Software Status Bulletin
OpenLAB CDS ChemStation Edition M83xxAA
Known Problem Report as of Feb 19 2013 9:56AM
Preface
-------------------------
This Software Status Bulletin (SSB) documents all known problems in the software
product designated above. The SSB is derived from Known Problem Reports (KPR)
which result from user problems that have been classified as documentation
problems or software defects. When a KPR is written, an identifying number is
assigned to it, and the KPR is added to the next edition of the SSB.
User inputs that have been classified as Enhancement Requests are not documented in the SSB. User problems that have been submitted, but that have not been classified by the time the SSB is generated are not included in the SSB.
How to use the SSB
-------------------------
When you experience a problem with a product, first check this SSB to see if the
problem has been reported already, and if there is a temporary workaround
available for the problem, or if the problem has already been fixed by a new
revision. If the problem is not listed in this SSB then you may wish to report
it to the Response Center or to your field support representative.
To determine if your problem is documented in this SSB, first look in the "Keyword Index" section of the SSB. Under each keyword is a listing of one-line descriptions of related KPRs. If any of these sound like yours, locate the KPR # in the "Known Problem Reports" section of the SSB, and read the full KPR. The KPRs in the "Known Problem Reports" section are sorted by KPR #.
There are two sections in the SSB:
Keyword Index: This index is categorized by keyword. For each KPR there is a
brief description and a KPR #. A KPR may be associated with more than one
keyword.
Known Problem Reports: This section contains KPRs, with all the available information relevant to the problem. KPRs in this section are sorted by KPR #.
Keyword Glossary
Keyword Index
Lampall off switches off lamps and pumps
KPR#:1601
Wrong firmware revision displayed in Instrument Configuration dialog after firmware upgrade
KPR#:1539
Cluster information can be overwritten from the offline session
KPR#:1720
Lampall off switches off lamps and pumps
KPR#:1601
Keyword: 1200 Control , Method Scouting Wizard
Cluster information can be overwritten from the offline session
KPR#:1720
35900E - Error The Current Event Name is the same for a special event
KPR#:1623
cannot start two instruments using the two channels of one ADC at the same time
KPR#:1677
An AD converter cannot be split into two instruments from a ChemStation client
KPR#:1702
35900: Unable to set up Timed Events
KPR#:1707
View ACQ method from Data Analysis does not list acquisition parameters for the 35900 AD converter.
KPR#:1721
ADC can be non-responsive after long reprocessing
KPR#:1739
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826
Keyword: 6850 Control , 6890 Control
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1826
6890 RC.NET driver: instrument diagram not updated to show configuration
KPR#:1412
Addition of new columns to the column catalog (database) is not possible after upgrade
KPR#:1464
Sequence continues while GC temperature is not ready
KPR#:1622
6890 always opens to Configuration tab when opening Edit Agilent 6890 GC Parameters
KPR#:1684
Column info is missing with 6890
KPR#:1685
6890 RC.NET: Edit a running Method by making the oven hold time longer and click ok, then the Sequence/Single run stops
KPR#:1784
7890A: ALS errors abort sequence; no Retry/Abort option offered
KPR#:914
7890A Runs cannot be started until the GC and injector are idle, view GC Status
KPR#:1239
ALS errors abort sequence; no Retry/Abort dialog box option offered
KPR#:1329
Sequence aborts if vial not found with Overlap injection enabled (7890A only)
KPR#:1569
Must restart data system for barcode reader availability change (7890A)
KPR#:1625
Not possible to modify data analysis parameters of method during a single run in the offline; cannot save
KPR#:1641
No special characters should be used for names of methods, data files or sequences
KPR#:1660
Terminal Service can close under high workload
KPR#:1701
Running dual tower sequence with all methods saved as Dual, get “Injector doesn’t match the sequence’s DUAL Injector”
KPR#:1732
Changes in sequence methods made during acquisition are not saved automatically
KPR#:1742
Queue Method vial range reflects range of loaded method, not the method browsed to
KPR#:1747
Acq.txt and acq.macaml is missing for 7890A back injector only sequences
KPR#:1811
"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763
Method Audit Trail does not include change of Injection source when HS/PAL is no longer configured
KPR#:1749
UIB II - set a timetable and control "Method Audit Trail"
KPR#:1819
manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676
Intelligent reporting does not support exponential baselines
KPR#:13
Insufficient contrast on b/w printers running in eco mode
KPR#:30
Signal Plot thickness appears inhomogenous in case of high data rates
KPR#:31
The VialTable check does not occur during the startup of the CE ChemSation
KPR#:1587
CE G1600 instrument cannot be launched
KPR#:1664
Amount is given as zero when it actually could not be calculated
KPR#:1311
Calibration method printout does not print compound level's Heights when quantitation is based on height
KPR#:1766
Re-establish the old layout for calibration reports
KPR#:1793
Changing calibration table in Batch review generates an error
KPR#:1799
Calibration method printout does not print compound level heights when quantitation is based on height
KPR#:1807
C.01.04 HotFix03 : Amount units in calibration report printout is not updated
KPR#:1809
Sequence run for bracketing standard which is started with blank completed without creating data files
KPR#:1817
C.01.04 Hot Fix03 : incorrect calibration curves legend in calibration curves printout
KPR#:1827
Keyword: Client Report Rendering
Method information partially hidden in intelligent reports exported to XLS file format
KPR#:15
"Report Printed" time stamp in report footer is several hours off in Print Preview
KPR#:16
Localized character in saved PDF report cannot be searched in Adobe Reader
KPR#:20
Page numbering start counting at 0 in case of .doc report file format
KPR#:21
Scaling a signal to "full scale" does not add a percent sign to the axis label
KPR#:23
Report shows unexpected blank pages
KPR#:24
Report Export to .TXT or .DOC may fail if report renderer creates an out-of-memory condition
KPR#:28
Ineractively selected report parameters switch back to default when printing to PDF
KPR#:29
Interactivly entered report parameters are not always updated
KPR#:34
Large matrix reports may produce additional empty pages
KPR#:628
Calculating the standard deviation of identical numbers may return NaN
KPR#:653
Changes in report templates are not immediately shown on report printouts
KPR#:657
Rendering error when using a template containing a large result matrix
KPR#:659
Keyword: Common Infrastructure
High load on CAN may cause “Spectra runbuffer overflow” entries and a loss of spectra
KPR#:18
Companion fails to run blank run for valved/ALS injections
KPR#:1609
Custom Field content might get lost when a sequence is edited
KPR#:1771
Custom fields are not transferred to new result set
KPR#:1789
Custom Fields table erased when opening the Sequence table while running
KPR#:1795
Missing Sample Custom fields in the report
KPR#:1810
Custom field values become blank in the sequence table
KPR#:1828
Empty Custom Fields are not printed in Intelligent Report
KPR#:1837
Relative Retention Time solution does not work for uncalibrated peaks
KPR#:980
Keyword: DAD/MWD Configuration
Installing a lamp with tag may invalidate "alway use lamp" setting in detector's control
KPR#:3
Autobalace executed although in Sequence table set to "OFF"
KPR#:26
Reprocess a sequence that is still in progress
KPR#:1356
Reprocessing copy starting as a background window.
KPR#:1578
Data System will not "Autoscale" a TCD data file to the correct peak
KPR#:1639
Summed Peaks calculation error in the Summed Peaks Report
KPR#:1651
Y-axis labeled incorrectly when loading GC signals acquired with MSD ChemStation
KPR#:1652
ChemStation B.04.03 [54] will not "Autoscale" the TCD data file
KPR#:1654
Red X mark displays in data explorer of offline data analysis
KPR#:1706
Error "An additional register must be added to the list of locked registers"
KPR#:1762
Baseline not displayed properly when using ChemStation AIC’s
KPR#:1768
In Recalculate Last Result mode no reports can be printed after method modification
KPR#:1785
Report template not found when reprocessing with a master method
KPR#:1796
Data analysis: Fractions are not displayed in UIB chromatogram
KPR#:1813
Wrong method displayed in navigation table after partial reprocessing
KPR#:1678
PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688
Keyword: Data Analysis , Navigation Table
Wrong method displayed in navigation table after partial reprocessing
KPR#:1678
PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
HTML Help requires Internet Explorer revision 6.0.28.00.1106CO or higher
KPR#:1575
Understanding Your ChemStation manual needs more information about peak width calculations
KPR#:1734
G2613A 7683A injector firmware is incorrect in the Firmware guide
KPR#:1783
Keyword: ECM Integration/OpenLab Option
OpenLAB Option: Difficulties to lock the ChemStation during reprocessing
KPR#:1212
Problems after moving items from ChemStation to a different LCDF path in ECM
KPR#:1215
no automatic "check-in" after deleting a checked-out method
KPR#:1267
Acquisition only runs do not write instrument and column parameters to ACAML
KPR#:1309
token "operatorname" uses full name including special characters
KPR#:1363
ACAML result file does not contain all peak performance parameters
KPR#:1428
OpenLAB Option: data not marked as modified after changing the method
KPR#:1449
chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode
KPR#:1453
Path definition in OpenLab Control Panel is not used in ChemStation
KPR#:1611
Data not transferred to ECM after resuming the queue in the ChemStation
KPR#:1629
The storage path for files in ECM is user-specific
KPR#:1659
Cannot switch current ECM authentication to a different ECM server
KPR#:1667
ChemStore legacy data doesn't get reprocessed automatically
KPR#:1714
Wrong error message when downloading a method from ECM
KPR#:1753
Audit entry generated in the ECM or ECM Queue when doing Partial Sequence
KPR#:1774
ECM user licenses not released when user is changed in OLCDS ChemStation
KPR#:1794
Incorrect file name suggested when uploading from Review view
KPR#:1797
Update methods from ECM dialog locks up
KPR#:1806
In file information, the Operator shows disconnected on client
KPR#:1814
No warning before taking over session for data analysis only instruments
KPR#:1815
User token in upload path is resolved to current user - not to submit user
KPR#:1818
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1833
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763
In localized CDS for ELSD detector G4218A no signal data is shown or stored
KPR#:4
ELSD G4218A can not handle change of Sample Time during run
KPR#:7
Easy Sequences can end up in unexpected locations
KPR#:1655
Queue Planner will not run sequences of differing injection source
KPR#:1670
Methods have to be in single folder
KPR#:1185
Sequence stops when pushing the GC STOP key during sequence
KPR#:1600
FetchCPVar Cannot Talk to Application Error
KPR#:1608
The ability to collect partial data for specified times is needed for RC.NET GC drivers
KPR#:1626
FetchCPVar Cannot Talk to Application Error
KPR#:1628
Column compensation cannot be started from the Data System with 68xx RC.NET
KPR#:1630
G1888A - Handshake Abort - Vial Status does not update
KPR#:1638
Partial storage of data is not an option for GC RC.Net drivers
KPR#:1657
Error "creation of a data object failed“ when acquiring runs with data rates 1 Hz and below with 6890 RC.NET
KPR#:1680
OpenLab CDS, C.01.02, 7820A run continues when STOP on 7820A is pressed
KPR#:1683
Pulsed Split/Splitless injection does not wait for ready 6890
KPR#:1693
7890A stops run, ChemStation keeps going; GC Connection state is MONITOR in Status
KPR#:1700
Capacity factor (k') is not calculated; the file acqres.reg on the .D directory contains no column information
KPR#:1759
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604
Self-assembled result sets do not have full functionality for GC data
KPR#:1741
Unclear error message: Undefined symbol: License$
KPR#:1738
The PDF report on a Japanese Windows XP system might look garbled.
KPR#:5
Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17
Keyword: Globalization , Sampler Configuration
Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17
B.04.03 [16], error, no string 8002, Abort this copy of Alice?
KPR#:1615
Sporadically launch of an online session will not proceed
KPR#:1618
User with corresponding privilege cant't launch instrument configured on the AIC
KPR#:1658
CE G1600 instrument cannot be launched
KPR#:1664
Keyword: Initialization , CE Control
CE G1600 instrument cannot be launched
KPR#:1664
missing link and no automatic installation of the localized readme.txt of the CDS ed.
KPR#:1589
AIC ports not opened for 7890A acquisition
KPR#:1627
Methods Must be Stored on the C Drive
KPR#:90
Determination of tangent point on the tailing part of the peak
KPR#:821
Integration unable to handle complex chromatogram for area sum integration event
KPR#:981
Peak areas may be calculated incorrectly when using "New Exponential" as tangent skim mode
KPR#:1580
Data lost when using area sum integration event, must have 0.06 minute space between groups
KPR#:1642
Misleading display of tangent peak width
KPR#:1744
Timed integration event "Height Reject" cannot be configured.
KPR#:1800
GC compound peaks might be labelled as solvent peaks
KPR#:1838
manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676
Keyword: Integration , Audit Trail
manual integration may become slow, if many calibrated peaks are not found and the Results Audit Trail is switched on
KPR#:1676
Keyword: Localization,Japanese
Japanese characters from saved pdf report are garbled when copied and pasted to another application
KPR#:1668
'Method not downloaded to device' message when PAL is present
KPR#:1637
LC/MS ChemStation cannot open Logbook files
KPR#:1649
Deconvolution tools are not available for LC/MS
KPR#:1665
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616
Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment
KPR#:1675
MSD Tune View will awlays load ATUNES.TUN if tune file loading is cancelled
KPR#:1686
LC/MS may remain in a run state after Method/Sequence termination due to an HPLC hardware error
KPR#:1730
Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681
"Error occurred during loading method" after rejecting method resolution
KPR#:1343
CTC portion of method printout does not include overlap settings
KPR#:1473
Revision B.04.02 SP1 [208]: Import eMethod on 6890 RC.NET driver does not work
KPR#:1522
ChemStation hangs up during Edit Entire Method when starting this process via the UI
KPR#:1593
Method printout for RID contains typos
KPR#:1602
Undefined symbol: PrDARep error displayed when selecting Export e-Method
KPR#:1624
Open GC edit parameters and close it without changes; method modification flag is set
KPR#:1643
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1650
OpenLAB CDS C.01.02 sequence table can't find currently loaded method
KPR#:1656
No information around handling for special characters
KPR#:1663
After a single run, switching from the Method & Run Control view to the Data analysis view of the ChemStation, the method loaded in Data Analysis in different than the one used to run the sample
KPR#:1689
C.01.01, modified method flag gets set intermittently in error when loading different methods
KPR#:1691
Method marked Dirty on initial "Specify Report ..." with DEF_GC.M
KPR#:1694
Switching between Injection source sequences does not load correct method
KPR#:1704
Cannot transfer an English method to a Japanese or Chinese ChemStation and vice versa
KPR#:1710
Add/Removal of 7697 HS from 7890 does not show a change occurred; not displayed in Method Load options; method not shown as modified
KPR#:1748
Cannot save offline method if configuration is different: 'Method is not consistent with the configuration'
KPR#:1750
"Chemmain has encountered a problem and needs to close" when selecting back/dual inj position for GC manual inj
KPR#:1765
C.01.03 upgraded to C.01.04: 'Edit Entire Method' not working
KPR#:1802
"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671
"Specify Report" may be overwritten with the settings from another method during sequence
KPR#:1671
Keyword: Method Scouting Wizard
Cluster information can be overwritten from the offline session
KPR#:1720
Wrong method displayed in navigation table after partial reprocessing
KPR#:1678
Changing the sample type in a sequence with cyclic calibration corrupts the sequence structure
KPR#:1752
Date Time entries can be incorrect for aborted fraction collection runs
KPR#:1836
Printing fails if printer name exceeds 100 characters
KPR#:1612
Peak Purity Report Template is not available in Intelligent report
KPR#:1829
AIC can experience communication loss if Application Experience service is running
KPR#:1617
The command "macro" does not list *.mac files
KPR#:1755
The command "macro " cannot be used to select *.mac files
KPR#:1761
Scheduler does not work on an AIC
KPR#:1831
Misrepresentation of Bottlefillings in Dashboad of PumpValveCluster
KPR#:27
Cannot Check “Pump and solvent selection valve installed” Box for Flex Cube if Once Clear it
KPR#:32
PumpValveCluster with ISET - Solvent composition: resulting %A is wrong (at 1 line in timetable)
KPR#:34
Keyword: Report Template Design
Wrong labeling of standard deviation and variance
KPR#:11
Re-calibration does no longer update the data analysis method change time stamp
KPR#:19
Last displayed digits might be cut off, in cases where the number is zero
KPR#:25
RSD calculation based on converted report item causes template corruption
KPR#:27
Error message when adding a chromatogram to a report template
KPR#:620
Internal grouping of tables and usage of calculation variables cannot be applied at the same time
KPR#:621
Summary Calculations cannot be applied to table columns containing calculation variables
KPR#:622
Adding a template description to a previously erased and saved description field causes a KeyNotFoundException
KPR#:623
Customized composite groups should have a unique name
KPR#:627
Filter definition does not always allow usage of OR operators
KPR#:662
Reporting time format does not offer an option of (dd/mm/yyyy and time)
KPR#:668
Different background colour in .pdf files generated with PDF XChange printer of ChemStation C.01.01
KPR#:6
Data field CalibCurve_CorrCoefficient is not properly labeld
KPR#:14
Warning message concerning missing MSPGothic font
KPR#:18
Export of Short_ESTD reports to .DOC produces two extra pages
KPR#:22
Quantified compound markers are shown at absolute position in case of ISTD calibrations
KPR#:26
Early Maintenance Feedback (EMF) counters may not appear on report
KPR#:33
Customized chromatogram or axis lables may not display correclty
KPR#:658
problem with sizing of the chromatogram in a Short Report
KPR#:72
Chromatogram Output in reports: meaning of "Portrait" and "Landscape"
KPR#:276
ChemStation does not report resolution as defined in European Pharmacopoeia
KPR#:1011
Sequence summary statistics with different methods
KPR#:1324
it is not possible to print report with chromatograms that have been overlaid from different runs
KPR#:1377
acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off
KPR#:1440
Batch mode does not allow usage of intelligent reports
KPR#:1572
Report Templates newly added to a Result Set do not appear in the list of available templates
KPR#:1579
Extended performance report (IR) reports "Infinity" for plates per meter
KPR#:1581
Mismatch in drawing of peak performance between Classic Report and Intelligent Report
KPR#:1582
System may hang when ".DIF" output is choosen as report type.
KPR#:1585
Calibration Peak Areas included in the Area sum
KPR#:1614
Intelligent Reports do not contain ECM specific information in Review mode
KPR#:1619
B.04.03 DSP1 [54] Data Analysis, Graphics, response range does not work
KPR#:1634
No Chromatogram displayed on Intelligent Report for GC
KPR#:1645
CSV, DIF, and EMF files are not generated for some Classic report styles
KPR#:1673
sample name not reported with Intelligent Reporting in non-result set data that have been acquired prior to C.01.01
KPR#:1679
Resolution of chromatograms in Intelligent Reporting is low
KPR#:1690
ChemMain.exe can crash when a long file path is used
KPR#:1703
Chinese/Japanese characters cannot be searched in the Report Viewer
KPR#:1705
Tangent peak width not correctly evaluated.
KPR#:1722
Mass- Based Purification (Fraction Collection) report missing information & shows inaccurate Fraction Tick marks (RC.NET only)
KPR#:1733
Reports might show different calibration time stamps for the same method.
KPR#:1743
eSignature data file privilege is missing from the ChemStation lab manager, analyst, and operator roles
KPR#:1751
The structure of the reports to file has changed
KPR#:1754
CP3: Number format error in: %d when printing report to screen in Data Analysis
KPR#:1756
Error 41265 occurred at end of run; Number format error in: %d shown in logbook
KPR#:1757
Result data sent to OpenLAB ECM Intelligent Reporter may not be updated.
KPR#:1758
Preview of classic reports fails with a macro error. 'Number Format error' in Sequence log file.
KPR#:1775
Chromatogram in On-Screen Classic report is not displayed properly
KPR#:1776
'Number Format error' in Sequence log file after printing report to screen
KPR#:1778
Peaks are calculated by area only or height only in a report
KPR#:1779
Sequence summary report sorts the data by injection time
KPR#:1780
A Sequence Summary Report after Partial Sequence reports all the runs of the full sequence
KPR#:1786
Macro error occurs (Parameter 2: Table not found) when printing a calibration table
KPR#:1791
C.01.03: XML results: The 'ESTD %' and 'ISTD %' result types in the Quant Calc node are incorrectly 'ESTD' and 'ISTD'
KPR#:1801
Acquisition method name is not displayed in an intelligent report
KPR#:1803
Missing Report Style Options for LCMS and LCMS Qualitative
KPR#:1820
Classic detailed report does not show calibration curves after encountering a compound with invalid calib curve
KPR#:1821
ChemStation freezes while generating report preview
KPR#:1832
Calibration data modification time stamp is missing on reports printed from review view
KPR#:1834
Calibration data modification time stamp is set-off by 12 hours
KPR#:1835
Empty Custom Fields are not printed in Intelligent Report
KPR#:1837
report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620
Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715
Keyword: Reports , Custom Fields
Empty Custom Fields are not printed in Intelligent Report
KPR#:1837
Keyword: Reports , Data Analysis
report printed to screen hides behind the chromatogram window when DA.M is loaded
KPR#:1595
incorrect plate numbers calculated under certain conditions when using tangent method
KPR#:1620
Keyword: Reports , MSD Control
Incorrect Polarity Switch Delay parameter listed in ACQ.TXT
KPR#:1681
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715
No other Windows application can run, when online ChemStation is busy
KPR#:1597
PDF reports might not be updated when the file is viewed with PDF reader
KPR#:1662
HeadSpace sequence method will not load when it follows a Valve sequence
KPR#:1604
Sequence Summary Report does not do statistics when the first compound is not present
KPR#:1715
Change of sample type during sequence acquisition is ignored
KPR#:778
Method changes not used after pause/edit/save/resume sequence using unique folder ON
KPR#:1247
Mismatch for sequence completed time stamp in sequence and instrument logbook
KPR#:1317
ChemStation crash when trying to open Partial Sequence with non-exisiting data path in the Preferences
KPR#:1443
The path to a data file has a limited length
KPR#:1514
Cannot edit Method in Offline session while it is running in an online sequence
KPR#:1548
Missing Vial during sequences causes CTC ChemStation to hang
KPR#:1571
Cannot run partial sequence acquisition with headspace samplers
KPR#:1603
Headspace or PAL Sampler sequence with methods saved to a location other than C:\ does not run
KPR#:1636
Data system uses the currently loaded method to validate the vial range in the Sequence table
KPR#:1640
Sequence parameter subdirectory overwrites sample info parameters during run
KPR#:1648
Missing ‘Reprocessing Only’ in the Dropdown List for Unique Folder Creation OFF
KPR#:1687
Acq method for 7890A back injector is not displayed in data analysis view
KPR#:1696
Missing vial in Headspace sequence obtained one, 1 less data file than expected
KPR#:1699
C.01.03: Partial Sequence Acquisition Does Not Work with Headspace with the Back Sequence Table
KPR#:1709
Opening the sequence table, click on a cell, make no changes, and select OK causes sequence to stop with 7697A Headspace Sampler
KPR#:1713
Sample statistics are not done, when the first compond in the calibration table is not detected.
KPR#:1725
Partial sequence with methods saved to a location other than C:\ does not run with headspace and PAL Samplers
KPR#:1736
GC: sequence does not continue if there is a missing vial, uploads data to ECM and goes to next sequence
KPR#:1767
Sequence stops when pushing the GC STOP key during sequence
KPR#:1787
Custom field window not prompted for Back injector on first run of the sequence
KPR#:1788
Sequence Reprocessing: Not possible to stop a sequence that is reprocessing
KPR#:1798
Sequence Import Mapping is missing LIMS ID columns
KPR#:1805
Sequence parameters: Changes to Fraction start location ignored
KPR#:1822
After upgrade, Sequence modified flag is set even if no changes are made before reprocessing
KPR#:1824
'Number Format error' in Sequence log file after reprocessing
KPR#:1830
Lampall off switches off lamps and pumps
KPR#:1601
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763
Keyword: SEQ , 1100 Control , 1200 Control
Lampall off switches off lamps and pumps
KPR#:1601
Keyword: Sampler Configuration
Flexible Cube G4227A does not work with HiP ALS G1367E
KPR#:17
Control of Tray illumination missing for G1329B & G4303A
KPR#:5
Parameter mismatch for custom-defined Well Plates
KPR#:6
AFC Cluster: Edit Wellplates Dialog does not display imported wellplates.
KPR#:29
Use Vial as Fraction collections
KPR#:36
Control of Tray illumination missing for new G1329A
KPR#:9
Keyword: Sampler Configuration , Sampler Status
Control of Tray illumination missing for new G1329A
KPR#:9
Control of tray illumination missing for G1329A/B & G4303A in Japanese and Chinese ChemStation
KPR#:16
Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15
Keyword: Sampler Driver , Sampler Status
Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15
ICF A.01.04 - No method download, if method contains pretreatment/injector program steps.
