instrument diagram.

Classic 68xx driver works correctly.
Fix Information:

No fix Planned.


KPR#: 1464    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: 6890 Control

One-line Description:

Addition of new columns to the column catalog (database) is not possible after upgrade

Problem:
Upgrading from B.01.03 to B.04.02, it is no longer possible to add new columns to the GC column catalog (database) in Rev. B.04.02.

The following error is shown:

The Catalog File cannot be opened.  If needed, you may obtain an appropriate replacement file from Agilent.
C:\Chem32\CORE\...\DRIVERS\catalog.mdb
Temporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this.

The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP HP-5 in the database.  Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) 

In theory, customers should not be creating more than one catalog entry with the same model number.  It’s like having two items in the LLBean catalog with the same order number!  But, of course, the older software allowed the duplicates.

Temporary Fix:

Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example.  Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.
Fix Information:

Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example.  Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.


KPR#: 1622    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 6890 Control

One-line Description:

Sequence continues while GC temperature is not ready

Problem:
When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run.  This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period.
Temporary Solution:
n/a
Fix Information:

Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1[54]. 

Fixed in C.01.03


KPR#: 1684    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: 6890 Control

One-line Description:

6890 always opens to Configuration tab when opening Edit Agilent 6890 GC Parameters

Problem:
Select Edit Agilent 6890 GC Parameters…
The tab will always open to the configuration tab.
Temporary Solution:
n/a
Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#: 1685    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: 6890 Control

One-line Description:

Column info is missing with 6890

Problem:
The column specified as being used is missing information: description, Manufacturer and Model Number.
Temporary Solution:
n/a
Fix Information:

This is fixed in 68xx Driver 6.11 which is included in B.04.03 SP1 and C.01.03.


KPR#: 1784    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: 6890 Control

One-line Description:

6890 RC.NET: Edit a running Method by making the oven hold time longer and click ok, then the Sequence/Single run stops

Problem:
During a run, the GC method parameters are opened.  Select OK after increasing the oven parameter hold time.  The sequence or single run will just stop and the method is dirty. 

Logbook shows:
Attempt to download method to GC failed.
Instrument Error - Method/Sequence stopped
Method Aborted
Instrument run aborted
Method Aborted
Sequence terminated due to an error
Temporary Solution:
n/a
Fix Information:

Issue fixed in OpenLAB C.01.04.


KPR#: 914    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01 SR1.1

Keyword: 7890A Control

One-line Description:

7890A: ALS errors abort sequence; no Retry/Abort option offered

Problem:
7890 with ALS. Create an ALS error, such as a Vial not found event.  Instead of offering a pop-up that offers a second chance to Retry/Abort with the named error, this simply aborts the sequence. The error is buried deep in the sequence log of that instrument:

'Method       Method started: (F) line# 2 at 2 inj# 1        09:11:51 09/20/07
  67 41e0 46f27198    0
Method       Instrument running sample Vial 2 (front)       09:11:52 09/20/07
   0 1da5 46f271b2  300
GC         1 There was no bottle in the gripper.            09:12:18 09/20/07
   0 1da6 46f271b2    0
Instrument   GCI1 did not start a run                       09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Instrument Error - Method/Sequence stopped     09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Method aborted                                 09:12:18 09/20/07
 43e 41df 46f271b2    0
Sequence     FRONTALS.S terminated due to an error          09:12:18 09/20/07'
Temporary Solution:
n/a
Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#: 1239    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: 7890A Control

One-line Description:

7890A Runs cannot be started until the GC and injector are idle, view GC Status

Problem:
Occassionly, when trying to start a run or sequence on a GC whose front display and ALS are Ready, fails to start. Logbook shows:
'Runs cannot be started until the GC and injector are idle'.
This can occur after an abort or stop of sequence, or connecting to a GC that is controlled by another PC.
Temporary Solution:
Pressing "stop" on the GC front panel enabled the GC to makes runs again.  Power cycling the GC may be necessary.
Fix Information:

Fixed in revision B.04.03.


KPR#: 1329    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01

Keyword: 7890A Control

One-line Description:

ALS errors abort sequence; no Retry/Abort dialog box option offered

Problem:
7890A with ALS. 

If an ALS error, occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort.  The sequence simply aborts. The error is shown in the sequence log of that instrument:

'Method       Method started: (F) line# 2 at 2 inj# 1        09:11:51 09/20/07
  67 41e0 46f27198    0
Method       Instrument running sample Vial 2 (front)       09:11:52 09/20/07
   0 1da5 46f271b2  300
GC         1 There was no bottle in the gripper.                09:12:18 09/20/07
   0 1da6 46f271b2    0
Instrument   GCI1 did not start a run                              09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Instrument Error - Method/Sequence stopped     09:12:18 09/20/07
  67 41e0 46f271b2    0
Method       Method aborted                                 09:12:18 09/20/07
43e 41df 46f271b2    0
Sequence     FRONTALS.S terminated due to an error          09:12:18 09/20/07'

GC goes back to "ready" state instead, and vials remain in the position they were when the injector failed (not returned to tray).
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1569    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: 7890A Control

One-line Description:

Sequence aborts if vial not found with Overlap injection enabled (7890A only)

Problem:
7890A system using overlap injection feature

1. Define a sample overlap of 1 min.
2. Define a sequence of 5 runs
3. Insert 4 vials only

The 7890A GC starts the acquisition of Vial 4 correctly.  When it starts to prepare and pick up vial 5 which is missing, the acquisition of vial 4 is stopped and the sequence aborts.

When sample overlap is enabled and the vial to be prepared is not found, the currently running sample in the sequence will abort.  The sequence will stop and there will be errors displayed in the sequence logbook and GC status.

The error messages displayed are as follows:

- Command to Tray failed:  Tray Prep Failure Code: 250: Front injector has no 
  vail in turret...
- User stopped pretreatment program, finalizing.
- Instrument Error - Method/Sequence stopped
- Method aborted
- Sequence.S terminated due to an error
- Sequence.S completed
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1625    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01

Keyword: 7890A Control

One-line Description:

Must restart data system for barcode reader availability change (7890A)

Problem:
If the presence of the barcode reader is changed while the data system is connected to a 7890A, the availability of barcode options in the Sequence Parameters screen does not change until the data system is restarted.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB revision C.01.03.


KPR#: 1641    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: ACQ

One-line Description:

Not possible to modify data analysis parameters of method during a single run in the offline; cannot save

Problem:
It is not possible to change integration parameters in the offline for a single run and save the new parameters in the method file.
Temporary Solution:
Data analysis parameters can be updated and saved if using a sequence.  It is recommended to prepare a single run sequence.
Fix Information:

 Fixed in revision C.01.04


KPR#: 1660    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 LC DSP

Keyword: ACQ

One-line Description:

No special characters should be used for names of methods, data files or sequences

Problem:
Names for data files, methods and sequences should only contain alphanumeric characters. Special characters like ",#,$, %, &, @, ? and others should not be used and can in general cause trouble. The same is true for trailing blanks.
While this is outlined in the written documentation several times, it is not enforced for all dialogs inside the software. There are still the occasional loop hole where such a character can be entered or imported.
Temporary Solution:
The possiblity of entering special characters does by no means imply that it is safe to do it.
Fix Information:

n/a


KPR#: 1701    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: ACQ

One-line Description:

Terminal Service can close under high workload

Problem:
It has been observed that the terminal service on an AIC can terminate under high workload.
This is a limitation of Windows 2008 server. Since this only affects the remote session control to the ChemStation Client, no data will be lost.
Temporary Solution:
No fix available.
Fix Information:

Will not be fixed for Windows 2008.


KPR#: 1732    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: ACQ

One-line Description:

Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”

Problem:
1. Create seq table, with one run (Method A) for the front table, one run (Method B) for the back table this must be done first
2. Before line one of the sequence completes, add a run2 to the back table only (BACK.M) by append
3. When the back run is in process, add a sequence line to the front table, use a different method (method A) 
4. Message stating that Injector doesn’t match the sequence’s  DUAL Injector occurs even though all methods are saved with Dual injection selected as the injection source
Temporary Solution:
n/a
Fix Information:

Fixed in Revision C.01.04.

Point patch available for B.04.03 SP1.


KPR#: 1742    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: ACQ

One-line Description:

Changes in sequence methods made during acquisition are not saved automatically

Problem:
Run a sequence and change the method during the acquisition (e.g. change the runtime). There is no possibility to save the method. The method should get saved at the end of the run automatically, but it does not. This issue occurs in C.01.01 - C.01.03 versions.
Temporary Solution:
n/a
Fix Information:

Resolved in Revision C.01.04.


KPR#: 1747    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ACQ

One-line Description:

Queue Method vial range reflects range of loaded method, not the method browsed to

Problem:
Load a method in ChemStation. Select Queue Method. Within that dialog, Browse... to load another method with a different vial range. The vial range (Sample Parameters: Vial/Location) is not updated to reflect the range of the method loaded in Queue Method, though the Signal (1 and/or 2) Data File info is correctly loaded from the browsed to method.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1811    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: ACQ

One-line Description:

Acq.txt and acq.macaml is missing for 7890A back injector only sequences

Problem:
Acq.txt and acq.macaml is missing for 7890A back injector only sequences
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1749    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Audit Trail

One-line Description:

Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured

Problem:
GC with HS or PAL as the injection source. Close the session and remove the external device. Open the session and select Compare. No differences are seen in the method.

Download the original method. The  injection source has been changed to Manual. No change is logged to the Method Audit Trail.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1819    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Audit Trail

One-line Description:

UIB II - set a timetable and control "Method Audit Trail"

Problem:
If a timetable of UIB II  is set and stored via "Save Method as..", the time of the change should be stored in the "Audit Trail".
But the time of the method change is not correct.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 13    Product: Unionsquare     Unionsquare    A.01.02

Keyword: Bitmaps

One-line Description:

Intelligent reporting does not support exponential baselines

Problem:
The intelligent reporting of OpenLAB CDS rev. A.01.02 does not support displaying exponential baseline constructions in chromatograms.
A straight line is shown instead of an exponential curve.
Peak area and height calculations are not affected.
Temporary Solution:
Classical reports of ChemStation and EZCHrom show the correct baseline position.
Fix Information:

A hot fix for OpenLAB CDS ChemStation is currently under investigation.


KPR#: 30    Product: Unionsquare     Unionsquare    A.01.05

Keyword: Bitmaps

One-line Description:

Insufficient contrast on b/w printers running in eco mode

Problem:
The chromatogram plots of Intelligent Reporting of OpenLAB CDS A.01.04 and earlier do not show satisfying contrast on black & white printers. This is especially true if the printer defaults are set to eco mode.
Temporary Solution:
n/a
Fix Information:

Intelligent Reporting of OpenLAB CDS A.01.05 offers the option to print all signals in black in order to increase the contrast on b/w printers.


KPR#: 31    Product: Unionsquare     Unionsquare    A.01.06

Keyword: Bitmaps

One-line Description:

Signal Plot thickness appears inhomogenous in case of high data rates

Problem:
The signal plot thickness of Intelligent Reporting of OpenLAB CDS A.01.04 and earlier may appear inhomogenous and peaks show different line thickness compared to rnages with no peaks.  The quality of the chromatogram print out doesn't look satisfying. 

The problem is caused by a data compression algorithm affecting chromatogram plots where the signal data rate is too high in comparison to the size of the printout. E.g. in case there are 3 or more data points per pixel repesenting a time interval.
Temporary Solution:
The plot compression can be prevented by choosing a larger scaling of the time axis and stretching the signal in the direction of the chromatogram's time axis.
Chromatograms showing larger run times (> 60 min) should be plotted on landscape paper format or the chromatogram could be cut in 2 or 3 time slices to fit on a portrait paper format.
Fix Information:

If you encounter the described problem and do not want to enlarge the time axis scaling you can use the following procedure.

Intelligent Reporting of OpenLAB CDS A.01.05 introduces a new custom parameter 
to turn  data compression off. This parameter needs to be edited in the RDL file directly.

New templates generated with OpenLAB CDS A.01.05:

-  Create a backup copy of your template RDL file
-  Open the template RDL with a text editor, like Microsoft Notepad
-  Search the string "AlwaysDrawAllPoints"
-  For each chromatogram plot inside a report template you will find a section

                  
                    AgtPlot:ChromPlot:AlwaysDrawAllPoints
                    False
                  

-  Change  False to  True
- Save the template



Report templates generated with OpenLAB CDS A.01.04 or earlier:

-  Create a backup copy of your template RDL file
-  Open the template RDL with a text editor, like Microsoft Notepad
-  Search the string "ChromPlot:IsAnnotateAllPeakMarkers"
-  after each section

                  
                    AgtPlot:ChromPlot:IsAnnotateAllPeakMarkers
                    True
                  

 you should append a section

                  
                    AgtPlot:ChromPlot:AlwaysDrawAllPoints
                    True
                  

- Save the template


KPR#: 1311    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01

Keyword: Calibration

One-line Description:

Amount is given as zero when it actually could not be calculated

Problem:
In case of e.g. a quadratic calibration curve it may happen that an amount cannot be calculated for the area in a chromatogram (if the area is outside the calibrated area range).
In these cases the amount will be reported as zero instead of "-" as an indication that the amount could not be calculated. However, with the default report styles, the report contains a warning "invalid calibration curve".
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1766    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Calibration

One-line Description:

Calibration method printout does not print compound level's Heights when quantitation is based on height

Problem:
Calibration method printout does not print compound level's Heights when quantitation is based on height
Temporary Solution:
Available fix in C.01.04 Hotfix03
Fix Information:

Fixed in C.01.05.


KPR#: 1793    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Calibration

One-line Description:

Re-establish the old layout for calibration reports

Problem:
Re-establish the old layout for calibration reports
Temporary Solution:
n/a
Fix Information:

Fixed in C.01.04 Hotfix03.


KPR#: 1799    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: Calibration

One-line Description:

Changing calibration table in Batch review generates an error

Problem:
When run is spooled in Batch Review, click on menu Calibration –> New Caibration Table, select OK and overwrite existing table
Click OK on calibration table when it is displayed
For a brief moment an error is displayed
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1807    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Calibration

One-line Description:

Calibration method printout does not print compound level heights when quantitation is based on height

Problem:
Can't show calibration table based on Height when printing calibration table curves
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1809    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04 HF03

Keyword: Calibration

One-line Description:

C.01.04 HotFix03 : Amount units in calibration report printout is not updated

Problem:
Enter “ug/ml” for “Amount Units” in the “Calibration settings”. The calibration table was updated with “Amt (ug/ml)” but the calibration table printout for the Overview table indicated “Amt (wt%)”. This also happens with “Amt (kg)” and the calibration report printout still reflects as “Amt (wt%)”.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1817    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Calibration

One-line Description:

Sequence run for bracketing standard which is started with blank completed without creating data files

Problem:
When a sequence is started with blank data files, data files are not generated.  

When the sequence is started with standard as bracketing std. (line 1) and sample (line 2), sequence completed and data files are generated
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1827    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04 HF03

Keyword: Calibration

One-line Description:

C.01.04 Hot Fix03 : incorrect calibration curves legend in calibration curves printout

Problem:
When using "ISTD" as "Calculation mode" and "Area" for "Based on", the calibration curves x-axis and y-axis shown on the calibration curves were :
"x-axis=amount ratio"
"y-axis=area ratio”

but the report printout for calibration curves have the following incorrect legend in English version: 
"x=amount"
"y=area"

And incorrect legend in Japanese version:
"x= amount"
"y= height"
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1587    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: CE Control

One-line Description:

The VialTable check does not occur during the startup of the CE ChemSation

Problem:
During the startup of the CE ChemStation, the System VialTable is not loaded therefore when a method is loaded conflict checks with the method vialtable is not performed.
Temporary Solution:
If you are using an Agilent 1600 Capillary Electrophoresis instrument, the following post installations steps are required.

1. Insert the OpenLAB CDS Disk 6 
2. Open the folder called \Release Updates\OpenLAB CDS ChemStation Edition\CE\
3. Copy the file HPCE.MAC to the folder \CE\
   Note> The default chemStation path is c:\Chem32

It's recommended to create a backup copy of the original HPCE.MAC before replacing it.
Fix Information:

This issue is resolved in the software revision C.01.02


KPR#: 15    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Client Report Rendering

One-line Description:

Method information partially hidden in intelligent reports exported to XLS file format

Problem:
The method information part intelligent reports exported to XLS file format is partially hidden.

This issue occurs due to the limitation of the Excel renderer. Basically any report item while being rendered causes Excel's cells to merge. To suppress the undesired merging of cells the parameter ‘CanGrow’ is set to true, and  'AutoSize’ of the excel row fails as a consequence. 

To read more about this refer to the ‘Merging Cells’ paragraph on the link http://msdn.microsoft.com/en-us/library/ms159836.aspx 
Also find a suggested workaround for the issue at http://connect.microsoft.com/SQLServer/feedback/details/163236/reporting-services-row-height-error-on-export-to-excel
Temporary Solution:
Select all cells in the Excel sheet and double click on the separator between the 1st and the 2nd row. Excel automaically sizes all cells so that the content fits.
Fix Information:

There is currently no fix planned.


KPR#: 16    Product: Unionsquare     Unionsquare    A.01.01

Keyword: Client Report Rendering

One-line Description:

"Report Printed" time stamp in report footer is several hours off in Print Preview

Problem:
In the report preview, when switched to Print Preview, the print time displayed in the footer is several hours off. In interactive mode, the time is correct.

The problem is caused by Microsoft ReportViewer 2008 SP1.
Temporary Solution:
Print the report to PDF to review the final layout of the printed report.
Fix Information:

There is currently no fix planned.


KPR#: 20    Product: Unionsquare     Unionsquare    A.01.03

Keyword: Client Report Rendering

One-line Description:

Localized character in saved PDF report cannot be searched in Adobe Reader

Problem:
Localized character in saved PDF report cannot be searched in Adobe Reader.
Temporary Solution:
The wrong behavior is caused by a 3rd party component (Aspose). As a workaround it is recommended to use the ChemStation  PDF printer.
Fix Information:

There is currently no fix planned.


KPR#: 21    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Client Report Rendering

One-line Description:

Page numbering start counting at 0 in case of .doc report file format

Problem:
Intelligent Reports exported to .DOC file format show a page numbering starting at 0 when report is viewed in Microsoft Word application.
The defect is caused by a 3rd party report renderer and cannot be fixed in the reporter code.
Temporary Solution:
The footer section of the report template can be modified to translate the page numbering to word document parameter that could be updated by pressing the F9 key.