KPR#:31
Dua Loop Sampler: ACQ method shows wrong loop with overlap injection
KPR#:33
Control of Tray illumination missing for new G1329A
KPR#:9
Larger injection volume with 900µl head may be incorrectly displayed after injection
KPR#:15
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616
B.04.01 - Cannot remove (de-install) an instrument once configured
KPR#:1172
Error Fetching CP Variable observed sometimes when starting the chemstation
KPR#:731
OpenLAB Control Panel does not prevent connecting to an instrument already in use
KPR#:1574
Slow performances observed while loading datafiles when a networked printer is not accessible
KPR#:1577
Additional information about the [ChemStation:Sequence][Save Sequence] privilege
KPR#:1583
PAL Sampling diagram always shows 'Vial Status: Normal' for every vial
KPR#:1644
7820A, 7890A missing sampling diagram when only a rear injector is installed
KPR#:1647
GC instrument diagram appears in offline ChemStation data analysis
KPR#:1653
Distributed ChemStation: some dropdown lists are displayed only after a certain delay
KPR#:1666
Graphic artifacts/software crashes caused by graphic settings
KPR#:1674
7890A, 7820A missing sampling diagram when only a rear injector is installed
KPR#:1682
Change Access Level... is not available from menu bar
KPR#:1695
490 Micro GC is not a valid option for a 7890A GC
KPR#:1698
The logged-on user is not displayed in lock dialog
KPR#:1740
Finish Queue Sequence Help is hidden behind dialog
KPR#:1745
Sampling diagram for 68xx Classic driver appears, but if selected and reselected, disappears; also when switching betw Meth&Run and DA
KPR#:1746
Run-time error ‘282’: No foreign application responded to a DDE initiate
KPR#:1764
Missing dialog items when using ChemStation AIC’s
KPR#:1769
Change user, does not update the Connected user field in OLCP when using CS AIC
KPR#:1773
Values under Overview Table are not aligned with the parameter heading
KPR#:1808
Fraction Collector - F5 and F* not working
KPR#:1816
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763
Keyword: User Interface , Data Analysis
Preferences set in offline overwrite online Preferences and vice versa
KPR#:1688
Keyword: User Interface , SEQ , ACQ , ECM Integration/OpenLab Option
In Sequence Parameters dialog a missing backslash of remote data path needs to be corrected.
KPR#:1763
link to the base reference file IQTREF.XML on IQT Report not working
KPR#:1118
Cleanup on Shutdown and Data Cleanup not working with Data Store or ECM backend
KPR#:1770
FPD Checkout Macros no longer work since there is no "autoint.mac" with OL CDS ChemStation C.01.04
KPR#:1772
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616
Keyword: Utilities , Spectra Analysis , MSD Analysis
Shortcut to ChemStation Real-Time Display not included in OpenLAB CDS ChemStation Edition program group
KPR#:1616
6890 Method printout does not include units for NPD 'Adjust offset' or 'Equilibration time', which should be pA and min, respectively.Temporary Solution:
n/aFix Information:
n/a
With Intelligent Reporting, it is not possible to report the sequence log book for sequences which have been acquired with ChemStation revisions prior C.01.01.Temporary Solution:
Users who need to report the sequence log book for old sequences from B- or A- ChemStation can use the classic reporting which is still available in ChemStation C.01.01.Fix Information:
There is currently no fix planned.
During run a stop is requested to the Software. Opening the User Interface where the system is shown the Initial pump parameters were not loaded. It was still A: 99.84% B: 0.16% instead of A:100% and B:0%. Loading the Default method still the gradient is wrong displayed in UI.Temporary Solution:
n/aFix Information:
there's any fix for this issue. The simptoms was very seldom and random. Restarting the PC and the Modules will erase the wrong information displayed on the User Interface.
When the 6850 or 6890 Oven Cryo On is unchecked (turned off) the Cryo setpoints are not grayed out.Temporary Solution:
n/aFix Information:
n/a
The columns that are missing include: LC-3D and MSD: Target Masses (number, range, list) and AutoBalance (blank or ‘OFF’) CE: Voltage (number), StopTime (number), and User1-10 (1-48 or blank)Temporary Solution:
n/aFix Information:
n/a
6890: Add\Advanced Tower Selection Syring Volume does not update until session is closed/reopened. For any Add step, select Advanced. The volume determines which tower will be used if automatically select by volume is selected (default). The Syringe Volume of that injector is displayed read-only below it. If you go to the Configuration and change the syringe size for any tower, then add a new Add step, save and return, this syringe Volume change is not detected or displayed. It is only refreshed by closing and reopening the online session of ChemStation.Temporary Solution:
Close and reopen the session to update the syringe volume.Fix Information:
n/a
When designing a new report template with Intelligent Reporting, it is inconvenient to access the data for the internal standard amounts. If the user drags and drops the report item “DAInternalStandardAmounts” on the report, text will be reported. This text contains a ';'-separated list of the ISTD values. If a name for a specific value is given, a name-values pair is reported like 'name1=value1; name2=value2'. If only a number is given for the ISTD, a 'ISTD-" prefix is added to the specific number: 'ISTD-1=value1; ISTD-2=value2’. So the field reports the name and the amount of each ISTD, separated by semicolon.Temporary Solution:
To report the amount of a specific ISTD, an expression needs to be set for the text field.
Example:
=choose(1,split(First(Sample_DAInternalStandardAmount),";"))
reports the amount of the first ISTD in the list.
Fix Information:
There is currently no design change planned
No Injection Calibration Blanks from Easy Sequence using an Easy SamplePrep program with 'Use for unknowns only' checked reserves resources, even though the Easy SamplePrep program does not run in this case.Temporary Solution:
Supply an excess of resources to allow the sequence to be submitted to the Queue.Fix Information:
n/a
When creating an Easy Sequence template and select the sequence location funcion, the software displays the normal windows function for selecting a directory / sub-directlry. This is wrong. The software should show the selection of sequence locations as specified in the sequence preferences.Temporary Solution:
n/aFix Information:
n/a
If methods ISTD_EX.m or DEF_GC.M are loaded onto a configured 7890 offline session an error message (Exception in command "translateMethod$" occured). These 2 methods are 6890 specific and should not be loaded onto a 7890.Temporary Solution:
n/aFix Information:
Fixed in GC driver 4.01 [047]. This driver was included in the following datasystem packages: GC ChemStation B.04.03 SP1 [87] and OpenLab CDS C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
In the Easy Sequence Setup, a method is selected and then the vial numbers are entered. These vial numbers are passed to the data system to be validated. The validation occurs against the currently loaded method, not the method used in the template that will actually use the vials.Temporary Solution:
n/aFix Information:
Fixed in Easy Sequence 2.4[002].
Calibration Blanks with injection run Sample Prep when 'Use for unknowns only' is checkedTemporary Solution:
Use a different methods for blanks that has 'Use Sample Prep program' uncheckedFix Information:
n/a
Sample Custom Fields and Compound Custom Fields are missing for Extended Parameters.Temporary Solution:
n/aFix Information:
n/a
Minimum Peak Width and Save Partial Signal not available in Signal SourceTemporary Solution:
n/aFix Information:
Fixed in driver revision 6.10.
If a GSV and/or MSV is configured on the GC but the hardware is not present, the GC status will report not ready but not display the reason for the not readiness.Temporary Solution:
No current fix.Fix Information:
n/a
When bracketing is used with Easy Sequence with Unique Folder Creation OFF, data file names can not be generated for the calibration runs. The potential exists to overwrite files when using Bracketing with Unique Folder Creation OFF with Easy Sequence.Temporary Solution:
When using Bracketing with Easy Sequence, use Unique Folder Creation ON from Preferences.Fix Information:
n/a
It may occur that PDF Summary Reports with 0 Byte file size are created for "complex" or long (around 1000pages) summary reports. This may also occur after rendering summary reports multiple times.Temporary Solution:
If such report are expected, it is recommended to created the report using a newly started Offline session.Fix Information:
n/a
If there is an ‘Add’ step in Easy SamplePrep program with just the ‘Solvent Wash’; even though the Back injector is to be used, it may not use the Back injector.Temporary Solution:
n/aFix Information:
n/a
If a 6890 method is loaded onto a 7890 with identical hardware some of the common parameters are not loaded. Typical parameters ignored are: Front PTV: Split ratio Split flow Total flow Column 1 (Agilent 19091J-413: Average velocity Column 2 (same column but not part of common sample pathway): Pressure Nominal initial flow Average velocity Front detector (NPD): Hydrogen flow state (6890 method was off, came in as on) Makeup flow state (6890 method was off, came in as on) BAck detector (FID): Hydrogen flow state (6890 method was off, came in as on)Temporary Solution:
No current fixFix Information:
n/a
If the Aux Detector 1 and Valve Box temp are checked in the ignore ready checkbox, they are still included in the readiness calculation.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[23]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Resource layout on 7890A GC instrument dialog is not translated into Chinese and Japanese.Temporary Solution:
n/aFix Information:
Fixed in revision A.02.03.
Easy SamplePrep Method printout is not translated.Temporary Solution:
n/aFix Information:
Fixed in Revision A.02.03.
6890 with no ALS or tray. Configured to a HeadSpace. Create a front valve method, a back manual method, and a HeadSpace (front) method. Depending on the method loaded, load an .est and this fails with: '--------------------------- Inconsistent Easy Sequence Setup --------------------------- Sequence requires a back injector not present on the instrument. --------------------------- OK ---------------------------' The message really does not advise the user of what should be done to correct this situation.Temporary Solution:
This is current ChemStation behavior. The template created in Easy Sequence must use the currently loaded method.Fix Information:
n/a
Using the RC.NET Drivers to control the Autosampler Chiller the systems is injecting with the Autosampler Chiller turned off. This is a different behavior compared to the legacy Rapid Control drivers for EZChrom which forced an ON of the Autosampler Chiller.Temporary Solution:
n/aFix Information:
n/a
Method Edit screen for AIB does not offer the 'Subtract from Signal: Column Compensation Curves' options.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[24]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Method transfer (7890 to 7890) loads the values for positional detectors/inlets even when they do not match; e.g. FID to TCD. Invalid flows will be entered into the method.Temporary Solution:
No current fix.Fix Information:
n/a
6890 with a G4515A BCR Mixer/Heater installed. Select Method Edit\Injector\Tray screen. The labels for these items are misleading for the 6890. Although the three items, reader, mixer, heater are a single unit, they are not all related to barcode reading. They are there for use with the Sample Prep Program. The labels should not include the word 'barcode' for this reason in the title (Barcode heating and mixing) or label for Enable Heater (Enable barcode heater). Enable Mixer is already correct.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[008].
Resources replenished on 7890A GC instrument dialog not translatedTemporary Solution:
n/aFix Information:
Fixed in revision A.02.03.
The software allows the user to enter invalid parameters for the draw and dispense speeds for the syringe. Using an "Add" command it generates a "Command to XXX Tower failed: XXX Tower Prep Failure Code: 142: Autosampler setpoint error..." The error message is limited and it takes effort to find the "invalid" parameter.Temporary Solution:
n/aFix Information:
Fixed in A.03.01.
Open online session to a 6890 with dual ALS (Plasticman 6890A/7693). DEF_GC.M is configured by default to front injector. Set injection source to Dual and save as a new method, now loaded. Select Easy Sequence Setup. No injection sources are displayed, (presumably because of the injection source selection of the previously loaded method had one injector.) Load the Dual method. Still no injection sources.Temporary Solution:
The user must select 'Clear All' to refresh the screen in order to display the Front/Back injection source checkboxes after loading the dual injector method.Fix Information:
n/a
The Mix command needs some more detailed control. The mixer to is used to shake a vial (Mix 2500 rpm 1 sec, stop 1 sec , and change direction mix 2500 1 sec, repeat), this cannot be done with the current mix command.Temporary Solution:
n/aFix Information:
Fixed in revision 2.3.
The displayed sampler error is the same for all sampler error scenarios. Error appears only in the Instrument Activity log. Error reads "6890 GC Error # 1 Injection Error: stop pressed or sampler error. (410)"Temporary Solution:
n/aFix Information:
Fixed in driver revision 6.10.
Shutdown events, such as a thermal or pneumatic shutdown, displayed on the GC Keyboard are not reported to the software and therefore not displayed on the “Agilent GC Status” screen.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. This driver is included with the following datasystems: GC ChemStationB.04.03 DSP1 [52] and OpenLAB C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
When the 6890 GC encounters a missing vial, the IALog displays Stop key pressed error #1. The error should read Missing Vial Error.Temporary Solution:
NoneFix Information:
None
HeadSpace Method loaded in ChemStation. The Injection Source is set to HeadSpace Sampler Easy Sequence Setup Sample Information Starting Vial Location Any number entered into this location will error with the following error message: The vial position could not be validated against the current configurationTemporary Solution:
n/aFix Information:
Fixed in Easy Sequence 2.4[012].
When a barcode mismatch occurs with overlapped injections (option of do not inject on barcode mismatch), the mismatched vial is taken to the turret and chemstation stalls with the sequence running and waiting. An abort is required to recover from this error.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
The signal 2 data cannot be saved independantly as signal 1 check box cannot be unchecked.Temporary Solution:
If only one signal needs to be saved, you must assign it as signal 1.Fix Information:
Fixed in 68xx driver revision 6.11[008].
Method with a single add step of 1000 uL. Syringe size is 10 uL. Vials are moved to turret and then error message displays: method not running. This is a mismatch in # injections/volume/syringe size/solvent allowed by firmware.Temporary Solution:
n/aFix Information:
n/a
Create a method which has the "Use Sample Prep Program" checkbox enabled. Use this method to create a method which does not have the "Use Sample Prep Program" checkbox enabled. Load the method which does not have the checkbox enabled. Error message displays: Configuration mismatch: Default parameters loaded for 'Agilent 7890'.Temporary Solution:
n/aFix Information:
Fixed in 3.1[21]. This revision will only fix with Openlab Chemstation C.01.03.
If the ALS faults during a run, no pop up error messages appear on the Chemstation screen. A message will be entered in the instrument log book and on the GC display.Temporary Solution:
None.Fix Information:
n/a
When using the multimode injection port, all parameters relating to cryo are available whether cryo on is checked or unchecked.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[23]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
The Instrument diagram of the GC instrumentation indicates a different column configuration than is set in the method. The GC keyboard will reflect the new changes, only the software instrument diagram is affected.Temporary Solution:
Beginning with B.04.03 DSP1 [54], after the configuration change is made to the columns, the workaround is to save the method and then load it. This will update the displayed column configuration. Another workaround is to shut down the ChemStation software and restart to reflect the new configuration changes. These workarounds may not work for all configurations. Classic 68xx driver works correctly.Fix Information:
n/a
Even with the 7693A tray disabled on the GC Config display, the config display when using method editing shows the tray as enabled. Enabling or disabling the 7693 Tray does not evoke method resolution.Temporary Solution:
n/aFix Information:
Fixed in Revision 6.02.
It is possible to the software edit parameters screen and change the oven hold time or oven temp and select OK. The GC does not accept the updates. The method in the software is changed to the new values; the GC ignores all.Temporary Solution:
n/aFix Information:
n/a
Edit instrument parameters; columns Applies to GC’s: 7890A, 6890N, 6850A The window displayed is incomplete; it shows the column type and flow/pressure parameters, but does not give the length, ID or film thickness. These are critical parameters, and should be displayed just as they were in the previous ChemStation revisions. To find these parameters, the user now has to look on the configuration window.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[008].
The 7890 pressure units are set to kPa from the GC display. If an auto configuration is performed from EZChrom the pressure units are reset to psi. This error occurs in EZChrom only.Temporary Solution:
Change the pressure units from EZChrom and download the configuation to the GC.Fix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Create an Easy Sequence Template for a G1888/6890. There is no error checking for Injection volume for this injection source in either the Setup or Sample List. All values are accepted, including negative numbers. There is no place to setup an 'injection volume' for a HeadSpace. There is only a time and pressure.Temporary Solution:
n/aFix Information:
n/a
Starting Vial Location and Number of Samples do not do any range checking against any method or configuration.Temporary Solution:
n/aFix Information:
Fixed in revision 2.3
GC is ready; Injection source = manual It is not possible to walk up with a sample and manually inject into the GC. The run can not be started from the GC Keyboard start button.Temporary Solution:
The user must first start the run from the Chemstation menu RunControl > Run Method and then press start at the GC.Fix Information:
Fixed in driver revision 6.10.
Easy Sequence Setup: Entry of text in the Sample Name combo box does not check for and advise that a counter token is needed if the user also uses the token 'Sample Name' for the Data File Name. The user will find out when sending to the queue 'Error: Sample Name xxxx occurs more than once.'Temporary Solution:
n/aFix Information:
n/a
Lamps and cells of G1314F show incorrect dates in the RFID tagsTemporary Solution:
n/aFix Information:
Fixed with the driver version A.02.06
7890/7693: Create methods for front, back, and dual ALS injection sources. Easy Seqeunce Setup does not refresh to show the correct injection location of the loaded method until user loads the corresponding method in ChemStation. Even loading a different method, the previous injection source is kept.Temporary Solution:
If the injection source has been changed, use the "Clear All" button and reload the method.Fix Information:
n/a
There is no error checking within Easy Sequence Setup to determine if the vial range of the loaded method's injection source is accurate. The method can be saved and submitted to the queue. It fails at this time with: '--------------------------- Error --------------------------- Invalid location found in FRONT sequence, line 1 There may be others --------------------------- OK ---------------------------' This defect occurs because there is a difference between the loaded method's injection source and the method that is selected in the Easy Sequence Template.Temporary Solution:
Load the method in ChemStation from Method > Load Method menu that will be used to create the Easy Sequence Setup Template.Fix Information:
n/a
Run a sequence with 11 runs and the 11th vial is missing. The sequence aborts and SSIZip file was not uploaded to ECM as the 11th vial was missing. The ignore missing vial option is not present in GC.Temporary Solution:
n/aFix Information:
n/a
Send a sequence to the Queue and while it is running, edit it. This modified/changed sequence pauses the Queue, but the first blue button is still in a 'Pause||' state. Click the button and change it to Resume, but it does not do anything as it is waiting for the current sequence to be saved before starting the next sequence. In this case, because the sequence has been modified, the ChemStation is not accepting sequences. If you click pause || button, you pause the queue and it does not do any thing. The message on the active queue is red and the message says the sequence has been modified.Temporary Solution:
Save the sequence that has been modified. Once the sequence has been saved, the queue will resume running the next sequence.Fix Information:
n/a
Load an Easy Sequence.es for a valve or HeadSpace Injection Source and the Sample Name cell in the Sample List shows a Read barcode button.Temporary Solution:
n/aFix Information:
n/a
In the 6890, the legend for the Instrument Setup graph says Column 2, irrespective of which column is highlighted. The graph is correct, it is just the legend that is wrong.Temporary Solution:
n/aFix Information:
n/a
The 6890 has an split-splitless front inlet and PCM rear inlet. The Instrument Setup’s Inlet tab shows that the rear inlet is a PCM and that it can be controlled. However, the column tab shows no ability to control the PCM rear inlet and it is called an Unknown Inlet. This is incorrect and inconsistent behavior. PCM is a legitimate inlet whose pressure can be controlled.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[008].
HeadSpace software will not run a sequence without a vial location. Therefore, the Easy Sequence Setup\Calibration\Simple Calibration options should only allow Blanks with Injection and not allow blanks without injection.Temporary Solution:
n/aFix Information:
n/a
Create a setup with a blank with injection and a blank with no injection. Save and load to Easy Sequence. The Injection volume for the Blank with no injection is entered as 0. Native sequence table also shows this as zero. This is not consistent with the volume for the blank with injectionTemporary Solution:
n/aFix Information:
Fixed in revision 2.2
System configuration: 1 DAD 1 UIB II 2 AFCs -> 1 AFC Cluster Analog out of DAD connected to Analog In of UIB II Simulated signal of DAD was used Method settings for modules (expect threshold for UIB II signal): see attached method 1. Created a data file with the simulated signal recorded by the UIB II 2. Loaded data file into Fraction Preview 3. Used Fraction Preview to collect peakbased fractions with threshold > 0.12 V 4. The value 0.12 V is written into the table on the left side without conversion (the table shows mV as unit) Using this method will cause that the AFC will fractioning all the time. Cause: The firmware interface expects the threshold in mV but the driver writes 0.12 (V) to the AFC This causes that the AFC is fractioning all the time.Temporary Solution:
n/aFix Information:
Fixed in UIB II signal is now delivered as mV signal such like in the UIB I
I worked with one AFC. Configured 2 wps +10 vials and 15 vial tray for recovery. Started multiple runs. For each run a new Recovery location was used. Then removed the recovery tray and entered it again. Next run started with first recovery location --> this was already used.Temporary Solution:
If you want to reset the Funnel Trays only, use "Reset Funnel Tray Volume" If you want to reset Wellplates and vials (all except funnels), use "Reset Fraction Volume"Fix Information:
Issue fixed with the LC and CE driver revision A.02.07
The setting is selected by opening the Tray/Overlap tab in Method Editor. The Instrument Status must be opened to view the dialog box. Pause is the default setting for ChemStation and MassHunter. Skip is the default setting for EZChrom.Temporary Solution:
n/aFix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
6890A RC.NET driver with ALS. If an ALS error occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort. The sequence simply aborts.Temporary Solution:
n/aFix Information:
n/a
Create a Setup with Calibration. Return to Sample tab and load the same or a different method with the same injection location. Return to Calibration tab and all has been cleared. Reloading a method does not clear entries made on the Samples tab.Temporary Solution:
n/aFix Information:
n/a
Start a 3 line headspace sequence from the queue, but remove the FIRST vial in the sequence. Queue does not update correctly, showing the first vial as running, though the HeadSpace has moved on to vial 2.Temporary Solution:
n/aFix Information:
n/a
HeadSpace aborts, but ChemStation does not. Select the Abort Run button from the Waiting for ready dialog and ChemStation aborts the sequence. Queue History shows this run as 'Complete' instead of 'Aborted'.Temporary Solution:
n/aFix Information:
n/a
The Use Inject the volume from the Method is not enabled.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
There is no advisory about the invalid vial locations until the sequence has been added to the queue.Temporary Solution:
n/aFix Information:
Fixed in revision 2.3
Sequence submitted to the queue but will not start and cannot be deleted.Temporary Solution:
As a workaround, delete the following file: SequenceQueue.que from the C:\Chem32\4\TEMP directory.Fix Information:
Fixed in revision 2.4.
Create a sequence using a method specifying sample overlap. If a non-initial vial is not found, the currently running sample will abort, the sequence will abort, and there will be only an error displayed in the logbook and GC Status. 68xx RC.NETTemporary Solution:
n/aFix Information:
n/a
The Easy Sequence Setup dialog should show Injection volume units, which in ChemStation always seem to be uL. When the user uses EasySequence with a Valved injection, the volume, which for valves is mL, must be multiplied by 1000 to get a value that will be accepted.Temporary Solution:
n/aFix Information:
n/a
Enter a long Sample Name in Easy Sequence Setup and add a counter. Save and load to Easy Sequence. Fill Samples. Not until then is the user warned (with a red X) in the Sample list that there is a 40 character limit. It is possible to save this but it will not be possible to add to the Queue.Temporary Solution:
n/aFix Information:
Fixed in Easy Sequence 2.4[014].
Chiller/Heater: On/Off should only appear in the method printout if the there is a chiller tray installed.Temporary Solution:
n/aFix Information:
Fixed in driver revision 6.10.
A small box should be displayed in lower right corner in graphical plot field.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Select Easy Sequence Setup. Enter a number greater than 1 in the Inj/vial cell on the right side and then tab. All labels and menu items disappear.Temporary Solution:
Restart the session.Fix Information:
n/a
Run a single vial using Run Method. Send sequence to queue vials 2-4. When the run method completes and the sequence loads and then the first vial in the sequence starts, Vial status shows the first line (vial 2) as immediately 'completed'. The system stalls at Sequence running. The GC is in prerun state and the system cannot start another run until it is shutdown and restarted.Temporary Solution:
n/aFix Information:
n/a
There were no barcode options in the GC Parameters for the 68xx RC.NET driver with a 7683B ALS with a barcode reader.Temporary Solution:
Turn the barcode on from the GC.Fix Information:
Fixed in driver revision 6.10.
Any changes done on Easy Sequence are not included in the audit trail; there is no separate audit log present for Easy Sequence.Temporary Solution:
n/aFix Information:
n/a
Run an Easy Sequence using a HeadSpace injection source. When a Vial not Found event happens, the final Status for this run, which did not start a run, is reported as 'Completed'. The status being reported by the HS device is 'Run Aborted'.Temporary Solution:
n/aFix Information:
n/a
Easy Sequence 2.01.41 on Chemstation B.04.03 DSP1 will not accept plate designations of PAL injection source in the Setup location combo box. (P1A1) The user should be able to choose the location designation.Temporary Solution:
n/aFix Information:
Resolved in revision 2.4 [12].
68xx RC.net Quickly setup a method and started a run. Error: Code #370 occurred in module #1 happened because there is no signal saved. This is not obvious.Temporary Solution:
n/aFix Information:
This issue no longer occurs in driver revision 6.10.
Configure 6850 RC.NET or 6890N RC.NET GC. Open an online session, initial or any subsequent try. This results in a benign pop-up error: 'FetchCPVar cannot talk to application'.Temporary Solution:
Go to view menu and turn off the sampling and instrument diagrams. This will keep hpgcui00.exe from starting up.Fix Information:
Fixed in driver revision 6.02. Driver included in OpenLab C.01.01.