The correct syntax for the page count footer is:
=PageNumber & " of " & TotalPages
Fix Information:

There is currently no fix planned.


KPR#: 23    Product: Unionsquare     Unionsquare    A.01.01.113

Keyword: Client Report Rendering

One-line Description:

Scaling a signal to "full scale" does not add a percent sign to the axis label

Problem:
When changing the signal axis of a chromatogram from "all the same scale" to "each in full scale", the signal axis range changes from 0 - 100%.
However, the signal axis labels do not show the percent sign and it's not immediatly clear that the scaling is relative as opposed to absolute.
Temporary Solution:
N/A
Fix Information:

A fix a planned for the next minor release after A.01.04.


KPR#: 24    Product: Unionsquare     Unionsquare    N/A

Keyword: Client Report Rendering

One-line Description:

Report shows unexpected blank pages

Problem:
Working with report template editor of OpenLAB CDS A.01.03 or ealier it happens frequently that reports show additional empty pages.
They are typically introduced by dragging and dropping a large table to the center of a template or by adding more columns to an existing table.
The width of the template editing panel gets expanded in order to show all the content added.
However, when the table gets shrinked or moved the additional width does not get automatically released again.
Temporary Solution:
Check for a thick black vertical line at the right hand side of the editing pane. Move this line to the left until it hits the header box.

If this still does not remove the empty page, the RDL template file needs to be examined with a simple text editor, like Microsoft Notepad.
Search the entire template for data fields called "" and check for values greater 8 inches. Depending on the paper format and the page margins those values are potential troublemakers and should be reduced to e.g. 7.8in.
Fix Information:

The auto-shrinking of the editing pane has been improved with revision A.01.04 of OpenLAB CDS A.01.04.


KPR#: 28    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Client Report Rendering

One-line Description:

Report Export to .TXT or .DOC may fail if report renderer creates an out-of-memory condition

Problem:
Exporting large sequence summary reports (700+ injections) to multiple output files (PDF, XLS, TXT, DOC) may generate an out-of-memory condition and the last report file in the queue may not be generated.

The TXT report shows the following warnign message: The report template definition is invalid. The requested operation cannot be completed.
Temporary Solution:
This problem has been observed on a 32-bit Windows 7 operating system and is less likely to happen on  64-bit operating systems.

- Choose less output formats to be generated at the same time
- switch to 64-bit operating systems
Fix Information:

There is no fix planned.


KPR#: 29    Product: Unionsquare     Unionsquare    A.01.05

Keyword: Client Report Rendering

One-line Description:

Ineractively selected report parameters switch back to default when printing to PDF

Problem:
Intelligent Reporting of OpenLAB CDS offers the option to add or select user defined report parameters in the report preview.
By refreshing the preview these parameters can be applied to e.g. calculation or view filters.

When generating a PDF file based on the report preview of OpenLAB CDS ChemStation, the report gets rendered again and the parameters switch back to their default values.
Temporary Solution:
For user of OpenLAB Data Analysis:
This problem can be avoided by printing the report to a PDF printer (e.g. ChemStation PDF printer)

If external report parameters are used in calculations it is generally recommended to include the plain values in a report template section in order to allow a result verification and to ensure result consistency.
Fix Information:

A fix is currently under investigation.


KPR#: 34    Product: Unionsquare     Unionsquare    A.01.06

Keyword: Client Report Rendering

One-line Description:

Interactivly entered report parameters are not always updated

Problem:
Intelligent Reporting of OpenLAB cDS A.01.04 or higher offers the possibilty to 
change report parameters interactively, if the report template is setup accordingly.
When changing one or more parameters and switching from "Report Preview" to 
"Print Preview" (or vice versa), not all changed parameters are applied to the report.

In case of multi-select boxes the mouse focus needs to moved to another field in
order to submit the changed value. If the mouse focus remains in a multi-select box, 
the value does not get applied when switch to "Print Preview".
Temporary Solution:
The recommend workflow when using interactive report parameters is to refresh the 
report preview before printing or switching to print preview layout.
Fix Information:

There is no fix planned.


KPR#: 628    Product: RTE     RTE    A.02.04

Keyword: Client Report Rendering

One-line Description:

Large matrix reports may produce additional empty pages

Problem:
Large matrix reports may produce additional empty pages.
The issue is in the Microsoft Report Viewer control and cannot be fixed in the current product.
Temporary Solution:
Matrix reports easily exceed A4 or Letter sized paper formats and printing to a larger PDF paper format could be used as alternative.
Fix Information:

There is currently no fix planned.


KPR#: 653    Product: Open Lab Intelligence Reporter     G4635AA    A.02.01 SP1 HF01

Keyword: Client Report Rendering

One-line Description:

Calculating the standard deviation of identical numbers may return NaN

Problem:
When calculating the standard deviation or relative standard deviation of multiple identical numbers, the calculation may return "NaN" rather than 0.

The problem is caused by the small delta derived from the rounded result of the average calculation and the individual values. Summing up these small numbers may result in a floating point exception.
Temporary Solution:
Do not apply standard deviation calculations to series of identical  values.
Fix Information:

There is currently no fix planned.


KPR#: 657    Product: Open Lab Intelligence Reporter     G4635AA    A.02.04

Keyword: Client Report Rendering

One-line Description:

Changes in report templates are not immediately shown on report printouts

Problem:
Changes in report templates are not immediately  shown on the printout if the user does not preview the report and save the template first.
Temporary Solution:
Always preview the report printout after applying changes to the template.
Or save and reload the template from disk, if a preview should be avoided.
Fix Information:

There is no fix planned.


KPR#: 659    Product: Open Lab Intelligence Reporter     G4635AA    A.02.04

Keyword: Client Report Rendering

One-line Description:

Rendering error when using a template containing a large result matrix

Problem:
A template having a matrix inside a list may show a rendering error - The value 'x' is invalid. Valid values are between '0' and 'x'
This error occurs when there is large data used for rendering and due to this the matrix data spans across multiple pages.
Temporary Solution:
For large result matrices it is recommended to print the results to PDF files, rather than paper.
The paper format should adjusted to display large matrices on a single page.
Fix Information:

There is currently no fix planned.


KPR#: 18    Product: LC and CE Drivers     RC 1200.NET    N/A

Keyword: Common Infrastructure

One-line Description:

High load on CAN may cause “Spectra runbuffer overflow” entries and a loss of spectra

Problem:
If LAN is connected not to LC module generating most data (i.e. to Pump in spite of DAD's presence) then interconnecting CAN connection between modules might get overloaded. Potential consecutive loss of spectra may be observed and is flagged as “Spectra runbuffer overflow” in logbook and on report.
Temporary Solution:
Installation documentation for LC HW states that the LAN access should be in the module with the highest workload, i.e. by DAD/MWD – FLD – VWD – RID – Pump – ALS (in this order) and independent whether it is a 1100 – 1200 – 1290 – 1260 system.
Fix Information:

n/a


KPR#: 1609    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.02

Keyword: Companion

One-line Description:

Companion fails to run blank run for valved/ALS injections

Problem:
Create a run table that uses valid locations and some that use 'blank' as the injection location (a valid choice in Companion).
The runs associated with this location fail to run with a pop-up:
'Vial range must be in range 1-32'.
This also happened for an ALS method.
Temporary Solution:
n/a
Fix Information:

Fixed in Revision B.04.01.


KPR#: 1771    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Custom Fields

One-line Description:

Custom Field content might get lost when a sequence is edited

Problem:
An issie with a sequence has been observed, when it uses methods with custom fields.
All fields have been populated before the sequence started. When the sequence is edited by adding runs or when the custom fields are checked during a run, the content of the fields is wiped.
The behaviour was partially fixed in revision C.01.03. When the master methods are stored in a subfolder of the methods directory, the problem still occurred.
Temporary Solution:
For OpenLAB CDS ChemStation C.01.02 a hotfix04 is available.
Fix Information:

Fixed in revision C.01.04


KPR#: 1789    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Custom Fields

One-line Description:

Custom fields are not transferred to new result set

Problem:
When a new result set is created from a sequence with compound custom fields, the new sequence is missing the compound custom fields.
Temporary Solution:
n/a
Fix Information:

Issues fixed in OpenLAB C.01.05.


KPR#: 1795    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Custom Fields

One-line Description:

Custom Fields table erased when opening the Sequence table while running

Problem:
When you open the sequence table to add new samples while the sequence is running, the custom fields table is erased. If you don't update it while adding the new samples, the custom fields table will remain empty and you won't be able to access it in Data Analysis.
Temporary Solution:
Fixed in C.01.02 Hotfix02.
Fix Information:

Fixed in C.01.03.


KPR#: 1810    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Custom Fields

One-line Description:

Missing Sample Custom fields in the report

Problem:
When printing the final Intelligent Report, not all the fields are reported (for example the field for injection line 13 is present in the Custom Fields but not in the report).
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1828    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Custom Fields

One-line Description:

Custom field values become blank in the sequence table

Problem:
Below are the steps to reproduce the behavior:

1. Create a method with custom field setup.
2. In the custom field set up menu, keep the Default Value field blank.
3. Create a sequence and add sample lines.
4. Select the method created with the custom fields in the sequence table.
5. Click on Custom Fields button at Sequence table and input custom field values -> Click OK.
6. Sequence table -> OK button to close table.
7. Open same sequence table again and click on Custom Field button again.
8. All of custom field values are erased and it appears blank.
9. Even if the sequence is saved after entering the custom field values, the custom field values are empty.
10. The issue occurs with both mandatory and not mandatory custom values.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.03.


KPR#: 980    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.02.01 SR2

Keyword: Custom Reports

One-line Description:

Relative Retention Time solution does not work for uncalibrated peaks

Problem:
The Relative Retention Time solution available on the Agilent web site in the FAQ section only works for calibrated peaks.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 3    Product: LC and CE Drivers     RC 1200.NET    A.02.01

Keyword: DAD/MWD Configuration

One-line Description:

Installing a lamp with tag may invalidate "alway use lamp" setting in detector's control

Problem:
Installing a lamp with tag may invalidate the detector's control setting "always use lamp" and silently modify it to "use lamp only if lamp tag is available".
Temporary Solution:
Check detector's Control setting after installing a new lamp.
Fix Information:

Fixed in A.02.04


KPR#: 26    Product: LC and CE Drivers     RC 1200.NET    A.02.05

Keyword: DAD/MWD Configuration

One-line Description:

Autobalace executed although in Sequence table set to "OFF"

Problem:
1) A method XYZ.M is saved with following DAD settings: Autobalance -> Prerun
2) A sequence table is saved using the method XYZ.M and AutoBalace set to "OFF"
3) The sequence gets started and although AutoBalace set to "OFF" it is executed.
Temporary Solution:
still under investigation
Fix Information:

none


KPR#: 1356    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: Data Analysis

One-line Description:

Reprocess a sequence that is still in progress

Problem:
It is not possible to reprocess data while the sequence is still in progress.
Temporary Solution:
n/a
Fix Information:

In OpenLAB CDS ChemStation C.01.01, recalculate mode was added and samples can be reprocessed after the first run is completed.


KPR#: 1578    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Data Analysis

One-line Description:

Reprocessing copy starting as a background window.

Problem:
When using "Reprocessing Copy" from the view menu, it may happen that the reprocessing session is hidden behind the Online session.
Temporary Solution:
You can minimize the online session to access the Reprocessing Session
Fix Information:

n/a


KPR#: 1639    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Data Analysis

One-line Description:

Data System will not "Autoscale" a TCD data file to the correct peak

Problem:
Data analysis will not "Autoscale" a TCD data file.
AUTOSCALE autoscales to the first (smallest) peak. 

NOTE, there is no solvent, first peak at 0.798 minutes is very small, software identifies peak 4 at 2.158 minutes as the solvent.
Temporary Solution:
Full Scale works correctly.

Use Ranges can be made to work correctly.
Fix Information:

n/a


KPR#: 1651    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Data Analysis

One-line Description:

Summed Peaks calculation error in the Summed Peaks Report

Problem:
A Peak Summing Table containing two segments and each segment has a Use Reference for the response factor and an ISTD Reference.  The Chemstation does not report the amounts for each segment correctly.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1652    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Data Analysis

One-line Description:

Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation

Problem:
When loading GC-signals acquired with the MSD ChemStation, the Y-axis is labeled: mAu. This should be pA.

The signal units are read directly from the rawdata file "FID1A.ch"
which are written by the acquiring software.

The acquiring software shows that mAU was written.
In order to be correct this, it must be fixed by the acquiring SW.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1654    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: Data Analysis

One-line Description:

ChemStation B.04.03 [54] will not "Autoscale" the TCD data file

Problem:
There is no solvent peak and the first peak is very small so the software identifies the wrong peak as the solvent.  The Chromatogram is not autoscaled correctly.
Temporary Solution:
Full Scale works correctly.

Use Ranges can be made to work correctly.
Fix Information:

n/a


KPR#: 1706    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Data Analysis

One-line Description:

Red X mark displays in data explorer of offline data analysis

Problem:
A red X mark displays in data explorer of offline data analysis during sequence.
Temporary Solution:
During data acquisition, don't display the subdirectory in the offline data explorer window.
Fix Information:

 Fixed in revision C.01.04


KPR#: 1762    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Data Analysis

One-line Description:

Error "An additional register must be added to the list of locked registers"

Problem:
During various operations in the ChemStation (e.g. repeated manual integration of data files) a lock-up situation could occur. Internal registers were no longer accessible and the software might either no longer respond or produce error messages like "An additional register must be added to the list of locked registers"
Temporary Solution:
A fix for this issue is available for C.01.03 with Hotfix03 and any other Hotfixes released after  Hotfix03
Fix Information:

Fixed in C.01.04


KPR#: 1768    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Data Analysis

One-line Description:

Baseline not displayed properly when using ChemStation AIC’s

Problem:
A nearly horizontal baseline is not displayed or printed completely when doing a Remote Desktop connection to a ChemStation WorkStation or to a ChemStation AIC.
Temporary Solution:
None
Fix Information:

A fix for this issue is available in C.01.03 Hotfix 04 and higher

The graphics library from the ChemStation revision C.01.04, which uses the latest GDI+ API and a new GDI smoothing mode, has been implemented. A variable in the ChemStation.ini file is used to toggle between the GDI smoothing mode only or the full GDI+ features. The fix affects the appearance of chromatograms. The changes do not affect results, only the display and printing of chromatographic curves and baselines.


KPR#: 1785    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Data Analysis

One-line Description:

In Recalculate Last Result mode no reports can be printed after method modification

Problem:
As opposed to earlier workflows, it is no longer possible to change the method in Recalculate Last Result mode and create a report afterwards.
Temporary Solution:
n/a
Fix Information:

This is the intended behavior.


KPR#: 1796    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Data Analysis

One-line Description:

Report template not found when reprocessing with a master method

Problem:
When a result set is analyzed with a master method, i.e a method from the method path, it will use a report template from the result set.
The template name is taken from the method. Since the template is not copied to the result set automatically, no report will be created.
Temporary Solution:
N/A
Fix Information:

That is the intended workflow. Users who intend to use master methods might want to use report templates that are not part of the result set.
The results of the analysis with different methods can be compared in a common template in the Review mode.


KPR#: 1813    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Data Analysis

One-line Description:

Data analysis: Fractions are not displayed in UIB chromatogram

Problem:
Completed a single run with manual fraction collection.
In Chemstation Data Analysis, loaded the datafile and switched to "Purify".
The UIB1 chromatogram does not show the fractions.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1678    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Data Analysis , Navigation Table

One-line Description:

Wrong method displayed in navigation table after partial reprocessing

Problem:
When a sequence is partially reprocessed with a new method, the Navigation Table does not reflect the changed method. The changed method will only be displayed immediately after reprocessing, but not when loading the sequence set into the Navigation Table again at a later point in time. This implies that also the old sequence method is loaded when loading the affected runs with the option "use sequence method".

Reprocessing of the complete sequence is not affected by this issue.
Temporary Solution:
n/a
Fix Information:

 Fixed in C.01.03.


KPR#: 1662    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Data Analysis , SEQ

One-line Description:

PDF reports might not be updated when the file is viewed with PDF reader

Problem:
A pdf report file file cannot be updated when it is open in Acrobat reader or any other PDF viewer. It can also not be uploaded to ECM while it is open.
Temporary Solution:
Close PDF viewers before reporocessing or printing reports.
Fix Information:

C.01.03 comes with a report file viewer that works on a copy of the report file.


KPR#: 1575    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Documentation

One-line Description:

HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher

Problem:
The HTML Help requires Internet Explorer Rev. 6.0.2.800.1106CO or higher.
If a lower IE revision is used, it is possible that executing the 
'Reference' button on the top of the Help screen gives an 
error message: '...a Script Error ...'
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1734    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Documentation

One-line Description:

Understanding Your ChemStation manual needs more information about peak width calculations

Problem:
The manual needs updated to include more information on the peak width calculation. Is the peak width calculation meant to approach, approximate, or enhance the standard half-height method of calculation.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1783    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Documentation

One-line Description:

G2613A 7683A injector firmware is incorrect in the Firmware guide

Problem:
The CDS Supported Instruments and Firmware Guide does not list the correct firmware for the 7683A injector.  The correct firmware is A.10.07.
Temporary Solution:
n/a
Fix Information:

CDS Supported Instruments and Firmware Guide has been updated with the correct firmware revision in OpenLAB C.01.05.


KPR#: 1655    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Easy Sequence

One-line Description:

Easy Sequences can end up in unexpected locations

Problem:
Sequences that have been added to the queue may end up in an unexpected location on disk, when the Easy Sequence file name contains the string ".es".
The expected file name ***.es.es was not handled correctly.
Temporary Solution:
When saving an *.es file, do not include .es in the file name. The extension .es will be appended automatically.
Fix Information:

Fixed in revisions B.04.03 SP1 and C.01.03, by not allowing .es in an Easy Sequence file name.


KPR#: 1670    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Easy Sequence

One-line Description:

Queue Planner will not run sequences of differing injection source

Problem:
The Queue Planner will not run sequences using methods with different injection sources such as PAL Autosampler and valve.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1212    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

OpenLAB Option: Difficulties to lock the ChemStation during reprocessing

Problem:
During reprocessing (especially with complex reports as e.g. the Performance reports) the menu items and icon for locking the ChemStation are frequently not accessible because ChemStation is busy. The user has to wait for a point of time where ChemStation is not busy.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.05.


KPR#: 1215    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

Problems after moving items from ChemStation to a different LCDF path in ECM

Problem:
If a ChemStation master method, sequence template, or data set is downloaded from ECM and afterwards the copy in ECM is moved to a different LCDF path, the local copy in the ChemStation will not be linked to the item in ECM anymore. This will lead to errors during automatic data upload, local method/sequence update from ECM, viewing the ECM Properties, etc.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1267    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

no automatic "check-in" after deleting a checked-out method

Problem:
When a method is downloaded from ECM and available in ChemStatioon Explorer. In this case it is possible to delete the method from the local disk via the ChemStation File menu. In case the method was currently checked out, it will not be checked in automatically.
Temporary Solution:
Check in the method interactively.
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02


KPR#: 1309    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

Acquisition only runs do not write instrument and column parameters to ACAML

Problem:
Acquisition only runs do not write instrument and column parameters to the sequence.acaml or sample.acaml file (which are needed for reporting with OpenLAB Intelligence Reporter). The .acaml files are only written when data analysis is performed, too.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.02


KPR#: 1363    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

token "operatorname" uses full name including special characters

Problem:
The "Operator Name" token for the ECM LCDF uses the user's full name. This name can contain any special characters (@, &, ., etc., or local language characters). Many of these characters cannot be used for LCDF folders in ECM, which results in data not being transfered to ECM.
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1428    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

ACAML result file does not contain all peak performance parameters

Problem:
The ChemStation OpenLAB option offers the option to calculate the peak performance parameters each time a ChemStation report gets generrated.
Without having this option selected, ChemStation calculates those numbers on demand only. 
This is heavily dependant on the ChemStation report style selected (Short, Performance+Noise, Extended Performance report, etc.)

If the option is selected, the performance parameters aof a standard perfromance report are calculated and written to teh ACAML result file. However, the noise and extended performance calculation results are not generated.
Temporary Solution:
The calculation of all extended performance and noise parameters can be triggered by the following procedure:

- turn off the "Always calculate performance parameters" option in the ChemStation Open LAB configuration 

-  add the following Customized Data Analysis Command to the run time checklist of the method in use:

macro C:\Chem32\REPSTYLE\shortutl.mac; print USR_CalcPerfNoise$()

(Where c:\Chem32 needs to be replaced with the current ChemStation
installation path)

-  select a report style e.g. SHORT, that does not trigger the performance or noise calculation on it's own.  

This procedure would generate the maximum content of the ACAML result file in respect to peak perfromance parameters.
Fix Information:

This is fixed in OpenLAB CDS ChemStattion Edition Rev C.01.01


KPR#: 1449    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: ECM Integration/OpenLab Option

One-line Description:

OpenLAB Option: data not marked as modified after changing the method

Problem:
The following scenario can occur when autoamtic upload to ECM after "any data modification" is not enabled.
When the method is modified, but the data set is not uploaded when switching to the next sequence container, the modified container will not be marked as changed.
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02.


KPR#: 1453    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode

Problem:
The chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1611    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

Path definition in OpenLab Control Panel is not used in ChemStation

Problem:
The storage path that is entered in the control panel is not supposed to be the default storage location in ChemStation.
The storage path is related to EZChrom CDS projects.
Temporary Solution:
None
Fix Information:

No fix planned.


KPR#: 1629    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

Data not transferred to ECM after resuming the queue in the ChemStation

Problem:
When a Sequence or Single Injection Run cannot be transferred to the OpenLAB ECM
storage, the transfer process is paused and an error message is displayed in the
Queue Management tool for the first item in the queue.
Users with the necessary privilege can start the Queue Management tool and resume
the transfer process. After a successful transfer the item is removed from the 
queue otherwise it remains in the queue.

We discovered that items with the following error message “Object Reference not set 
to an instance of an object” appear to the end user as being transferred after 
resuming the queue but they are in fact deleted from the queue although the transfer
is failing again.
Subsequent reprocess of such sequence(s) will not pause the queue and will not be 
transferred to the OpenLAB ECM storage. 

If the option "Cleanup Data on Shutdown" is selected, the local copy of the data will
also be deleted when shutting down the ChemStation. 
In such a case, the data is lost and can't be recovered.

The error “Object Reference not set to an instance of an object” can be caused 
among others by:

  • Incorrectly configuring the ECM Authentication in the OpenLAB Control Panel
  • Manually appending text at the end of the Server string in Transfer Settings
    of the ECM Preferences
  • Using “Tokens” (Instrument Name, Instrument No, Operator or Computer Name) 
    for the Drawer and/or Folder or manually typing the Drawer and Folder in
    the Transfer Settings and not having the privilege to add folders in OpenLAB ECM.
  • Users having a project role in OpenLAB Control Panel where the privilege 
    “Save to ECM” is not granted.
  • Selecting in Sequence Parameters a Remote Data Path having a “/” in the name
   (e.g. 2011/07)
Temporary Solution:
As an immediate action and until a solution is available, we recommend on all 
OpenLAB CDS ChemStation Edition systems (Workstations, Networked Workstations and AIC’s)
connected to ECM to: 

  • Start the ChemStation Administration tool 
   (Start | Programs | Agilent Technologies | OpenLAB CDS ChemStation Edition)
  • Select the option “Use these settings for all Instances on this computer
  • DO NOT select any of the checkboxes in the Automatic Data Transfer Settings 
    and in the Transfer Management Settings.
  • Select the two checkboxes in the Audit Trail section

This will disable the transfer to ECM for all Instruments configured in the computer.
Fix Information:

A solution for the defect is currently being worked on. The fix will be made available
as soon as possible.


KPR#: 1659    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: ECM Integration/OpenLab Option

One-line Description:

The storage path for files in ECM is user-specific

Problem:
The LCDF storage path for saving methods to ECM is user specific.A user that logs on to ECM from different instruments will always save to "his" path by default.
While this is the intended behaviour, it can be confusing. Method developers would prefer a workflow that makes the path instrument specific.
Temporary Solution:
Check the folder path before uploading methods.
Consider including the instrument name in the folder name.
Fix Information:

Fix considered for a future revision.


KPR#: 1667    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: ECM Integration/OpenLab Option

One-line Description:

Cannot switch current ECM authentication to a different ECM server

Problem:
On a stand alone installation, the ECM authentication mode has been set.  Due to a hardware problem (or something else), the current ECM authentication mode must be switched to a different ECM server with a different account.
When the user tries to change the ECM authentication to a different ECM server and enters the new hostname, the administrator cannot be selected. The user is back on the previously configured ECM system.
Temporary Solution:
A workaround is to set authentication mode to none and then go to the new ECM server.
Fix Information:

n/a


KPR#: 1714    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

ChemStore legacy data doesn't get reprocessed automatically

Problem:
Migrated ChemStore data acquired with ChemStation rev. A..01.03 and earlier does not get automatically reprocessed when downloading the files from ECM.
The automatic reprocessing is requried to populate the ChemStation navigation table. The reprocessing stops with error message: 
Sequence was aborted due to a file access error. Please restart your sequence.

Restarting the sequence reprocessing does not resolve the error.
Temporary Solution:
Affected sequences cannot be reporcessed automatically, but the runs are available as single runs.
Re-calibration using the calibration samples is possible interactively. Result review can be automated by using the auto-stepping feature of recalculation mode.
Fix Information:

This defect will be fixed in ChemStation rev. B.04.03 Service Pack 1 as well as rev. C.01.04.


KPR#: 1753    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Wrong error message when downloading a method from ECM

Problem:
When a method is downloaded from ECM and this download is repeated in the same session an error message appears:

The Server Path is not identical.

This message is incorrect. One of the two path strings has a leading backslash, whole the other does not. 
There is also a warning that the method already exists. This message is correct.
Temporary Solution:
Repeat the download attempt. It works the second time.
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1774    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

Audit entry generated in the ECM or ECM Queue when doing Partial Sequence

Problem:
If the OpenLAB ECM is disconnected or not reachable. The upload from the ChemStation “Import/Commit commands” end up in the queue, the execution from the queue creates new document centric audit trail entries at the end of the upload process. These entries can't be executed successfully, remain in the queue and stop further uploads. This behavior is mainly seen with partial sequence acquisition.
Temporary Solution:
N/A
Fix Information:

This issue affects the chemStation revisions C.01.02, C.01.03 and C.01.04


KPR#: 1794    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

ECM user licenses not released when user is changed in OLCDS ChemStation

Problem:
ECM user licenses not released when user is changed in OLCDS ChemStation
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1797    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Incorrect file name suggested when uploading from Review view

Problem:
When e.g PDF files are manually uploaded to ECM, the file type is *.*
This default type is appended to the suggested file name, when the back end storage location is changed. Upload of a file with a name like " testing*.* " will fail.
Temporary Solution:
Delete the *.* extension from the file name
Fix Information:

N/A


KPR#: 1806    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

Update methods from ECM dialog locks up

Problem:
When selecting ECM -> Update Methods.. in the ChemStation Method and Run Control View, the dialog box locks up if more than a certain number of methods are linked to ECM which can be as low as 20.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1814    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: ECM Integration/OpenLab Option

One-line Description:

In file information, the Operator shows disconnected on client

Problem:
There is no distinction between a connected user and a logged-on user on the client.  In file information, the Operator shows disconnected.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1815    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

No warning before taking over session for data analysis only instruments

Problem:
For data analysis only instruments, when a session is taken over, no warning message pops up. This creates problems in the lab because all of a sudden the session is lost in the middle of work.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1818    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: ECM Integration/OpenLab Option

One-line Description:

User token in upload path is resolved to current user - not to submit user

Problem:
If user A submits multiple sequences into the queue and the user is switched to B while the first sequence executes
AND
the sequence parameter contains a user token for the upload path
THEN
the first sequence is uploaded to a path for user A (user is resolved at beginning of sequence acquisition) and all other sequences are uploaded to the path for user B (these acquisitions begin after the user switch).
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1833    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: ECM Integration/OpenLab Option

One-line Description:

In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Problem:
By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition.

The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically.

Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.
Temporary Solution:
Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.
Fix Information:

n/a


KPR#: 4    Product: LC and CE Drivers     RC 1200.NET    A.02.02

Keyword: ELSD Driver

One-line Description:

In localized CDS for ELSD detector G4218A no signal data is shown or stored

Problem:
On a localized chromatograpy data system the driver for ELSD detector throws an exception and fails to show or store the detector's signal data. On an English system the ELSD driver's operation is not affected.
Temporary Solution:
n/a
Fix Information:

For a localized (e.g. Chinese, Japanese...) system a patch for the ELSD driver is made available.


KPR#: 7    Product: LC and CE Drivers     RC 1200.NET    A.02.02

Keyword: ELSD Driver

One-line Description:

ELSD G4218A can not handle change of Sample Time during run

Problem:
With Rc.Net driver internal filters of ELSD G4218A get confused by change of "Sample Time" during run.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1185    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: GC ACQ

One-line Description:

Methods have to be in single folder

Problem:
Customer uses paths like: C:\Chem32\1\methods\customer1
and C:\Chem32\1\methods\customer2, etc.

If samples of two or more customers have to be analyzed, it is not possible to use methods from different folders.
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
Subfolders for Methods are supported in this release.


KPR#: 1600    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01 SP1

Keyword: GC ACQ

One-line Description:

Sequence stops when pushing the GC STOP key during sequence

Problem:
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence.  

This occurs on ChemStation ver.
- B.04.01 SP1
- B.04.02
- B.04.02 SP1
- B.04.03
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1608    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ

One-line Description:

FetchCPVar Cannot Talk to Application Error

Problem:
Start up Chemstation, but no instrument or sampling diagrams are shown.  
FetchCPVar Cannot Talk to Application Error occurs.

Systems use 7683 ALS.
Temporary Solution:
n/a
Fix Information:

Fixed in revision B.04.03 DSP1.


KPR#: 1626    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: GC ACQ

One-line Description:

The ability to collect partial data for specified times is needed for RC.NET GC drivers

Problem:
On the signals screen, need the ability to collect partial data for specified times.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1628    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ

One-line Description:

FetchCPVar Cannot Talk to Application Error

Problem:
Open an online session, initial or any subsequent try. This results in a benign pop-up error:  'FetchCPVar cannot talk to application'.
Temporary Solution:
n/a
Fix Information:

Fixed in B.04.03 DSP1 [54].


KPR#: 1630    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ

One-line Description:

Column compensation cannot be started from the Data System with 68xx RC.NET

Problem:
Column compensation (comp) cannot be started from the Data System with 68xx RC.NET
Temporary Solution:
Column compenstation must be started via the GC keyboard.
Fix Information:

n/a


KPR#: 1638    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: GC ACQ

One-line Description:

G1888A - Handshake Abort - Vial Status does not update

Problem:
G1888A configuration 
Set Handshake to Abort

Start Sequence
Open GC oven door to prior to end of vial equilibration time.

GC Logbook records:
G1888 - Job Stopped prematurely.

However,
Detailed Vial status never updates for vial one.
It will say Preparing for injection is 'In Process' and is stuck
at 40%.

Status will go to Ready.

ChemStation software has to be manually aborted.

An error in headspace (HS) that should abort the datasystem, no longer aborts.  

Abort does work using a 7697A.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1657    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: GC ACQ

One-line Description:

Partial storage of data is not an option for GC RC.Net drivers

Problem:
Unlike classical drivers the RC.Net drivers did not support partial storage of data for a specified run time.
Temporary Solution:
It is possible to load data file partially, by using Signal Options from the Calibration menu.
Fix Information:

Fixed in C.01.03


KPR#: 1680    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: GC ACQ

One-line Description:

Error "creation of a data object failed“ when acquiring runs with data rates 1 Hz and below with 6890 RC.NET

Problem:
For runs using data rates 1 Hz and below , you get a macro error "creation of a data object failed" when trying to load the data.

In addition, there is the following error in the sequence log during acquisition:
CP Macro     ========================================       13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     In Command deltab in macro ExecHook            13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     at line 1176 in file C:\CHEM32\CORE\SIGNAL.M>  13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     CX                                             13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     Error 16548 occurred:                          13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     An additional register must be added to the >  13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     list of locked registers.                      13:06:40 06/27/11
  70 41e0 4e086440    0
CP Macro     ========================================       13:06:40 06/27/11
Temporary Solution:
n/a
Fix Information:

Fixed in revisions B.04.03 SP1 and C.01.03


KPR#: 1683    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: GC ACQ

One-line Description:

OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed

Problem:
A manual injection is done and if the STOP on the 7820A front panel is selected, the GC stops (and resets the initial GC settings) but the chemstation continues.

When the STOP button is pushed on the 7820A, the "current logbook" reports the end of run correctly, and reports the system is running data analysis but does not go into Data Analysis right away.

If this is a 30 minute run, and the STOP button is pressed on the 7820A after 10 minutes, AND YOU WAIT 20 more minutes until the Chemstation gets to the end of the run, a report is generated.  

The system appears to hang up waiting on the expected run time.
Temporary Solution:
Workaround is to use STOP run from the software.
Fix Information:

n/a


KPR#: 1693    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: GC ACQ

One-line Description:

Pulsed Split/Splitless injection does not wait for ready 6890

Problem:
The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample prior to the time at which the system had met its defined inlet pressure.

Pulsed split/split less inlet types
Temporary Solution:
If the 6890 is operated from the GC keyboard, the GC has the expected result.
Fix Information:

Fixed in revision C.01.03.


KPR#: 1700    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ

One-line Description:

7890A stops run, ChemStation keeps going; GC Connection state is MONITOR in Status

Problem:
7890A with A.01.12.1 firmware connected to Chemstation B.04.03 and C.01.02 and when the 7890A reached the run time (10 minutes) the GC stopped and reset the setpoints, but the chemstation continued and did not stop. It is possible to abort the run.  Created a new method, same problem. Noticed that the dirty method flag was set every time even when there were no changes.

The 7890A is finished the run and GC ready state is Ready but ChemStation Status is BLUE, Run in progress, DATA ACQUISITION, but nRdy Wait.

GC Connection state shown in Status is Online MONITOR which is WRONG if the ChemStation is collecting data.
Temporary Solution:
Workaround:  Power cycle the 7890A, shut down chemstation, power on GC, restart chemstation and everything worked.
Fix Information:

This issue could not be reproduced in C.01.03.


KPR#: 1759    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ

One-line Description:

Capacity factor (k') is not calculated; the file acqres.reg on the .D directory contains no column information

Problem:
From the Configure Instrument user interface, there are configurable modules such as 7890A, 7697A, 7694B.  If the GC’s location in the “Selected Modules” table in not listed first or as module 1, k’ is not calculated.
 
For example, if the 7697 is module 1 and the 7890A is module 2 then the requests for column info would fail and k' (Capacity factor) is not calculated.  Column information will not be included in reports.
Temporary Solution:
The order in the module table is set by the order in configuration. The workaround  is to re-order configuration so that 78xx is first.
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1604    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: GC ACQ , SEQ

One-line Description:

HeadSpace sequence method will not load when it follows a Valve sequence

Problem:
The valid vial range is determined by the injection source of the loaded method, not the method in the sequence.

As soon as the HS sequence loads, it fails with:
'---------------------------
Error
---------------------------
Invalid location found in FRONT sequence, line 1
There may be others
---------------------------
OK   
---------------------------'
The Valve method has reloaded which is invalid for this HS sequence
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1741    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: GC Analysis

One-line Description:

Self-assembled result sets do not have full functionality for GC data

Problem:
A result set that is built from individual runs does not provide the full functionality for GC runs. The resulting sequence does not support front or back injections. Reprocessing is not possible, because there is a mismatch with the method.
Temporary Solution:
N/A
Fix Information:

Fixed in revision C.01.04.


KPR#: 1738    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: GC Autosampler

One-line Description:

Unclear error message: Undefined symbol: License$

Problem:
When a GC method is run that uses a vial number not valid for the configuration, an error message "Undefined symbol:  License$" appears.
Temporary Solution:
The error message is misleading. Check the vial range whether it makes sense for the selected injection source.
Fix Information:

Fixed in revision C.01.04.


KPR#: 5    Product: Unionsquare     Unionsquare    A.01.01

Keyword: Globalization

One-line Description:

The PDF report on a Japanese Windows XP system might look garbled.