Bar Code support was not available in B.04.02 for 68xx RC.NET – only heating and mixing function was supported. Bar Code support was added in B.04.02 SP1 for 68xx RC.NET and is implemented as follows: Vial -> BCR -> vial back to tray -> injector. Note: the gripper remains over the vial until it is picked up to move to the turret Making an intermediate stop at the tray location added from 9 to 11 seconds with the newer circular tray, and 8.2 to 9.3 seconds with the rectangular tray.Temporary Solution:
The 68xx classic driver works as follows: Vial -> BCR -> injectorFix Information:
Fixed in driver revision 6.10.
In B.04.02 [96] it is not possible to start the 6890 RC.NET from the front panel Start key. Software configuration is Injection source = manual. 6890 is READY, press start and get error message "ERROR Start Key disabled".Temporary Solution:
One workaround is for the user to start the run from the "run control" then "run Method" in the Chemstation software and then press start at the GC. Second workaround is for the user to switch to the "Classic" driver.Fix Information:
Fixed in driver revision 6.10.
1. Method and Run control screen> easy sequence settings tab > calibration tab 2. Select "simple calibration" for calibration mode, add Blank to the sequence. click print button to print preview easy sequence settings. Print preview screen does not appear.Temporary Solution:
n/aFix Information:
Fixed in Easy Sequence 2.4[30].
CIS4 inlet is not recognized with 6890 GC RC.NETTemporary Solution:
Currently, only the 68xx Classic driver supports the CIS4 inlet.Fix Information:
n/a
When the 7890A goes into hydrogen safety shutdown the ChemStation shows "Front inlet flow shutdown" and other messages, but not "Hydrogen Safety Shutdown". The chemstation does not report the hydrogen shutdown but it does list the other causes. In addition, the heated zones can be turned on from the chemstation. Since a hydrogen shutdown should require operator intervention AT THE GC remote control from the (any) chemstation should be disabled.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The 4.01 driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
It is not possible to edit when to save the start and stop time for GC signals. On the signals screen, need the ability to collect partial data for specified times.Temporary Solution:
n/aFix Information:
Fixed in driver revision 6.10 [27].
Adjust offset of the NPD bead is not available from the ChemStation softwareTemporary Solution:
Adjust offset of the NPD bead must be done via the GC Keyboard.Fix Information:
n/a
All setpoints are off for the NPD, but the Instrument Actuals continue to show GC Not Ready and lists all the temps, utility gas flow, and H2 flows for the detector as not ready and temp 'waiting'. This should not be seen as Not Ready.Temporary Solution:
n/aFix Information:
n/a
Select to add an Easy Sample Prep program, select an Add step. Enter any injection volume and select the Advanced>> button to open the dialog. Syringe size is always noted as 10 uL for both front and back syringes. It is not possible to refresh this from the checkbox tower selection. Configuration and method edit show this correctly. This did NOT happen for a 6890N, 6890 Plus, or 7890A.Temporary Solution:
If you change the syringe size, to get the new size reflected in Easy SamplePrep program editor, restart the instrument session.Fix Information:
Fixed in driver revision 6.10.
With the Keyboard lock set to on, it is possible to make changes to the GC set points at the GC keyboard, even during a run.Temporary Solution:
n/aFix Information:
Fixed in 68xx Driver version 6.02 [11].
Purged Packed Inlet includes parameters: 'Pressure On 20.261 psi' 'Total Flow, On, 5.4077 mL/min' 'Septum Purge Flow On -0.0089673 mL/min' Pressure, Total Flow, and Septum Purge Flow are parameters that are not available for this GC.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. This driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Start a 3 line sequence using a method where the Sequence Actions, System Not Ready @0.00 min is set to Continue. As soon as the first vial starts to process, the GC goes to a Not Ready state. First sample starts the GC, as expected, all subsequent runs are continued, but the GC never starts for the rest of the sequence. Seq logbook shows 'Data not available'. The 68xx will not start a run while being controlled by Chemstation, but will start a run if it is running stand alone.Temporary Solution:
n/aFix Information:
Fixed in driver revision 6.10.
n/aTemporary Solution:
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence. This occurs on ChemStation ver. - B.04.01 SP1 - B.04.02 - B.04.02 SP1 - B.04.03Fix Information:
n/a
If the 7890 gets into an error status (such as flame off) this is NOT reported within the 7890 status window The flame failed to ignite at the FID but the status window ONLY displays that the GC is Not Ready.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[009]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
The NPD Offset adjust to adjust the target current is not available.Temporary Solution:
n/aFix Information:
n/a
On the column configuration dialog (Instrument -> Agilent 7820A Configuration -> Configuration tab -> column tab), 6 columns can be configured, but the 7820 hardware allows only 4 columns.Temporary Solution:
Configure only 4 columns since that is what is allowed.Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Run online 7820 GC with 7683/7693 ALS. Select View -->Sampling Diagram, Sampling Diagram is displayed. Click [switch screen layout], "Agilent 7820ATray" diagram is displayed. There are two "Sampling Diagram" in Method & Run Control view.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
The fast plunger speed is only available when using multiple PTV Solvent Vent injections. Most users of multiple PTV Solvent Vent injections need to use Slow or Variable plunger speed, not Fast.Temporary Solution:
n/aFix Information:
Fixed in driver revision 6.10.
The 6890 RC.NET injects before it is ready only if the instrument has been sitting idle in between runs. For example, the GC becoming Not Ready because a new method was downloaded with new thermal setpoints during the sequence. (The sequence didn’t use the same method in every line.) This defect occurs while in a sequence. This defect does not occur if running isothermal or for systems that run the next sample soon after the last run was completed. This defect only occurs in 68xx RC.NET Driver systems only.Temporary Solution:
n/aFix Information:
Fixed in GC driver revision 6.10 [027].
In the 7820 GC Configuration, Any inlets are labelled the following way: SSZ Inlet PPZ InletTemporary Solution:
n/aFix Information:
Fixed in 78XX driver revision 4.02. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
From Control panel\Regional settings: change the decimal separator from a period to a comma. It is not possible to open a 6890/6850 ChemStation.Temporary Solution:
1. Backup data, methods, sequences, column databases (found in C:\Chem32\DRIVERS), etc Record GC IP addresses 2. Delete IQ1.CFG located under the Chem32\1 directory for example 3. Uninstall Chemstation 4. Reinstall ChemstationFix Information:
Period is required in Regional Settings.
Start up Chemstation, but no instrument or sampling diagrams are shown. FetchCPVar Cannot Talk to Application Error occurs. Systems use 7683 ALS.Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03 DSP1.
On the signals screen, need the ability to collect partial data for specified times.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01 [31]. This driver version is included with the following data systems: GC ChemStation B.04.03 DSP1 [54] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
The ability to do digital switching of detector signals in the Runtime events table is needed. It is only possible to do analog switching.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. Driver version 4.01 is included with the following data systems: GC ChemStation B.04.03 DSP1 [52] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Column 2 inlet configuration will never be correct if column 2 is connected to anything other than inlet 2.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.10[044].
The 'return all' command of Sample Prep was not performed.Temporary Solution:
Workaround is to use the "Move" command to return vials instead of the "Return all" command.Fix Information:
This is fixed in software driver 6.10 [028].
Connect to 6890N GC. Within seconds of successfully opening an online session, logbook logs the following simultaneously with the loading of the method. 'Logbook Welcome to ChemStation 12:17:12 06/28/10 0 1da6 4c28cb1f 0 6890 1 Timed out waiting for method download to fin> 12:17:35 06/28/10 0 1da6 4c28cb1f 0 6890 1 ish. 12:17:35 06/28/10 7da 41e0 4c28cb1f 0 Method Loading Method LCV_GCCONTROL.M 12:17:35 06/28/10'Temporary Solution:
n/aFix Information:
Fixed in GC Driver revision 6.10 [027].
The "Method modified after loading" flag occurs after lauching the sequence. Started the sequence from the ChemStation software. The method modification status indicates that the method has changed when it was not changed and the report will indicate 'modified after loading'. The GC had a calibrated capillary column that was changed to a packed column. When using this method that now has a packed column, the GC reports an error in the length parameter, and that is causing the method download not to reset the 'modified after loading' flag.Temporary Solution:
There is a workaround: bring Method into MethodEdit UI, and "reconfigure" column (Packed to Packed would be OK) to get new structure.Fix Information:
Fixed in GC Driver revision 6.10 [027]. Note, this fix will only apply to new methods, not to current methods.
7693A is mounted on a 6890 PLUS using B.04.03 chemstation software, the driver being used is the RC.NET. Every time a sequence is started it aborts with the error "set point out of range". This usually means the 6890 is configured as "classic" with a 7693 (not supported) but in this case the driver is RC.NET. 6890 firmware is correct (A.03.08) and all of the ALS firmware is the latest revision.Temporary Solution:
Build a new method with no pre or post washes or sample washes or pumps, and only do a 1ul injection. The ALS ran an injection with no errors. Methods without post washes run correctly, no errors.Fix Information:
n/a
The 7820A configuration displays 8 valve positions. The 7820A GC only supports 2 valve positions.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
When a run is started by a user press of the GC's Start button or activation of the remote start input without being prepared by the data system by a call to PerformInjection(), PerformPretreatInjection(), or StartRun(), the run log contents reported at the end of the run contain the contents reported on the previous run followed by any events in the run log for the new run.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[003].
It is not possible to add new columns to the GC column catalog (database) using the RC.NET driver. It is possible to add columns to the same column database using the Classic driver. User switched back to classic on same system and could add columns without error. The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdb The RC.NET driver is prevented from opening the db because you can’t have two column entries with the same model number!Temporary Solution:
Start with a new, “stock” db column catalog file from another system.Fix Information:
The column model number should be unique. The newer RC.NET drivers enforce this behavior.
Running ChemStation B.04.02.96. Start a sequence with overlapped injection set to immediately after injection. While the sample is being delivered to the turret, the Park button is pressed on the 7693A Tray. After delivering the vial, the tray is put in the Park position and the Injector completes the injection. At this point the ChemStation is unresponsive. (For example, when the ChemStation is in this state, any report printed to the screen can not be closed.) The ChemStation responds after the Park button is pushed and vial is returned.Temporary Solution:
The ChemStation responds after the Park button is pushed and vial is returned.Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run. This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period. This could result in a (retention time) RT shift.Temporary Solution:
n/aFix Information:
Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1 [54]. Fixed beginning in: B.04.03 SP1 [87] and OpenLAB C.01.03 [37].
Method download fails, sequence aborts after start; no injection made Error reported in logbook: Attempt to download method to GC failed. The Data System will not allow the run to continue when the method download has failed.Temporary Solution:
n/aFix Information:
This fix is in the 68xx driver version 6.10.032 and 6.11.001.
After successfully running a sequence with a back method, another back method is loaded. The back injector is not present in the edit method screen.Temporary Solution:
Change a parameter at the GC keyboard and click "upload from instrument" in the edit method screen. The back injector will show in the edit method screen.Fix Information:
n/a
Load a method which was developed for both detectors and has the front detector signal chosen under signals tab. Load this method on a GC with no front detector - method resolution. Changes made to the detector signal tab are not detected, tab remains yellow (invalid).Temporary Solution:
Click on a different tab and modify the value of any field to a valid value or click on the configuration tab and then click away. Change the signals tab to a valid value or turn it off to resolve the method.Fix Information:
n/a
RTL: New calculated pressure doesn’t update column 2 - 6 on current 7890A methods. Only Column 1 is updatedTemporary Solution:
n/aFix Information:
Fixed in driver revision 4.02[034]. This driver is included with the following datasystems: GC OpenLab CDS C.01.01 GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.04
RTL is not supported on a 7820A GC System. However, the RTL menus are shown in the 7820A software user interface menus in GC driver versions prior to 4.01.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
During a run, the GC method parameters are opened via the instrument diagram icon. Select OK after looking at the oven parameter to close the screen. The sequence or single run will just stop and the method is dirty. See Log File for error message. Log File: 67 41e2 4e69f8b3 0 Method Method started 07:29:55 09/09/11 67 41e2 4e69f8b4 0 Method Instrument running sample Vial 1 (front) 07:29:56 09/09/11 67 41e2 4e69f8b4 0 Method Instrument running sample Vial 2 (back) 07:29:56 09/09/11 0 1da5 4e69f91a 0 6890 1 Attempt to download method to GC failed. 07:31:38 09/09/11 67 41e2 4e69f91a 0 Method Instrument Error - Method/Sequence stopped 07:31:38 09/09/11 67 41e2 4e69f91a 0 Method Method aborted 07:31:38 09/09/11Temporary Solution:
Click the Cancel button rather than the OK button if the intent was to check the values of some of the instrument setpoints. This would not interfere with any run which is in progress.Fix Information:
Fixed in 68xx driver revision 6.11[008].
Once it detected a vial not found, everything would abort, including samples that were in process (that have already been injected and are being analyzed). The new behavior, is that the vial that is currently being processed will be finished. Everything else will abort, if the abort option is selected.Temporary Solution:
n/aFix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Sequence terminates with an error if user attempts to edit a parameter such as inlet temp while sequence is running. Logbook reports: Instrument Error - Method/Sequence stopped.Temporary Solution:
n/aFix Information:
Fixed in 68XX driver revision 6.11.
Column compensation (comp) cannot be started from the Data System with 68xx RC.NETTemporary Solution:
Column compenstation must be started via the GC keyboard.Fix Information:
n/a
The 7820A does not support the septum purge and it should be inactive. In driver versions before 4.01 septum purge was active and should not have been.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Method download fails, sequence aborts after start when trying to run a 6890 with a packed column. Error reported in logbook: Attempt to download method to GC failed. The Data System will not allow the run to continue when the method download has failed.Temporary Solution:
This can be run by setting the column as a capillary column but that changes the chromatography.Fix Information:
Fixed in driver revision 6.10[032] and 6.11[001].
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen. Only the first event, which is 'not ready', is reported to the Agilent GC Status screen. The second event, which is the shutdown, is not displayed.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Within the chemstation software, there is no display for the BCR heater actual or tray temperature actual. The setpoint is displayed and the instrument can be configured for that temperature to affect readiness.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample at the same time, then starts data acquisition and increase of inlet pressure to pulsed pressure.Temporary Solution:
If the 6890 is operated from the GC keyboard, the GC has the expected result.Fix Information:
Fixed in 68xx driver revision 6.11[007].
Upon upgrading from B.01.03 to B.04.02 it is no longer possible to add new columns to the GC column catalog (database). The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdbTemporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this. The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP, HP-5 in the database. Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) Temporary Fix: Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example. Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.Fix Information:
This is fixed in GC Driver version 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
After a stop, if there is a vial in the gripper it is not returned to its original location, it is kept in the gripper until the start of a new run. In the interim, if a new vial is placed in the open location, during the next run the system runs into an error because the gripper wants to return the vial to its original location which is now occupied.Temporary Solution:
n/aFix Information:
n/a
Saved a method. Open the GC edit parameters to review a setpoint and made no changes or changed the view. Selected OK and the method modification flag indicates a change was made. No changes were made. This does not occur with the Classic Driver.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[007].
Load a method from a 6890 GC with a fast oven on a 6890 GC with different hardware and a normal oven. Method resolution occurs. The oven temperature program from the fast oven GC has not been loaded on the normal oven GC.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[002].
The column specified as being used is missing information: description, Manufacturer and Model Number.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[005].
1. First Connection to an online GC and save the method OR connect to an instrument and save the method, open method editor, click on ‘Upload from instrument’. Click ‘Ok’. 3. Click ‘Edit entire Method’, go on clicking ‘Ok’ without making any change. 4. Method is dirty at the end. This only occurs on 6850 GCs with G4513A injectors.Temporary Solution:
Save the method as a new name after initial connection or subsequent method upload.Fix Information:
Fixed in 68xx driver revision 6.11[008].
Freeze Signal Value and Resume Signal Value missing from runtime eventsTemporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
The override column max temp checkbox in the oven parameter screen is automaticaly checked by default in versions of the driver prior to 5.01. In driver version 5.01, it will not be checked and the box will by highlighted in yellow if the column maximum temperature exceeds the value that is recommended for the column that is installed. To make the box not yellow, either check the "override column max temp" checkbox, or enter a value that is lower than the recommended maximum temperature for the installed column.Temporary Solution:
n/aFix Information:
Change of Behavior in Driver Version 5.01 This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
When a thermal aux zone is configured as a Cryo Trap, the cooling of the trap is accomplished via a run time event entitled Cryo trap cooling with an ON an OFF state.Temporary Solution:
n/aFix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
6890N GC with COC inlet connected to FID. The COC inlet pressure setpoint reflects the correct pressure set for the column, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.Temporary Solution:
Enable setup actuals (inlet) in the setup method screen, the inlet pressure is properly displayed.Fix Information:
Fixed in 68xx driver revision 6.11[018].
The ECD anode purge flow setpoint reflects the correct value, however the instrument actual value (blue text) in the setup method screen is displayed as 0.0.Temporary Solution:
Enable setup actuals in the setup method screen, the actual anode purge flow displayed properly.Fix Information:
Fixed in 68xx driver revision 6.11[018].
6890 configured with the classic driver. Selected valve as the injection source. Sampling diagram shows a tray when none is present. When injection source is set for Injector, diagram shows the front and back 7683 injectors, each with a three vial turret.Temporary Solution:
n/aFix Information:
This functionality is available in the .NET driver, improvements to the classic driver will not be addressed.
Edit GC Configuration, unchecked the Keyboard Lock checkbox to unlock the GC Keypad. Save the method in the software, and then make a change on the GC. The change will upload to the software. Any subsequent changes from the keyboard are not allowed. Open Configuration screen: Keyboard Lock checkbox is rechecked. Only one change can be made from the GC before the keyboard locks. This occurs on Windows 7 and Windows Vista PCs.Temporary Solution:
n/aFix Information:
Fixed in 68xx driver revision 6.11[10].
Create a resource layout. Select menu item Reset Resources... From the list, select one or several resources and OK. The logbook always logs 'User reset all resources' no matter how many resources are actually reset.Temporary Solution:
n/aFix Information:
n/a
The tray "Park" status overrides the start of a sequence.Temporary Solution:
n/aFix Information:
The G4514A arm from the tray must be UN-PARKED in order for a sequence to proceed. User must press Park button to UN-PARK arm.
It is not clear that the check box for the Control mode for column pressure/flow is linked to the inlet pressure. When the control mode check box in turned off in the column screen, the inlet pressure/flow is also turned off. This is not obvious to most users that this happens.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Extended Parameters for both the native sequence Parameters and for Easy Sequence Setup does not disable the Bar Code Reader section when no such hardware exists.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
For the split/splitless inlet, it is possible to run the septum purge in either standard mode or "switched" mode. In the switched mode, the septum purge is turned off during the injection cycle so that the only flow out of the inlet is through the column. The MMI would benefit from having this mode as well, particularly for splitless and solvent vent modes. In both modes on the MMI, the septum purge does not remain off until the inlet switches back to the purge state.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
When using a 7890A GC with Split Splitless inlet to perform backflush the Inlet Setpoint and Actual Pressure assume negative values on the GC and Chemstation display during backflush. This occurs with a splitter (>2 columns) or with a purged union (Primary column plus restrictor). The columns must be running in "Flow" mode. A similar symptom is the inlet pressure gets set to "0.00" PSI during backflush rather than the wizard recommended value of "1.0" PSI.Temporary Solution:
The workaround is to set the columns to "pressure" mode then back to flow and re-save the method.Fix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
It is difficult to set up backflush parameters by using the backflush method editor. After applying the settings, these are not correctly filled in the CFT and oven fields. After repeating the procedure (sometimes more than twice) it will be correct.Temporary Solution:
n/aFix Information:
Fixed in driver revision 3.01[38] This driver is included in the following datasystem packages: GC ChemStation B.04.02 SP1 [208] Ezchrom Elite 332 SP2 GCMSD ChemStation E.02.02
Started a run and observed that the inlet's temperature increased. Sometimes the SS inlet’s temperature is increasing at the start of a run. Looking at the xml contents of the methods, there was an initial hold time of 0.05 min and a ramp associated with that inlet, even though this ramp did not show on either the GC’s front panel or in the chemstation’s method editor. The behavior observed was due to using a method originally developed for an MMI on a SS Inlet in that same location. Especially given that an initial hold time of 0.05 min, a typical time at which MMI operation would be to begin ramping its temperature.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
7890A Runtime Events allow entries controlling valves which are not configured. The 7890A display doesn't show information about the Valve status (upon pressing Valve#) since the 7890A is not aware of the valve presence.Temporary Solution:
Not ApplicableFix Information:
Not Applicable
The Column compensation menu is available during the sequence and if selected will stop the sequence.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
User interface needs Constant Flow, Constant Pressure, Ramped Pressure, and Constant Flow for the Control Mode selections.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. This driver was included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
Setup 7693 injector and Tray. Put Tray in Park and then tried to start a run. The ChemStation showed that the GC went offline momentariily. The Event Log showed the following events : Disconnecting because a System.Web.Services.Protocols.SoapException was caught when sending a command; 29: GC refused to do an injection The connection state of the GC at 130.30.245.60 is now Offline Attempting to connect to 130.30.245.60 In DataChannel[1].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[2].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 In DataChannel[3].SetRunStartTime, nTime (321.148156 seconds) should be >= time sent last time 391.300000 The connection state of the GC at IP address is now Online Connected to 7890 at 130.30.245.60, using license key Cannot start runs until GC and injector are idle.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
Inlet pressure shutdown advisories indicating a problem with the instrument not shown in software status user interface.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. This driver is included with the following datasystems: GC ChemStation B.04.03 DSP1 [52] and OpenLab CDS C.01.02 GCMSD ChemStation E.02.02 OpenLab CDS EzChrom A.04.02
In configuration we now offer the option to configure how the keypad lock will behave.Temporary Solution:
n/aFix Information:
The driver with this new behavior is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
In previous versions of software, only the temperature of the detectors was shown in instrument status. In driver version 5.01, we show all GC detector status in instrument status.Temporary Solution:
n/aFix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
7890A has the round 100 position tray but the 7693A tray is shown instead.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.02[035]. This driver is included with the following datasystems: GC OpenLab CDS C.01.01 GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.04
If the colum is configured from one inlet to other (in case for a non-purged CFT or other applications) it's not possible to define the pressure of the outlet. It's possble to set it from the GC keyboard.Temporary Solution:
n/aFix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
The GC status and instrument actuals list the signal value as 0.0 but does not provide feedback if the flame is out.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[001] This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
7890A has a uECD detector with flows configured.
With the following method settings...
Makeup Flow(N2) : 30 mL/min
Constant Column + Makeup : 31 mL/min
... the method printout reads:
Makeup Flow(N2) : 31 mL/min
Constant Column + Makeup : 0 mL/min
Temporary Solution:
n/aFix Information:
Fixed in 7890A SW Driver 4.01 [048]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
If PCM is configured with the inlet heater as PCM A or PCM B controlling the column flows/pressures, the temperature and pressure/flow setpoints cannot be set. CS shows a septum purge value and the instrument goes into shutdown, due to "septum purge not ready".Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[007]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
By design column compensation is not available on the 7820.Temporary Solution:
Not AvailableFix Information:
Not Available
A sequence was run using auto zero at the beginning of each run. At run 4 in the sequence, autozero did not occur. This is due to a firmware issue wherein only 15 signal events (including auto-zero) would be processed before failure. The number 15 is cumulative since the last reboot which means that after 15 runs, the signal event will not be done.Temporary Solution:
n/aFix Information:
This issue is fixed in 7890A firmware revision A.01.12.1. No software update is required.
User makes a manual injection and waits until the peaks of interest elute, then hits the STOP on the 7820 front panel the GC stops (and resets the initial GC settings) but the chemstation continues and reports the system is running data analysis but does not go into data analysis. System appears to hang up waiting on the expected run time and report is generated at that point rather than directly after STOP is pressed.Temporary Solution:
Use STOP run from the software.Fix Information:
Fixed in driver revision 4.02[11]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Run Log entries (i.e. oven door opened) occurring during a run remain in the logbook until the restart of the GC.Temporary Solution:
n/aFix Information:
No software update is required. This is fixed in GC firmware revision A.01.12.1.
Saved a method. Open the GC edit parameters to review a setpoint and made no changes or changed the view. Selected OK and the method modification flag indicates a change was made. No changes were made.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[019]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
7693A ALS errors, e.g. turret, plunger, door open, etc. show a non-specific logbook entry: GCI1 did not start a run.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
The LTM column's equilibration time is shown as 0. Change the equilibration time by entering 0.5 and downloading the method. This does not change the value at the GC.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[005] and GC Firmware revision A.01.13. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
The diagnostic signal can be selected for the LTM actual temperature. Upon download and upload it is returned as oven temperature.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[007]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Open the instrument parameter from GC Instrument Diagram and confirm sample prep is off. Set inlet heater - save changes. Open the instrument parameter from the GC Instrument Diagram and confirm sample prep is off. Change the inlet heater temp and save as a new method. Load the new method - a configuration mismatch message is displayed. Requires Easy Sample PrepTemporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[004]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
After modifying and saving the method the injector screen will not accept the syringe size of 10 uL. The menu will pull down and 10uL is accepted but when the screen is OK'ed the size changes to blank.Temporary Solution:
Enter the syringe size from the GC front panel, upload the configuration and save the method.Fix Information:
Fixed in 78xx driver revision 4.02[019]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
After making a change to an LTM ramp, switch to a different LTM subtab or oven subtab. When user goes back to the original LTM subtab the change is gone and everything has reverted to the initial settings.Temporary Solution:
Make a change to an LTM ramp; and then switch to a different main tab (signals for example).Fix Information:
Fixed in driver revision 4.02[001]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
Other and Unknown Inlet cannot be selected on the config panel for inlets or on the instrument.Temporary Solution:
n/aFix Information:
Fixed in GC Firmware revision A.01.013. No software update is required.
Enter a new pressure on the PTV inlet. The number is not accepted, the value reverts back to the existing value.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[19]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
After modifying and saving the method the injector screen will not accept the syringe size of 10 uL. The menu will pull down and 10uL is accepted but when the screen is OK'ed the size changes to blank.Temporary Solution:
Enter the syringe size from the GC front panel, upload the configuration and save the method.Fix Information:
Fixed in 78xx driver revision 4.02[019]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
When the outlet of the column configuration is set as "Other", there is no backflush icon displayed in the 7890 Edit Parameter Screen.Temporary Solution:
n/aFix Information:
Fixed in 78xx driver revision 4.02[009]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03 Please note, when the configuration is changed from backflush not available to backflush available, the user must exit the method editor and begin the edit a second time in order for the backflush icon to be shown.
If you run Backflush from Chemstation with segmented columns (usually associated with LTM), only the in-segment portion of the configuration is used in the backflush flow rate and void volume calculations. This leaves the user unable to calculate the backflush time.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.02[023]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
In the Edit Parameter screen for the oven and the inlets, the display of this screen is small. Adjusting the method setup screen will not resize this screen.Temporary Solution:
Right click on parameter setting screen, and go to "show graphical plots". The display will resize to a normal size.Fix Information:
Fixed in 78xx driver revision 4.02[006]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
In driver revision 4.01, after editing the GC method and selecting ok, a "Validation Error" error message flashes in the left hand corner of the screen. User must select cancel to exit the GC parameter screen. No method resolution occurs when loading method Software loads fine and can make runs. Methods cannot be edited in online or offline mode. The configuration includes 4 packed columns that flow to the Back TCD Detector. Column 1's inlet is PCM C-2 (@ const pressure); 4's inlet is Aux EPC 5 (@ const pressure); and both 2's and 3's inlet is PCM C-1. Column 2's mode is constant flow; and 3, since it is also connected to PCM C-1, is reported as being controlled by #2.Temporary Solution:
Workaround for customers with driver revision 4.01: replace columns #2 and #3 with a single undefined (e.g., packed) column, suitably described as being a parallel combination. The user can then set either a flow or a pressure setpoint to the new combined column #2.Fix Information:
Fixed in driver revision 4.02. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03 OpenLab CDS ChemStation C.01.04
On a 7820, with only a Back Injector Installed the following message is displayed when launching the 7820: Upon launching the 7820A, the following message appears: DEF_GC.M: 7820 Resetting to Manual injection. For method 'DEF_GC.M, injection source 'GC Injector' conflicts with injection location. This does not happen when there is only a Front Injector installed.Temporary Solution:
Open the injection source and change it to GC Injector.Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Start on oven page, switch to channel partner detector tab, ok method. Restart editor - editor displays as a blank window with buttons at the bottom.Temporary Solution:
Use the instrument diagram to start the editor on any tab except the detector tab. Then proceed to navigate to the detector tab.Fix Information:
This is fixed in 78xx driver revision 4.02[022]. This driver is included with the following datasystems: GC ChemStation B.04.03 SP1 [87] GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.03
7890 GC with PCM controlled inlet heater. Open Edit Parameters, enable PCM inlet heater checkbox and make a change to the heater settings. Download the method to the GC or hit apply and the inlet heater is turned OFF at the GC. If the user closes edit parameters, and then re-opens edit parameters the PCM inlet heater checkbox is disabled.Temporary Solution:
Create your method, save, and download to the GC (the PCM inlet heater will shutdown). At the GC, set the PCM inlet heater to the desired setting. Upload the method from the GC (Instrument, Upload Method), and save your method as a new method (Save method as …). This method can now be used and loaded successfully in the future. The workaround will need to be performed any time changes are made to the method.Fix Information:
This driver is included with the following datasystems: OpenLab GC ChemStation C.01.04 GCMSD ChemStation E.02.02 SP1 OpenLab CDS EzChrom A.04.04
The set position of the gas streams used for a multiposition valve is controlled in the sequence. The user can run from different gas streams by entering the different numbers in the sequence table. The position is not stored as part of the method. The MPV set position is always reset to start from position #1 between runs.Temporary Solution:
n/aFix Information:
This is fixed in 7890A firmware revision A.01.14. No software update is required.
MMI in Split or Pulsed Split Mode does not allow editing of Total Flow value. The Total Flow combo box is active, but any entry made there reverts to its original entry.Temporary Solution:
Total Flow is editable in this mode from the GC keypad in driver versions prior to 5.01.Fix Information:
Fixed in the 5.01 Driver. This driver is included in: A.01.05 OpenLAB CDS B.07.00 MassHunter
Click on the Maintenance item in the Instrument drop down menu to open or close the Maintenance screen. There is not a red X to close this menu.Temporary Solution:
n/aFix Information:
n/a
The Maintenance View May Not Change Until Software is Closed and Re-OpenedTemporary Solution:
In the Maintenance menu, the user has the option to change the view of the menu by changing it under the Options button. Sometimes the change may not go into effect until the software is closed and re-opened.Fix Information:
n/a
If a sequence is aborted in the middle of a sequence, the vial will be left in the turret until the next run is started. At that time, it will be put back in the tray.Temporary Solution:
This is expected behavior.Fix Information:
n/a
The7890B GC driver included in the new data systems (A.01.05 OpenLAB CDS and B.07.00 MassHunter) supports up to 12 valves. The GC driver in previous data systems only reported 8 valves. Because of this difference, you cannot develop a method on a new data system and run it on an old data sytem. You can take a method loaded on an old data system and run it on the new data system. If you try to load a new data system method on an old data system, the behavior that you will see is that after starting a run, the software and the GC just sit there waiting for prerun to start, but nothing happens.Temporary Solution:
n/aFix Information:
Documented Behavior for GC Methods Developed Using A.01.05 OpenLAB CDS and B.07.00 MassHunter.
The peak inflection point position is drawn different compared to the classic reporting. This is because there are multiple valid positions where the point can be drawn. The effect can be observed for peaks that show only very small changes of the slope over a "longer" time range and the 2nd derivative is close to 0 over that range.Temporary Solution:
The reported peak performance values are the same in classic and Intelligent Reporting. Users who prefer the old drawing need to use classic reporting which is still available in ChemStation C.01.01.Fix Information:
There is currently no fix planned
Exponential calibration curve plots cannot be reported with Intelligent Reporting. An empty plot is shown instead.Temporary Solution:
If exponential calibration curves plots need to be reported, the (old) classic reporting needs to be used in ChemStation. EZChrom users need to use the EZChrom standard reporting.Fix Information:
A potentially fix is currently under investigation.
The software will automatically lower the oven max temperature when a low temp column is installed. This is to protect the column. The software does not rely on the User to do this to avoid damaging expensive columns. The requested enhancement is to automatically increase the oven maximum to match the higher oven max of the newly installed column. Currently, the user is required to manually update the oven max if it needs to be higher; the oven max is not automatically increased to match the higher temp column. This is true for 68xx and 7890.Temporary Solution:
n/aFix Information:
Fixed in 7890A Software Driver Version 4.01 [044]. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
The thermal inlet shutdown is not reported in the software “Agilent GC Status” screen. Only the first event, which is 'not ready', is reported to the Agilent GC Status screen. The second event, which is the shutdown, is not displayed.Temporary Solution:
n/aFix Information:
Fixed in driver revision 4.01. 7890A requires a minimum Firmware revision of A.01.12. The driver is included in the following datasystem packages: GC ChemStation B.04.03 DSP1 [52] OpenLab CDS EzChrom A.04.02 GCMSD ChemStation E.02.02
After an LC/MS spray chamber has been opened and closed, temperature will no longer be applied to the Spray Chamber heated zones until the spray chamber parameters are updated (by clicking "ok" in the Spray Chamber Parameters dialog box or loading a method) or the system is turned "off" or to "standby" and then back to on. This occurs with the APCI, APPI and Agilent Jet Stream spray chambers. The MultiMode source will set the system to "Standby" if opened & closed, however.Temporary Solution:
Turn the instrument to "off" or to "standby" in ChemStation, then back to "on" if this condition is encountered.Fix Information:
n/a
Intelligent Reporter templates designed for multi-signal methods may need to be adjusted after updating to ChemStation C.01.04. Non-calibrated peaks of signals calibrated in a main signal do no longer appear as calibrated compounds (as they are not quantified on the 2nd signal) but as un-identified peaks. Note: Calibration tables setup to "calculate signals separatly" are not affected by this design change.Temporary Solution:
In order to display the results of the mentioned peaks, the peak filter of the report template should be turned off completely (remove all check marks in the "show peak types" screen).Fix Information:
In rev. C.01.05. the mentioned peaks are treated as identified, but not quantified peaks.
The "copy to" function of the intelligent report method setup does not support usage of UNC paths.Temporary Solution:
Map the network path as a drive letter and select the new drive letter as a report destination.Fix Information:
There is no fix planned.
Queue method generated status “error” when using 35900ETemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
The built-in reporting snippet "Samples > Column Info" of intelligent reporting of OpenLAB CDS shows inaccurate units and lables. The inner diameter unit is printed as milimeters, where it should be micrometers. The column lenght unit is printed as milimeters where it should be meters. The lable for Inner diameter, Inventory number does not match the classical report.Temporary Solution:
Use report template editor to modify the settings of the report items inide the Column info snippet Inner diameter: Change the value definition from =FORMATNUMBER(First(Injection_ColumnDiameter),2,True,False,False) & " mm" to =FORMATNUMBER(First(Injection_ColumnDiameter),2,True,False,False) & " um" in order to replace mm by um. Column length: Change the value definition from =FormatNumber(First(Injection_ColumnLength),1,0,0,0) & " " & First(Injection_ColumnLengthUnit) to =FormatNumber(First(Injection_ColumnLength),1,0,0,0) & " " & "m" in order to replace mm by m. The columns void time can be added as an individual reporting parameter (Fields > Injection > ColumnVoidTime)Fix Information:
There is currently no fix planned.
OpenLAB CDS print services may stop unexpectedly when a redirected printer is configured for use with OpenLAB Shared Services. This results in the inability to print reports automatically from acquisition when using this printer.Temporary Solution:
Redirected printers are created during remote desktop or terminal service sessions and are intended only for use by the application that created them. Redirected printers are not supported with OpenLAB CDS.Fix Information:
Use only directly configured printers when adding printers to OpenLAB Shared Services.
If any printer configured for use on a CDS system is in an error condition, such as a paper jam, all printing ceases on the system. Print operations back up on the print server and no print operations will take place on any printer in the system. Printing of all compiled reports occurs when the printer error is cleared.Temporary Solution:
Check for printer errors and clear them if printing ceases to function on the CDS system.Fix Information:
Fix is currently under investigation.
When upgrading an A.01.02 CDS system that was installed in a location other than the default location (C:\Program Files\Agilent Technologies\) the IQ fails. This happens as a result of the A.01.03 version installing in the default location although A.01.02 was not in that location and the IQ failure is a valid result.Temporary Solution:
If OpenLAB CDS A0102 was installed in a non-default directory do not use the automatic upgrade. Uninstall A.01.02 using the A.01.02 installation media prior to performing installation of an A.01.03 system. If the upgrade was already performed and IQ is failing uninstall all OpenLAB CDS software and perform the A.01.03 installation again.Fix Information:
Will be fixed in next minor revision. Update - Resolved with OpenLAB CDS A.01.04 product revisions.
When defining printers in the OpenLAB Control Panel interface an unhandled exception may occur on a system that has User Account Control (UAC) enabled when attempting to start or stop the print service.Temporary Solution:
Open the control panel with a right click > as administrator or disable UAC to define and enable CDS print service configurations.Fix Information:
n/a
After a firmware upgrade for LC modules using RC.NET drivers, the old firmware revision (prior to upgrade) is still displayed in the Instrument Configuration dialog.Temporary Solution:
Remove the affected modules from the configuration and re-add them.Fix Information:
Fixed in OpenLAB C.01.05.
Changes in a TCC or pump cluster configuration (like e.g. changing a solvent name) usually cause method resolution. The changes will also be updated, even when the configuration dialog is closed by pressing Cancel. This is true for online and offline. This will consequently modify methods, although nothing has actually been changed.Temporary Solution:
NoneFix Information:
Fixed in revision B.04.03 SP1.
B.04.03 GC ChemStation with a 35900E configured. Select the CHANGE, DEFINE EVENTS tab in device configuration and you get an error "The Current Event Name is the same for a special event. Please change the current event name." This error is reported on a clean install of B.04.03 [16]. ChemStation B.03.02 works correctly.Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03 SP1.
When two instruments are configured that use channel A and channel B of the same ADC converter, respectively, the two instruments cannot be running at the same time. When the second instrument is started, both ChemStation instruments will freeze and can no longer be operated.Temporary Solution:
The two instruments will work when using only one of the instruments at a time.Fix Information:
Fixed in OpenLAB C.01.03.
The two channels of a 35900 ADC cannot be part of two different instruments when controlled from a ChemStation client.Temporary Solution:
When the 35900 is connected to an AIC and its two channels need to be part of two different instruments, they need to be setup from a networked workstation.Fix Information:
No fix available.
In the 35900 configuration, the Timed Events table is populated with default values. In the method, with the injection source set to 35900 Pre-Run Events, the Setup ADC Timed Events window opened but would not permit setup of the events. All buttons and drop down menus were gray and inaccessible.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
The acquisition parameters of an AD converter are not listed in the method printout from Data Analysis view.Temporary Solution:
Use the Print Method... option from Method&Run Control view.Fix Information:
No fix available.
The ADC might become NOT READY when the software is busy for a long time (e.g sequence reprocessing with IR to screen). When the system is no longer busy it cannot communicate with the ADC.Temporary Solution:
Software and instrument need to be restarted.Fix Information:
Fixed in revision C.01.04.
Column compensation (comp) cannot be started from the Data System with 68xx RC.NETTemporary Solution:
Column compenstation must be started via the GC keyboard.Fix Information:
n/a
In the 6890 RC.NET driver, changed the GC column configuration to have the back inlet go to the back column to the back detector. This change was not reflected in the instrument diagram as it is still showing the back inlet going to the front column to the back detector.Temporary Solution:
After the configuration change is made to the column connections, the workaround is to shut down the ChemStation software and restart to reflect the new configuration changes. Beginning with B.04.03 DSP1 [54], after updating the column configuration, save and reload the method will update the instrument diagram. Classic 68xx driver works correctly.Fix Information:
No fix Planned.
Upgrading from B.01.03 to B.04.02, it is no longer possible to add new columns to the GC column catalog (database) in Rev. B.04.02. The following error is shown: The Catalog File cannot be opened. If needed, you may obtain an appropriate replacement file from Agilent. C:\Chem32\CORE\...\DRIVERS\catalog.mdbTemporary Solution:
The problem is that there are several columns that have the same manufacturer and model number. Our new database software doesn’t handle this. The newer database software (in B.04.02 for example) no longer allows duplicates of manufacturer + model number. i.e., there can only be one HP HP-5 in the database. Note: the new software will inform the user that the model number “xxx” is already in use and request that a different value be chosen. (The user is returned to the original entry screen with all the info filled in.) In theory, customers should not be creating more than one catalog entry with the same model number. It’s like having two items in the LLBean catalog with the same order number! But, of course, the older software allowed the duplicates. Temporary Fix: Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example. Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.Fix Information:
Before upgrading, fix the problem with the column database duplicates while running B.01.03 for example. Review the column catalogs and make sure there are no duplicates and fix any that exist before the upgrade.
When running a sequence where consecutive runs use the same method using the ALS on a data system containing 6890 driver version 6.10.027 or earlier, the data system may proceed to start a new run before the GC has equilibrated after the conclusion of the previous run. This is more likely if the method contains a post run period of sufficient length that the GC equilibrates to the post-run conditions and the data system has completed data analysis well before the end of the post-run period.Temporary Solution:
n/aFix Information:
Patch available for RC.NET GC driver revision 6.10 [027] for B.04.03 DSP1[54]. Fixed in C.01.03
Select Edit Agilent 6890 GC Parameters… The tab will always open to the configuration tab.Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03 SP1.
The column specified as being used is missing information: description, Manufacturer and Model Number.Temporary Solution:
n/aFix Information:
This is fixed in 68xx Driver 6.11 which is included in B.04.03 SP1 and C.01.03.
During a run, the GC method parameters are opened. Select OK after increasing the oven parameter hold time. The sequence or single run will just stop and the method is dirty. Logbook shows: Attempt to download method to GC failed. Instrument Error - Method/Sequence stopped Method Aborted Instrument run aborted Method Aborted Sequence terminated due to an errorTemporary Solution:
n/aFix Information:
Issue fixed in OpenLAB C.01.04.
7890 with ALS. Create an ALS error, such as a Vial not found event. Instead of offering a pop-up that offers a second chance to Retry/Abort with the named error, this simply aborts the sequence. The error is buried deep in the sequence log of that instrument: 'Method Method started: (F) line# 2 at 2 inj# 1 09:11:51 09/20/07 67 41e0 46f27198 0 Method Instrument running sample Vial 2 (front) 09:11:52 09/20/07 0 1da5 46f271b2 300 GC 1 There was no bottle in the gripper. 09:12:18 09/20/07 0 1da6 46f271b2 0 Instrument GCI1 did not start a run 09:12:18 09/20/07 67 41e0 46f271b2 0 Method Instrument Error - Method/Sequence stopped 09:12:18 09/20/07 67 41e0 46f271b2 0 Method Method aborted 09:12:18 09/20/07 43e 41df 46f271b2 0 Sequence FRONTALS.S terminated due to an error 09:12:18 09/20/07'Temporary Solution:
n/aFix Information:
Issue resolved in OpenLAB C.01.05.
Occassionly, when trying to start a run or sequence on a GC whose front display and ALS are Ready, fails to start. Logbook shows: 'Runs cannot be started until the GC and injector are idle'. This can occur after an abort or stop of sequence, or connecting to a GC that is controlled by another PC.Temporary Solution:
Pressing "stop" on the GC front panel enabled the GC to makes runs again. Power cycling the GC may be necessary.Fix Information:
Fixed in revision B.04.03.
7890A with ALS. If an ALS error, occurs such as a Vial not found event, there is no dialog box option presented to Retry/Abort. The sequence simply aborts. The error is shown in the sequence log of that instrument: 'Method Method started: (F) line# 2 at 2 inj# 1 09:11:51 09/20/07 67 41e0 46f27198 0 Method Instrument running sample Vial 2 (front) 09:11:52 09/20/07 0 1da5 46f271b2 300 GC 1 There was no bottle in the gripper. 09:12:18 09/20/07 0 1da6 46f271b2 0 Instrument GCI1 did not start a run 09:12:18 09/20/07 67 41e0 46f271b2 0 Method Instrument Error - Method/Sequence stopped 09:12:18 09/20/07 67 41e0 46f271b2 0 Method Method aborted 09:12:18 09/20/07 43e 41df 46f271b2 0 Sequence FRONTALS.S terminated due to an error 09:12:18 09/20/07' GC goes back to "ready" state instead, and vials remain in the position they were when the injector failed (not returned to tray).Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
7890A system using overlap injection feature 1. Define a sample overlap of 1 min. 2. Define a sequence of 5 runs 3. Insert 4 vials only The 7890A GC starts the acquisition of Vial 4 correctly. When it starts to prepare and pick up vial 5 which is missing, the acquisition of vial 4 is stopped and the sequence aborts. When sample overlap is enabled and the vial to be prepared is not found, the currently running sample in the sequence will abort. The sequence will stop and there will be errors displayed in the sequence logbook and GC status. The error messages displayed are as follows: - Command to Tray failed: Tray Prep Failure Code: 250: Front injector has no vail in turret... - User stopped pretreatment program, finalizing. - Instrument Error - Method/Sequence stopped - Method aborted - Sequence.S terminated due to an error - Sequence.S completedTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
If the presence of the barcode reader is changed while the data system is connected to a 7890A, the availability of barcode options in the Sequence Parameters screen does not change until the data system is restarted.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB revision C.01.03.
It is not possible to change integration parameters in the offline for a single run and save the new parameters in the method file.Temporary Solution:
Data analysis parameters can be updated and saved if using a sequence. It is recommended to prepare a single run sequence.Fix Information:
Fixed in revision C.01.04
Names for data files, methods and sequences should only contain alphanumeric characters. Special characters like ",#,$, %, &, @, ? and others should not be used and can in general cause trouble. The same is true for trailing blanks. While this is outlined in the written documentation several times, it is not enforced for all dialogs inside the software. There are still the occasional loop hole where such a character can be entered or imported.Temporary Solution:
The possiblity of entering special characters does by no means imply that it is safe to do it.Fix Information:
n/a
It has been observed that the terminal service on an AIC can terminate under high workload. This is a limitation of Windows 2008 server. Since this only affects the remote session control to the ChemStation Client, no data will be lost.Temporary Solution:
No fix available.Fix Information:
Will not be fixed for Windows 2008.
1. Create seq table, with one run (Method A) for the front table, one run (Method B) for the back table this must be done first 2. Before line one of the sequence completes, add a run2 to the back table only (BACK.M) by append 3. When the back run is in process, add a sequence line to the front table, use a different method (method A) 4. Message stating that Injector doesn’t match the sequence’s DUAL Injector occurs even though all methods are saved with Dual injection selected as the injection sourceTemporary Solution:
n/aFix Information:
Fixed in Revision C.01.04. Point patch available for B.04.03 SP1.
Run a sequence and change the method during the acquisition (e.g. change the runtime). There is no possibility to save the method. The method should get saved at the end of the run automatically, but it does not. This issue occurs in C.01.01 - C.01.03 versions.Temporary Solution:
n/aFix Information:
Resolved in Revision C.01.04.
Load a method in ChemStation. Select Queue Method. Within that dialog, Browse... to load another method with a different vial range. The vial range (Sample Parameters: Vial/Location) is not updated to reflect the range of the method loaded in Queue Method, though the Signal (1 and/or 2) Data File info is correctly loaded from the browsed to method.Temporary Solution:
n/aFix Information:
n/a
Acq.txt and acq.macaml is missing for 7890A back injector only sequencesTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
GC with HS or PAL as the injection source. Close the session and remove the external device. Open the session and select Compare. No differences are seen in the method. Download the original method. The injection source has been changed to Manual. No change is logged to the Method Audit Trail.Temporary Solution:
n/aFix Information:
n/a
If a timetable of UIB II is set and stored via "Save Method as..", the time of the change should be stored in the "Audit Trail". But the time of the method change is not correct.Temporary Solution:
n/aFix Information:
n/a
The intelligent reporting of OpenLAB CDS rev. A.01.02 does not support displaying exponential baseline constructions in chromatograms. A straight line is shown instead of an exponential curve. Peak area and height calculations are not affected.Temporary Solution:
Classical reports of ChemStation and EZCHrom show the correct baseline position.Fix Information:
A hot fix for OpenLAB CDS ChemStation is currently under investigation.
The chromatogram plots of Intelligent Reporting of OpenLAB CDS A.01.04 and earlier do not show satisfying contrast on black & white printers. This is especially true if the printer defaults are set to eco mode.Temporary Solution:
n/aFix Information:
Intelligent Reporting of OpenLAB CDS A.01.05 offers the option to print all signals in black in order to increase the contrast on b/w printers.
The signal plot thickness of Intelligent Reporting of OpenLAB CDS A.01.04 and earlier may appear inhomogenous and peaks show different line thickness compared to rnages with no peaks. The quality of the chromatogram print out doesn't look satisfying. The problem is caused by a data compression algorithm affecting chromatogram plots where the signal data rate is too high in comparison to the size of the printout. E.g. in case there are 3 or more data points per pixel repesenting a time interval.Temporary Solution:
The plot compression can be prevented by choosing a larger scaling of the time axis and stretching the signal in the direction of the chromatogram's time axis. Chromatograms showing larger run times (> 60 min) should be plotted on landscape paper format or the chromatogram could be cut in 2 or 3 time slices to fit on a portrait paper format.Fix Information:
If you encounter the described problem and do not want to enlarge the time axis scaling you can use the following procedure.
Intelligent Reporting of OpenLAB CDS A.01.05 introduces a new custom parameter
to turn data compression off. This parameter needs to be edited in the RDL file directly.