Problem:
Windows XP SP3 (and earlier) is not compatible with a JIS2004 font. Since Intelligent Reporting creates PDF reports with such types of fonts, the content of those PDF files is not shown properly on such systems.
Temporary Solution:
N/A
Fix Information:

Windows XP needs to have the Microsoft fix ‘KB927489’ installed to support JIS2004 fonts properly. KB927489 can be downloaded from 
http://www.microsoft.com/japan/windows/products/windowsvista/jp_font/jis04/default.mspx


KPR#: 17    Product: LC and CE Drivers     RC 1200.NET    A.02.02 SP1

Keyword: Globalization , Sampler Configuration

One-line Description:

Flexible Cube G4227A does not work with HiP ALS G1367E

Problem:
On Chinese or Japanese CDS only the 1290 Infinity Flexible Cube G4227A does not work with High Performance Autosampler G1367E. 

On English CDS the operation of 1290 Infinity Flexible Cube G4227A with High Performance Autosampler G1367E is not affected.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1615    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Initialization

One-line Description:

B.04.03 [16], error, no string 8002, Abort this copy of Alice?

Problem:
Installed B.04.03 on a DELL connected to  instrument-1, a 6890, instrument 2, a 7890, instrument 3, a second 6890, and instrument 4 a second 7890. 

Instruments 1,2, and 4 work correctly.

Instrument 3 gives the following errors on start up,  Error "no string 8002" when starting ChemStation.  After pressing OK debug message "Abort this copy of alice? Yes is normal abort, No is Fatal Error, Cancel continues" comes. No closes ChemStation.
Temporary Solution:
Workaround is to delete all of the files in the HPCHEM\3\TEMP DIRECTORY.
Fix Information:

 Fiixed in C.01.03


KPR#: 1618    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Initialization

One-line Description:

Sporadically launch of an online session will not proceed

Problem:
Sporadically the launch of an online session will not proceed in a networked setup.
Temporary Solution:
Launch the online session a second time. The second attempt usually works.
Fix Information:

No fix planned


KPR#: 1658    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Initialization

One-line Description:

User with corresponding privilege cant't launch instrument configured on the AIC

Problem:
Users with the privileges 'View instrument or location' and 'Run Instrument' is configured in OpenLAB Control Panel can't start instruments configured on ChemStation AIC's. The following error message is displayed: " Can't launch the instrument"
Temporary Solution:
As a temporary workaround, assign the role "Instrument Administrator" or the privilege "Manage Instrument or Location"
Fix Information:

This issue is fixed with the software revision C.01.03


KPR#: 1664    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Initialization , CE Control

One-line Description:

CE G1600 instrument cannot be launched

Problem:
In ChemStation revision C.01.02, an instrument configured as G1600 CE cannot be launched. The startup process hangs during "Initialize Instrument".
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1589    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Installation

One-line Description:

missing link and no automatic installation of the localized readme.txt of the CDS ed.

Problem:
The localized readme.txt of the CDS ed. are present on the root of the CD/DVD holding the software (OpenLAB CDS ChemStation ) .
Nevertheless for the localized readme.txt  no automatic installation of the file to the harddisk will take place, the link in the programs folder links to the english readme.txt.
Temporary Solution:
n/a
Fix Information:

 Fixed in revision C.01.04


KPR#: 1627    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Installation

One-line Description:

AIC ports not opened for 7890A acquisition

Problem:
It was not possible to run any 7890A from either the client or AIC. It was discovered that communication ports were not opened upon installation of the AIC version of C.01.02. 

file "Exceptions.AgtChemRCNetAdapter.Online"
Temporary Solution:
Manually create and open a port.
Fix Information:

n/a


KPR#: 90    Product: Headspace Driver Software     G292xAA    B.01.01

Keyword: Instrument Control

One-line Description:

Methods Must be Stored on the C Drive

Problem:
The headspace sampler will run a method, but it will not run a sequence if the methods are saved to a location other than C:\ 
This defect affects all headspace hardware models running with B.0x.0x series headspace Sampler Software.
Temporary Solution:
The C Drive must be selected.
Fix Information:

Point patch available for B.04.03 SP1 [87].

This is fixed in C.01.04 OpenLab CDS.


KPR#: 821    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.02.01 SR1

Keyword: Integration

One-line Description:

Determination of tangent point on the tailing part of the peak

Problem:
In the tailing part of a peak, it may occur that the inflection point (needed for the calculation of the tangent peak width) is not correctly determined.  The problem results from the algorithm applied to find the inflection point: It is done by the first/second derivate of the peak shape but at the tail of the peak, the derivates of the "noise" on the signal may have about the same value as the tail it self. As the calculation looks for the minimum/maximum value of the second derivative, a wrong point may be identified for the inflection point.
Temporary Solution:
n/a
Fix Information:

Fiixed in C.01.03


KPR#: 981    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01

Keyword: Integration

One-line Description:

Integration unable to handle complex chromatogram for area sum integration event

Problem:
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1580    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 Hotfix-1

Keyword: Integration

One-line Description:

Peak areas may be calculated incorrectly when using "New Exponential" as tangent skim mode

Problem:
When using "New Exponential" as tangent skim mode, incorrect areas may be calculated in case of tangent-skimmed peaks. In these cases the tangent-skimmed peak will not be integrated at all, while the main peak is reported with an area that is too large. The defect is most pronounced when the tangent-skimmed peak is very small.

The likelihood of being impacted by this issue also decreases substantially with increasing number of data points collected per peak. If detector settings are chosen according to the recommendations in the ChemStation on-line help (data rate, detector peak width) the peaks are defined by 15+ data points and data collected with these methods are not affected by this issue.
The settings of the instrument specific default methods (DEF_LC.M, DEF_GC.M) take care of this, with the exception of older instrument hardware (e.g. GC 5890, 5 Hz default sampling rate), high data volumes  (LC/MSD, full scan mode), and fast chromatography techniques (Fast GC, LC Rapid Resolution).
The high information density produced by these detection techniques always require adjustment of the default detector sampling rates according to the recommendations in the ChemStation on-line help.
Temporary Solution:
Do not use "New Exponential" tangent skim mode, but choose a different mode.
Fix Information:

Point patches for software revision C.01.01, B.04.03 and B.04.02 are available
for download on
www.chem.agilent.com/EN-US/SUPPORT/DOWNLOADS/PATCHES/MULTIINSTRUMENTCHEMSTATION/Pages/default.aspx Fiixed in C.01.03


KPR#: 1642    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Integration

One-line Description:

Data lost when using area sum integration event, must have 0.06 minute space between groups

Problem:
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.
Temporary Solution:
n/a
Fix Information:

 Fixed in revision C.01.04


KPR#: 1744    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Integration

One-line Description:

Misleading display of tangent peak width

Problem:
The annotation of the tangent peak width in the performance tool is done by highlighting the part the baseline defined by the intersection with the tangents. It has been considered as misleading, because the drawing suggests that the tangent width is the length of the baseline segment. The mathematical truth is the projection of the segment on the time axis. This is what the tool calculates.
Temporary Solution:
The calculation is done correctly. 
All other peaks widths are drawn as horizontal lines. Only the tangent width is drawn following the baseline  which better describes how the tangent width is constructed.
Fix Information:

N/A


KPR#: 1800    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Integration

One-line Description:

Timed integration event "Height Reject" cannot be configured.

Problem:
In C.01.04 one new timed event appeared: Height Reject.
The expected behaviour for an event parameter would be a numeric value for the minimum peak height. Instead a selection box allows values TAN and DROP. 
The timed event Height Reject should not have been in the selection list. Do not use Timed height reject in C.01.04.
Temporary Solution:
Don't use this event. Timed Area Reject works as expected.
Fix Information:

Fixed in C..01.05 by adding the event with the expected functionality.


KPR#: 1838    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Integration

One-line Description:

GC compound peaks might be labelled as solvent peaks

Problem:
The integrator algorithm annotates many compound peaks with Peak Code S, which should only be given to solvent peaks.
Temporary Solution:
n/a
Fix Information:

No fix planned


KPR#: 1676    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Integration , Audit Trail

One-line Description:

manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on

Problem:
Manual integration may become slow, if many calibrated peaks are not found in the chromatogram and the Results Audit Trail is switched on. This behavior is especially pronounced in distributed ChemStation systems.
Temporary Solution:
Fixed in C.01.02 [17] HF03
Fix Information:

Fixed in revision C.01.03.


KPR#: 1668    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Localization,Japanese

One-line Description:

Japanese characters from saved pdf report are garbled when copied and pasted to another application

Problem:
Japanese characters from a saved ChemStation .pdf report are garbled when copied and pasted to another application (e.g. Notepad, Word, Paint, etc.).
Temporary Solution:
n/a
Fix Information:

No fix is planned for this issue.


KPR#: 1637    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.01.01

Keyword: Logbook

One-line Description:

'Method not downloaded to device' message when PAL is present

Problem:
Configure 7697A to 7890A. Close and reopen session and configure a PAL, as well.

With the headspace (HS) method still loaded, Run Method to the still cabled device. This runs, but logs message to the logbook:
'G3382AA    1 Method not downloaded to the device, because>  12:12:53 03/18/11
   0 1da9 4d838485    0
G3382AA    1  the settings are not compatible with the de>  12:12:53 03/18/11
   0 1da9 4d838485    0
G3382AA    1 vice configuration.                            12:12:53 03/18/11
  67 41e2 4d838485    0
Method       Instrument running sample Vial 1 (front)       12:12:53 03/18/11
   0 1da7 4d83848f    0
7697A      1 Sequence Started                               12:13:03 03/18/11'
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1649    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Logbook

One-line Description:

LC/MS ChemStation cannot open Logbook files

Problem:
Intermittently, the LC/MS ChemStation may not be able to open Logbook files, either through the GUI or from the "View | Logbook" menu item.  The inability to open Logbook files has been traced to an incorrect Manufacturing Date entry on the LC/MS instrument SmartCard.
Temporary Solution:
As a workaround fix, the Logbook file can be opened independently from ChemStation as a text file: in the Windows Explorer environment, browse to the C:\Chem32\1\TEMP directory and open the INSTR1.LOG file.
Fix Information:

To correct this issue, ensure that the LC/MS instrument SmartCard has the proper Manufacturing Date encoded in the proper format - an Agilent representative may be required to complete this task. Fiixed in C.01.03


KPR#: 1343    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: Method

One-line Description:

"Error occurred during loading method" after rejecting method resolution

Problem:
When a method is loaded that requires resolution, the user has the possibility to cancel loading of this method and return to the previous method.
In this case, a macro error is observed "error occurred during loading method", and the previous method is loaded, but is marked as changed.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1473    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: Method

One-line Description:

CTC portion of method printout does not include overlap settings

Problem:
CTC portion of the method printout does not include the Setup Injector Overlap settings.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1522    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: Method

One-line Description:

Revision B.04.02 SP1 [208]: Import eMethod on 6890 RC.NET driver does not work

Problem:
Description: 
1. Start GC ChemStation with 6890 RC driver
2. Select Method -> Export eMethod...
3. Generate eMethod file
4. Select Method -> Export eMethod...
5. Import eMethod file created by step3

The error message is displayed and can’t import eMethod file.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.01.


KPR#: 1593    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Method

One-line Description:

ChemStation hangs up during Edit Entire Method when starting this process via the UI

Problem:
ChemStation hangs up during Edit Entire Method when starting this process via the UI (system diagram). This happens after the Signal Details dialog.
Temporary Solution:
The problem is not observed when starting Edit Entire Method via the corresponding entry in the Method menu.
Fix Information:

Fiixed in C.01.03


KPR#: 1602    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.01.03

Keyword: Method

One-line Description:

Method printout for RID contains typos

Problem:
The printout for methods containing an RID can contain incorrect values.
For a selected peak width of 0.05 or 0.01 the report will read 0.5 or 0.001 respectively.
Temporary Solution:
No temporary fix available . This is a typo in the value list. Peak widths of 0.5 and 0.001 do not exist in the selection list.
Fix Information:

Fixed in C.01.03.


KPR#: 1624    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: Method

One-line Description:

Undefined symbol: PrDARep error displayed when selecting Export e-Method

Problem:
An error occurs when selecting “Export e-method”.  The error that is displayed on the bottom left side of software is “Undefined symbol: PrDARep”.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1643    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Method

One-line Description:

Open GC edit parameters and close it without changes; method modification flag is set

Problem:
Saved a method.  Open the GC edit parameters to review a setpoint and made no changes or changed the view.  Selected OK and the method modification flag indicates a change was made.  No changes were made.

This does not occur with the Classic Driver.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1650    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: Method

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run

Problem:
Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler sequence does not start when method is saved to the D drive

This defect affects all of the B.0x.0x series headspace and PAL Sampler software.

It is possible to run a method.
Temporary Solution:
Workaround is save and use methods from the C:\ drive.
Fix Information:

Point patch available for B.04.03 SP1 [87]

Patch has been tested with:

G2181BA ChemStore rev. B.04.02 SP2
G2183BA Security Pack   rev. B.04.02 SP2
Headspace Control Software B.01.02.1


KPR#: 1656    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Method

One-line Description:

OpenLAB CDS C.01.02 sequence table can't find currently loaded method

Problem:
Create a method. It works correctly. Save method as "TEST1 .M".  Note there is a space after the "1" and before the dot. Run the method and it works. Select Sequence, then Sequence Table and in the method name use the down arrow to list the methods, all the methods appear including "TEST1 " with a space. Select "TEST1 " and click on Inj/Vial and you get an error message. . .  Method TEST1.M (no space)  in sequence table does not exist in current method directory. 

TECHNICALLY this is true, but the chemstation changed the name and took out the blank.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1663    Product: OpenLAB CDS ChemStation Edition     M83xxAA    A.09.01

Keyword: Method

One-line Description:

No information around handling for special characters

Problem:
ChemStation online/offline handles special characters in the following way:

(1) The (dot) '.' is accepted when entered once at the end of the name. However, more than one ‘.’ at the end of the file name is not accepted.

(2) The following characters are not accepted but there is no error message: '*' and '?'. This could be because they are wildcard characters, but an error message is encountered when trying to accept '%', '<', '>', '+', '=' and ',' as special characters.

Details will be outlined in the ChemStation Manual.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1689    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Method

One-line Description:

After a single run, switching from the Method & Run Control view to the Data analysis view of the ChemStation, the method loaded in Data Analysis in different than the one used to run the sample

Problem:
A user runs a single sample. When switching from the Method & Run Control view to the Data analysis view of the ChemStation, the method name displayed in the toolbar is DEF_GC.M, for example. The method used to run the sample in Method & Run Control is not loaded in data analysis.  A different method is loaded in Data Analysis.  Methods of Method & Run Control view and Data Analysis view are kept separately.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1691    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Method

One-line Description:

C.01.01, modified method flag gets set intermittently in error when loading different methods

Problem:
The OpenLAB Chemstation is very, very slow to load a method or save a method.  The method modification flag takes sometimes 5 or more seconds to display.  So if you select save before the ‘dirty’ flag is displayed, and load another method, this newly loaded method becomes dirty (it was not the one that was edited). 

Also, when the system is busy, the menu items were not grayed preventing a user from doing an operation.
Temporary Solution:
n/a
Fix Information:

Fixed in C.01.03.


KPR#: 1694    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Method

One-line Description:

Method marked Dirty on initial "Specify Report ..." with DEF_GC.M

Problem:
This problem is reproducible if you load DEF_GC.M. Once loaded, execute 'Edit Entire Method' and check 'Data Anylsis'. Close all dialogs with OK without doing changes. Then method becomes dirty [Error!].

Once the modifed DEF_GC.M is saved, the problem does not occur.

Also happens in DA View. If DEF-GC.M is loaded and "Specify report" dialog is entered and closed without a change, then the Method dirty flag is set.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1704    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: Method

One-line Description:

Switching between Injection source sequences does not load correct method

Problem:
Install Headspace B.01.01.

Create a three line Sequence using the ALS
Run - all is okay

Create and run a three line Headspace sequence
ChemStation hangs.
ALS method displays in Headspace Status
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1710    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: Method

One-line Description:

Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa

Problem:
It is not possible to take a method from a ChemStation running one language and move it to a ChemStation running a different language (English -> Japanese, for instance) without visiting this user interface and checking / fixing any values.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1748    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Method

One-line Description:

Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified

Problem:
Removed the HS in Control Panel. Open online session and select Compare at method download. The HS is not seen as a module. Method Load Options Compare does not include the HS as a module or detect any difference in the methods.  Download to instrument and no method resolution occurs. The method is not marked as modified, though the Instr.log shows the configuration changes.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1750    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Method

One-line Description:

Cannot save offline method if configuration is different: 'Method is not consistent with the configuration'

Problem:
GC with 7693 tray and dual injectors: create and save a method for that configuration. Close session, unplug the tray, select to Configure... in Control Panel and Get GC Configuration.

Open an offline session of the instrument. User cannot save any method now. OK to the Method Editor and get:
'---------------------------

---------------------------
Method is not consistent with the configuration. Reason: Configuration mismatch:  
Tray

---------------------------
OK   
---------------------------'
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.05.


KPR#: 1765    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Method

One-line Description:

"Chemmain has encountered a problem and needs to close" when selecting back/dual inj position for GC manual inj

Problem:
On a GC system with no injectors (manual injection), select injection source from menu and select Back or Dual,  ChemMain.exe error occurs.  7820A , 7890A, 6890 have no injectors installed.
Temporary Solution:
Fix included in C.01.04 Hotfix03
Fix Information:

Fixed in OpenLAB Revision C.01.05.


KPR#: 1802    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Method

One-line Description:

C.01.03 upgraded to C.01.04: 'Edit Entire Method' not working

Problem:
When the user select any items from “Check method sections to edit” and click on “OK”, there is no action.  The dialog box remains at its position.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1671    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Method , ACQ

One-line Description:

"Specify Report" may be overwritten with the settings from another method during sequence

Problem:
If running a sequence, the "Specify Report" dialog can be opened from the dashboard. If this dialog is open for a longer time, it can happen that the sequence has already loaded the next method. If the dialog then is closed, the currently loaded method is overwritten with the other method's report settings without any warning.
Temporary Solution:
n/a
Fix Information:

This issue will be fixed in revision C.01.03: as long as the dialog is open the method is paused.


KPR#: 1665    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: MSD Analysis

One-line Description:

Deconvolution tools are not available for LC/MS

Problem:
The Deconvolution tools are not available in LC/MS ChemStation even if BioAnalysis Tools are configured for the instrument.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1675    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: MSD Control

One-line Description:

Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment

Problem:
Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the  RC.Net environment.  In the Classic Environment, Automated Delay Calibration of the MS signal for Fraction Collection is supported and accessible through the Diagnosis View from the Maintenance | Fraction Collector ... | Delay Volume Calibration ... menu item.
Temporary Solution:
To perform a Delay Calibration of the MS signal for Fraction Collection, configure the system to operate in the Classic Environment.  This may require setting some LC Modules into "emulation mode" to be recognized in the Classic Environment.  Once the Delay Calibration(s) is/are completed, make note of the Delay Calibration value(s) and revert the system to RC.Net environment if so desired.

For systems that cannot be reverted to operate in the Classic Environment to perform Delay Calibration, please contact Agilent support for assistance.
Fix Information:

This issue will be resolved in a future release.


KPR#: 1686    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: MSD Control

One-line Description:

MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled

Problem:
In MSD Tune View, it is possible to manually load a tune file from the File | Load MSD Tune File... menu item.  If the "Cancel" button is clicked in the subsequent Dialog Box, the tune file ATUNES.TUN will be loaded, regardless of what tune file was previously loaded.  The loaded tune file is displayed in a drop-down box below the menu bar.
Temporary Solution:
In the event that the tune file loading process is cancelled as in the Problem Description, please ensure that the desired tune file is subsequently loaded in MSD Tune View.  Note that this problem does not impact data acquistion, as each unique data acquistion method will load the tune file that is specified as part of the method.
Fix Information:

n/a


KPR#: 1730    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: MSD Control

One-line Description:

LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error

Problem:
After the HPLC experiences a hardware error (such as a liquid leak), the acquistion method/sequence will terminate.  However, the LC/MS  may remain in a run state.
Temporary Solution:
If the LC/MS remains in a run state after the HPLC experiences a hardware error, the ChemStation software must be closed and re-openned.  This will reset the LC/MS status to ready and bring it out of the run state.
Fix Information:

 26-Apr-2012 Susanne Kramer, Heiko Fessenmayr, Achim Teubner, Uwe Nassal, Katja Johann


KPR#: 1752    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Navigation Table

One-line Description:

Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure

Problem:
When a sequence has been run with cyclic calibration, it is not possible to change the sample type for the calibration runs to Sample.
When this is done and the sequence is reprocessed, the sequence structure is corrupted. The sequence cannot be loaded in the Navigatio Table. The Navigation Table will not be usable and the ChemStation has to be closed and reopened.
In Cyclic calibrations the structure of the sequence, the order of runs and the file naming depend very much on the sequence table setup. Later changes will cause conflicts.
Temporary Solution:
To reconstruct the original sequence the batch file needs to be removed from the sequence folder. Clear the Navigation Table and load the individual runs from the file menu. Then create a new result set from the runs using the functionality from the Sequence menu.
Fix Information:

N/A


KPR#: 1836    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Navigation Table

One-line Description:

Date Time entries can be incorrect for aborted fraction collection runs

Problem:
The column Date Time in a Navigation Table for single runs can have incorrect entries, if the run has been aborted conditions like "No valid next position".
Aborted runs do not contain the file that stores the acquisition time. The time stamp in the Navigation Table and reports is taken from the last correct run state from the instrument.
The situation has not been observed for runs that have been aborted by user interaction or hardware errors.
Temporary Solution:
Make sure to not exceed the range of available vials.
Use time date tokens for the data file name.
Fix Information:

No fix available.


KPR#: 1612    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Output

One-line Description:

Printing fails if printer name exceeds 100 characters

Problem:
Printing fails without error message, when the length of printer name plus port name exceeds 100 characters. For a TCP/IP port the printer name should not exceed 87 characters.
Temporary Solution:
Use a shorter printer name.
Fix Information:

n/a


KPR#: 1829    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Peak Purity

One-line Description:

Peak Purity Report Template is not available in Intelligent report

Problem:
Peak Purity Report Template is not available in Intelligent report
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1617    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Programming

One-line Description:

AIC can experience communication loss if Application Experience service is running

Problem:
The likelihood of an irrecoverable communication loss between instrument, AIC and OLSS will increase, when the Application Experience service is running on the AIC.
Temporary Solution:
To increase stability of communication it is recommended to disable the Aero look on the AIC, install the SP1 for Windows Server 2008 R2 and disable the Application Experience Service 
See also:  http://support.microsoft.com/kb/902196
Fix Information:

N/A


KPR#: 1755    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Programming

One-line Description:

The command "macro" does not list *.mac files

Problem:
The internal command "Macro" can be used to load a macro file from disk. Typing macro from the command line opens a dialog that points to the Chem32\Core\ folder and allows selecting *.mac and *.mcx files.
In C.01.04 *.mac files are no longer selectable.
Temporary Solution:
Use the macro name and its full location as parameter.
Example: 
Macro shutdown.mac   (equivalent to Macro C:\chem32\core\shutdown.mac)
Macro C:\chem32\core\addon\install.mac
Fix Information:

Fixed in revision C.01.05.


KPR#: 1761    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: Programming

One-line Description:

The command "macro " cannot be used to select *.mac files

Problem:
Occasionally the command MACRO that is used to select and load macro files from disk is not working properly.
In some casesonly *.mcx files could be loaded, but no *.mac files. There have been other cases where typing MACRO from the command line caused the ChemStation program to terminate.
Temporary Solution:
Use the command with the extended syntax. Rather than selecting the file from a dialog box, use the full path of the macro in the syntax.
Examples:
macro shutdown.mac
or
macro U:\Addons\mymacro.mac
Fix Information:

No fix planned on the B revisions.


KPR#: 1831    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Programming

One-line Description:

Scheduler does not work on an AIC

Problem:
The ChemStation scheduler cannot establish a connection with instruments on an AIC. 
When a command is processed, the message "Instrument not found" will be listed in the result column.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB revision C.01.05


KPR#: 27    Product: LC and CE Drivers     RC 1200.NET    A.02.05

Keyword: Pump Configuration

One-line Description:

Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster

Problem:
When a Valve is configured for Channel A in PVC the visual presentations of the BottleFillings of ChannelB, ChannelC and ChannelD are wrong.
It will not show the BottleFillings of the corresponding Channel but the BottleFilling of Channel A Valve Pos 2 etc.
Temporary Solution:
none
Fix Information:

Will be fixed with the Rc.NET Drivers Rev . A.02.06


KPR#: 32    Product: LC and CE Drivers     RC 1200.NET    A.02.06

Keyword: Pump Configuration

One-line Description:

Cannot Check “Pump and solvent selection valve installed” Box for Flex Cube if Once Clear it

Problem:
LC Driver version: A.02.06 [363] (Test application/Install on Japanese ChemStation)
1. Open Instrument configuration dialog
2. Select Flexible Cube
3. Open Flexible Cube Configuration dialog
4. Clear the “Pump and Solvent selection valve installed” check box, then press OK
5. Re-open Flexible Cube Configuration dialog
6. Thick that check box and press OK
7. Open Flexible Cube Configuration dialog again -> check mark is gone
Temporary Solution:
Workaround: use autoconfig to set up the FlexCube
Fix Information:

Fix in the next rev RC.Net A.02.07.


KPR#: 34    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: Pump Method

One-line Description:

PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)

Problem:
1. Working with a G4220A pump with ISET enabled.
2. The emulated pump was: G1312B no mixer.
3. Set up a timetable.
4. Added a G1160A valve into the system.
5. Reconfigured the instrument and setup a pump-valve-cluster.
6. During startup selected "Save instrument settings in new method".
7. Enabled ISET and selected G1311A pump this time.

-> the timetable of previous configuration still exists, but on one line "95.1%A and 5%B" is displayed instead of "95%A, 5%B
Temporary Solution:
Select Method -> New Method
Fix Information:

Solved with RC.Net A.02.07


KPR#: 11    Product: Unionsquare     Unionsquare    A.01.01

Keyword: Report Template Design

One-line Description:

Wrong labeling of standard deviation and variance

Problem:
The following defect affects the usage of statistical functions as part of aggregator functions:

When using report template editor of OpenLAB Intelligence Reporter rev. A.02.01 (with or without SP1) or OpenLAB CDS A.01.01 (ChemStation or EZChrome Edition) the results of standard deviation and variance calculations are switched.
The standard deviation function calculates the variance, whereas the variance calculation returns the standard deviation.

Note: This defect does not affect simple table column statistics, but the aggregator functions only.
Temporary Solution:
Add the variance calculation and change the field label to STDV and vice versa.
Note:  After applying the hot fix described in the next section, the report templates need to be adjusted if the workaround has been applied.
Fix Information:

This defect is fixed in 
OpenLAB Intelligence Reporter A.02.01 Hot Fix 4 and
OpenLAB CDS A.01.01 Hot Fix 2.

The hotfixes are available for download on
https://www.chem.agilent.com/en-US/Support/Downloads/Pages/default.aspx


KPR#: 19    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Report Template Design

One-line Description:

Re-calibration does no longer update the data analysis method change time stamp

Problem:
With OpenLAB CDS ChemStation A.01.04 a new time stamp for calibration table changes got introduced.
The data analysis method (DAmethod) timestamp does no longer get update by a simple re-calibration.
The introduction of the new time stamp allows a more granular tracing of DA method changes.
Temporary Solution:
n/a
Fix Information:

Existing calibration report templates need to be updated to show the calibration table modification time stamp instead of or in addition to the DAmethod change time stamp.


KPR#: 25    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Report Template Design

One-line Description:

Last displayed digits might be cut off, in cases where the number is zero

Problem:
If the number formatting scheme "1234.000" is used the selected number of digits is not shown in case the last digit is zero. The number 1.340 is displayed as 1.34, which does not contain the same information as initially intended.
This defect afects also the rounding mechanism used for peak labels.
Temporary Solution:
Use formatting scheme "1,234.000", which is not affected by the defect.

For chromatogram labeling (e.g. retention time), the expression 
FORMATNUMBER(Fields!PEAK_RETENTIONTIME.Value,3,True,False,True)
can be copied to the expression editor of the peak lable.

This syntax is taken from the RDL template file containg a result table with the correct formatting of the retention time column. The FORMATNUMBER command is not available in the expression builder, but still a valid and supported command. It can also be applied to other data fields containing numbers.
Fix Information:

There is currently no fix planned.


KPR#: 27    Product: Unionsquare     Unionsquare    A.01.05

Keyword: Report Template Design

One-line Description:

RSD calculation based on converted report item causes template corruption

Problem:
The following defect affects the Report Template Editor of OpenLAB CDS A.01.04, OpenLAB ECM Intelligent Reporter A.02.04, as well as OpenLAB Data Analysis A.01.01:

When using a string to number conversion function (val()) on a result table column, the RSD calculation on this column does not return a number.
On top of that, the report temaplate gets partially corrupted and  cannot be loaded to the Report Template Editor.

This defect mainly affects calculations using custom fields/custom parameters as these numbers are stored as string and need to converted to numbers, first.
Temporary Solution:
- Create a backup copy of the affected report template file (*.RDL
-   Open the RDL file with a text editor, like Microsoft Notepad 
-   Search for the string "/Avg(val"  and verify the expression reflects a division of a standard deviation by the average of a data field
-   Delete everything from       "/Avg(val"    until     ")*100"
- Search for the next string "/Avg(val"  and repeat the delete operation
- save the template

The corruption of the temaplte is resolved now.  As the calculation of the RSD has been modified it appears now as the absolute standard deviation in the column properties/Summary calculations dialog of the Report Template Editor.

To calculate the RSD:
- select the AVG calulation and save it as a variable, e.g. A 
- repeatt the same for the STDV. Save it as e.g. B

Add a new field under the table using the expression =B/A.  If you format the result as percentage it gets automatically mutiplied by 100.
Fix Information:

A hot fix is currently investigated.


KPR#: 620    Product: RTE     RTE    A.02.07

Keyword: Report Template Design

One-line Description:

Error message when adding a chromatogram to a report template

Problem:
The following defect may affect the report template editor of OpenLAB CDS installations on Windows XP operating systems:

Whenever adding a chromatogram to a template or loading or applying a template with chromatograms, the following error message is shown:

Could not load file or assembly 'Agilent.ChromPlot.Infrastructure, Version=1.0.0.0, Culture=neutral, PublicKeyToken=26d0ba2f646bead1' or one of its dependencies. The located assembly's manifest definition does not match the assembly reference.
Temporary Solution:
n/a
Fix Information:

Install all updates from Windows Update for the .NET Framework 3.5 SP1 and .NET Framework 2.0 Service Pack 2 on Windows XP Service Pack 3 and Windows Server 2003 Service Pack 2 , in particular the security update 2518864 (MS11-044). 
Visit  http://update.microsoft.com to download and install the mentioned updates.


KPR#: 621    Product: RTE     RTE    A.02.07

Keyword: Report Template Design

One-line Description:

Internal grouping of tables and usage of calculation variables cannot be applied at the same time

Problem:
Internal grouping of tables and usage of calculation variables cannot be applied at the same time.

In cases where the table contains an expression that is saved as a calculation variable it is not possible to define a table internal group. Whenever you setup a group it is not saved and not applied.

In cases where a grouped table gets extended by column that defines a new expression saved as a calculation variable, the expression is not calculated and the variable is not saved when previewing the report.
Temporary Solution:
In order to apply statistical calculations (average, standard deviation, RSD, etc.) on calculated results, the results need to be stored in an array rather than a single variable.
Consult the report template editor on-line on how to use category aggregators to define a result array.
Fix Information:

There is currently no fix planned.


KPR#: 622    Product: RTE     RTE    A.02.01 M3

Keyword: Report Template Design

One-line Description:

Summary Calculations cannot be applied to table columns containing calculation variables

Problem:
Enhanced calculations often require storing results in variables to be used in further calculations.
If a table contains a calculation expression with previously defined variables, the statistical summary calculations on the column return "NaN".
Temporary Solution:
In order to calculate statistics on table columns including calculation expressions with variables, the results of this line-by-line calculation need to be stored in an array for further processing.
Consult the on-line help for the usage of "category aggregator" to set-up such an array.
Fix Information:

There is currently no fix planned.


KPR#: 623    Product: RTE     RTE    A.02.07

Keyword: Report Template Design

One-line Description:

Adding a template description to a previously erased and saved description field causes a KeyNotFoundException

Problem:
Whenever a template description gets deleted from the template properties dialog a new description can't be entered if the template has been saved with the empty description field.

The user will see an error message:

--- System.Collections.Generic.KeyNotFoundException ---
   at System.ThrowHelper.ThrowKeyNotFoundException()
   at System.Collections.Generic.Dictionary`2.get_Item(TKey key)
Temporary Solution:
It is recommended to have the template description field always being populated with text.
Fix Information:

There is currently no fix planned.


KPR#: 627    Product: RTE     RTE    A.02.04

Keyword: Report Template Design

One-line Description:

Customized composite groups should have a unique name

Problem:
Whenever naming a customized composite groups with the name of an existing, pre-defined snipped, the composite group name cannot not be deleted from the tree view.
The composite group itself is deleted, but name remains under the composite group fields.
Temporary Solution:
Use an individual and unique name for a customized compound group.
Fix Information:

There is currently no fix planned.


KPR#: 662    Product: Open Lab Intelligence Reporter     G4635AA    A.02.03

Keyword: Report Template Design

One-line Description:

Filter definition does not always allow usage of OR operators

Problem:
The usage of OR operators in report filter conditions is only allowed when combining several different report parameters to one filter expression.
In cases where a report parameter may fulfill multiple filter conditions (e.g. Instrument Name = Instrument A or Instrument B), the report template editor does not allow using an OR operator.

This is a limitation due to the definition of the operators available for filtering in report rendering (Microsoft Reporting Services). http://msdn.microsoft.com/en-us/library/ms154029(v=sql.90).aspx.  In particular, there is no "OR" at all.
Temporary Solution:
The exception is the "IN" operator, which allows a FilterExpression to be compared against a list of FilterValues. 
The desired filtering may also be achieved by using wildcards (*) in combination with a CONTAINS condition.
For number, the usage of > and < conditions could be used to achieve the desired filtering.

For external report parameters (e.g. filtering for multiple sample types) the following JOIN statement could be used in the expression editor of the filter:
[=Join(Report_Parameter_0, " ").Contains(Sample_Type   1)    = True]   =  "True"
Fix Information:

There is currently no fix possible.


KPR#: 668    Product: Open Lab Intelligence Reporter     G4635AA    A.02.03

Keyword: Report Template Design

One-line Description:

Reporting time format does not offer an option of (dd/mm/yyyy and time)

Problem:
The various time stamp formats offered by the Report Template Editor does not offer the combination of dd/mm/yyyy plus time stamp.
Temporary Solution:
The format of the time stamps can be modified by using Microsoft Business Intelligence Studio or simple a text editor, like Microsoft Notepad.

- open the report temaplte file (*.RDL)
- search the string "toString"
- Modify the following section like
      ToString("MM/dd/yyyy HH:mm:ss")
 to e.g.
     ToString("dd/mm/yyyy HH:mm:ss")

- save the RDL file and validate the changes.

Note:  The Report Template Editor cannot interpret the time stamp format correctly and will show an error filed in the editor window.  However, the time stamp format is correctly applied on the reports.
Fix Information:

RTE Support of the (dd/mm/yyyy and time) time stamp format is considered for a future relaese.


KPR#: 6    Product: Unionsquare     Unionsquare    A.01.01

Keyword: Report Templates

One-line Description:

Different background colour in .pdf files generated with PDF XChange printer of ChemStation C.01.01

Problem:
In generated .pdf files from SequenceSummary_Standard.rdl and SequenceSummary_Extended.rdl by using PDF XChange printer installed with ChemStation C.01.01, a different background colour may appear in chromatography.
Temporary Solution:
Use other virtual printers  or physical printers to generate/print reports.
Fix Information:

Fix planned for future release.


KPR#: 14    Product: Unionsquare     Unionsquare    A.01.01.113

Keyword: Report Templates

One-line Description:

Data field CalibCurve_CorrCoefficient is not properly labeld

Problem:
OpenLAB CDS ChemStation and OpenLAB CDS EZCHrom populate the result field "CalibCurve_CorrCoefficient" with the coefficient of determination "r^2", rather than the correlation coefficient "r".
Temporary Solution:
Change the the data field label of your templates to
Corr. Coeff. r^2  or
r^2     or
Correlation r^2   or
Goodness of fit (r^2) like EZ Chrom does.
Fix Information:

The default labelling is planned to be changed with the next minor revision.