New templates generated with OpenLAB CDS A.01.05:
- Create a backup copy of your template RDL file
- Open the template RDL with a text editor, like Microsoft Notepad
- Search the string "AlwaysDrawAllPoints"
- For each chromatogram plot inside a report template you will find a section
AgtPlot:ChromPlot:AlwaysDrawAllPoints
False
- Change False to True
- Save the template
Report templates generated with OpenLAB CDS A.01.04 or earlier:
- Create a backup copy of your template RDL file
- Open the template RDL with a text editor, like Microsoft Notepad
- Search the string "ChromPlot:IsAnnotateAllPeakMarkers"
- after each section
AgtPlot:ChromPlot:IsAnnotateAllPeakMarkers
True
you should append a section
AgtPlot:ChromPlot:AlwaysDrawAllPoints
True
- Save the template
In case of e.g. a quadratic calibration curve it may happen that an amount cannot be calculated for the area in a chromatogram (if the area is outside the calibrated area range). In these cases the amount will be reported as zero instead of "-" as an indication that the amount could not be calculated. However, with the default report styles, the report contains a warning "invalid calibration curve".Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
Calibration method printout does not print compound level's Heights when quantitation is based on heightTemporary Solution:
Available fix in C.01.04 Hotfix03Fix Information:
Fixed in C.01.05.
Re-establish the old layout for calibration reportsTemporary Solution:
n/aFix Information:
Fixed in C.01.04 Hotfix03.
When run is spooled in Batch Review, click on menu Calibration –> New Caibration Table, select OK and overwrite existing table Click OK on calibration table when it is displayed For a brief moment an error is displayedTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Can't show calibration table based on Height when printing calibration table curvesTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Enter “ug/ml” for “Amount Units” in the “Calibration settings”. The calibration table was updated with “Amt (ug/ml)” but the calibration table printout for the Overview table indicated “Amt (wt%)”. This also happens with “Amt (kg)” and the calibration report printout still reflects as “Amt (wt%)”.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
When a sequence is started with blank data files, data files are not generated. When the sequence is started with standard as bracketing std. (line 1) and sample (line 2), sequence completed and data files are generatedTemporary Solution:
n/aFix Information:
n/a
When using "ISTD" as "Calculation mode" and "Area" for "Based on", the calibration curves x-axis and y-axis shown on the calibration curves were : "x-axis=amount ratio" "y-axis=area ratio” but the report printout for calibration curves have the following incorrect legend in English version: "x=amount" "y=area" And incorrect legend in Japanese version: "x= amount" "y= height"Temporary Solution:
n/aFix Information:
n/a
During the startup of the CE ChemStation, the System VialTable is not loaded therefore when a method is loaded conflict checks with the method vialtable is not performed.Temporary Solution:
If you are using an Agilent 1600 Capillary Electrophoresis instrument, the following post installations steps are required. 1. Insert the OpenLAB CDS Disk 6 2. Open the folder calledFix Information:\Release Updates\OpenLAB CDS ChemStation Edition\CE\ 3. Copy the file HPCE.MAC to the folder \CE\ Note> The default chemStation path is c:\Chem32 It's recommended to create a backup copy of the original HPCE.MAC before replacing it.
This issue is resolved in the software revision C.01.02
The method information part intelligent reports exported to XLS file format is partially hidden. This issue occurs due to the limitation of the Excel renderer. Basically any report item while being rendered causes Excel's cells to merge. To suppress the undesired merging of cells the parameter ‘CanGrow’ is set to true, and 'AutoSize’ of the excel row fails as a consequence. To read more about this refer to the ‘Merging Cells’ paragraph on the link http://msdn.microsoft.com/en-us/library/ms159836.aspx Also find a suggested workaround for the issue at http://connect.microsoft.com/SQLServer/feedback/details/163236/reporting-services-row-height-error-on-export-to-excelTemporary Solution:
Select all cells in the Excel sheet and double click on the separator between the 1st and the 2nd row. Excel automaically sizes all cells so that the content fits.Fix Information:
There is currently no fix planned.
In the report preview, when switched to Print Preview, the print time displayed in the footer is several hours off. In interactive mode, the time is correct. The problem is caused by Microsoft ReportViewer 2008 SP1.Temporary Solution:
Print the report to PDF to review the final layout of the printed report.Fix Information:
There is currently no fix planned.
Localized character in saved PDF report cannot be searched in Adobe Reader.Temporary Solution:
The wrong behavior is caused by a 3rd party component (Aspose). As a workaround it is recommended to use the ChemStation PDF printer.Fix Information:
There is currently no fix planned.
Intelligent Reports exported to .DOC file format show a page numbering starting at 0 when report is viewed in Microsoft Word application. The defect is caused by a 3rd party report renderer and cannot be fixed in the reporter code.Temporary Solution:
The footer section of the report template can be modified to translate the page numbering to word document parameter that could be updated by pressing the F9 key. The correct syntax for the page count footer is: =PageNumber & " of " & TotalPagesFix Information:
There is currently no fix planned.
When changing the signal axis of a chromatogram from "all the same scale" to "each in full scale", the signal axis range changes from 0 - 100%. However, the signal axis labels do not show the percent sign and it's not immediatly clear that the scaling is relative as opposed to absolute.Temporary Solution:
N/AFix Information:
A fix a planned for the next minor release after A.01.04.
Working with report template editor of OpenLAB CDS A.01.03 or ealier it happens frequently that reports show additional empty pages. They are typically introduced by dragging and dropping a large table to the center of a template or by adding more columns to an existing table. The width of the template editing panel gets expanded in order to show all the content added. However, when the table gets shrinked or moved the additional width does not get automatically released again.Temporary Solution:
Check for a thick black vertical line at the right hand side of the editing pane. Move this line to the left until it hits the header box. If this still does not remove the empty page, the RDL template file needs to be examined with a simple text editor, like Microsoft Notepad. Search the entire template for data fields called "Fix Information:" and check for values greater 8 inches. Depending on the paper format and the page margins those values are potential troublemakers and should be reduced to e.g. 7.8in.
The auto-shrinking of the editing pane has been improved with revision A.01.04 of OpenLAB CDS A.01.04.
Exporting large sequence summary reports (700+ injections) to multiple output files (PDF, XLS, TXT, DOC) may generate an out-of-memory condition and the last report file in the queue may not be generated. The TXT report shows the following warnign message: The report template definition is invalid. The requested operation cannot be completed.Temporary Solution:
This problem has been observed on a 32-bit Windows 7 operating system and is less likely to happen on 64-bit operating systems. - Choose less output formats to be generated at the same time - switch to 64-bit operating systemsFix Information:
There is no fix planned.
Intelligent Reporting of OpenLAB CDS offers the option to add or select user defined report parameters in the report preview. By refreshing the preview these parameters can be applied to e.g. calculation or view filters. When generating a PDF file based on the report preview of OpenLAB CDS ChemStation, the report gets rendered again and the parameters switch back to their default values.Temporary Solution:
For user of OpenLAB Data Analysis: This problem can be avoided by printing the report to a PDF printer (e.g. ChemStation PDF printer) If external report parameters are used in calculations it is generally recommended to include the plain values in a report template section in order to allow a result verification and to ensure result consistency.Fix Information:
A fix is currently under investigation.
Intelligent Reporting of OpenLAB cDS A.01.04 or higher offers the possibilty to change report parameters interactively, if the report template is setup accordingly. When changing one or more parameters and switching from "Report Preview" to "Print Preview" (or vice versa), not all changed parameters are applied to the report. In case of multi-select boxes the mouse focus needs to moved to another field in order to submit the changed value. If the mouse focus remains in a multi-select box, the value does not get applied when switch to "Print Preview".Temporary Solution:
The recommend workflow when using interactive report parameters is to refresh the report preview before printing or switching to print preview layout.Fix Information:
There is no fix planned.
Large matrix reports may produce additional empty pages. The issue is in the Microsoft Report Viewer control and cannot be fixed in the current product.Temporary Solution:
Matrix reports easily exceed A4 or Letter sized paper formats and printing to a larger PDF paper format could be used as alternative.Fix Information:
There is currently no fix planned.
When calculating the standard deviation or relative standard deviation of multiple identical numbers, the calculation may return "NaN" rather than 0. The problem is caused by the small delta derived from the rounded result of the average calculation and the individual values. Summing up these small numbers may result in a floating point exception.Temporary Solution:
Do not apply standard deviation calculations to series of identical values.Fix Information:
There is currently no fix planned.
Changes in report templates are not immediately shown on the printout if the user does not preview the report and save the template first.Temporary Solution:
Always preview the report printout after applying changes to the template. Or save and reload the template from disk, if a preview should be avoided.Fix Information:
There is no fix planned.
A template having a matrix inside a list may show a rendering error - The value 'x' is invalid. Valid values are between '0' and 'x' This error occurs when there is large data used for rendering and due to this the matrix data spans across multiple pages.Temporary Solution:
For large result matrices it is recommended to print the results to PDF files, rather than paper. The paper format should adjusted to display large matrices on a single page.Fix Information:
There is currently no fix planned.
If LAN is connected not to LC module generating most data (i.e. to Pump in spite of DAD's presence) then interconnecting CAN connection between modules might get overloaded. Potential consecutive loss of spectra may be observed and is flagged as “Spectra runbuffer overflow” in logbook and on report.Temporary Solution:
Installation documentation for LC HW states that the LAN access should be in the module with the highest workload, i.e. by DAD/MWD – FLD – VWD – RID – Pump – ALS (in this order) and independent whether it is a 1100 – 1200 – 1290 – 1260 system.Fix Information:
n/a
Create a run table that uses valid locations and some that use 'blank' as the injection location (a valid choice in Companion). The runs associated with this location fail to run with a pop-up: 'Vial range must be in range 1-32'. This also happened for an ALS method.Temporary Solution:
n/aFix Information:
Fixed in Revision B.04.01.
An issie with a sequence has been observed, when it uses methods with custom fields. All fields have been populated before the sequence started. When the sequence is edited by adding runs or when the custom fields are checked during a run, the content of the fields is wiped. The behaviour was partially fixed in revision C.01.03. When the master methods are stored in a subfolder of the methods directory, the problem still occurred.Temporary Solution:
For OpenLAB CDS ChemStation C.01.02 a hotfix04 is available.Fix Information:
Fixed in revision C.01.04
When a new result set is created from a sequence with compound custom fields, the new sequence is missing the compound custom fields.Temporary Solution:
n/aFix Information:
Issues fixed in OpenLAB C.01.05.
When you open the sequence table to add new samples while the sequence is running, the custom fields table is erased. If you don't update it while adding the new samples, the custom fields table will remain empty and you won't be able to access it in Data Analysis.Temporary Solution:
Fixed in C.01.02 Hotfix02.Fix Information:
Fixed in C.01.03.
When printing the final Intelligent Report, not all the fields are reported (for example the field for injection line 13 is present in the Custom Fields but not in the report).Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Below are the steps to reproduce the behavior: 1. Create a method with custom field setup. 2. In the custom field set up menu, keep the Default Value field blank. 3. Create a sequence and add sample lines. 4. Select the method created with the custom fields in the sequence table. 5. Click on Custom Fields button at Sequence table and input custom field values -> Click OK. 6. Sequence table -> OK button to close table. 7. Open same sequence table again and click on Custom Field button again. 8. All of custom field values are erased and it appears blank. 9. Even if the sequence is saved after entering the custom field values, the custom field values are empty. 10. The issue occurs with both mandatory and not mandatory custom values.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.03.
The Relative Retention Time solution available on the Agilent web site in the FAQ section only works for calibrated peaks.Temporary Solution:
n/aFix Information:
n/a
Installing a lamp with tag may invalidate the detector's control setting "always use lamp" and silently modify it to "use lamp only if lamp tag is available".Temporary Solution:
Check detector's Control setting after installing a new lamp.Fix Information:
Fixed in A.02.04
1) A method XYZ.M is saved with following DAD settings: Autobalance -> Prerun 2) A sequence table is saved using the method XYZ.M and AutoBalace set to "OFF" 3) The sequence gets started and although AutoBalace set to "OFF" it is executed.Temporary Solution:
still under investigationFix Information:
none
It is not possible to reprocess data while the sequence is still in progress.Temporary Solution:
n/aFix Information:
In OpenLAB CDS ChemStation C.01.01, recalculate mode was added and samples can be reprocessed after the first run is completed.
When using "Reprocessing Copy" from the view menu, it may happen that the reprocessing session is hidden behind the Online session.Temporary Solution:
You can minimize the online session to access the Reprocessing SessionFix Information:
n/a
Data analysis will not "Autoscale" a TCD data file. AUTOSCALE autoscales to the first (smallest) peak. NOTE, there is no solvent, first peak at 0.798 minutes is very small, software identifies peak 4 at 2.158 minutes as the solvent.Temporary Solution:
Full Scale works correctly. Use Ranges can be made to work correctly.Fix Information:
n/a
A Peak Summing Table containing two segments and each segment has a Use Reference for the response factor and an ISTD Reference. The Chemstation does not report the amounts for each segment correctly.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
When loading GC-signals acquired with the MSD ChemStation, the Y-axis is labeled: mAu. This should be pA. The signal units are read directly from the rawdata file "FID1A.ch" which are written by the acquiring software. The acquiring software shows that mAU was written. In order to be correct this, it must be fixed by the acquiring SW.Temporary Solution:
n/aFix Information:
n/a
There is no solvent peak and the first peak is very small so the software identifies the wrong peak as the solvent. The Chromatogram is not autoscaled correctly.Temporary Solution:
Full Scale works correctly. Use Ranges can be made to work correctly.Fix Information:
n/a
A red X mark displays in data explorer of offline data analysis during sequence.Temporary Solution:
During data acquisition, don't display the subdirectory in the offline data explorer window.Fix Information:
Fixed in revision C.01.04
During various operations in the ChemStation (e.g. repeated manual integration of data files) a lock-up situation could occur. Internal registers were no longer accessible and the software might either no longer respond or produce error messages like "An additional register must be added to the list of locked registers"Temporary Solution:
A fix for this issue is available for C.01.03 with Hotfix03 and any other Hotfixes released after Hotfix03Fix Information:
Fixed in C.01.04
A nearly horizontal baseline is not displayed or printed completely when doing a Remote Desktop connection to a ChemStation WorkStation or to a ChemStation AIC.Temporary Solution:
NoneFix Information:
A fix for this issue is available in C.01.03 Hotfix 04 and higher The graphics library from the ChemStation revision C.01.04, which uses the latest GDI+ API and a new GDI smoothing mode, has been implemented. A variable in the ChemStation.ini file is used to toggle between the GDI smoothing mode only or the full GDI+ features. The fix affects the appearance of chromatograms. The changes do not affect results, only the display and printing of chromatographic curves and baselines.
As opposed to earlier workflows, it is no longer possible to change the method in Recalculate Last Result mode and create a report afterwards.Temporary Solution:
n/aFix Information:
This is the intended behavior.
When a result set is analyzed with a master method, i.e a method from the method path, it will use a report template from the result set. The template name is taken from the method. Since the template is not copied to the result set automatically, no report will be created.Temporary Solution:
N/AFix Information:
That is the intended workflow. Users who intend to use master methods might want to use report templates that are not part of the result set. The results of the analysis with different methods can be compared in a common template in the Review mode.
Completed a single run with manual fraction collection. In Chemstation Data Analysis, loaded the datafile and switched to "Purify". The UIB1 chromatogram does not show the fractions.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
When a sequence is partially reprocessed with a new method, the Navigation Table does not reflect the changed method. The changed method will only be displayed immediately after reprocessing, but not when loading the sequence set into the Navigation Table again at a later point in time. This implies that also the old sequence method is loaded when loading the affected runs with the option "use sequence method". Reprocessing of the complete sequence is not affected by this issue.Temporary Solution:
n/aFix Information:
Fixed in C.01.03.
A pdf report file file cannot be updated when it is open in Acrobat reader or any other PDF viewer. It can also not be uploaded to ECM while it is open.Temporary Solution:
Close PDF viewers before reporocessing or printing reports.Fix Information:
C.01.03 comes with a report file viewer that works on a copy of the report file.
The HTML Help requires Internet Explorer Rev. 6.0.2.800.1106CO or higher. If a lower IE revision is used, it is possible that executing the 'Reference' button on the top of the Help screen gives an error message: '...a Script Error ...'Temporary Solution:
n/aFix Information:
n/a
The manual needs updated to include more information on the peak width calculation. Is the peak width calculation meant to approach, approximate, or enhance the standard half-height method of calculation.Temporary Solution:
n/aFix Information:
n/a
The CDS Supported Instruments and Firmware Guide does not list the correct firmware for the 7683A injector. The correct firmware is A.10.07.Temporary Solution:
n/aFix Information:
CDS Supported Instruments and Firmware Guide has been updated with the correct firmware revision in OpenLAB C.01.05.
Sequences that have been added to the queue may end up in an unexpected location on disk, when the Easy Sequence file name contains the string ".es". The expected file name ***.es.es was not handled correctly.Temporary Solution:
When saving an *.es file, do not include .es in the file name. The extension .es will be appended automatically.Fix Information:
Fixed in revisions B.04.03 SP1 and C.01.03, by not allowing .es in an Easy Sequence file name.
The Queue Planner will not run sequences using methods with different injection sources such as PAL Autosampler and valve.Temporary Solution:
n/aFix Information:
n/a
During reprocessing (especially with complex reports as e.g. the Performance reports) the menu items and icon for locking the ChemStation are frequently not accessible because ChemStation is busy. The user has to wait for a point of time where ChemStation is not busy.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.05.
If a ChemStation master method, sequence template, or data set is downloaded from ECM and afterwards the copy in ECM is moved to a different LCDF path, the local copy in the ChemStation will not be linked to the item in ECM anymore. This will lead to errors during automatic data upload, local method/sequence update from ECM, viewing the ECM Properties, etc.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
When a method is downloaded from ECM and available in ChemStatioon Explorer. In this case it is possible to delete the method from the local disk via the ChemStation File menu. In case the method was currently checked out, it will not be checked in automatically.Temporary Solution:
Check in the method interactively.Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02
Acquisition only runs do not write instrument and column parameters to the sequence.acaml or sample.acaml file (which are needed for reporting with OpenLAB Intelligence Reporter). The .acaml files are only written when data analysis is performed, too.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.02
The "Operator Name" token for the ECM LCDF uses the user's full name. This name can contain any special characters (@, &, ., etc., or local language characters). Many of these characters cannot be used for LCDF folders in ECM, which results in data not being transfered to ECM.Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
The ChemStation OpenLAB option offers the option to calculate the peak performance parameters each time a ChemStation report gets generrated. Without having this option selected, ChemStation calculates those numbers on demand only. This is heavily dependant on the ChemStation report style selected (Short, Performance+Noise, Extended Performance report, etc.) If the option is selected, the performance parameters aof a standard perfromance report are calculated and written to teh ACAML result file. However, the noise and extended performance calculation results are not generated.Temporary Solution:
The calculation of all extended performance and noise parameters can be triggered by the following procedure: - turn off the "Always calculate performance parameters" option in the ChemStation Open LAB configuration - add the following Customized Data Analysis Command to the run time checklist of the method in use: macro C:\Chem32\REPSTYLE\shortutl.mac; print USR_CalcPerfNoise$() (Where c:\Chem32 needs to be replaced with the current ChemStation installation path) - select a report style e.g. SHORT, that does not trigger the performance or noise calculation on it's own. This procedure would generate the maximum content of the ACAML result file in respect to peak perfromance parameters.Fix Information:
This is fixed in OpenLAB CDS ChemStattion Edition Rev C.01.01
The following scenario can occur when autoamtic upload to ECM after "any data modification" is not enabled. When the method is modified, but the data set is not uploaded when switching to the next sequence container, the modified container will not be marked as changed.Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02.
The chromatogram data not included in the ACAML file when a sequence is acquired in "Acquisition only" mode.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
The storage path that is entered in the control panel is not supposed to be the default storage location in ChemStation. The storage path is related to EZChrom CDS projects.Temporary Solution:
NoneFix Information:
No fix planned.
When a Sequence or Single Injection Run cannot be transferred to the OpenLAB ECM
storage, the transfer process is paused and an error message is displayed in the
Queue Management tool for the first item in the queue.
Users with the necessary privilege can start the Queue Management tool and resume
the transfer process. After a successful transfer the item is removed from the
queue otherwise it remains in the queue.
We discovered that items with the following error message “Object Reference not set
to an instance of an object” appear to the end user as being transferred after
resuming the queue but they are in fact deleted from the queue although the transfer
is failing again.
Subsequent reprocess of such sequence(s) will not pause the queue and will not be
transferred to the OpenLAB ECM storage.
If the option "Cleanup Data on Shutdown" is selected, the local copy of the data will
also be deleted when shutting down the ChemStation.
In such a case, the data is lost and can't be recovered.
The error “Object Reference not set to an instance of an object” can be caused
among others by:
• Incorrectly configuring the ECM Authentication in the OpenLAB Control Panel
• Manually appending text at the end of the Server string in Transfer Settings
of the ECM Preferences
• Using “Tokens” (Instrument Name, Instrument No, Operator or Computer Name)
for the Drawer and/or Folder or manually typing the Drawer and Folder in
the Transfer Settings and not having the privilege to add folders in OpenLAB ECM.
• Users having a project role in OpenLAB Control Panel where the privilege
“Save to ECM” is not granted.
• Selecting in Sequence Parameters a Remote Data Path having a “/” in the name
(e.g. 2011/07)
Temporary Solution:
As an immediate action and until a solution is available, we recommend on all
OpenLAB CDS ChemStation Edition systems (Workstations, Networked Workstations and AIC’s)
connected to ECM to:
• Start the ChemStation Administration tool
(Start | Programs | Agilent Technologies | OpenLAB CDS ChemStation Edition)
• Select the option “Use these settings for all Instances on this computer
• DO NOT select any of the checkboxes in the Automatic Data Transfer Settings
and in the Transfer Management Settings.
• Select the two checkboxes in the Audit Trail section
This will disable the transfer to ECM for all Instruments configured in the computer.
Fix Information:
A solution for the defect is currently being worked on. The fix will be made available as soon as possible.
The LCDF storage path for saving methods to ECM is user specific.A user that logs on to ECM from different instruments will always save to "his" path by default. While this is the intended behaviour, it can be confusing. Method developers would prefer a workflow that makes the path instrument specific.Temporary Solution:
Check the folder path before uploading methods. Consider including the instrument name in the folder name.Fix Information:
Fix considered for a future revision.
On a stand alone installation, the ECM authentication mode has been set. Due to a hardware problem (or something else), the current ECM authentication mode must be switched to a different ECM server with a different account. When the user tries to change the ECM authentication to a different ECM server and enters the new hostname, the administrator cannot be selected. The user is back on the previously configured ECM system.Temporary Solution:
A workaround is to set authentication mode to none and then go to the new ECM server.Fix Information:
n/a
Migrated ChemStore data acquired with ChemStation rev. A..01.03 and earlier does not get automatically reprocessed when downloading the files from ECM. The automatic reprocessing is requried to populate the ChemStation navigation table. The reprocessing stops with error message: Sequence was aborted due to a file access error. Please restart your sequence. Restarting the sequence reprocessing does not resolve the error.Temporary Solution:
Affected sequences cannot be reporcessed automatically, but the runs are available as single runs. Re-calibration using the calibration samples is possible interactively. Result review can be automated by using the auto-stepping feature of recalculation mode.Fix Information:
This defect will be fixed in ChemStation rev. B.04.03 Service Pack 1 as well as rev. C.01.04.
When a method is downloaded from ECM and this download is repeated in the same session an error message appears: The Server Path is not identical. This message is incorrect. One of the two path strings has a leading backslash, whole the other does not. There is also a warning that the method already exists. This message is correct.Temporary Solution:
Repeat the download attempt. It works the second time.Fix Information:
Fixed in OpenLAB C.01.05.
If the OpenLAB ECM is disconnected or not reachable. The upload from the ChemStation “Import/Commit commands” end up in the queue, the execution from the queue creates new document centric audit trail entries at the end of the upload process. These entries can't be executed successfully, remain in the queue and stop further uploads. This behavior is mainly seen with partial sequence acquisition.Temporary Solution:
N/AFix Information:
This issue affects the chemStation revisions C.01.02, C.01.03 and C.01.04
ECM user licenses not released when user is changed in OLCDS ChemStationTemporary Solution:
n/aFix Information:
n/a
When e.g PDF files are manually uploaded to ECM, the file type is *.* This default type is appended to the suggested file name, when the back end storage location is changed. Upload of a file with a name like " testing*.* " will fail.Temporary Solution:
Delete the *.* extension from the file nameFix Information:
N/A
When selecting ECM -> Update Methods.. in the ChemStation Method and Run Control View, the dialog box locks up if more than a certain number of methods are linked to ECM which can be as low as 20.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
There is no distinction between a connected user and a logged-on user on the client. In file information, the Operator shows disconnected.Temporary Solution:
n/aFix Information:
n/a
For data analysis only instruments, when a session is taken over, no warning message pops up. This creates problems in the lab because all of a sudden the session is lost in the middle of work.Temporary Solution:
n/aFix Information:
n/a
If user A submits multiple sequences into the queue and the user is switched to B while the first sequence executes AND the sequence parameter contains a user token for the upload path THEN the first sequence is uploaded to a path for user A (user is resolved at beginning of sequence acquisition) and all other sequences are uploaded to the path for user B (these acquisitions begin after the user switch).Temporary Solution:
n/aFix Information:
n/a
By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition. The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically. Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.Temporary Solution:
Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.Fix Information:
n/a
On a localized chromatograpy data system the driver for ELSD detector throws an exception and fails to show or store the detector's signal data. On an English system the ELSD driver's operation is not affected.Temporary Solution:
n/aFix Information:
For a localized (e.g. Chinese, Japanese...) system a patch for the ELSD driver is made available.