KPR#: 18    Product: Unionsquare     Unionsquare    A.01.03

Keyword: Report Templates

One-line Description:

Warning message concerning missing MSPGothic font

Problem:
The following defect affects OpenLAB CDS ChemStation (Japanese) only.
When loading the SequenceSummary_Standard report template, an error message indicates that the font MSPGothic is missing. By default MSPGothic is not available on Windows XP systems.
Temporary Solution:
Download the font MSPGothic from Microsoft and follow the installation instructions.
Fix Information:

There is currently no fix planned.


KPR#: 22    Product: Unionsquare     Unionsquare    A.01.04

Keyword: Report Templates

One-line Description:

Export of Short_ESTD reports to .DOC produces two extra pages

Problem:
Export of reports based on Short_ESTD.rdl to .doc format creates two unnecessary empty pages.
Temporary Solution:
Modify the Short_ESTD.rdl template, reduce the size of the chromatogram picture, and shift the report items together.
Fix Information:

There is no fix planned.


KPR#: 26    Product: Unionsquare     Unionsquare    A.01.05

Keyword: Report Templates

One-line Description:

Quantified compound markers are shown at absolute position in case of ISTD calibrations

Problem:
The quantified compound response and amount are positioned incorrectly in the related calibration curve plot. The markers are showing absolute responses, rather than relative responses.
Temporary Solution:
Do not use this feature when working with ISTD calibrations.
Fix Information:

This is currently no fix planned.


KPR#: 33    Product: Unionsquare     Unionsquare    A.01.06

Keyword: Report Templates

One-line Description:

Early Maintenance Feedback (EMF) counters may not appear on report

Problem:
Intelligent Reporting of OpenLAB CDS A.01.05 introduces the possibility to report EMF counters.
These numbers are initially available for a few Agilent instruments only, like e.g. GC 7890B.

For other instruments the instrument's firmware will need to be updated in order to enable this feature.
Temporary Solution:
n/a
Fix Information:

Please visit the technical support section of www.chem.agilent.com and browse for recent firmware updates for your instruments.


KPR#: 658    Product: Open Lab Intelligence Reporter     G4635AA    A.02.04

Keyword: Report Templates

One-line Description:

Customized chromatogram or axis lables may not display correclty

Problem:
The correct syntax for using the and "&" function requires a blank space in front and behind the ampersand sign.
Example: An expression like 
=FieldName&”Response” needs to be written like 
=FieldName & ”Response”
if it is used as a label in a chromatogram plot.

Expressions like  =Str(FieldName)&”Response” are not supported in reporting service and it results in an  “#Error” in preview.
Temporary Solution:
Use a blank space in front and behind the ampersand sign.
Fix Information:

There is currently no fix planned.


KPR#: 72    Product: OpenLAB CDS ChemStation Edition     M83xxAA    A.10.02

Keyword: Reports

One-line Description:

problem with sizing of the chromatogram in a Short Report

Problem:
Sizing of the chromatogram output for a Short report does not work in the 20 - 40% range for the Response axis. The 40 - 100% range works correctly, values smaller than 20% are not allowed.
Temporary Solution:
n/a
Fix Information:

This issue will not be fixed in the "classic" reporting from the ChemStation.
OpenLAB CDS ChemStation Rev C.01.01 introduces a new Reporting engine "Intelligent Reporting".


KPR#: 276    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.01.01

Keyword: Reports

One-line Description:

Chromatogram Output in reports: meaning of "Portrait" and "Landscape"

Problem:
Edit the Specify Report section of a method. In the Chromatogram Output, you can select e.g. "Portrait" or "Landscape".
It would be expected that "Landscape" means that the chromatogram is printed in landscape format. But actually it means that is printed rotated by 90 degrees in comparison to the orientation of the printer setup. I.e., if you set Landscape in File -> Printer Setup, the chromatograms are printed in Portrait.
Temporary Solution:
n/a
Fix Information:

It is not planned to change this behavior in the classic ChemStation reporting engine.


KPR#: 1011    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.02

Keyword: Reports

One-line Description:

ChemStation does not report resolution as defined in European Pharmacopoeia

Problem:
ChemStation does not report resolution as defined in European Pharmacopoeia. EP requires the following formula to be used:
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)


Only the following definitions are available in ChemStation for resolution:

USP, ASTM (in Extended Performance Report):
R = 2 * (Tr(b)-Tr(a)) / WB(b)+WB(a)

Half-width method as used in Performance Report:
R = (2.35/2) * (Tr(b)-Tr(a)) / W50(b)+W50(a)

JP (in Report Layout view only):
R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)
Temporary Solution:
In Report Layout view, use Resolution (JP) which uses the same formula.
Fix Information:

Fixed in revision C.01.04


KPR#: 1324    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.02

Keyword: Reports

One-line Description:

Sequence summary statistics with different methods

Problem:
If a sequence includes runs with another method apart from the method that should be used for statistics calculation (e.g. a wash method), the statiscal calculations in the sequence summary report cannot be performed for all the runs. The statistical calculations will be done separately for each section of the sequence between the runs with the second method.
Temporary Solution:
Mark the second method as "Acquisition only".
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1377    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: Reports

One-line Description:

it is not possible to print report with chromatograms that have been overlaid from different runs

Problem:
It is not possible to print report with chromatograms that have been overlaid from different runs.
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1440    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01 SP1

Keyword: Reports

One-line Description:

acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off

Problem:
The acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off. Only the run-time checklist, method information and data analysis parameters are reported.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1572    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

Batch mode does not allow usage of intelligent reports

Problem:
If user works in batch mode and the batch method is setup to use intelligent reports, this selection is ignored, even if the batch options are set to use the batch method s report settings.
Temporary Solution:
N/A
Fix Information:

There is no plan to fix this issue in future.


KPR#: 1579    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

Report Templates newly added to a Result Set do not appear in the list of available templates

Problem:
When you add more than one report template to a result set, only the seletcted template in the browse dialog will be visible in the Specify Report list box under Report Templates or Report Templates for calibration samples
Temporary Solution:
After closing and reopening the Specify Report menu, all Report Templates added to the result set are visible.
Fix Information:

n/a


KPR#: 1581    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

Extended performance report (IR) reports "Infinity" for plates per meter

Problem:
In case when no column information is available report shows"infinity" instead of the expected "--".
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02


KPR#: 1582    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

Mismatch in drawing of peak performance between Classic Report and Intelligent Report

Problem:
The peak performance drawing in the Classic Reporting and in the Intelligent Reporting are slightly different. Thisi is only a graphical representation and this is not affecting the performance
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1585    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

System may hang when ".DIF" output is choosen as report type.

Problem:
Due to a memory leak in the DIF file creation component, when long sequences are running and the ".DIF" output is choosen as report type, the System may hang and abort running instruments. When this occur, the computer needs to be rebooted.
Temporary Solution:
This sporadic hang can be prevented by not selecting ".DIF" output.
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02


KPR#: 1614    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 DSP1

Keyword: Reports

One-line Description:

Calibration Peak Areas included in the Area sum

Problem:
Calibration Peaks are counted in the Area sum which causes the results to be incorrect.
Temporary Solution:
n/a
Fix Information:

 Fiixed in C.01.03


KPR#: 1619    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Reports

One-line Description:

Intelligent Reports do not contain ECM specific information in Review mode

Problem:
ECM specific fields like path, server, operator or version number are not printed in the Intelligent Report templates in the Review mode.
Temporary Solution:
Print the report from the Data Analysis view.
Fix Information:

No fix available.


KPR#: 1634    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: Reports

One-line Description:

B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work

Problem:
B.04.03, data analysis, load a data signal  (demo\100FID20 for example) and select Graphics. Select Use Ranges and set the time range to 0 to 6 minutes and the response range to 0 to 100 pA for example, then OK.  Time scale works (6 minutes) but the response range does not work.
Temporary Solution:
WORKAROUND 
Even though this is a single chromatogram, in Multi Chromatogram, LAYOUT select separated, and SCALE, select All the same scale and it works.
Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#: 1645    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

No Chromatogram displayed on Intelligent Report for GC

Problem:
When intelligent report is used for report style for 6890 acquisition, the result chromatogram is not displayed.
Temporary Solution:
n/a
Fix Information:

Fixed in Revision C.01.02.


KPR#: 1673    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

CSV, DIF, and EMF files are not generated for some Classic report styles

Problem:
The CSV, DIF, and EMF files are not generated for some Classic report styles:
CSV and DIF files are not created for Library Search and Performance related report styles.
For the Extended Performance report style, also the EMF report is not created.
Temporary Solution:
n/a
Fix Information:

It is not planned to fix this issue.


KPR#: 1679    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01

Problem:
The sample name is not reported when using Intelligent Reporting when working on data that have been required in revisions prior to OpenLAB CDS ChemStation Edition C.01.01.
Temporary Solution:
Use the "result set migration" tool of the Chemstation, to migrate the sequence to container format.
Fix Information:

 Fixed in revision C.01.04


KPR#: 1690    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

Resolution of chromatograms in Intelligent Reporting is low

Problem:
The rendering for chromatograms to be printed in Intelligent Reports provided a low resolution. While this makes rendering faster, the level of detail in longer chromatograms was not high enough.
Temporary Solution:
A Hotfix for ChemStation C.01.02 has been released.
Fix Information:

Fixed in C.01.03.


KPR#: 1703    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

ChemMain.exe can crash when a long file path is used

Problem:
The file path in ChemStation has a hard coded limit of 256 characters.
In an ECM environment there are also hidden characters added to ensure unique file naming. As a consequence the complete lenth for  +  +  +  must not exceed a length of 175 characters.
Temporary Solution:
Make sure that the path length of a report from  -->  C:\Chem32\... to  ...pdf  <-- does not exceed 175 characters.
Fix Information:

No fix available.


KPR#: 1705    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

Chinese/Japanese characters cannot be searched in the Report Viewer

Problem:
The report viewer that was introduced with C.01.03 allows searching for text within a report.
Chinese and Japanese characters, however, cannot be searched.
Temporary Solution:
No fix available.
Fix Information:

n/a


KPR#: 1722    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Reports

One-line Description:

Tangent peak width not correctly evaluated.

Problem:
The peak performance parameters have been modified in ChemStation revision B.04.03 to provide better peak width evaluation for chromatograms with high data rates.
This ne algorithm sometimes fails to find the correct inflection points. As a result peak width at baseline (tangent peak widths) can be reported incorrectly. Consequently parameters calculated based on this peak width (tangent resolution and tangent plates) are incorrect as well.
Temporary Solution:
A patch for B.04.03 is available on request.
Fix Information:

Fixed in revision B.04.03 SP1.


KPR#: 1733    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Mass- Based Purification (Fraction Collection) report missing information & shows inaccurate Fraction Tick marks (RC.NET only)

Problem:
The Mass-Based Purification (Fraction Collection) report is missing information in RC.NET mode; this same information is available in Classic mode.

The "Mass" column in the report should specify which detected mass(es) triggered the fraction collection; this feature is currently available in Classic mode only.

Additionally, the Fraction Tick Marks displayed are incorrect as is the BeginTime & EndTime information in the report Fraction Information section. The Fraction Tick Marks do not line up accurately with the collected fractions triggered by the LC/MS; these values have the MS Delay Time added to them, which is incorrect.
Temporary Solution:
To determine which masses triggered Fraction Collection or what the actual Fraction Collection BeginTime and EndTime values are, open the affected  ChemStation data file (.D folder) and open the file MBFC.ini.  This MBFC.INI file contains the Mass-Based Fraction Collection triggering events, including Mass, Collection ON, and Collection OFF information.  Do not modify this file.

The format of MBFC.INI must be interpreted linearly.
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1743    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Reports might show different calibration time stamps for the same method.

Problem:
In Intelligent Reporting two different variables can be used to report the time of the last recalibration.
The two variables are populated in different locations in the code. Both actions should happen a few milliseconds apart, which can lead to a (rounded) difference of one second.
Temporary Solution:
None
Fix Information:

N/A


KPR#: 1751    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles

Problem:
ChemStation / Data Store integration:  the "eSignature data file" privilege is missing from the ChemStation lab manager, ChemStation analyst, and ChemStation operator roles.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1754    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

The structure of the reports to file has changed

Problem:
The structure of the reports in xls, csv and dif format has changed.
New columns have been introduced to hold additional information provided since C.01.04.
Temporary Solution:
Users who perform their own spreadsheet calculations based on those reports need to be aware that the order and content of columns is different.
Fix Information:

N/A


KPR#: 1756    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: Reports

One-line Description:

CP3: Number format error in: %d when printing report to screen in Data Analysis

Problem:
Completed a single run and a report was printed to screen.

Loaded a different method in DA
Select Report > Print Report
No report is printed to screen
Error shown in left hand corner: CP3: Number format error in: %d
Temporary Solution:
The problem was resolved by doing a clean uninstall and new installation.
Fix Information:

Not reproducible.


KPR#: 1757    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Error 41265 occurred at end of run; Number format error in: %d shown in logbook

Problem:
Method run completed. The following error message appeared at the end of the run in the Logbook.

CP Macro     ========================================       14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     In Command sprintf$ in macro GetRunPdfRptFil>  14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     eNameRoot$                                     14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     at line 196 in file C:\CHEM32\CORE\REPORT.MC>  14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     X                                              14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     Error 41265 occurred:                          14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     Number format error in: %d                     14:46:33 02/13/12
  70 41e0 4f396899    0
CP Macro     ========================================       14:46:33 02/13/12

Default report setting was used.
Temporary Solution:
The problem was resolved by doing a clean uninstall and new installation.
Fix Information:

Not reproducible.


KPR#: 1758    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

Result data sent to OpenLAB ECM Intelligent Reporter may not be updated.

Problem:
When reprocessing data acquired with ChemStation B.04.0x or earlier, the result data result data sent to OpenLAB ECM Intelligent Reporter may not be updated. 
This is onlz the case if ChemStation operates with automated ECM upload after reprocessing. The interactive "Save data to ECM" function is not affected.
Temporary Solution:
Untill a point patch gets available it is recommended to use the interactive  "Save data to ECM" function an addition to the automated ECM upload.
Fix Information:

This defect is fixed in OpenLAB CDS ChemStation C.01.04.
 A fix for this issue is available for C.01.03 with Hotfix02 and any other Hotfixes released after  Hotfix02


KPR#: 1775    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: Reports

One-line Description:

Preview of classic reports fails with a macro error. 'Number Format error' in Sequence log file.

Problem:
Doing preview of a report with the method settings to screen only works fine for 25 days. After that, the number used in the filename of  the temporary.pdf report file shown in the report preview is getting larger than MAX int (2.147.483.647) and causes a formatting error during the conversion to a string value. After again about 25 days it starts over with 0 and works again for 25 days.
Temporary Solution:
The number used for the temp file name is the windows OS tick count, and it gets larger than MAX int after 24 days. It is reset after every reboot.
Fix Information:

This issue is fixed with C.01.03 [45] HF06. Fixed with C.01.05. Fixed with B.04.03 SP2.


KPR#: 1776    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

Chromatogram in On-Screen Classic report is not displayed properly

Problem:
On systems accessed via  Remote Desktop, the on screen classic report may not be displayed properly. Part of the baseline, x-axis or y-axis labels may be missing. This is only affecting the chromatogram in the on-screen classic report. The printed report or display in the data analysis view will correctly show the baseline, axis markers and labels.

Affected revisions: C.01.03 and C.01.04.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB revision C.01.05.


KPR#: 1778    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: Reports

One-line Description:

'Number Format error' in Sequence log file after printing report to screen

Problem:
An error message appears when a report  is printed to screen on computers that have not been rebooted for more than 24 days.
Temporary Solution:
For the OpenLAB CDS ChemStation C.01.03 a hotfix 06 is available.
Fix Information:

Fixed in B.04.03 SP2. Fixed in C.01.05.


KPR#: 1779    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.01.03

Keyword: Reports

One-line Description:

Peaks are calculated by area only or height only in a report

Problem:
It is not possible to report area for one peak and height for another peak in one report.  It is one selection for all compounds.
Need to be able to select height or area for each compound.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1780    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Sequence summary report sorts the data by injection time

Problem:
In case of self-assembled result sets, the order of acquisition sequence lines numbers might appear in a random order. This is caused by the default sorting criteria, which is the injection time.
Temporary Solution:
Customize the report template and adjust the sorting criteria to your needs and preferences.
In case of self-assembled result sets, the e.g. processing time stamp is a suitable sorting criteria.
Fix Information:

There is no fix planned.


KPR#: 1786    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence

Problem:
In IR Sequence Summary reports always the full sequence is reported, even in partial sequence.
Classic SSR reacts on the selection of partial sequence.
Temporary Solution:
To have only a subset of the sequence in the IR use the Review mode or create a self-assembled result set.
Fix Information:

n/a


KPR#: 1791    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Macro error occurs (Parameter 2: Table not found) when printing a calibration table

Problem:
Macro error occurs (Parameter 2: Table not found) when printing a calibration table
Temporary Solution:
n/a
Fix Information:

Fixed in Hotfix03 for C.01.04.


KPR#: 1801    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

C.01.03: XML results: The 'ESTD %' and 'ISTD %' result types in the Quant Calc node are incorrectly 'ESTD' and 'ISTD'

Problem:
ChemStation C.01.03: XML results:  The 'ESTD %' and  'ISTD %' result types in the Quant Calc node are incorrectly reported as 'ESTD' and 'ISTD'. This is confusing the receiving LIMS system. The printed reports have correct titles.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1803    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Acquisition method name is not displayed in an intelligent report

Problem:
Acquisition method name is not displayed in an intelligent report
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1820    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Missing Report Style Options for LCMS and LCMS Qualitative

Problem:
Report Style LCMS and LCMS Qualitative:  Missing Report Style Options:

-Background Subtract
-Spectrum Average Peak
-EIC
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1821    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Reports

One-line Description:

Classic detailed report does not show calibration curves after encountering a compound with invalid calib curve

Problem:
The calibration user interface shows the calibration curve of compounds.  But on the detailed report, the printing of the calibration curves is stopped when one compound with an invalid calibration curve is encountered.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1832    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: Reports

One-line Description:

ChemStation freezes while generating report preview

Problem:
The ChemStation stays busy and the status line says: Calculating results.
The application hangs for at least 10 minutes and needs to be terminated.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.02.