With Rc.Net driver internal filters of ELSD G4218A get confused by change of "Sample Time" during run.Temporary Solution:
n/aFix Information:
n/a
Customer uses paths like: C:\Chem32\1\methods\customer1 and C:\Chem32\1\methods\customer2, etc. If samples of two or more customers have to be analyzed, it is not possible to use methods from different folders.Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01. Subfolders for Methods are supported in this release.
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence. This occurs on ChemStation ver. - B.04.01 SP1 - B.04.02 - B.04.02 SP1 - B.04.03Temporary Solution:
n/aFix Information:
n/a
Start up Chemstation, but no instrument or sampling diagrams are shown. FetchCPVar Cannot Talk to Application Error occurs. Systems use 7683 ALS.Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03 DSP1.
On the signals screen, need the ability to collect partial data for specified times.Temporary Solution:
n/aFix Information:
n/a
Open an online session, initial or any subsequent try. This results in a benign pop-up error: 'FetchCPVar cannot talk to application'.Temporary Solution:
n/aFix Information:
Fixed in B.04.03 DSP1 [54].
Column compensation (comp) cannot be started from the Data System with 68xx RC.NETTemporary Solution:
Column compenstation must be started via the GC keyboard.Fix Information:
n/a
G1888A configuration Set Handshake to Abort Start Sequence Open GC oven door to prior to end of vial equilibration time. GC Logbook records: G1888 - Job Stopped prematurely. However, Detailed Vial status never updates for vial one. It will say Preparing for injection is 'In Process' and is stuck at 40%. Status will go to Ready. ChemStation software has to be manually aborted. An error in headspace (HS) that should abort the datasystem, no longer aborts. Abort does work using a 7697A.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Unlike classical drivers the RC.Net drivers did not support partial storage of data for a specified run time.Temporary Solution:
It is possible to load data file partially, by using Signal Options from the Calibration menu.Fix Information:
Fixed in C.01.03
For runs using data rates 1 Hz and below , you get a macro error "creation of a data object failed" when trying to load the data. In addition, there is the following error in the sequence log during acquisition: CP Macro ======================================== 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro In Command deltab in macro ExecHook 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro at line 1176 in file C:\CHEM32\CORE\SIGNAL.M> 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro CX 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro Error 16548 occurred: 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro An additional register must be added to the > 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro list of locked registers. 13:06:40 06/27/11 70 41e0 4e086440 0 CP Macro ======================================== 13:06:40 06/27/11Temporary Solution:
n/aFix Information:
Fixed in revisions B.04.03 SP1 and C.01.03
A manual injection is done and if the STOP on the 7820A front panel is selected, the GC stops (and resets the initial GC settings) but the chemstation continues. When the STOP button is pushed on the 7820A, the "current logbook" reports the end of run correctly, and reports the system is running data analysis but does not go into Data Analysis right away. If this is a 30 minute run, and the STOP button is pressed on the 7820A after 10 minutes, AND YOU WAIT 20 more minutes until the Chemstation gets to the end of the run, a report is generated. The system appears to hang up waiting on the expected run time.Temporary Solution:
Workaround is to use STOP run from the software.Fix Information:
n/a
The GC does not wait for the inlet pressure to increase to and stabilize at the pulsed pressure before injection and data acquisition. The GC starts and injects sample prior to the time at which the system had met its defined inlet pressure. Pulsed split/split less inlet typesTemporary Solution:
If the 6890 is operated from the GC keyboard, the GC has the expected result.Fix Information:
Fixed in revision C.01.03.
7890A with A.01.12.1 firmware connected to Chemstation B.04.03 and C.01.02 and when the 7890A reached the run time (10 minutes) the GC stopped and reset the setpoints, but the chemstation continued and did not stop. It is possible to abort the run. Created a new method, same problem. Noticed that the dirty method flag was set every time even when there were no changes. The 7890A is finished the run and GC ready state is Ready but ChemStation Status is BLUE, Run in progress, DATA ACQUISITION, but nRdy Wait. GC Connection state shown in Status is Online MONITOR which is WRONG if the ChemStation is collecting data.Temporary Solution:
Workaround: Power cycle the 7890A, shut down chemstation, power on GC, restart chemstation and everything worked.Fix Information:
This issue could not be reproduced in C.01.03.
From the Configure Instrument user interface, there are configurable modules such as 7890A, 7697A, 7694B. If the GC’s location in the “Selected Modules” table in not listed first or as module 1, k’ is not calculated. For example, if the 7697 is module 1 and the 7890A is module 2 then the requests for column info would fail and k' (Capacity factor) is not calculated. Column information will not be included in reports.Temporary Solution:
The order in the module table is set by the order in configuration. The workaround is to re-order configuration so that 78xx is first.Fix Information:
Fixed in OpenLAB C.01.05.
The valid vial range is determined by the injection source of the loaded method, not the method in the sequence. As soon as the HS sequence loads, it fails with: '--------------------------- Error --------------------------- Invalid location found in FRONT sequence, line 1 There may be others --------------------------- OK ---------------------------' The Valve method has reloaded which is invalid for this HS sequenceTemporary Solution:
n/aFix Information:
n/a
A result set that is built from individual runs does not provide the full functionality for GC runs. The resulting sequence does not support front or back injections. Reprocessing is not possible, because there is a mismatch with the method.Temporary Solution:
N/AFix Information:
Fixed in revision C.01.04.
When a GC method is run that uses a vial number not valid for the configuration, an error message "Undefined symbol: License$" appears.Temporary Solution:
The error message is misleading. Check the vial range whether it makes sense for the selected injection source.Fix Information:
Fixed in revision C.01.04.
Windows XP SP3 (and earlier) is not compatible with a JIS2004 font. Since Intelligent Reporting creates PDF reports with such types of fonts, the content of those PDF files is not shown properly on such systems.Temporary Solution:
N/AFix Information:
Windows XP needs to have the Microsoft fix ‘KB927489’ installed to support JIS2004 fonts properly. KB927489 can be downloaded from http://www.microsoft.com/japan/windows/products/windowsvista/jp_font/jis04/default.mspx
On Chinese or Japanese CDS only the 1290 Infinity Flexible Cube G4227A does not work with High Performance Autosampler G1367E. On English CDS the operation of 1290 Infinity Flexible Cube G4227A with High Performance Autosampler G1367E is not affected.Temporary Solution:
n/aFix Information:
n/a
Installed B.04.03 on a DELL connected to instrument-1, a 6890, instrument 2, a 7890, instrument 3, a second 6890, and instrument 4 a second 7890. Instruments 1,2, and 4 work correctly. Instrument 3 gives the following errors on start up, Error "no string 8002" when starting ChemStation. After pressing OK debug message "Abort this copy of alice? Yes is normal abort, No is Fatal Error, Cancel continues" comes. No closes ChemStation.Temporary Solution:
Workaround is to delete all of the files in the HPCHEM\3\TEMP DIRECTORY.Fix Information:
Fiixed in C.01.03
Sporadically the launch of an online session will not proceed in a networked setup.Temporary Solution:
Launch the online session a second time. The second attempt usually works.Fix Information:
No fix planned
Users with the privileges 'View instrument or location' and 'Run Instrument' is configured in OpenLAB Control Panel can't start instruments configured on ChemStation AIC's. The following error message is displayed: " Can't launch the instrument"Temporary Solution:
As a temporary workaround, assign the role "Instrument Administrator" or the privilege "Manage Instrument or Location"Fix Information:
This issue is fixed with the software revision C.01.03
In ChemStation revision C.01.02, an instrument configured as G1600 CE cannot be launched. The startup process hangs during "Initialize Instrument".Temporary Solution:
n/aFix Information:
n/a
The localized readme.txt of the CDS ed. are present on the root of the CD/DVD holding the software (OpenLAB CDS ChemStation ) . Nevertheless for the localized readme.txt no automatic installation of the file to the harddisk will take place, the link in the programs folder links to the english readme.txt.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
It was not possible to run any 7890A from either the client or AIC. It was discovered that communication ports were not opened upon installation of the AIC version of C.01.02. file "Exceptions.AgtChemRCNetAdapter.Online"Temporary Solution:
Manually create and open a port.Fix Information:
n/a
The headspace sampler will run a method, but it will not run a sequence if the methods are saved to a location other than C:\ This defect affects all headspace hardware models running with B.0x.0x series headspace Sampler Software.Temporary Solution:
The C Drive must be selected.Fix Information:
Point patch available for B.04.03 SP1 [87]. This is fixed in C.01.04 OpenLab CDS.
In the tailing part of a peak, it may occur that the inflection point (needed for the calculation of the tangent peak width) is not correctly determined. The problem results from the algorithm applied to find the inflection point: It is done by the first/second derivate of the peak shape but at the tail of the peak, the derivates of the "noise" on the signal may have about the same value as the tail it self. As the calculation looks for the minimum/maximum value of the second derivative, a wrong point may be identified for the inflection point.Temporary Solution:
n/aFix Information:
Fiixed in C.01.03
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
When using "New Exponential" as tangent skim mode, incorrect areas may be calculated in case of tangent-skimmed peaks. In these cases the tangent-skimmed peak will not be integrated at all, while the main peak is reported with an area that is too large. The defect is most pronounced when the tangent-skimmed peak is very small. The likelihood of being impacted by this issue also decreases substantially with increasing number of data points collected per peak. If detector settings are chosen according to the recommendations in the ChemStation on-line help (data rate, detector peak width) the peaks are defined by 15+ data points and data collected with these methods are not affected by this issue. The settings of the instrument specific default methods (DEF_LC.M, DEF_GC.M) take care of this, with the exception of older instrument hardware (e.g. GC 5890, 5 Hz default sampling rate), high data volumes (LC/MSD, full scan mode), and fast chromatography techniques (Fast GC, LC Rapid Resolution). The high information density produced by these detection techniques always require adjustment of the default detector sampling rates according to the recommendations in the ChemStation on-line help.Temporary Solution:
Do not use "New Exponential" tangent skim mode, but choose a different mode.Fix Information:
Point patches for software revision C.01.01, B.04.03 and B.04.02 are available for download on www.chem.agilent.com/EN-US/SUPPORT/DOWNLOADS/PATCHES/MULTIINSTRUMENTCHEMSTATION/Pages/default.aspx Fiixed in C.01.03
When creating area sum integration event, the 'off' time for one group and the 'on' time for the next group must be separated by a minimum of 0.06 minutes. In a complex chromatogram this is a significant loss of data.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
The annotation of the tangent peak width in the performance tool is done by highlighting the part the baseline defined by the intersection with the tangents. It has been considered as misleading, because the drawing suggests that the tangent width is the length of the baseline segment. The mathematical truth is the projection of the segment on the time axis. This is what the tool calculates.Temporary Solution:
The calculation is done correctly. All other peaks widths are drawn as horizontal lines. Only the tangent width is drawn following the baseline which better describes how the tangent width is constructed.Fix Information:
N/A
In C.01.04 one new timed event appeared: Height Reject. The expected behaviour for an event parameter would be a numeric value for the minimum peak height. Instead a selection box allows values TAN and DROP. The timed event Height Reject should not have been in the selection list. Do not use Timed height reject in C.01.04.Temporary Solution:
Don't use this event. Timed Area Reject works as expected.Fix Information:
Fixed in C..01.05 by adding the event with the expected functionality.
The integrator algorithm annotates many compound peaks with Peak Code S, which should only be given to solvent peaks.Temporary Solution:
n/aFix Information:
No fix planned
Manual integration may become slow, if many calibrated peaks are not found in the chromatogram and the Results Audit Trail is switched on. This behavior is especially pronounced in distributed ChemStation systems.Temporary Solution:
Fixed in C.01.02 [17] HF03Fix Information:
Fixed in revision C.01.03.
Japanese characters from a saved ChemStation .pdf report are garbled when copied and pasted to another application (e.g. Notepad, Word, Paint, etc.).Temporary Solution:
n/aFix Information:
No fix is planned for this issue.
Configure 7697A to 7890A. Close and reopen session and configure a PAL, as well. With the headspace (HS) method still loaded, Run Method to the still cabled device. This runs, but logs message to the logbook: 'G3382AA 1 Method not downloaded to the device, because> 12:12:53 03/18/11 0 1da9 4d838485 0 G3382AA 1 the settings are not compatible with the de> 12:12:53 03/18/11 0 1da9 4d838485 0 G3382AA 1 vice configuration. 12:12:53 03/18/11 67 41e2 4d838485 0 Method Instrument running sample Vial 1 (front) 12:12:53 03/18/11 0 1da7 4d83848f 0 7697A 1 Sequence Started 12:13:03 03/18/11'Temporary Solution:
n/aFix Information:
n/a
Intermittently, the LC/MS ChemStation may not be able to open Logbook files, either through the GUI or from the "View | Logbook" menu item. The inability to open Logbook files has been traced to an incorrect Manufacturing Date entry on the LC/MS instrument SmartCard.Temporary Solution:
As a workaround fix, the Logbook file can be opened independently from ChemStation as a text file: in the Windows Explorer environment, browse to the C:\Chem32\1\TEMP directory and open the INSTR1.LOG file.Fix Information:
To correct this issue, ensure that the LC/MS instrument SmartCard has the proper Manufacturing Date encoded in the proper format - an Agilent representative may be required to complete this task. Fiixed in C.01.03
When a method is loaded that requires resolution, the user has the possibility to cancel loading of this method and return to the previous method. In this case, a macro error is observed "error occurred during loading method", and the previous method is loaded, but is marked as changed.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
CTC portion of the method printout does not include the Setup Injector Overlap settings.Temporary Solution:
n/aFix Information:
n/a
Description: 1. Start GC ChemStation with 6890 RC driver 2. Select Method -> Export eMethod... 3. Generate eMethod file 4. Select Method -> Export eMethod... 5. Import eMethod file created by step3 The error message is displayed and can’t import eMethod file.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.01.
ChemStation hangs up during Edit Entire Method when starting this process via the UI (system diagram). This happens after the Signal Details dialog.Temporary Solution:
The problem is not observed when starting Edit Entire Method via the corresponding entry in the Method menu.Fix Information:
Fiixed in C.01.03
The printout for methods containing an RID can contain incorrect values. For a selected peak width of 0.05 or 0.01 the report will read 0.5 or 0.001 respectively.Temporary Solution:
No temporary fix available . This is a typo in the value list. Peak widths of 0.5 and 0.001 do not exist in the selection list.Fix Information:
Fixed in C.01.03.
An error occurs when selecting “Export e-method”. The error that is displayed on the bottom left side of software is “Undefined symbol: PrDARep”.Temporary Solution:
n/aFix Information:
n/a
Saved a method. Open the GC edit parameters to review a setpoint and made no changes or changed the view. Selected OK and the method modification flag indicates a change was made. No changes were made. This does not occur with the Classic Driver.Temporary Solution:
n/aFix Information:
n/a
Save method on D drive, and start sequence with headspace or PAL Sampler.
GC stays in "Prep Run".
headspace and PAL Samplers do not work.
Headspace or PAL Sampler sequence does not start when method is saved to the D drive
This defect affects all of the B.0x.0x series headspace and PAL Sampler software.
It is possible to run a method.
Temporary Solution:
Workaround is save and use methods from the C:\ drive.Fix Information:
Point patch available for B.04.03 SP1 [87] Patch has been tested with: G2181BA ChemStore rev. B.04.02 SP2 G2183BA Security Pack rev. B.04.02 SP2 Headspace Control Software B.01.02.1
Create a method. It works correctly. Save method as "TEST1 .M". Note there is a space after the "1" and before the dot. Run the method and it works. Select Sequence, then Sequence Table and in the method name use the down arrow to list the methods, all the methods appear including "TEST1 " with a space. Select "TEST1 " and click on Inj/Vial and you get an error message. . . Method TEST1.M (no space) in sequence table does not exist in current method directory. TECHNICALLY this is true, but the chemstation changed the name and took out the blank.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
ChemStation online/offline handles special characters in the following way: (1) The (dot) '.' is accepted when entered once at the end of the name. However, more than one ‘.’ at the end of the file name is not accepted. (2) The following characters are not accepted but there is no error message: '*' and '?'. This could be because they are wildcard characters, but an error message is encountered when trying to accept '%', '<', '>', '+', '=' and ',' as special characters. Details will be outlined in the ChemStation Manual.Temporary Solution:
n/aFix Information:
n/a
A user runs a single sample. When switching from the Method & Run Control view to the Data analysis view of the ChemStation, the method name displayed in the toolbar is DEF_GC.M, for example. The method used to run the sample in Method & Run Control is not loaded in data analysis. A different method is loaded in Data Analysis. Methods of Method & Run Control view and Data Analysis view are kept separately.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
The OpenLAB Chemstation is very, very slow to load a method or save a method. The method modification flag takes sometimes 5 or more seconds to display. So if you select save before the ‘dirty’ flag is displayed, and load another method, this newly loaded method becomes dirty (it was not the one that was edited). Also, when the system is busy, the menu items were not grayed preventing a user from doing an operation.Temporary Solution:
n/aFix Information:
Fixed in C.01.03.
This problem is reproducible if you load DEF_GC.M. Once loaded, execute 'Edit Entire Method' and check 'Data Anylsis'. Close all dialogs with OK without doing changes. Then method becomes dirty [Error!]. Once the modifed DEF_GC.M is saved, the problem does not occur. Also happens in DA View. If DEF-GC.M is loaded and "Specify report" dialog is entered and closed without a change, then the Method dirty flag is set.Temporary Solution:
n/aFix Information:
n/a
Install Headspace B.01.01. Create a three line Sequence using the ALS Run - all is okay Create and run a three line Headspace sequence ChemStation hangs. ALS method displays in Headspace StatusTemporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
It is not possible to take a method from a ChemStation running one language and move it to a ChemStation running a different language (English -> Japanese, for instance) without visiting this user interface and checking / fixing any values.Temporary Solution:
n/aFix Information:
n/a
Removed the HS in Control Panel. Open online session and select Compare at method download. The HS is not seen as a module. Method Load Options Compare does not include the HS as a module or detect any difference in the methods. Download to instrument and no method resolution occurs. The method is not marked as modified, though the Instr.log shows the configuration changes.Temporary Solution:
n/aFix Information:
n/a
GC with 7693 tray and dual injectors: create and save a method for that configuration. Close session, unplug the tray, select to Configure... in Control Panel and Get GC Configuration. Open an offline session of the instrument. User cannot save any method now. OK to the Method Editor and get: '--------------------------- --------------------------- Method is not consistent with the configuration. Reason: Configuration mismatch: Tray --------------------------- OK ---------------------------'Temporary Solution:
n/aFix Information:
Fixed in revision C.01.05.
On a GC system with no injectors (manual injection), select injection source from menu and select Back or Dual, ChemMain.exe error occurs. 7820A , 7890A, 6890 have no injectors installed.Temporary Solution:
Fix included in C.01.04 Hotfix03Fix Information:
Fixed in OpenLAB Revision C.01.05.
When the user select any items from “Check method sections to edit” and click on “OK”, there is no action. The dialog box remains at its position.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
If running a sequence, the "Specify Report" dialog can be opened from the dashboard. If this dialog is open for a longer time, it can happen that the sequence has already loaded the next method. If the dialog then is closed, the currently loaded method is overwritten with the other method's report settings without any warning.Temporary Solution:
n/aFix Information:
This issue will be fixed in revision C.01.03: as long as the dialog is open the method is paused.
The Deconvolution tools are not available in LC/MS ChemStation even if BioAnalysis Tools are configured for the instrument.Temporary Solution:
n/aFix Information:
n/a
Automated Delay Calibration of the MS signal for Fraction Collection is not supported in the RC.Net environment. In the Classic Environment, Automated Delay Calibration of the MS signal for Fraction Collection is supported and accessible through the Diagnosis View from the Maintenance | Fraction Collector ... | Delay Volume Calibration ... menu item.Temporary Solution:
To perform a Delay Calibration of the MS signal for Fraction Collection, configure the system to operate in the Classic Environment. This may require setting some LC Modules into "emulation mode" to be recognized in the Classic Environment. Once the Delay Calibration(s) is/are completed, make note of the Delay Calibration value(s) and revert the system to RC.Net environment if so desired. For systems that cannot be reverted to operate in the Classic Environment to perform Delay Calibration, please contact Agilent support for assistance.Fix Information:
This issue will be resolved in a future release.
In MSD Tune View, it is possible to manually load a tune file from the File | Load MSD Tune File... menu item. If the "Cancel" button is clicked in the subsequent Dialog Box, the tune file ATUNES.TUN will be loaded, regardless of what tune file was previously loaded. The loaded tune file is displayed in a drop-down box below the menu bar.Temporary Solution:
In the event that the tune file loading process is cancelled as in the Problem Description, please ensure that the desired tune file is subsequently loaded in MSD Tune View. Note that this problem does not impact data acquistion, as each unique data acquistion method will load the tune file that is specified as part of the method.Fix Information:
n/a
After the HPLC experiences a hardware error (such as a liquid leak), the acquistion method/sequence will terminate. However, the LC/MS may remain in a run state.Temporary Solution:
If the LC/MS remains in a run state after the HPLC experiences a hardware error, the ChemStation software must be closed and re-openned. This will reset the LC/MS status to ready and bring it out of the run state.Fix Information:
26-Apr-2012 Susanne Kramer, Heiko Fessenmayr, Achim Teubner, Uwe Nassal, Katja Johann
When a sequence has been run with cyclic calibration, it is not possible to change the sample type for the calibration runs to Sample. When this is done and the sequence is reprocessed, the sequence structure is corrupted. The sequence cannot be loaded in the Navigatio Table. The Navigation Table will not be usable and the ChemStation has to be closed and reopened. In Cyclic calibrations the structure of the sequence, the order of runs and the file naming depend very much on the sequence table setup. Later changes will cause conflicts.Temporary Solution:
To reconstruct the original sequence the batch file needs to be removed from the sequence folder. Clear the Navigation Table and load the individual runs from the file menu. Then create a new result set from the runs using the functionality from the Sequence menu.Fix Information:
N/A
The column Date Time in a Navigation Table for single runs can have incorrect entries, if the run has been aborted conditions like "No valid next position". Aborted runs do not contain the file that stores the acquisition time. The time stamp in the Navigation Table and reports is taken from the last correct run state from the instrument. The situation has not been observed for runs that have been aborted by user interaction or hardware errors.Temporary Solution:
Make sure to not exceed the range of available vials. Use time date tokens for the data file name.Fix Information:
No fix available.
Printing fails without error message, when the length of printer name plus port name exceeds 100 characters. For a TCP/IP port the printer name should not exceed 87 characters.Temporary Solution:
Use a shorter printer name.Fix Information:
n/a
Peak Purity Report Template is not available in Intelligent reportTemporary Solution:
n/aFix Information:
n/a
The likelihood of an irrecoverable communication loss between instrument, AIC and OLSS will increase, when the Application Experience service is running on the AIC.Temporary Solution:
To increase stability of communication it is recommended to disable the Aero look on the AIC, install the SP1 for Windows Server 2008 R2 and disable the Application Experience Service See also: http://support.microsoft.com/kb/902196Fix Information:
N/A
The internal command "Macro" can be used to load a macro file from disk. Typing macro from the command line opens a dialog that points to the Chem32\Core\ folder and allows selecting *.mac and *.mcx files. In C.01.04 *.mac files are no longer selectable.Temporary Solution:
Use the macro name and its full location as parameter. Example: Macro shutdown.mac (equivalent to Macro C:\chem32\core\shutdown.mac) Macro C:\chem32\core\addon\install.macFix Information:
Fixed in revision C.01.05.
Occasionally the command MACRO that is used to select and load macro files from disk is not working properly. In some casesonly *.mcx files could be loaded, but no *.mac files. There have been other cases where typing MACRO from the command line caused the ChemStation program to terminate.Temporary Solution:
Use the command with the extended syntax. Rather than selecting the file from a dialog box, use the full path of the macro in the syntax. Examples: macro shutdown.mac or macro U:\Addons\mymacro.macFix Information:
No fix planned on the B revisions.
The ChemStation scheduler cannot establish a connection with instruments on an AIC. When a command is processed, the message "Instrument not found" will be listed in the result column.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB revision C.01.05
When a Valve is configured for Channel A in PVC the visual presentations of the BottleFillings of ChannelB, ChannelC and ChannelD are wrong. It will not show the BottleFillings of the corresponding Channel but the BottleFilling of Channel A Valve Pos 2 etc.Temporary Solution:
noneFix Information:
Will be fixed with the Rc.NET Drivers Rev . A.02.06
LC Driver version: A.02.06 [363] (Test application/Install on Japanese ChemStation) 1. Open Instrument configuration dialog 2. Select Flexible Cube 3. Open Flexible Cube Configuration dialog 4. Clear the “Pump and Solvent selection valve installed” check box, then press OK 5. Re-open Flexible Cube Configuration dialog 6. Thick that check box and press OK 7. Open Flexible Cube Configuration dialog again -> check mark is goneTemporary Solution:
Workaround: use autoconfig to set up the FlexCubeFix Information:
Fix in the next rev RC.Net A.02.07.
1. Working with a G4220A pump with ISET enabled. 2. The emulated pump was: G1312B no mixer. 3. Set up a timetable. 4. Added a G1160A valve into the system. 5. Reconfigured the instrument and setup a pump-valve-cluster. 6. During startup selected "Save instrument settings in new method". 7. Enabled ISET and selected G1311A pump this time. -> the timetable of previous configuration still exists, but on one line "95.1%A and 5%B" is displayed instead of "95%A, 5%BTemporary Solution:
Select Method -> New MethodFix Information:
Solved with RC.Net A.02.07
The following defect affects the usage of statistical functions as part of aggregator functions: When using report template editor of OpenLAB Intelligence Reporter rev. A.02.01 (with or without SP1) or OpenLAB CDS A.01.01 (ChemStation or EZChrome Edition) the results of standard deviation and variance calculations are switched. The standard deviation function calculates the variance, whereas the variance calculation returns the standard deviation. Note: This defect does not affect simple table column statistics, but the aggregator functions only.Temporary Solution:
Add the variance calculation and change the field label to STDV and vice versa. Note: After applying the hot fix described in the next section, the report templates need to be adjusted if the workaround has been applied.Fix Information:
This defect is fixed in OpenLAB Intelligence Reporter A.02.01 Hot Fix 4 and OpenLAB CDS A.01.01 Hot Fix 2. The hotfixes are available for download on https://www.chem.agilent.com/en-US/Support/Downloads/Pages/default.aspx
With OpenLAB CDS ChemStation A.01.04 a new time stamp for calibration table changes got introduced. The data analysis method (DAmethod) timestamp does no longer get update by a simple re-calibration. The introduction of the new time stamp allows a more granular tracing of DA method changes.Temporary Solution:
n/aFix Information:
Existing calibration report templates need to be updated to show the calibration table modification time stamp instead of or in addition to the DAmethod change time stamp.
If the number formatting scheme "1234.000" is used the selected number of digits is not shown in case the last digit is zero. The number 1.340 is displayed as 1.34, which does not contain the same information as initially intended. This defect afects also the rounding mechanism used for peak labels.Temporary Solution:
Use formatting scheme "1,234.000", which is not affected by the defect. For chromatogram labeling (e.g. retention time), the expression FORMATNUMBER(Fields!PEAK_RETENTIONTIME.Value,3,True,False,True) can be copied to the expression editor of the peak lable. This syntax is taken from the RDL template file containg a result table with the correct formatting of the retention time column. The FORMATNUMBER command is not available in the expression builder, but still a valid and supported command. It can also be applied to other data fields containing numbers.Fix Information:
There is currently no fix planned.
The following defect affects the Report Template Editor of OpenLAB CDS A.01.04, OpenLAB ECM Intelligent Reporter A.02.04, as well as OpenLAB Data Analysis A.01.01: When using a string to number conversion function (val()) on a result table column, the RSD calculation on this column does not return a number. On top of that, the report temaplate gets partially corrupted and cannot be loaded to the Report Template Editor. This defect mainly affects calculations using custom fields/custom parameters as these numbers are stored as string and need to converted to numbers, first.Temporary Solution:
- Create a backup copy of the affected report template file (*.RDL - Open the RDL file with a text editor, like Microsoft Notepad - Search for the string "/Avg(val" and verify the expression reflects a division of a standard deviation by the average of a data field - Delete everything from "/Avg(val" until ")*100" - Search for the next string "/Avg(val" and repeat the delete operation - save the template The corruption of the temaplte is resolved now. As the calculation of the RSD has been modified it appears now as the absolute standard deviation in the column properties/Summary calculations dialog of the Report Template Editor. To calculate the RSD: - select the AVG calulation and save it as a variable, e.g. A - repeatt the same for the STDV. Save it as e.g. B Add a new field under the table using the expression =B/A. If you format the result as percentage it gets automatically mutiplied by 100.Fix Information:
A hot fix is currently investigated.
The following defect may affect the report template editor of OpenLAB CDS installations on Windows XP operating systems: Whenever adding a chromatogram to a template or loading or applying a template with chromatograms, the following error message is shown: Could not load file or assembly 'Agilent.ChromPlot.Infrastructure, Version=1.0.0.0, Culture=neutral, PublicKeyToken=26d0ba2f646bead1' or one of its dependencies. The located assembly's manifest definition does not match the assembly reference.Temporary Solution:
n/aFix Information:
Install all updates from Windows Update for the .NET Framework 3.5 SP1 and .NET Framework 2.0 Service Pack 2 on Windows XP Service Pack 3 and Windows Server 2003 Service Pack 2 , in particular the security update 2518864 (MS11-044). Visit http://update.microsoft.com to download and install the mentioned updates.
Internal grouping of tables and usage of calculation variables cannot be applied at the same time. In cases where the table contains an expression that is saved as a calculation variable it is not possible to define a table internal group. Whenever you setup a group it is not saved and not applied. In cases where a grouped table gets extended by column that defines a new expression saved as a calculation variable, the expression is not calculated and the variable is not saved when previewing the report.Temporary Solution:
In order to apply statistical calculations (average, standard deviation, RSD, etc.) on calculated results, the results need to be stored in an array rather than a single variable. Consult the report template editor on-line on how to use category aggregators to define a result array.Fix Information:
There is currently no fix planned.
Enhanced calculations often require storing results in variables to be used in further calculations. If a table contains a calculation expression with previously defined variables, the statistical summary calculations on the column return "NaN".Temporary Solution:
In order to calculate statistics on table columns including calculation expressions with variables, the results of this line-by-line calculation need to be stored in an array for further processing. Consult the on-line help for the usage of "category aggregator" to set-up such an array.Fix Information:
There is currently no fix planned.
Whenever a template description gets deleted from the template properties dialog a new description can't be entered if the template has been saved with the empty description field. The user will see an error message: --- System.Collections.Generic.KeyNotFoundException --- at System.ThrowHelper.ThrowKeyNotFoundException() at System.Collections.Generic.Dictionary`2.get_Item(TKey key)Temporary Solution:
It is recommended to have the template description field always being populated with text.Fix Information:
There is currently no fix planned.
Whenever naming a customized composite groups with the name of an existing, pre-defined snipped, the composite group name cannot not be deleted from the tree view. The composite group itself is deleted, but name remains under the composite group fields.Temporary Solution:
Use an individual and unique name for a customized compound group.Fix Information:
There is currently no fix planned.
The usage of OR operators in report filter conditions is only allowed when combining several different report parameters to one filter expression. In cases where a report parameter may fulfill multiple filter conditions (e.g. Instrument Name = Instrument A or Instrument B), the report template editor does not allow using an OR operator. This is a limitation due to the definition of the operators available for filtering in report rendering (Microsoft Reporting Services). http://msdn.microsoft.com/en-us/library/ms154029(v=sql.90).aspx. In particular, there is no "OR" at all.Temporary Solution:
The exception is the "IN" operator, which allows a FilterExpression to be compared against a list of FilterValues. The desired filtering may also be achieved by using wildcards (*) in combination with a CONTAINS condition. For number, the usage of > and < conditions could be used to achieve the desired filtering. For external report parameters (e.g. filtering for multiple sample types) the following JOIN statement could be used in the expression editor of the filter: [=Join(Report_Parameter_0, " ").Contains(Sample_Type 1) = True] = "True"Fix Information:
There is currently no fix possible.
The various time stamp formats offered by the Report Template Editor does not offer the combination of dd/mm/yyyy plus time stamp.Temporary Solution:
The format of the time stamps can be modified by using Microsoft Business Intelligence Studio or simple a text editor, like Microsoft Notepad.
- open the report temaplte file (*.RDL)
- search the string "toString"
- Modify the following section like
ToString("MM/dd/yyyy HH:mm:ss")
to e.g.
ToString("dd/mm/yyyy HH:mm:ss")
- save the RDL file and validate the changes.
Note: The Report Template Editor cannot interpret the time stamp format correctly and will show an error filed in the editor window. However, the time stamp format is correctly applied on the reports.
Fix Information:
RTE Support of the (dd/mm/yyyy and time) time stamp format is considered for a future relaese.
In generated .pdf files from SequenceSummary_Standard.rdl and SequenceSummary_Extended.rdl by using PDF XChange printer installed with ChemStation C.01.01, a different background colour may appear in chromatography.Temporary Solution:
Use other virtual printers or physical printers to generate/print reports.Fix Information:
Fix planned for future release.
OpenLAB CDS ChemStation and OpenLAB CDS EZCHrom populate the result field "CalibCurve_CorrCoefficient" with the coefficient of determination "r^2", rather than the correlation coefficient "r".Temporary Solution:
Change the the data field label of your templates to Corr. Coeff. r^2 or r^2 or Correlation r^2 or Goodness of fit (r^2) like EZ Chrom does.Fix Information:
The default labelling is planned to be changed with the next minor revision.
The following defect affects OpenLAB CDS ChemStation (Japanese) only. When loading the SequenceSummary_Standard report template, an error message indicates that the font MSPGothic is missing. By default MSPGothic is not available on Windows XP systems.Temporary Solution:
Download the font MSPGothic from Microsoft and follow the installation instructions.Fix Information:
There is currently no fix planned.
Export of reports based on Short_ESTD.rdl to .doc format creates two unnecessary empty pages.Temporary Solution:
Modify the Short_ESTD.rdl template, reduce the size of the chromatogram picture, and shift the report items together.Fix Information:
There is no fix planned.
The quantified compound response and amount are positioned incorrectly in the related calibration curve plot. The markers are showing absolute responses, rather than relative responses.Temporary Solution:
Do not use this feature when working with ISTD calibrations.Fix Information:
This is currently no fix planned.
Intelligent Reporting of OpenLAB CDS A.01.05 introduces the possibility to report EMF counters. These numbers are initially available for a few Agilent instruments only, like e.g. GC 7890B. For other instruments the instrument's firmware will need to be updated in order to enable this feature.Temporary Solution:
n/aFix Information:
Please visit the technical support section of www.chem.agilent.com and browse for recent firmware updates for your instruments.
The correct syntax for using the and "&" function requires a blank space in front and behind the ampersand sign. Example: An expression like =FieldName&”Response” needs to be written like =FieldName & ”Response” if it is used as a label in a chromatogram plot. Expressions like =Str(FieldName)&”Response” are not supported in reporting service and it results in an “#Error” in preview.Temporary Solution:
Use a blank space in front and behind the ampersand sign.Fix Information:
There is currently no fix planned.
Sizing of the chromatogram output for a Short report does not work in the 20 - 40% range for the Response axis. The 40 - 100% range works correctly, values smaller than 20% are not allowed.Temporary Solution:
n/aFix Information:
This issue will not be fixed in the "classic" reporting from the ChemStation. OpenLAB CDS ChemStation Rev C.01.01 introduces a new Reporting engine "Intelligent Reporting".
Edit the Specify Report section of a method. In the Chromatogram Output, you can select e.g. "Portrait" or "Landscape". It would be expected that "Landscape" means that the chromatogram is printed in landscape format. But actually it means that is printed rotated by 90 degrees in comparison to the orientation of the printer setup. I.e., if you set Landscape in File -> Printer Setup, the chromatograms are printed in Portrait.Temporary Solution:
n/aFix Information:
It is not planned to change this behavior in the classic ChemStation reporting engine.
ChemStation does not report resolution as defined in European Pharmacopoeia. EP requires the following formula to be used: R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a) Only the following definitions are available in ChemStation for resolution: USP, ASTM (in Extended Performance Report): R = 2 * (Tr(b)-Tr(a)) / WB(b)+WB(a) Half-width method as used in Performance Report: R = (2.35/2) * (Tr(b)-Tr(a)) / W50(b)+W50(a) JP (in Report Layout view only): R = 1.18 * (Tr(b)-Tr(a)) / W50(b)+W50(a)Temporary Solution:
In Report Layout view, use Resolution (JP) which uses the same formula.Fix Information:
Fixed in revision C.01.04
If a sequence includes runs with another method apart from the method that should be used for statistics calculation (e.g. a wash method), the statiscal calculations in the sequence summary report cannot be performed for all the runs. The statistical calculations will be done separately for each section of the sequence between the runs with the second method.Temporary Solution:
Mark the second method as "Acquisition only".Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
It is not possible to print report with chromatograms that have been overlaid from different runs.Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
The acquisition parameters missing in the Method section of the sequence summary report when Unique Folder Creation is switched off. Only the run-time checklist, method information and data analysis parameters are reported.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
If user works in batch mode and the batch method is setup to use intelligent reports, this selection is ignored, even if the batch options are set to use the batch method s report settings.Temporary Solution:
N/AFix Information:
There is no plan to fix this issue in future.
When you add more than one report template to a result set, only the seletcted template in the browse dialog will be visible in the Specify Report list box under Report Templates or Report Templates for calibration samplesTemporary Solution:
After closing and reopening the Specify Report menu, all Report Templates added to the result set are visible.Fix Information:
n/a
In case when no column information is available report shows"infinity" instead of the expected "--".Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02
The peak performance drawing in the Classic Reporting and in the Intelligent Reporting are slightly different. Thisi is only a graphical representation and this is not affecting the performanceTemporary Solution:
n/aFix Information:
n/a
Due to a memory leak in the DIF file creation component, when long sequences are running and the ".DIF" output is choosen as report type, the System may hang and abort running instruments. When this occur, the computer needs to be rebooted.Temporary Solution:
This sporadic hang can be prevented by not selecting ".DIF" output.Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.02
Calibration Peaks are counted in the Area sum which causes the results to be incorrect.Temporary Solution:
n/aFix Information:
Fiixed in C.01.03
ECM specific fields like path, server, operator or version number are not printed in the Intelligent Report templates in the Review mode.Temporary Solution:
Print the report from the Data Analysis view.Fix Information:
No fix available.
B.04.03, data analysis, load a data signal (demo\100FID20 for example) and select Graphics. Select Use Ranges and set the time range to 0 to 6 minutes and the response range to 0 to 100 pA for example, then OK. Time scale works (6 minutes) but the response range does not work.Temporary Solution:
WORKAROUND Even though this is a single chromatogram, in Multi Chromatogram, LAYOUT select separated, and SCALE, select All the same scale and it works.Fix Information:
Fixed in revision B.04.03 SP1.
When intelligent report is used for report style for 6890 acquisition, the result chromatogram is not displayed.Temporary Solution:
n/aFix Information:
Fixed in Revision C.01.02.
The CSV, DIF, and EMF files are not generated for some Classic report styles: CSV and DIF files are not created for Library Search and Performance related report styles. For the Extended Performance report style, also the EMF report is not created.Temporary Solution:
n/aFix Information:
It is not planned to fix this issue.
The sample name is not reported when using Intelligent Reporting when working on data that have been required in revisions prior to OpenLAB CDS ChemStation Edition C.01.01.Temporary Solution:
Use the "result set migration" tool of the Chemstation, to migrate the sequence to container format.Fix Information:
Fixed in revision C.01.04
The rendering for chromatograms to be printed in Intelligent Reports provided a low resolution. While this makes rendering faster, the level of detail in longer chromatograms was not high enough.Temporary Solution:
A Hotfix for ChemStation C.01.02 has been released.Fix Information:
Fixed in C.01.03.
The file path in ChemStation has a hard coded limit of 256 characters. In an ECM environment there are also hidden characters added to ensure unique file naming. As a consequence the complete lenth forTemporary Solution:+ + + must not exceed a length of 175 characters.
Make sure that the path length of a report from --> C:\Chem32\... to ...pdf <-- does not exceed 175 characters.Fix Information:
No fix available.
The report viewer that was introduced with C.01.03 allows searching for text within a report. Chinese and Japanese characters, however, cannot be searched.Temporary Solution:
No fix available.Fix Information:
n/a
The peak performance parameters have been modified in ChemStation revision B.04.03 to provide better peak width evaluation for chromatograms with high data rates. This ne algorithm sometimes fails to find the correct inflection points. As a result peak width at baseline (tangent peak widths) can be reported incorrectly. Consequently parameters calculated based on this peak width (tangent resolution and tangent plates) are incorrect as well.Temporary Solution:
A patch for B.04.03 is available on request.Fix Information:
Fixed in revision B.04.03 SP1.
The Mass-Based Purification (Fraction Collection) report is missing information in RC.NET mode; this same information is available in Classic mode. The "Mass" column in the report should specify which detected mass(es) triggered the fraction collection; this feature is currently available in Classic mode only. Additionally, the Fraction Tick Marks displayed are incorrect as is the BeginTime & EndTime information in the report Fraction Information section. The Fraction Tick Marks do not line up accurately with the collected fractions triggered by the LC/MS; these values have the MS Delay Time added to them, which is incorrect.Temporary Solution:
To determine which masses triggered Fraction Collection or what the actual Fraction Collection BeginTime and EndTime values are, open the affected ChemStation data file (.D folder) and open the file MBFC.ini. This MBFC.INI file contains the Mass-Based Fraction Collection triggering events, including Mass, Collection ON, and Collection OFF information. Do not modify this file. The format of MBFC.INI must be interpreted linearly.Fix Information:
Fixed in OpenLAB C.01.05.
In Intelligent Reporting two different variables can be used to report the time of the last recalibration. The two variables are populated in different locations in the code. Both actions should happen a few milliseconds apart, which can lead to a (rounded) difference of one second.Temporary Solution:
NoneFix Information:
N/A
ChemStation / Data Store integration: the "eSignature data file" privilege is missing from the ChemStation lab manager, ChemStation analyst, and ChemStation operator roles.Temporary Solution:
n/aFix Information:
n/a
The structure of the reports in xls, csv and dif format has changed. New columns have been introduced to hold additional information provided since C.01.04.Temporary Solution:
Users who perform their own spreadsheet calculations based on those reports need to be aware that the order and content of columns is different.Fix Information:
N/A
Completed a single run and a report was printed to screen. Loaded a different method in DA Select Report > Print Report No report is printed to screen Error shown in left hand corner: CP3: Number format error in: %dTemporary Solution:
The problem was resolved by doing a clean uninstall and new installation.Fix Information:
Not reproducible.
Method run completed. The following error message appeared at the end of the run in the Logbook. CP Macro ======================================== 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro In Command sprintf$ in macro GetRunPdfRptFil> 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro eNameRoot$ 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro at line 196 in file C:\CHEM32\CORE\REPORT.MC> 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro X 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro Error 41265 occurred: 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro Number format error in: %d 14:46:33 02/13/12 70 41e0 4f396899 0 CP Macro ======================================== 14:46:33 02/13/12 Default report setting was used.Temporary Solution:
The problem was resolved by doing a clean uninstall and new installation.Fix Information:
Not reproducible.
When reprocessing data acquired with ChemStation B.04.0x or earlier, the result data result data sent to OpenLAB ECM Intelligent Reporter may not be updated. This is onlz the case if ChemStation operates with automated ECM upload after reprocessing. The interactive "Save data to ECM" function is not affected.Temporary Solution:
Untill a point patch gets available it is recommended to use the interactive "Save data to ECM" function an addition to the automated ECM upload.Fix Information:
This defect is fixed in OpenLAB CDS ChemStation C.01.04. A fix for this issue is available for C.01.03 with Hotfix02 and any other Hotfixes released after Hotfix02
Doing preview of a report with the method settings to screen only works fine for 25 days. After that, the number used in the filename of the temporary.pdf report file shown in the report preview is getting larger than MAX int (2.147.483.647) and causes a formatting error during the conversion to a string value. After again about 25 days it starts over with 0 and works again for 25 days.Temporary Solution:
The number used for the temp file name is the windows OS tick count, and it gets larger than MAX int after 24 days. It is reset after every reboot.Fix Information:
This issue is fixed with C.01.03 [45] HF06. Fixed with C.01.05. Fixed with B.04.03 SP2.
On systems accessed via Remote Desktop, the on screen classic report may not be displayed properly. Part of the baseline, x-axis or y-axis labels may be missing. This is only affecting the chromatogram in the on-screen classic report. The printed report or display in the data analysis view will correctly show the baseline, axis markers and labels. Affected revisions: C.01.03 and C.01.04.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB revision C.01.05.
An error message appears when a report is printed to screen on computers that have not been rebooted for more than 24 days.Temporary Solution:
For the OpenLAB CDS ChemStation C.01.03 a hotfix 06 is available.Fix Information:
Fixed in B.04.03 SP2. Fixed in C.01.05.
It is not possible to report area for one peak and height for another peak in one report. It is one selection for all compounds. Need to be able to select height or area for each compound.Temporary Solution:
n/aFix Information:
n/a
In case of self-assembled result sets, the order of acquisition sequence lines numbers might appear in a random order. This is caused by the default sorting criteria, which is the injection time.Temporary Solution:
Customize the report template and adjust the sorting criteria to your needs and preferences. In case of self-assembled result sets, the e.g. processing time stamp is a suitable sorting criteria.Fix Information:
There is no fix planned.
In IR Sequence Summary reports always the full sequence is reported, even in partial sequence. Classic SSR reacts on the selection of partial sequence.Temporary Solution:
To have only a subset of the sequence in the IR use the Review mode or create a self-assembled result set.Fix Information:
n/a
Macro error occurs (Parameter 2: Table not found) when printing a calibration tableTemporary Solution:
n/aFix Information:
Fixed in Hotfix03 for C.01.04.
ChemStation C.01.03: XML results: The 'ESTD %' and 'ISTD %' result types in the Quant Calc node are incorrectly reported as 'ESTD' and 'ISTD'. This is confusing the receiving LIMS system. The printed reports have correct titles.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Acquisition method name is not displayed in an intelligent reportTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Report Style LCMS and LCMS Qualitative: Missing Report Style Options: -Background Subtract -Spectrum Average Peak -EICTemporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
The calibration user interface shows the calibration curve of compounds. But on the detailed report, the printing of the calibration curves is stopped when one compound with an invalid calibration curve is encountered.Temporary Solution:
n/aFix Information:
n/a
The ChemStation stays busy and the status line says: Calculating results. The application hangs for at least 10 minutes and needs to be terminated.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.02.
The Short_ESTD.RDL report template (and all related rdl templates) contains a data field "Calibration data modification time" which is not printed if reporting is triggered from the review view. This time stamp is not stored in the result file but it is the change time stamp of the currently loaded method in the Data Analysis view. Hence, this time stamp is not available in the review view.Temporary Solution:
n/aFix Information:
OpenLAB CDS ChemStation C.01.04 introduced a new compound specific and more granular time stamp "CalibrationCurveLastModifiedDate". This time stamp gets stored in the result file and remains with the result. Report templates need to be adjusted in order to print this new time stamp.
The Short_ESTD.RDL report template (and all related rdl templates) contains a data field Calibration data modification time. The time stamp always shows the appendix AM, even in case recalibrations processed in the afternoon. Switching the time format to 24:00 does not correct the problem.Temporary Solution:
N/AFix Information:
This defect if fixed in HF 7 for OpenLAB CDS ChemStation C.01.03 and OpenLAB CDS ChemStation C.01.04 and higher. OpenLAB CDS ChemStation C.01.04 also introduced a new compound specific and more granular time stamp "CalibrationCurveLastModifiedDate". This time stamp gets stored in the result file and remains with the result. Report templates need to be adjusted in order to print this new time stamp.
If the content of a Custom Fields is empty this Custom Field will not appear in an IR at all.Temporary Solution:
Use default values during CF setup.Fix Information:
No fix Planned
When a report is printed to screen, it hides behind the chromatogram window as soon as the individual method DA.M of a data file is loaded. This could occur e.g. at the end of reprocessing when the review option "Use Method from Data File" is selected. In this case, DA.M is loaded again directly after the end of reprocessing.Temporary Solution:
n/aFix Information:
Fiixed in C.01.03
Under certain conditions the number of theoretical plates according to the tangent method may be calculated incorrectly in ChemStation B.04.03. This may especially occur when the peak has a rather wide baseline compared to the peak width or when the peak exhibits distortions at the end of the peak.Temporary Solution:
n/aFix Information:
Fiixed in C.01.03
The Polarity Switch Delay parameter is incorrectly recorded in the ACQ.TXT file that is generated with the run. Only the default parameter of 50 msec is shown for 6130, 6140 and 6150 systems. This gives the impression that the method values are being ignored. However, when Hyperterminal was used to query the "MPST" value directly, it was found that the correct Polarity Switch Delay value was being used. Therefore, the correct Polarity Switch Delay parameter is being used during acquition; the recorded value is simply incorrect.Temporary Solution:
N/AFix Information:
To be fixed in a future release. Fixed in C.01.03.
When the first compound in a calibration table is not found, the sequence summary report does not perform any statistics for all the other compounds.Temporary Solution:
Reporocess with a method that does not have this compound in the calibration table, or with a method that uses a manual integration event at the expected retention time.Fix Information:
Fix planned for C.01.04. Fixed in revision B.04.03 SP1.
It has been observerved under Windows Vista and Windows 7 that during acquisition of large sequences the system is so busy that no other application would start. The ChemStation acquisition was still active in these cases. This has been attributed to the Windows Explorer that is assigning resources to the ChemStation files.Temporary Solution:
The Windows Explorer needs to be closed from the Windows Task Manager. Close the Windows Explorer before starting long sequences. Disabling File Indexing can help.Fix Information:
Cannot be fixed from the ChemStation side.