KPR#: 1834    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: Reports

One-line Description:

Calibration data modification time stamp is missing on reports printed from review view

Problem:
The Short_ESTD.RDL report template (and all related rdl templates) contains a data field "Calibration data modification time"
which is not printed if reporting is triggered from the review view.
This time stamp is not stored in the result file but it is the change time stamp of the currently loaded method in the Data Analysis view. Hence, this time stamp is not available in the review view.
Temporary Solution:
n/a
Fix Information:

OpenLAB CDS ChemStation C.01.04 introduced a new compound specific and more granular time stamp  "CalibrationCurveLastModifiedDate". This time stamp gets stored in the result file and remains with the result.
Report templates need to be adjusted in order to print this new time stamp.


KPR#: 1835    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Reports

One-line Description:

Calibration data modification time stamp is set-off by 12 hours

Problem:
The Short_ESTD.RDL report template (and all related rdl templates) contains a data field Calibration data modification time.
The time stamp always shows the appendix AM, even in case recalibrations processed in the afternoon.
Switching the time format to 24:00 does not correct the problem.
Temporary Solution:
N/A
Fix Information:

This defect if fixed in HF 7 for  OpenLAB CDS ChemStation C.01.03 and OpenLAB CDS ChemStation C.01.04 and higher.

OpenLAB CDS ChemStation C.01.04 also introduced a new compound specific and more granular time stamp  "CalibrationCurveLastModifiedDate". This time stamp gets stored in the result file and remains with the result. Report templates need to be adjusted in order to print this new time stamp.


KPR#: 1837    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: Reports , Custom Fields

One-line Description:

Empty Custom Fields are not printed in Intelligent Report

Problem:
If the content of a Custom Fields is empty this Custom Field will not appear in an IR at all.
Temporary Solution:
Use default values during CF setup.
Fix Information:

No fix Planned


KPR#: 1595    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Reports , Data Analysis

One-line Description:

report printed to screen hides behind the chromatogram window when DA.M is loaded

Problem:
When a report is printed to screen, it hides behind the chromatogram window as soon as the individual method DA.M of a data file is loaded. This could occur e.g. at the end of reprocessing when the review option "Use Method from Data File" is selected. In this case, DA.M is loaded again directly after the end of reprocessing.
Temporary Solution:
n/a
Fix Information:

 Fiixed in C.01.03


KPR#: 1620    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Reports , Data Analysis

One-line Description:

incorrect plate numbers calculated under certain conditions when using tangent method

Problem:
Under certain conditions the number of theoretical plates according to the tangent method may be calculated incorrectly in ChemStation B.04.03. This may especially occur when the peak has a rather wide baseline compared to the peak width or when the peak exhibits distortions at the end of the peak.
Temporary Solution:
n/a
Fix Information:

Fiixed in C.01.03


KPR#: 1681    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: Reports , MSD Control

One-line Description:

Incorrect Polarity Switch Delay parameter listed in ACQ.TXT

Problem:
The Polarity Switch Delay parameter is incorrectly recorded in the ACQ.TXT file that is generated with the run.  Only the default parameter of 50 msec is shown for 6130, 6140 and 6150 systems.  This gives the impression that the method values are being ignored.  
However, when Hyperterminal was used to query the "MPST" value directly, it was found that the correct Polarity Switch Delay value was being used.  Therefore, the correct Polarity Switch Delay parameter is being used during acquition; the recorded value is simply incorrect.
Temporary Solution:
N/A
Fix Information:

To be fixed in a future release. Fixed in C.01.03.


KPR#: 1715    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Reports , SEQ

One-line Description:

Sequence Summary Report does not do statistics when the first compound is not present

Problem:
When the first compound in a calibration table is not found, the sequence summary report does not perform any statistics for all the other compounds.
Temporary Solution:
Reporocess with a method that does not have this compound in the calibration table, or with a method that uses a manual integration event at the expected retention time.
Fix Information:

Fix planned for C.01.04. Fixed in revision B.04.03 SP1.


KPR#: 1597    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: Run Control

One-line Description:

No other Windows application can run, when online ChemStation is busy

Problem:
It has been observerved under Windows Vista and Windows 7 that during  acquisition of large sequences the system is so busy that no other application would start. The ChemStation acquisition was still active in these cases.
This has been attributed to the Windows Explorer that is assigning resources to the ChemStation files.
Temporary Solution:
The Windows Explorer needs to be closed from the Windows Task Manager. 
Close the Windows Explorer before starting long sequences. Disabling File Indexing can help.
Fix Information:

Cannot be fixed from the ChemStation side.


KPR#: 5    Product: LC and CE Drivers     RC 1200.NET    A.02.02

Keyword: Sampler Configuration

One-line Description:

Control of Tray illumination missing for G1329B & G4303A

Problem:
RC.Net driver for high performance autosampler does not present possibility to switch on or off tray illumination of G1329B or G4303A.
Temporary Solution:
On ChemStation's command line enter:
print SendModule$(LALS,"LGHT 1")
to switch on sampler's illumination and
print SendModule$(LALS,"LGHT 0")
to switch it off.

In ChemStation the use of a configuration with 'Cassic' drivers might also be an option.
Fix Information:

fixed in A.02.02 SP1


KPR#: 6    Product: LC and CE Drivers     RC 1200.NET    A.02.02

Keyword: Sampler Configuration

One-line Description:

Parameter mismatch for custom-defined Well Plates

Problem:
The parameters that were offered to define customized well plates did not intuitively match the dimensions of a well plate.
It was not possible to define whether a plate contains wells or vials.
This could also lead to errors in the injected volume.
Temporary Solution:
No temporary fix available.
Fix Information:

fixed in A.02.02 SP1


KPR#: 29    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: Sampler Configuration

One-line Description:

AFC Cluster: Edit Wellplates Dialog does not display imported wellplates.

Problem:
Edit Wellplates Dialog does not display imported wellplates.
Temporary Solution:
In Edit wellplates dialog Import some custom specific wellplates. 
Click on Imprt, browse to saved *.wlp file, click on open. 
File browser closes. 
Imported wellplate definitions are not visible.
Leave Edit Wellplates Dialog with OK. 
Reopen dialog. 
Now the imported items are visible.
Fix Information:

n/a


KPR#: 36    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: Sampler Configuration

One-line Description:

Use Vial as Fraction collections

Problem:
Started a run with fraction collection.

The fraction collector config does not show 10 vials but only trays.
Temporary Solution:
If you have installed a 2 WP-Tray, vials are allways forbidden to use for fractionation, regardless if there are wellplates inserted or not.
Fix Information:

Issue fixed with LC and CE Driver revision A.02.07


KPR#: 9    Product: LC and CE Drivers     RC 1200.NET    A.02.02 SP1

Keyword: Sampler Configuration , Sampler Status

One-line Description:

Control of Tray illumination missing for new G1329A

Problem:
New G1329A ALS contain a tray illumination now. RC.Net driver for standard autosampler does not present possibility to switch on or off tray illumination of these new G1329A ALS.
Temporary Solution:
Check if G1329A is a new one and already has tray illumination.

On ChemStation's command line manually enter:
print SendModule$(LALS,"LGHT 1")
to switch on sampler's illumination and
print SendModule$(LALS,"LGHT 0")
to switch it off.

In ChemStation the use of a configuration with 'Classic' drivers might also be an option.
Fix Information:

n/a


KPR#: 16    Product: LC and CE Drivers     RC 1200.NET    A.02.02 SP1

Keyword: Sampler Driver

One-line Description:

Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation

Problem:
The following problem occurs in Japanese and Chinese ChemStation only:

RC.Net drivers for Agilent 1260 Infinity Standard Autosampler G1329A/B or Agilent 1260 Infinity SFC Standard Autosampler G4303A do not provide any possibility - neither on ChemStation's menu nor on sampler's context menu - to switch on or off tray illumination inside G1329A/B or G4303A.

Control of Agilent 1260 Infinity Standard Autosampler G1329B by 'classic' drivers of ChemStation is not affected.

Agilent 1260 Infinity SFC Standard Autosampler G4303A requires the RC.Net Driver. No 'classic' driver for it exists.
Temporary Solution:
On ChemStation's command line enter:
print SendModule$(LALS,"LGHT 1")
to switch on sampler's illumination and:
print SendModule$(LALS,"LGHT 0")
to switch it off.
Fix Information:

The issue will be solved in driver version A.02.03.


KPR#: 15    Product: LC and CE Drivers     RC 1200.NET    A.02.02 SP1

Keyword: Sampler Driver , Sampler Status

One-line Description:

Larger injection volume with 900µl head may be incorrectly displayed after injection

Problem:
With G1329A/B, a 900µl head installed, and a large injection volume specified the RC.Net Drivers may display an irregular injection volume after the injection. Nevertheless the volume, which got injected, correctly represents the method's setpoint. It is a problem affecting only the presentation after the injection cycle. Neither the injected volume itself nor the display during the injection cycle are affected. For the duration of the injection cycle itself the correct injection volume is shown, and injected.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 31    Product: LC and CE Drivers     RC 1200.NET    A.02.04

Keyword: Sampler Method

One-line Description:

ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.

Problem:
Methods including pretreatment/injector steps are not downloaded to the instrument, prior to a run.
Temporary Solution:
n/a
Fix Information:

The defect is fixed with the Driver Release A.02.06

In relation to the Instrument Control Framework, a PointPatch is available.
Please contact your Agilent Support Team.


KPR#: 33    Product: LC and CE Drivers     RC 1200.NET    A.02.07

Keyword: Sampler Method

One-line Description:

Dua Loop Sampler: ACQ method shows wrong loop with overlap injection

Problem:
Pre Cond.: Overlapped injection wit overfill by factor with a DL with different loop sizes for upper and lower loop
Consequence: due FW automatism it happens that we will report the wrong loop and the wrong injection volume

Wrong injection volume reported in the case of 2 different loops
(does not make sense in the application).
A logbook entry: 
"Warning: Used injection volume can be different then specified in the method due to different loop size."
Temporary Solution:
no fix
Fix Information:

n/a


KPR#: 778    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.01

Keyword: SEQ

One-line Description:

Change of sample type during sequence acquisition is ignored

Problem:
If you change the sample type of a run (that has not yet been executed) of a running sequence, the change in sample type will be ignored.
Temporary Solution:
Change the sample type after the sequence has been completed and reprocess the sequence data.
Fix Information:

n/a


KPR#: 1247    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.02.01

Keyword: SEQ

One-line Description:

Method changes not used after pause/edit/save/resume sequence using unique folder ON

Problem:
During the sequence, select pause.
After the pause, open edit parameters and increased the oven hold time to 3 minutes and save the method.  

Resume the sequence. The method changes are not used when running with the Unique Folder ON.
Temporary Solution:
n/a
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
Changes made to the method during a pause of the sequense, will be applied to the next sample after resuming the sequence.


KPR#: 1317    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: SEQ

One-line Description:

Mismatch for sequence completed time stamp in sequence and instrument logbook

Problem:
The time stamps for sequence completion may differ between the sequence logbook and the instrument logbook:
The sequence logbook lists the time after completion of the acquisition, while the instrument logbook displays the time after completion of the sequence summary report.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1443    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: SEQ

One-line Description:

ChemStation crash when trying to open Partial Sequence with non-exisiting data path in the Preferences

Problem:
The following problem occurs when there are data paths specified in the Preferences dialog that no longer exist.
When opening the Partial Sequence dialog in this situation, ChemStation will crash.
Temporary Solution:
In the Preferences, remove all non-existing data paths.
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1514    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

The path to a data file has a limited length

Problem:
The current path that points to files inside a data folder has a limit of 260 characters.
With a data file name of 18 characters (16 + .d) and the known files inside a data folder the effective length of a data path, including the container name cannot exceed 170 characters.
Temporary Solution:
None
Fix Information:

No fix planned


KPR#: 1548    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.03.02

Keyword: SEQ

One-line Description:

Cannot edit Method in Offline session while it is running in an online sequence

Problem:
In G2070BA, if you have the sequence running with Method A, it is not possible to edit Method A in the Offline session and save it so it can be used for the next sample in the sequence table.
Temporary Solution:
n/a
Fix Information:

Fixed in Revision C.01.01.


KPR#: 1571    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: SEQ

One-line Description:

Missing Vial during sequences causes CTC ChemStation to hang

Problem:
Create a five line sequence using Add-on CTC.
Run sequence, removing vial three.

CTC will report missing vial and it is shown in the log book for CTC.

CTC will reset and go back to ready state, however, ChemStation will hang.

All menu items are grayed out and you cannot touch anything on the system.
The ChemStation must be ended in Task Manager.
Temporary Solution:
n/a
Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#: 1603    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.01.01

Keyword: SEQ

One-line Description:

Cannot run partial sequence acquisition with headspace samplers

Problem:
Partial sequence acquisition with headspace samplers has not been possible
with GC ChemStation.   

Running a partial sequence in the ChemStation with a Headspace sampler
processes the entire sequence instead of a partial sequence
Temporary Solution:
Beginning in headspace software revision B.01.01 and ChemStation revision B.04.03 DSP1 [54], a point patch is available to support partial sequence with the ChemStation when methods and sequences are saved to C:\.
Fix Information:

Fixed in Revision B.04.03 SP1 [87] and Revision C.01.03 [37].


KPR#: 1636    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run

Problem:
Save method on D drive, and start sequence with headspace or PAL Sampler.
    GC stays in "Prep Run".
    headspace and PAL Samplers do not work.

Headspace or PAL Sampler does not start when method is saved to a location other than C:\

This defect affects all headspace hardware models with the B.0x.0x series headspace and PAL Sampler Software.
Temporary Solution:
Workaround is to save and use methods from the C:\ drive.
Fix Information:

Fixed in revision C.01.04


KPR#: 1640    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Data system uses the currently loaded method to validate the vial range in the Sequence table

Problem:
Sequence does not run or will abort if the currently loaded method vial range is different than the vail range of the next method in the sequence table.

Set up two methods:
Method E with injection source set to headspace (valid vials 1-111)
Method A with injection source set to ALS sampler with tray. (valid vials 1-150)

Construct a sequence
Vial      Method
1            A
1            E
130        A
2            E

If method E is the currently loaded active method then this sequence will not validate because the first method in the sequence (method A) has a vial range that is valid for an ALS.  The vial numbers are checked against method E which has a vial range for the headspace. The sequence cannot be run.

If method A is the currently loaded active method, then this sequence will validate and start running. When the datasystem attempts to move to the 3rd line it validates the vial against E with a headspace injectorID and fails, aborting the sequence.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1648    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Sequence parameter subdirectory overwrites sample info parameters during run

Problem:
The sequence parameters (i.e subdirectory, file name, vial location and sample name) overwrites the sample info screen with sequence parameter information during the sequence run only.  

The Sample Info screen goes back to its original settings once the sequence finishes.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1687    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02 SP1

Keyword: SEQ

One-line Description:

Missing ‘Reprocessing Only’ in the Dropdown List for Unique Folder Creation OFF

Problem:
1. Start ChemStation (Online)
2. Select View -> Preferences -> Sequence tab
3. Select Unique Folder Creation OFF
4. Select Sequence -> Sequence Parameters
5. Click and open the dropdown list in the part of methods to run field
“Reprocessing Only” is missing.
Temporary Solution:
n/a
Fix Information:

Fixed in revision B.04.03.


KPR#: 1696    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: SEQ

One-line Description:

Acq method for 7890A back injector is not displayed in data analysis view

Problem:
Right click the data on navigation table and select “View ACQ method”
ACQ method can’t be displayed.

For dual injection method, the ACQ method for front data is displayed, but the Acq method for back injector is not displayed.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1699    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Missing vial in Headspace sequence obtained one, 1 less data file than expected

Problem:
For a vial not found in Headspace, the user expects the sequence to continue processing, skipping the non-existent vial.

Ran a Headspace sequence and obtained 1 less data file than expected. Attached is the log from the sequence.
 
Here are the parts of interest:

  69 41e1 4e2be3c4    0
Sequence     072411A.S started                              11:20:04 07/24/11
   0 1da7 4e2be3c5    0
7697A      1 Sequence Started                               11:20:05 07/24/11
   0 1da6 4e2be3c7    0
7890A      1 Attempt to download method to GC failed.       11:20:07 07/24/11 <--The GC driver attempts to stop the first run because the download failed
   0 1da6 4e2be3c7    0
7890A      1 Lost connection to Agilent 7890A at 10.100.1>  11:20:07 07/24/11
   0 1da6 4e2be3c7    0
7890A      1 00.11                                          11:20:07 07/24/11
   0 1da8 4e2be3c7    0
Instrument   HSS1 did not start a run                       11:20:07 07/24/11
   0 1da7 4e2be3c7    0
7697A      1 Vial 1 starting. Instrument time = 7/24/2011>  11:20:07 07/24/11
   0 1da7 4e2be3c7    0
7697A      1  12:17:33 PM.                                  11:20:07 07/24/11
7da 41e2 4e2be3c7    0
Method       Loading Method ALCOHOL.M                       11:20:07 07/24/11
   0 1da6 4e2be3ca    0
7890A      1 Connected to Agilent 7890A at 10.100.100.11    11:20:10 07/24/11
   0 1da6 4e2be3ca    0
7890A      1 Connection found GC in Pre-Run State. Forcin>  11:20:10 07/24/11
   0 1da6 4e2be3ca    0
7890A      1 g GC to idle.                                  11:20:10 07/24/11
   0 1da6 4e2be3ca    0
7890A      1 Runs cannot be started until the GC and inje>  11:20:10 07/24/11
   0 1da6 4e2be3ca    0
7890A      1 ctor are idle.                                 11:20:10 07/24/11
443 41e2 4e2be3ce 2010
Method       Method started: (B) line# 1 at 1 inj# 1        11:20:14 07/24/11<-- Chemstation starts the first run anyway.

Later in the log you'll see that Chemstation did not move onto the 2nd line of the sequence until AFTER the 2nd GC run. This caused the system to lose data.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1709    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: SEQ

One-line Description:

C.01.03: Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table

Problem:
It is not possible to run partial sequence acquisition using the back sequence table with Headspace.  