RC.Net driver for high performance autosampler does not present possibility to switch on or off tray illumination of G1329B or G4303A.Temporary Solution:
On ChemStation's command line enter: print SendModule$(LALS,"LGHT 1") to switch on sampler's illumination and print SendModule$(LALS,"LGHT 0") to switch it off. In ChemStation the use of a configuration with 'Cassic' drivers might also be an option.Fix Information:
fixed in A.02.02 SP1
The parameters that were offered to define customized well plates did not intuitively match the dimensions of a well plate. It was not possible to define whether a plate contains wells or vials. This could also lead to errors in the injected volume.Temporary Solution:
No temporary fix available.Fix Information:
fixed in A.02.02 SP1
Edit Wellplates Dialog does not display imported wellplates.Temporary Solution:
In Edit wellplates dialog Import some custom specific wellplates. Click on Imprt, browse to saved *.wlp file, click on open. File browser closes. Imported wellplate definitions are not visible. Leave Edit Wellplates Dialog with OK. Reopen dialog. Now the imported items are visible.Fix Information:
n/a
Started a run with fraction collection. The fraction collector config does not show 10 vials but only trays.Temporary Solution:
If you have installed a 2 WP-Tray, vials are allways forbidden to use for fractionation, regardless if there are wellplates inserted or not.Fix Information:
Issue fixed with LC and CE Driver revision A.02.07
New G1329A ALS contain a tray illumination now. RC.Net driver for standard autosampler does not present possibility to switch on or off tray illumination of these new G1329A ALS.Temporary Solution:
Check if G1329A is a new one and already has tray illumination. On ChemStation's command line manually enter: print SendModule$(LALS,"LGHT 1") to switch on sampler's illumination and print SendModule$(LALS,"LGHT 0") to switch it off. In ChemStation the use of a configuration with 'Classic' drivers might also be an option.Fix Information:
n/a
The following problem occurs in Japanese and Chinese ChemStation only: RC.Net drivers for Agilent 1260 Infinity Standard Autosampler G1329A/B or Agilent 1260 Infinity SFC Standard Autosampler G4303A do not provide any possibility - neither on ChemStation's menu nor on sampler's context menu - to switch on or off tray illumination inside G1329A/B or G4303A. Control of Agilent 1260 Infinity Standard Autosampler G1329B by 'classic' drivers of ChemStation is not affected. Agilent 1260 Infinity SFC Standard Autosampler G4303A requires the RC.Net Driver. No 'classic' driver for it exists.Temporary Solution:
On ChemStation's command line enter: print SendModule$(LALS,"LGHT 1") to switch on sampler's illumination and: print SendModule$(LALS,"LGHT 0") to switch it off.Fix Information:
The issue will be solved in driver version A.02.03.
With G1329A/B, a 900µl head installed, and a large injection volume specified the RC.Net Drivers may display an irregular injection volume after the injection. Nevertheless the volume, which got injected, correctly represents the method's setpoint. It is a problem affecting only the presentation after the injection cycle. Neither the injected volume itself nor the display during the injection cycle are affected. For the duration of the injection cycle itself the correct injection volume is shown, and injected.Temporary Solution:
n/aFix Information:
n/a
Methods including pretreatment/injector steps are not downloaded to the instrument, prior to a run.Temporary Solution:
n/aFix Information:
The defect is fixed with the Driver Release A.02.06 In relation to the Instrument Control Framework, a PointPatch is available. Please contact your Agilent Support Team.
Pre Cond.: Overlapped injection wit overfill by factor with a DL with different loop sizes for upper and lower loop Consequence: due FW automatism it happens that we will report the wrong loop and the wrong injection volume Wrong injection volume reported in the case of 2 different loops (does not make sense in the application). A logbook entry: "Warning: Used injection volume can be different then specified in the method due to different loop size."Temporary Solution:
no fixFix Information:
n/a
If you change the sample type of a run (that has not yet been executed) of a running sequence, the change in sample type will be ignored.Temporary Solution:
Change the sample type after the sequence has been completed and reprocess the sequence data.Fix Information:
n/a
During the sequence, select pause. After the pause, open edit parameters and increased the oven hold time to 3 minutes and save the method. Resume the sequence. The method changes are not used when running with the Unique Folder ON.Temporary Solution:
n/aFix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01. Changes made to the method during a pause of the sequense, will be applied to the next sample after resuming the sequence.
The time stamps for sequence completion may differ between the sequence logbook and the instrument logbook: The sequence logbook lists the time after completion of the acquisition, while the instrument logbook displays the time after completion of the sequence summary report.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
The following problem occurs when there are data paths specified in the Preferences dialog that no longer exist. When opening the Partial Sequence dialog in this situation, ChemStation will crash.Temporary Solution:
In the Preferences, remove all non-existing data paths.Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
The current path that points to files inside a data folder has a limit of 260 characters. With a data file name of 18 characters (16 + .d) and the known files inside a data folder the effective length of a data path, including the container name cannot exceed 170 characters.Temporary Solution:
NoneFix Information:
No fix planned
In G2070BA, if you have the sequence running with Method A, it is not possible to edit Method A in the Offline session and save it so it can be used for the next sample in the sequence table.Temporary Solution:
n/aFix Information:
Fixed in Revision C.01.01.
Create a five line sequence using Add-on CTC. Run sequence, removing vial three. CTC will report missing vial and it is shown in the log book for CTC. CTC will reset and go back to ready state, however, ChemStation will hang. All menu items are grayed out and you cannot touch anything on the system. The ChemStation must be ended in Task Manager.Temporary Solution:
n/aFix Information:
Issue resolved in OpenLAB C.01.05.
Partial sequence acquisition with headspace samplers has not been possible with GC ChemStation. Running a partial sequence in the ChemStation with a Headspace sampler processes the entire sequence instead of a partial sequenceTemporary Solution:
Beginning in headspace software revision B.01.01 and ChemStation revision B.04.03 DSP1 [54], a point patch is available to support partial sequence with the ChemStation when methods and sequences are saved to C:\.Fix Information:
Fixed in Revision B.04.03 SP1 [87] and Revision C.01.03 [37].
Save method on D drive, and start sequence with headspace or PAL Sampler.
GC stays in "Prep Run".
headspace and PAL Samplers do not work.
Headspace or PAL Sampler does not start when method is saved to a location other than C:\
This defect affects all headspace hardware models with the B.0x.0x series headspace and PAL Sampler Software.
Temporary Solution:
Workaround is to save and use methods from the C:\ drive.Fix Information:
Fixed in revision C.01.04
Sequence does not run or will abort if the currently loaded method vial range is different than the vail range of the next method in the sequence table. Set up two methods: Method E with injection source set to headspace (valid vials 1-111) Method A with injection source set to ALS sampler with tray. (valid vials 1-150) Construct a sequence Vial Method 1 A 1 E 130 A 2 E If method E is the currently loaded active method then this sequence will not validate because the first method in the sequence (method A) has a vial range that is valid for an ALS. The vial numbers are checked against method E which has a vial range for the headspace. The sequence cannot be run. If method A is the currently loaded active method, then this sequence will validate and start running. When the datasystem attempts to move to the 3rd line it validates the vial against E with a headspace injectorID and fails, aborting the sequence.Temporary Solution:
n/aFix Information:
n/a
The sequence parameters (i.e subdirectory, file name, vial location and sample name) overwrites the sample info screen with sequence parameter information during the sequence run only. The Sample Info screen goes back to its original settings once the sequence finishes.Temporary Solution:
n/aFix Information:
n/a
1. Start ChemStation (Online) 2. Select View -> Preferences -> Sequence tab 3. Select Unique Folder Creation OFF 4. Select Sequence -> Sequence Parameters 5. Click and open the dropdown list in the part of methods to run field “Reprocessing Only” is missing.Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03.
Right click the data on navigation table and select “View ACQ method” ACQ method can’t be displayed. For dual injection method, the ACQ method for front data is displayed, but the Acq method for back injector is not displayed.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
For a vial not found in Headspace, the user expects the sequence to continue processing, skipping the non-existent vial. Ran a Headspace sequence and obtained 1 less data file than expected. Attached is the log from the sequence. Here are the parts of interest: 69 41e1 4e2be3c4 0 Sequence 072411A.S started 11:20:04 07/24/11 0 1da7 4e2be3c5 0 7697A 1 Sequence Started 11:20:05 07/24/11 0 1da6 4e2be3c7 0 7890A 1 Attempt to download method to GC failed. 11:20:07 07/24/11 <--The GC driver attempts to stop the first run because the download failed 0 1da6 4e2be3c7 0 7890A 1 Lost connection to Agilent 7890A at 10.100.1> 11:20:07 07/24/11 0 1da6 4e2be3c7 0 7890A 1 00.11 11:20:07 07/24/11 0 1da8 4e2be3c7 0 Instrument HSS1 did not start a run 11:20:07 07/24/11 0 1da7 4e2be3c7 0 7697A 1 Vial 1 starting. Instrument time = 7/24/2011> 11:20:07 07/24/11 0 1da7 4e2be3c7 0 7697A 1 12:17:33 PM. 11:20:07 07/24/11 7da 41e2 4e2be3c7 0 Method Loading Method ALCOHOL.M 11:20:07 07/24/11 0 1da6 4e2be3ca 0 7890A 1 Connected to Agilent 7890A at 10.100.100.11 11:20:10 07/24/11 0 1da6 4e2be3ca 0 7890A 1 Connection found GC in Pre-Run State. Forcin> 11:20:10 07/24/11 0 1da6 4e2be3ca 0 7890A 1 g GC to idle. 11:20:10 07/24/11 0 1da6 4e2be3ca 0 7890A 1 Runs cannot be started until the GC and inje> 11:20:10 07/24/11 0 1da6 4e2be3ca 0 7890A 1 ctor are idle. 11:20:10 07/24/11 443 41e2 4e2be3ce 2010 Method Method started: (B) line# 1 at 1 inj# 1 11:20:14 07/24/11<-- Chemstation starts the first run anyway. Later in the log you'll see that Chemstation did not move onto the 2nd line of the sequence until AFTER the 2nd GC run. This caused the system to lose data.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
It is not possible to run partial sequence acquisition using the back sequence table with Headspace.
Exception is Found:
Exception
- Message: Object reference not set to an instance of an object.
- Source: AgtChemRCNetAdapter
- TargetSite: Void .ctor(PartialSequenceLine, SequenceLine, Int32, Int32, System.String, System.String)
- HelpLink:
- Data: System.Collections.ListDictionaryInternal
- StackTrace:
at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceEntryHelper..ctor(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatId, String OnSampleIDMismatch, String strInjectorID)
at Agilent.ChemStation.RapidControl.DataHelper.RCSequenceHelper.AddEntry(PartialSequenceLine partialSeqLine, SequenceLine seqLine, Int32 nMethodID, Int32 nPretreatID, String strInjectorID)
at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequence.CreateCollections(PartialSequenceTable partialSeqTable, SequenceContext sequenceContext, MethodCollection MethodCollection, PretreatCollection PretreatCollection, RCSequenceHelper SequenceHelper, Boolean bPreventMethodResolution)
at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.FullSequenceDownload.StartFullSequenceMode(String sampleIDMismatch)
at Agilent.ChemStation.RapidControl.AgtChemRCNetAdapter.ControlAdapter.DoPreSequencePhase()
Investigation:
The exception is the reason why the sequence is not downloaded to the instrument. It is related to back sequences. Due to a minor mistake, the attempt to resolve a partial Back-sequence line fails, because the ChemStation tries to resolve it from the Front-Sequence Table.
Temporary Solution:
Workaround is to use Front-Sequences only.Fix Information:
Fixed in revision C.01.04
If the sequence table is opened during a running sequence and the user clicks on a cell and makes no changes, there is a lock placed on the vial so that the instrument doesn't put a vial in the oven. If the user makes no changes and hits OK then the Chemstation never continues past the locked vial.Temporary Solution:
If the user makes a change to the sequence or hits the cancel button then the sequence works correctly.Fix Information:
Fixed in revision C.01.04
If the first calibrated compound is not found in the first sample run of a sequence, no sample statistics will be performed.Temporary Solution:
NoneFix Information:
Fixed in revision B.04.03 SP1.
A partial sequence run with methods saved to a location other than C:\ does not run. The logbook says it is running but the headspace or PAL Samplers do not run. This defect affects all headspace hardware modules with the B.0x.0x series headspace and PAL Sampler software.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.04
The GC Chemstation never had the ability to skip a missing vial (the classic GC driver posts a dialog box asking the user what to do, abort or retry). With RC.NET, if there is no vial on the tray the system stops the sequences, uploads the data to ECM and goes to the next sequence. Need to be able to specify what the system should do if there is a missing vial, skip or wait.Temporary Solution:
n/aFix Information:
Issue resolved in OpenLAB C.01.05.
Pushing the stop key at the GC keyboard will stop the current run and generate a report but does not continue to the next line in the sequence. This occurs beginning on ChemStation version B.04.01 SP1. In revisions B.03.01 - B.04.01, pushing the stop key at the GC Keyboard would stop the current run and generate a report and continue to the next line in the sequence.Temporary Solution:
n/aFix Information:
n/a
During the first run of the sequence, the custom field window is only prompted for the Front injector.Temporary Solution:
n/aFix Information:
Issue fixed in OpenLAB C.01.05.
When reprocessing the sequence, it is not possible to stop it. The stop button is present, but does not to work. The user has to wait till the whole sequence reprocess completes.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
The dialog box to map a sequence import file (Sequence -> Import Sequence Template) is missing the LimsID columns. The LimsID 2 and LimsID 3 columns specifically are read-only fields that need to be imported in order to ensure that a unique LimsID can later be used in the result export to the Lims to ensure results are matched to the correct sample.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
Sequence parameters: Changes to Fraction start location was ignored.Temporary Solution:
n/aFix Information:
n/a
Sequence modified flag is activated even if no changes are made before reprocessing for a dual method sequenceTemporary Solution:
n/aFix Information:
n/a
Migrated a sequence from B.04.03 SP1 with 'Unique folder creation OFF' to C.01.04. Used the 'Result set migration' tool to migrate the data to a result set. In online ChemStation, the result set was reprocessed. In the sequence.log file, 'CP Macro Number format error in: %d' was displayed.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
The Post-sequence command Lampall Off switches off the lamps of all LC modules, but also switches off all the pumps. This happens in RC.Net configuration.Temporary Solution:
NoneFix Information:
Fixed in B.04.03 Fiixed in C.01.03
With B.04.01 Chemstation software, it is impossible to delete, remove or deinstall an instrument.Temporary Solution:
The only workaround is to remove the entire chemstation from the SETUP.exe on the B.04.01 DVD and reinstall the chemstation minus the instrument(s) that is/are not required.Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
Occasionally, when the ChemStation is started or shutdown, these errors are observed: “FetchCPVar Cannot talk to Application” “Error Fetching CP Variable _GUIFInjUpperVial” “Error Fetching CP Variable _GUIBInjLowerVial” “Error Fetching CP Variable _GUIBInjUpperVial”Temporary Solution:
It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software. The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation. It should be possible to turn the views back on after startup.Fix Information:
n/a
With "Fast User Switching" enabled in Microsoft Windows, a user B can connect to an instrument started by user A in a different user session.Temporary Solution:
n/aFix Information:
Fast User Switching must be disabled in Microsoft Windows.
When a networked printer set a default printer is not responding, the ChemStation user interface may become slow when loading datafilles or when changing the scaling mode.Temporary Solution:
N/AFix Information:
N/A
When the "Save Sequence" privilege is granted to a user, this user will be able to create and save Sequence Template(s) or Easy Sequence Template(s). Without this privilege, the "Save Sequence" and "Save Sequence As..." menus will be greyed out. In addition, the "Easy Sequence Setup" tab will not be visible. It is also intended that users will be able to use the "Save Easy Sequence", "Save Easy Sequence as..." and "Save and Add to Queue" buttons from within the Easy Sequence tab.Temporary Solution:
N/AFix Information:
N/A
The tooltip over any circle (vial) in the CTC PAL Sampler Sampling Diagram reads ‘Vial Status: Normal’, no matter what its real status is.Temporary Solution:
n/aFix Information:
n/a
7820A, 7890A has an injector on the back. The sampling diagram is not shown when there is only a back injector (no front injector or tray).Temporary Solution:
n/aFix Information:
Fixed in C.01.03.
Occasionally elements from method and run control view can appear in data analysis view and vice versa.Temporary Solution:
Delete the files config.reg, conf_off.reg and instrument.config from the instrument directory.Fix Information:
Fixed in revision C.01.03.
On ChemStation CDS clients of an remote AIC there are issues with all toolbar's listboxes (signal list, calibration table configuration etc.). These dropdown lists are displayed only after a certain delay, the user has to either wait quite a long time (up to 15 sec) or move the mouse pointer over the area the list should appear to get it updated and repainted.Temporary Solution:
n/aFix Information:
Fixed in C.01.03.
Several problems have been observed on various ChemStation revisions that were the result of one common cause: Video settings that were adjusted for best appearance can cause graphic artifacts and can even cause the ChemStation software to crash.Temporary Solution:
Adjust the display settings for best performance rather than for best appearance. This will free resources and prevent video artifacts from anti-aliasing or filtering attempts by the video driver. The procedures below refer to Windows Classic screens and Control Panel in Classic View. The following steps are recommended depending on the operating system: Windows XP ----------------- 1) Go to Start > Control Panel > System. On the Advanced Tab press the settings button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button. 2) Go to Start > Control Panel > Display. On the Settings tab press the Advanced button. On the Troubleshooting tab set the slider for Hardware acceleration to None. Press the OK button. Windows Vista and Windows 7 ------------------------------------------- 1) Go to Start > Control Panel > System. Select Advanced system settings. On the Advanced Tab ress the Settings... button in the Performance section. Under Visual Effects select "Adjust for best performance". Press the OK button. 2) Go to Start > Control Panel > Administrative Tools. Select Services. Right click on the Desktop Window Manager Session Manager. In the context menu select Properties. Stop the service. Set the startup type to Disabled. This disables the Aero-Glass look.Fix Information:
n/a
7890A, 7820A has an injector on the back. The sampling diagram is not shown when there is only a back injector (no front injector or tray).Temporary Solution:
n/aFix Information:
Fixed in revision B.04.03 SP1.
Change Access Level... is grayed out and not available from the menu bar.Temporary Solution:
n/aFix Information:
n/a
The 490 Micro GC is valid as a standalone instrument, but is not valid as an option on a 7890A.Temporary Solution:
n/aFix Information:
Fixed in revision C.01.03.
Set authentication to Internal and create an 'everything' user Log into Control Panel and open a 6890 session, which fails to initialize This eventually times out and the lock dialog appears. The logged-on user is not displayed in lock dialog. Do not know who is logged into Control Panel.Temporary Solution:
The root cause is that the full name is not set.Fix Information:
Fixed in revision C.01.04.
Create a native sequence table; select RunControl\Queue Sequence... and walk through the sequence dialogs to get to the 'Finish Queue Sequence' dialog. Select Help from the button. Help opens but is hidden behind all the windows and will not come in front of the Finish Queue Sequence dialog.Temporary Solution:
n/aFix Information:
n/a
Connect to a 68xx Classic driver instrument. The old sampling diagram is correctly displayed in the top to bottom column. Select View\Sampling diagram to make it go away, then select again to make it reappear. Only the column reappears. The turret or tray display is not shown again in that session. If the user starts the ChemStation in online the sample turret and run start button are shown in Method and run control. But after the user goes to Data Analysis and back to Method and Run Control the turret picture and start button disappear.Temporary Solution:
Close and reopen the chemstation session and the turret displays again. The RC.NET driver does not have this problem. The user can use the RC.NET driver. Record the IP address, delete the 68xx GC and then add it back again using the RC.NET driver.Fix Information:
n/a
Run-time error ‘282’: No foreign application responded to a DDE initiateTemporary Solution:
It seems to be some timing issue during startup when the Sampling and Instrument diagrams are enabled in the software. The workaround is to turn off all sampling diagrams and instrument diagrams in all 3 views of Method&Run Control, and then shut down and re-start the Chemstation. It should be possible to turn the views back on after startup.Fix Information:
n/a
On ChemStation AIC’s (ChemStation Instrument Controllers), few dialogs are sometimes not properly displayed (e.g. Buttons or checkbox items missing). This has been mainly observed on the Print Method and Edit Entire Method dialogs.Temporary Solution:
n/aFix Information:
A fix for this issue is available in C.01.03 Hotfix04 and higher. To fix this issue, a delayed repaint of the dialogs and an enumeration of all controls within the dialogs are performed. This has been implemented for the following dialogs: Run Time Checklist, Method Information, Edit Entire method, and Print Method. The fix affects the display of the dialogs. Functionality of the items in the dialogs is not affected.
The “Connected User” is updated by the “Agilent OpenLAB ControlPanel ChemStation Plugin” when launching a (ChemStation) instrument from the OpenLAB Control Panel, but not when the user is changed “inside” the ChemStation.Temporary Solution:
n/aFix Information:
This issue is affecting systems with the revision C.01.03.
C.01.04 Chinese version, the values under Overview Table are not aligned with the parameter heading of : RT, Sig, Lvl, Amt[wt%], Area, Rsp.Factor ,Ref, ISTD, #, Compound.Temporary Solution:
n/aFix Information:
Fixed in OpenLAB C.01.05.
The F5 and F8 Key for Start / Stop is intermittently not working.Temporary Solution:
n/aFix Information:
n/a
Preferences set in offline overwrite online Preferences and vice versaTemporary Solution:
n/aFix Information:
n/a
By OpenLAB ChemStation Edition C.01.04 the requirement for ECM's remote data path (LCDF) became more strict than before. If for editing the Sequence Parameters dialog is opened, and a backslash at the LCDF's begin is missing, the builtin check of this dialog-box will demand a manual correction by the user. This can happen especially once loading a sequence template file from an earlier version of OpenLAB ChemStation Edition. The later release of OpenLAB ChemStation Edition (like e.g. C.01.05) will tolerate and correct the missing leading backslash at ECM's remote datat path (LCDF) in Sequence Parameters dialog automatically. Execution of sequences by OpenLAB ChemStaation Edition is not affected. I.e. sequence templates do NOT have to be corrected, neither for sequence execution nor for sequence reprocessing.Temporary Solution:
Leaving the opened Sequence Parameters Dialog by OK is interdicted until all parameters are in valid ranges. This now for OpenLAB ChemStation C.01.04 requires the remote data path (LCDF) for ECM to start with a leading backslash. If at some sequence template (e.g. from an earlier version of OpenLab ChemStation Edition) this character was missing, it now has to be inserted. Only then the Sequnece Parameters dialog can be successfully closed and left with OK-button.Fix Information:
n/a
The link to the base reference file IQTREF.XML on IQT Report is not working.Temporary Solution:
The file is located at \chem32\core.Fix Information:
This is fixed in OpenLAB CDS ChemStation Edition Rev C.01.01.
An execution error occurs when local data cleanup should be performed.
The problem is less frequent with ECM backend.
The Data cleanup fails with the following errors
Input string was not in a correct format.Couldn't store <2.0> in LocalVersion Column. Expected type is Int32
Execution error in macro VC_DataCleanup at level 1 in cmd/fct datacleanup
loaded from C:\\CHEM32\\CORE\\ECMMCX.MCX's line 1004 with CRC 6232
and the cleanup on shutdown with
Input string was not in a correct format.Couldn't store <2.0> in LocalVersion Column. Expected type is Int32
Exception in command 'DataCleanup' occurred! in VC_DataCleanup line 1004 code 999 at datacleanup
Temporary Solution:
Fixed in ChemStation C.01.04 Hotfix 1Fix Information:
Fixed in ChemStation C.01.05
FPD checkout macro used for FPD sample test. This FPD macro uses "autoint.mac" . OL CDS ChemStation C.01.04 no longer includes "autoint.mac" in the CORE directory. The following message occurs when loading the FPD Checkout macro: "Could not open file: C:\CHEM32\CORE\AUTOINT.MAC"Temporary Solution:
Copy "autoint.mac" from OL CDS ChemStation C.01.03 and add it to the Core directory of OL CDS ChemStation C.01.04 .Fix Information:
The macro code has been updated and does the following: •Checks to see if the “AutoIntegrate” macro has already been loaded. If so, it does nothing. •If “Autointegrate” macro has not been loaded, it looks for the file autoint.mac. If the file is found, it is loaded. •If autoint.mac is not found, it looks for the file sig_integ.mcx. If that file is found, it is loaded. The new macros work with or without autoint.mac. Autoint.mac is found in the CORE directory up until C.01.03. Beginning in C.01.04, autoint.mac is no longer included in Core. Autoint.mac is now part of sig_integ.mcx. The eLibrary User Contributed Software has been updated with the updated macros. They are also attached.
A shortcut to the ChemStation Real-Time Display is not included in the OpenLAB CDS ChemStation Edition program group. However, the Real-Time Display can still be accessed at: C:\Chem32\CORE\ChemRTD.exeTemporary Solution:
Manually create a shortcut pointing to C:\Chem32\CORE\ChemRTD.exe to access the ChemStation Real-Time DisplayFix Information:
To be resolved in a subsequent OpenLAB CDS ChemStation Edition release