Exception is Found:
                                Exception
                                - Message:    Object reference not set to an instance of an object.
                                - Source:     AgtChemRCNetAdapter
                                - TargetSite: Void .ctor(PartialSequenceLine, SequenceLine, Int32, Int32, System.String, System.String)
                                - HelpLink:   
                                - Data:       System.Collections.ListDictionaryInternal
                                - StackTrace:
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceEntryHelper..ctor(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatId, String OnSampleIDMismatch, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceHelper.AddEntry(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatID, String strInjectorID)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequence.CreateCollections(PartialSequenceTable partialSeqTable, SequenceContext sequenceContext, MethodCollection MethodCollection, PretreatCollection PretreatCollection, RCSequenceHelper SequenceHelper, Boolean bPreventMethodResolution)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequenceDownload.StartFullSequenceMode(String sampleIDMismatch)
                                   at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.ControlAdapter.DoPreSequencePhase()


Investigation:  

The exception is the reason why the sequence is not downloaded to the instrument.  It is related to back sequences. Due to a minor mistake, the attempt to resolve a partial Back-sequence line fails, because the ChemStation tries to resolve it from the Front-Sequence Table.
Temporary Solution:
Workaround is to use Front-Sequences only.
Fix Information:

 Fixed in revision C.01.04


KPR#: 1713    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: SEQ

One-line Description:

Opening the sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler

Problem:
If  the sequence table is opened during a running sequence and the user clicks on a cell and makes no changes, there is a lock placed on the vial so that the instrument doesn't put a vial in the oven. If the user makes no changes and hits OK then the Chemstation never continues past the locked vial.
Temporary Solution:
If the user makes a change to the sequence or hits the cancel button then the sequence works correctly.
Fix Information:

 Fixed in revision C.01.04


KPR#: 1725    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Sample statistics are not done, when the first compond in the calibration table is not detected.

Problem:
If the first calibrated compound is not found in the first sample run of a sequence, no sample statistics will be performed.
Temporary Solution:
None
Fix Information:

Fixed in revision  B.04.03 SP1.


KPR#: 1736    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Partial sequence with methods saved to a location other than C:\ does not run with headspace and PAL Samplers

Problem:
A partial sequence run with methods saved to a location other than C:\ does not run. The logbook says it is running but the headspace or PAL Samplers do not run.

This defect affects all headspace hardware modules with the B.0x.0x series headspace and PAL Sampler software.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.04


KPR#: 1767    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: SEQ

One-line Description:

GC: sequence does not continue if there is a missing vial, uploads data to ECM and goes to next sequence

Problem:
The GC Chemstation never had the ability to skip a missing vial (the classic GC driver posts a dialog box asking the user what to do, abort or retry).  With RC.NET, if there is no vial on the tray the system stops the sequences, uploads the data to ECM and goes to the next sequence. Need to be able to specify what the system should do if there is a missing vial, skip or wait.
Temporary Solution:
n/a
Fix Information:

Issue resolved in OpenLAB C.01.05.


KPR#: 1787    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: SEQ

One-line Description:

Sequence stops when pushing the GC STOP key during sequence

Problem:
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence.  

This occurs beginning on ChemStation version B.04.01 SP1.

In revisions B.03.01 - B.04.01, pushing the stop key at the GC Keyboard would stop the current run and generate a report and continue to the next line in the sequence.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1788    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: SEQ

One-line Description:

Custom field window not prompted for Back injector on first run of the sequence

Problem:
During the first run of the sequence, the custom field window is only prompted for the Front injector.
Temporary Solution:
n/a
Fix Information:

Issue fixed in OpenLAB C.01.05.


KPR#: 1798    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: SEQ

One-line Description:

Sequence Reprocessing: Not possible to stop a sequence that is reprocessing

Problem:
When reprocessing the sequence, it is not possible to stop it.  The stop button is present, but does not to work. The user has to wait till the whole sequence reprocess completes.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1805    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: SEQ

One-line Description:

Sequence Import Mapping is missing LIMS ID columns

Problem:
The dialog box to map a sequence import file (Sequence -> Import Sequence Template) is missing the LimsID columns.  The LimsID 2 and LimsID 3 columns specifically are read-only fields that need to be imported in order to ensure that a unique LimsID can later be used in the result export to the Lims to ensure results are matched to the correct sample.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1822    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: SEQ

One-line Description:

Sequence parameters: Changes to Fraction start location ignored

Problem:
Sequence parameters: Changes to Fraction start location was ignored.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1824    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.05

Keyword: SEQ

One-line Description:

After upgrade, Sequence modified flag is set even if no changes are made before reprocessing

Problem:
Sequence modified flag is activated even if no changes are made before reprocessing for a dual method sequence
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1830    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 SP1

Keyword: SEQ

One-line Description:

'Number Format error' in Sequence log file after reprocessing

Problem:
Migrated a sequence from B.04.03 SP1 with 'Unique folder creation OFF' to C.01.04.  

Used the 'Result set migration' tool to migrate the data to a result set.  In online ChemStation, the result set was reprocessed.
In the sequence.log file,  'CP Macro Number format error in: %d' was displayed.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1601    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.02

Keyword: SEQ , 1100 Control , 1200 Control

One-line Description:

Lampall off switches off lamps and pumps

Problem:
The Post-sequence command Lampall Off switches off the lamps of all LC modules, but also switches off all the pumps.
This happens in RC.Net configuration.
Temporary Solution:
None
Fix Information:

Fixed in B.04.03 Fiixed in C.01.03


KPR#: 1172    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: Uninstall

One-line Description:

B.04.01 - Cannot remove (de-install) an instrument once configured

Problem:
With B.04.01 Chemstation software, it is impossible to delete, remove or deinstall an instrument.
Temporary Solution:
The only workaround is to remove the entire chemstation from the SETUP.exe on the B.04.01 DVD and reinstall the chemstation minus the instrument(s) that is/are not required.
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 731    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.02.01 SR2

Keyword: User Interface

One-line Description:

Error Fetching CP Variable observed sometimes when starting the chemstation

Problem:
Occasionally, when the ChemStation is started or shutdown, these errors are observed:

 “FetchCPVar  Cannot talk to Application”

 “Error Fetching CP Variable   _GUIFInjUpperVial”

 “Error Fetching CP Variable   _GUIBInjLowerVial”

  “Error Fetching CP Variable  _GUIBInjUpperVial”
Temporary Solution:
It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software.

The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation.

It should be possible to turn the views back on after startup.
Fix Information:

n/a


KPR#: 1574    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

OpenLAB Control Panel does not prevent connecting to an instrument already in use

Problem:
With "Fast User Switching" enabled in Microsoft Windows, a user B can connect to an instrument started by user A in a different user session.
Temporary Solution:
n/a
Fix Information:

Fast User Switching must be disabled in Microsoft Windows.


KPR#: 1577    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

Slow performances observed while loading datafiles when a networked printer is not accessible

Problem:
When a networked printer set a default printer is not responding, the ChemStation user interface may become slow when loading datafilles or when changing the scaling mode.
Temporary Solution:
N/A
Fix Information:

N/A


KPR#: 1583    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

Additional information about the [ChemStation:Sequence][Save Sequence] privilege

Problem:
When the "Save Sequence" privilege is granted to a user, this user will be able to create and save Sequence Template(s) or Easy Sequence Template(s). 

Without this privilege, the "Save Sequence" and "Save Sequence As..." menus will be greyed out. In addition, the "Easy Sequence Setup" tab will not be visible. 
It is also intended that users will be able to use the "Save Easy Sequence", "Save Easy Sequence as..." and  "Save and Add to Queue" buttons from within the Easy Sequence tab.
Temporary Solution:
N/A
Fix Information:

N/A


KPR#: 1644    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: User Interface

One-line Description:

PAL Sampling diagram always shows 'Vial Status: Normal' for every vial

Problem:
The tooltip over any circle (vial) in the CTC PAL Sampler Sampling Diagram reads ‘Vial Status: Normal’, no matter what its real status is.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1647    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

7820A, 7890A missing sampling diagram when only a rear injector is installed

Problem:
7820A, 7890A has an injector on the back.  The sampling diagram is not shown when there is only a back injector (no front injector or tray).
Temporary Solution:
n/a
Fix Information:

 Fixed in C.01.03.


KPR#: 1653    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03 GC DSP

Keyword: User Interface

One-line Description:

GC instrument diagram appears in offline ChemStation data analysis

Problem:
Occasionally elements from method and run control view can appear in data analysis view and vice versa.
Temporary Solution:
Delete the files config.reg, conf_off.reg and instrument.config from the instrument directory.
Fix Information:

Fixed in revision C.01.03.


KPR#: 1666    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: User Interface

One-line Description:

Distributed ChemStation: some dropdown lists are displayed only after a certain delay

Problem:
On ChemStation CDS clients of an remote AIC there are issues with all toolbar's listboxes (signal list, calibration table configuration etc.). These dropdown lists are displayed only after a certain delay, the user has to either wait quite a long time (up to 15 sec) or move the mouse pointer over the area the list should appear to get it updated and repainted.
Temporary Solution:
n/a
Fix Information:

 Fixed in C.01.03.


KPR#: 1674    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.02

Keyword: User Interface

One-line Description:

Graphic artifacts/software crashes caused by graphic settings

Problem:
Several problems have been observed on various ChemStation revisions that were the result of one common cause: Video settings that were adjusted for best appearance can cause graphic artifacts and can even cause the ChemStation software to crash.
Temporary Solution:
Adjust the display settings for best performance rather than for best appearance. This will free resources and prevent video artifacts from anti-aliasing or filtering attempts by the video driver.

The procedures below refer to Windows Classic screens and Control Panel in Classic View.


The following steps are recommended depending on the operating system:

Windows XP
-----------------

1) Go to Start > Control Panel > System. On the Advanced Tab press the settings button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button.
2) Go to Start > Control Panel > Display. On the Settings tab press the Advanced button. On the Troubleshooting tab set the slider for Hardware acceleration to None. Press the OK button.

Windows Vista and Windows 7
-------------------------------------------

1) Go to Start > Control Panel > System. Select Advanced system settings. On the Advanced Tab ress the Settings... button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button.
2) Go to Start > Control Panel > Administrative Tools. Select Services. Right click on the Desktop Window Manager Session Manager. In the context menu select Properties. Stop the service. Set the startup type to Disabled. This disables the Aero-Glass look.
Fix Information:

n/a


KPR#: 1682    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.03

Keyword: User Interface

One-line Description:

7890A, 7820A missing sampling diagram when only a rear injector is installed

Problem:
7890A, 7820A has an injector on the back.  The sampling diagram is not shown when there is only a back injector (no front injector or tray).
Temporary Solution:
n/a
Fix Information:

 Fixed in revision B.04.03 SP1.


KPR#: 1695    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: User Interface

One-line Description:

Change Access Level... is not available from menu bar

Problem:
Change Access Level... is grayed out and not available from the menu bar.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1698    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

490 Micro GC is not a valid option for a 7890A GC

Problem:
The 490 Micro GC is valid as a standalone instrument, but is not valid as an option on a 7890A.
Temporary Solution:
n/a
Fix Information:

Fixed in revision C.01.03.


KPR#: 1740    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.01

Keyword: User Interface

One-line Description:

The logged-on user is not displayed in lock dialog

Problem:
Set authentication to Internal and create an 'everything' user
Log into Control Panel and open a 6890 session, which fails to initialize
This eventually times out and the lock dialog appears.  The logged-on user is not displayed in lock dialog. Do not know who is logged into Control Panel.
Temporary Solution:
The root cause is that the full name is not set.
Fix Information:

Fixed in revision C.01.04.


KPR#: 1745    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: User Interface

One-line Description:

Finish Queue Sequence Help is hidden behind dialog

Problem:
Create a native sequence table; select RunControl\Queue Sequence... and walk through the sequence dialogs to get to the 'Finish Queue Sequence' dialog. Select Help from the button. Help opens but is hidden behind all the windows and will not come in front of the Finish Queue Sequence dialog.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1746    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: User Interface

One-line Description:

Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA

Problem:
Connect to a 68xx Classic driver instrument. The old sampling diagram is correctly displayed in the top to bottom column. Select View\Sampling diagram to make it go away, then select again to make it reappear. Only the column reappears. The turret or tray display is not shown again in that session.

If the user starts the ChemStation in online the sample turret and run start button are shown in Method and run control. But after the user goes to Data Analysis and back to Method and Run Control the turret picture and start button disappear.
Temporary Solution:
Close and reopen the chemstation session and the turret displays again.

The RC.NET driver does not have this problem.  The user can use the RC.NET driver.  Record the IP address, delete the 68xx GC and then add it back again using the RC.NET driver.
Fix Information:

n/a


KPR#: 1764    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: User Interface

One-line Description:

Run-time error ‘282’: No foreign application responded to a DDE initiate

Problem:
Run-time error ‘282’: No foreign application responded to a DDE initiate
Temporary Solution:
It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software.

The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation.

It should be possible to turn the views back on after startup.
Fix Information:

n/a


KPR#: 1769    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: User Interface

One-line Description:

Missing dialog items when using ChemStation AIC’s

Problem:
On ChemStation AIC’s (ChemStation Instrument Controllers), few dialogs are sometimes not properly displayed (e.g. Buttons or checkbox items missing).
This has been mainly observed on the Print Method and Edit Entire Method dialogs.
Temporary Solution:
n/a
Fix Information:

A fix for this issue is available in C.01.03 Hotfix04 and higher.  

To fix this issue,  a delayed repaint of the dialogs and an enumeration of all controls within the dialogs are performed. This has been implemented for the following dialogs: Run Time Checklist, Method Information, Edit Entire method, and Print Method.
The fix affects the display of the dialogs. Functionality of the items in the dialogs is not affected.


KPR#: 1773    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.03

Keyword: User Interface

One-line Description:

Change user, does not update the Connected user field in OLCP when using CS AIC

Problem:
The “Connected User” is updated by the “Agilent OpenLAB ControlPanel ChemStation Plugin” when launching a (ChemStation) instrument from the OpenLAB Control Panel, but not when the user is changed “inside” the ChemStation.
Temporary Solution:
n/a
Fix Information:

This issue is affecting systems with the revision C.01.03.


KPR#: 1808    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: User Interface

One-line Description:

Values under Overview Table are not aligned with the parameter heading

Problem:
C.01.04 Chinese version, the values under Overview Table are not aligned with the parameter heading of : RT, Sig, Lvl, Amt[wt%], Area, Rsp.Factor ,Ref, ISTD, #, Compound.
Temporary Solution:
n/a
Fix Information:

Fixed in OpenLAB C.01.05.


KPR#: 1816    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: User Interface

One-line Description:

Fraction Collector - F5 and F* not working

Problem:
The F5 and F8 Key for Start / Stop is intermittently not working.
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1688    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: User Interface , Data Analysis

One-line Description:

Preferences set in offline overwrite online Preferences and vice versa

Problem:
Preferences set in offline overwrite online Preferences and vice versa
Temporary Solution:
n/a
Fix Information:

n/a


KPR#: 1763    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: User Interface , SEQ , ACQ , ECM Integration/OpenLab Option

One-line Description:

In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.

Problem:
By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition.

The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically.

Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.
Temporary Solution:
Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.
Fix Information:

n/a


KPR#: 1118    Product: OpenLAB CDS ChemStation Edition     M83xxAA    B.04.01

Keyword: Utilities

One-line Description:

link to the base reference file IQTREF.XML on IQT Report not working

Problem:
The link to the base reference file IQTREF.XML on IQT Report is not working.
Temporary Solution:
The file is located at \chem32\core.
Fix Information:

This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.


KPR#: 1770    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Utilities

One-line Description:

Cleanup on Shutdown and Data Cleanup not working with Data Store or ECM backend

Problem:
An execution error occurs when local data cleanup should be performed.
The problem is less frequent with ECM backend.

The Data cleanup fails with the following errors

    Input string was not in a correct format.Couldn't store <2.0> in LocalVersion Column.  Expected type is Int32
    Execution error in macro VC_DataCleanup at level 1 in cmd/fct datacleanup
    loaded from C:\\CHEM32\\CORE\\ECMMCX.MCX's line 1004 with CRC 6232

and the cleanup on shutdown with 

   Input string was not in a correct format.Couldn't store <2.0> in LocalVersion Column.  Expected type is Int32
  Exception in command 'DataCleanup'  occurred! in VC_DataCleanup line 1004 code 999 at datacleanup
Temporary Solution:
Fixed in ChemStation C.01.04 Hotfix 1
Fix Information:

Fixed in ChemStation C.01.05


KPR#: 1772    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Utilities

One-line Description:

FPD Checkout Macros no longer work since there is no "autoint.mac" with OL CDS ChemStation C.01.04

Problem:
FPD checkout macro used for FPD sample test.

This FPD macro uses "autoint.mac" .
OL CDS ChemStation C.01.04  no longer includes "autoint.mac" in the CORE directory.

The following message occurs when loading the FPD Checkout macro:

 "Could not open file: C:\CHEM32\CORE\AUTOINT.MAC"
Temporary Solution:
Copy "autoint.mac" from OL CDS ChemStation C.01.03 and add it to the Core directory of OL CDS ChemStation C.01.04 .
Fix Information:

The macro code has been updated and does the following:

•Checks to see if the “AutoIntegrate” macro has already been loaded.  If so, it does nothing. 
•If “Autointegrate” macro has not been loaded, it looks for the file autoint.mac.  If the file is found, it is loaded.
•If autoint.mac is not found, it looks for the file sig_integ.mcx.  If that file is found, it is loaded.

The new macros work with or without autoint.mac.  Autoint.mac is found in the CORE directory up until C.01.03.  

Beginning in C.01.04, autoint.mac is no longer included in Core.  Autoint.mac is now part of sig_integ.mcx.

The eLibrary User Contributed Software has been updated with the updated macros.  They are also attached.


KPR#: 1616    Product: OpenLAB CDS ChemStation Edition     M83xxAA    C.01.04

Keyword: Utilities , Spectra Analysis , MSD Analysis

One-line Description:

Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group

Problem:
A shortcut to the ChemStation Real-Time Display is not included in the OpenLAB CDS ChemStation Edition program group.  However, the Real-Time Display can still be accessed at:

C:\Chem32\CORE\ChemRTD.exe
Temporary Solution:
Manually create a shortcut pointing to

C:\Chem32\CORE\ChemRTD.exe

to access the ChemStation Real-Time Display
Fix Information:

To be resolved in a subsequent OpenLAB CDS ChemStation Edition release